Gilian 800i Air Sampling Pump

Transcription

Gilian 800i Air Sampling Pump
Gilian 800i
Air Sampling Pump
Operation Manual
1000 112TH Circle N, Suite 100 • St. Petersburg, Florida 33716 USA
(800) 451-9444 • (727) 530-3602 • (727) 539-0550 [FAX] • www.sensidyne.com
REF 360-0166-01 (Rev A)
Gilian 800i
Air Sampling Pump
Operation Manual
REF 360-0166-01 (Rev A)
Software Version L.1.02
1
Gilian 800i Air Sampling Pump
PROPRIETARY NOTICE
This manual was prepared exclusively for the owner of the Gilian 800i Live Flow Air Sampling Pump. The material
within this manual is proprietary information and is to be used only to understand, operate, and service the instrument.
By receiving this document, the recipient agrees that neither this document nor the information disclosed within nor
any part thereof shall be reproduced or transferred, physically, electronically or in any other form, or used or disclosed
to others for manufacturing or for any other purpose except as specifically authorized in writing by Sensidyne, LP.
COPYRIGHT NOTICE
© 2013 Sensidyne, LP All Rights Reserved. No part of this document may be reproduced, transmitted, transcribed,
stored in a retrieval system, or translated into any language in any form by any means without the prior written
permission of Sensidyne, LP.
TRADEMARK NOTICE
Sensidyne, the Sensidyne logo, Gilian, and the Gilian logo are registered trademarks of Sensidyne, LP. Gilian 800i and
the Gilian 800i logo are trademarks of Sensidyne, LP. These trademarks are protected through use and registration in
the United States. Other trademarks and servicemarks used in this document are the property of their respective
companies and are used only for informational and explanatory purposes.
SOFTWARE LICENSE
The software installed in the Gilian 800i pump is the property of Sensidyne, LP and shall remain the property of
Sensidyne, LP in perpetuity. The software is protected by U.S. and international copyright laws and is licensed for
specific use with the Gilian 800i pump. The user may NOT reverse-engineer, disassemble, decompile, or make any
attempt to discover the source code of the software. The software may NOT be translated, copied, merged or modified
in any way. The user may NOT sublicense, rent, or lease any portion of the software. The right to use the software
terminates automatically if any part of this license is violated.
DISCLAIMER
THE SELLER ASSUMES NO RESPONSIBILITY WHATSOEVER, TO ANY PARTY WHOSOEVER, FOR ANY
PROPERTY DAMAGE, PERSONAL INJURY, OR DEATH RECEIVED BY OR RESULTING FROM, IN WHOLE, OR
IN PART, THE IMPROPER USE, INSTALLATION, OR STORAGE OF THIS PRODUCT BY THE USER, PERSON,
FIRM, ENTITY, CORPORATION OR PARTY NOT ADHERING TO THE INSTRUCTIONS AND WARNINGS IN THIS
MANUAL, OR OTHERWISE PROVIDED BY THE SELLER OR FROM NOT ADHERING TO ALL FEDERAL, STATE,
AND LOCAL ENVIRONMENTAL AND OCCUPATIONAL HEALTH AND SAFETY LAWS AND REGULATIONS.
THE SELLER SHALL NOT BE LIABLE FOR DIRECT, INDIRECT, CONSEQUENTIAL, INCIDENTAL OR OTHER
DAMAGES RESULTING FROM THE SALE AND USE OF ANY GOODS AND SELLERS’ LIABILITY HEREUNDER
SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF ANY GOODS FOUND DEFECTIVE. THIS WARRANTY
IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR USE OR FOR A PARTICULAR PURPOSE
WHICH ARE EXPRESSLY DISCLAIMED.
360-0166-01 - (Rev A)
2
Gilian 800i Air Sampling Pump
TABLE OF CONTENTS
Preface
• Notices .............................................................................................................. 2
• Warnings ........................................................................................................... 4
• Certifications and Approvals ............................................................................. 5
SECTION ONE: Introduction
•
Components ................................................................................................. 6
SECTION TWO: Set-Up
2.1 Preparation .................................................................................................. 8
2.2 Pump Start-Up ............................................................................................. 9
2.3 Setting the Flow Rate ................................................................................ 10
2.4 Calibration .................................................................................................. 12
2.4.1 Sample Media Method Set-Up........................................................ 12
2.4.2 Display Calibration Procedure ........................................................ 13
SECTION THREE: Program Setup
3.1
3.2
3.3
Programming ............................................................................................. 16
Program enable/disable ............................................................................. 17
Program Edit .............................................................................................. 17
SECTION FOUR: Options
4.1
Option Settings........................................................................................... 18
SECTION FIVE: Operation
5.1
5.2
5.3
5.4
5.5
5.6
Starting the Sample Run ........................................................................... 21
Stopping the Sample Run ......................................................................... 23
Locking the Keypad.................................................................................... 24
Unlocking the Keypad ................................................................................ 24
Clearing the Run Data .............................................................................. 25
Flow Calculation ......................................................................................... 25
SECTION SIX: Maintenance
6.1
6.2
Battery Maintenance .................................................................................. 26
Pump Filter Maintenance ........................................................................... 27
SECTION SEVEN: Appendicies
Appendix A:
Appendix B:
Appendix C:
Appendix D:
Appendix E:
Troubleshooting Guide ................................................................. 28
Parts List ....................................................................................... 30
Specifications ............................................................................... 31
Fast Charger.................................................................................. 33
Factory Calibration and Service..................................................... 34
3
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
WARNINGS
READ AND UNDERSTAND ALL WARNINGS AND INSTRUCTIONS BEFORE USE
Failure to read, understand, and comply with ALL accompanying literature, product labels, and warnings could
result in property damage, severe personal injury, or death.
Read and understand ALL applicable environmental health and safety laws and regulations. Ensure complete
compliance with ALL applicable laws and regulations before and during use of this product.
DO NOT remove, cover, or alter any label or tag on this product, its accessories, or related products.
UNDER NO CIRCUMSTANCES should this product be used except by qualified, trained, technically
competent personnel.
The Gilian 800i portable Air Sampling Pump is intended for both indoor and outdoor use when protected from
splashed or wind blown liquids. The unit is not waterproof so NEVER submerge the unit in water. Pump failure
or faulting may result.
Pump is Intrinsically Safe when used with specified battery pack 783-0007-01. Refer to Certifications and
Approvals section for approval ratings.
DO NOT operate this product should it malfunction, require repair, or have a cracked or broken case.
DO NOT repair or modify, except as specified in Operation Manual. All user controls and adjustments are
made by sealed keypad on front of pump. The only user-replaceable parts are the Battery pack and Pump
Filter (See Section Six).
Use ONLY specified Sensidyne parts when performing maintenance procedures described in this manual.
Intrinsic safety certifications become void by substitution of components, unauthorized repair or alteration. All
other Service to be performed by Sensidyne Authorized Service Departments only. (See Appendix B for Parts
List. See Appendix F for Service Contact Information).
This product uses rechargeable Nickel-Metal-Hydride batteries. Always fully charge before use. DO NOT open
case, charge or replace batteries in an explosive atmosphere. Use only battery pack and chargers specified in
Parts List. Do not insert any foreign objects into the battery charging jack. Do not insert any foreign objects
into the battery connection. Shorting the contacts will blow the protective fuse. DO NOT operate pump while
charging. Caution: Both charger and battery become warm during charging.
If the equipment can come into contact with aggressive substances, it is the responsibility of the user to take
suitable precautions that prevent it from being adversely affected, thus ensuring that the Intrinsic Safety
protection is not compromised. Aggressive substances are acidic liquids or gases that may attack metals, or
solvents that may affect polymeric materials or other solvents or corrosives. Suitable precautions are regular
checks as part of routine inspections and establishing from material data sheets that chemicals known to be
present do not affect material of the pump (polycarbonate, polyphenylene, epoxy).
DO NOT operate with a dirty or blocked inlet filter or kinked tubing. Pump failure or faulting may result.
If further translation is required, please contact the Sensidyne EU Authorized Representative (see Back Cover
for contact information).
360-0166-01 - (Rev A)
4
Gilian 800i Air Sampling Pump
Certification and Approvals
Gilian 800i Air Sampling Pump REF 610-1301-01
NEC/CEC I.S. CL I, DIV 1, GPS A, B, C, D
CL II, GPS E, F, G,
CL III T4
Ta = -20 to +45°C
II 1 G Ex ia IIC T4
Ta = -20 to +45°C FM 07ATEX0018X
ATEX Explosion Protection:
II = Equipment Group
1 - Equipment Category
G = Hazardous Gases, vapors or mists
Ex = Explosion Protection equipment in compliance with EN/IEC/CAN/ANSI
60079 Standard Series
ia = Intrinsic safety protection method
IIC = Gas group
T4 = 135°C, maximum external surface temperature
Ta = -20 to +45°C ambient temperature range
Approval Standards:
United States:
FM 3600 1998
FM 3610 2007
ANSI/ISA 60079-0 2005
ANSI/ISA 60079-11 2002
Canada:
CSA/CAN C22.2 No. 157 1992
CSA/CAN C22.2 No. 157 1992
CAN/CSA E60079-0-02 2002
CAN/CSA E60079-11 2002
International
IEC 60079-0 Ed. 4.0 :2004
IEC 60079-11 Ed. 5.0 :2006
Europe:
EN 60079-0:2006
EN 60079-11:2007
Complies with EN1232:1977 Type G
5
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
SECTION ONE
Introduction
The Gilian 800i is a low flow rate sampling pump with extremely high back pressure capabilities. It offers user
programmability for easy, flexible preprogrammed sampling schedules, long battery life, and fast charge
capability.
This manual assumes that the pump is in the factory default state, with program and all options disabled.
Enabling the program or options can cause the pump to operate in a manner different than described in the
operation section of this manual. Operation with the program or options enabled is described in the applicable
section. The pump can be reset to the default state by the following procedure.
The features that can cause variation in the behavior of the system are the user program capability, An option
of disable restart retry, an option to enter keyboard lock when a sample starts, and the option to start a sample
immediately when power is applied.
To reset pump to factory initialization state:
With the pump power on, turn off pump power, when Off appears in display press
and hold Clr key, Clr will appear and flash, continue holding key until display blanks.
This procedure will disable all options, disable the program and return the display
calibration to the default setup.
360-0166-01 - (Rev A)
6
Gilian 800i Air Sampling Pump
LCD Details
(6.1) Four 7-Segment Characters, indicating
Flow Rate, Time, Volume Sampled and
Messages.
(6.2) VOL & L. Indicates number in display is
volume in liters.
(6.3) MIN. Indicates number in display is a time
in minutes.
(6.4) CC/MIN. Indicates number in display is a
flow rate in cc/minute
(6.5) SET. Indicates the Set/Cal button is active
to select the value shown in the numeric
display
(6.6) FAULT. Indicates a Fault Condition
FAULT appears when pump is not able
to maintain set flow rate.
(6.6) HOLD. If pump is in fault for 30 seconds
continuously, pump enters HOLD and the
indicator appears.
(6.7) HRS. Indicates number in display is
duration in hours
(6.8) Battery Indicator.
3 bars = High charge
2 bars = Medium charge
1 bar = Low charge
6.1
6.2
.
.
HOLD
6.8
HRS
6.7
VOL
L / MIN
CC / MIN
.
FAULT
6.3
6.4
IN MM
6.9
SET PROG
6.5
6.10
6.11
6.6
(6.9) NOT USED
(6.10) Indicates Program will run if pump is
started.
(6.11) Indicates a program is running. The
program controls the pump to be off or on,
depending on the program setup and
progress.
7
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
SECTION TWO
Setup
2.1 Preparation
The battery pack must be fully charged before using pump. It takes about three hours to
charge the battery from complete discharge. Refer to Section Six for full battery
maintenance.
Attach Sensidyne Charger PN 298-0013-01 to power source or Five-Unit Power station.
Plug charger into battery charging jack. Charging cycle will begin immediately and will
complete as indicated by charger LED. Refer to charger labeling or Appendix E for full
details of LED functions.
Battery pack may be charged through built-in jack while attached to pump or separately.
The pump should not be used during charging.
Caution
Both charger and battery pack become warm during charging.
DO NOT operate pump while charger is attached.
Do not short battery terminals. Shorting will blow internal fuse.
360-0166-01 - (Rev A)
8
Gilian 800i Air Sampling Pump
2.2 Pump Start-Up
Power Up
Press and release POWER button
Display will illuminate and run a Start-up
Sequence, then enter Ready Mode
Start-Up Sequence (approx. 10 seconds):
Screen Test
.
.
HOLD
This is a functional test of the entire display
VOL
L MIN
CC / MIN
.
FAULT
HRS
IN MM
SET PROG
Version No.
Indicates the Version of Software installed in
pump (Current version may vary from picture.
Should show version from page 1.)
.
Last Cal screen
Shows number of run hours since last
calibration. If more that 200 hours have passed
since last calibration, the value is displayed for
twice as long and blinks.
.
HRS
If the AutoStart option is enabled, the pump will
start a sample immediately. If “dCLr” is
displayed, AutoStart was selected, but unable to
start because data must be cleared.
Ready Mode
In Ready Mode, display cycles through following
screens:
CC / MIN
Flow Rate Set Point
MIN
Total Sample Time
VOL
.
Total Volume Sampled
L
If no buttons are pushed, Ready Mode continues cycling through screens for 75 minutes
then pump turns off.
9
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
Power Down
Press and hold Power button until display
shows “OFF” (3-4 seconds), then release.
Pump will show “OFF” for a few seconds until it
powers down.
2.3 Setting The Flow Rate
NOTE
This section is required only if you are changing pump flow rate. If you’re using previously
set flow rate, simply verify it using a Reference Meter (see Section 2.4.2).
SET
At Ready Mode, press SET/CAL button once.
“FLO” is displayed.
CC / MIN
Press ENTER button to begin setting the flow
rate.
360-0166-01 - (Rev A)
10
Gilian 800i Air Sampling Pump
CC / MIN
Press ▲ button to increase flow rate set point or
▼ button to decrease flow rate set point, in 1
cc/min increments.
Pressing and holding ▲ or ▼ button will change
setting rapidly after a short delay.
CC / MIN
MIN
When desired flow rate set point is reached
press ENTER button. Pump will return to Ready
Mode. Elapsed Time and Total Volume will be
cleared
VOL
.
11
L
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
2.4 Calibration
The pump should be calibrated every 200 run hours or 30 days, for optimum accuracy of the
displayed flow. Calibration is also recommended when a new flow is chosen. Entering
calibration will reset the hours since last calibration.
2.4.1 Set-Up
Set up a flow reference instrument to measure the pump flow rate, such as a Gilibrator-2 or
Challenger. The Gilibrator-2 is illustrated in this manual.
Choose a sample media of similar back pressure to that used in the field.
Attach 1/4” ID tubing from pump to media and from media to the reference meter.
To Reference Meter
360-0166-01 - (Rev A)
12
Gilian 800i Air Sampling Pump
2.4.2 Display Calibration
NOTE
To exit Calibration Mode without changing any values, press ENTER twice. This action
will also reset the hours since last calibration.
At Ready Mode, press SET/CAL button twice.
“CAL” is displayed
CC / MIN
Press ENTER button to enter Calibration Mode.
“SCAL” (Self-Calibration) is displayed for 10
seconds, allowing the pump to establish a zero
reference for the flow control system. Pump will
then start running, and display the set flow rate.
Measure flow rate using reference meter.
13
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
=
CC / MIN
SET
Adjust pump display to match actual
flow rate on reference meter. Press
▲ button to increase. Press and ▼
button to decrease.
CC / MIN
When pump display matches
reference meter press SET button.
Pump motor continues running and
adjusts speed to deliver adjusted
flow rate. Pump display returns to
the originally selected flow set point.
Continue to measure flow rate on
reference meter. If reference does
not match intended flow rate, you
may repeat Previous 2 Steps until
the actual flow rate is correct. When
the flow rate is correct, Proceed to
Next Step.
?
Press ENTER button to complete
calibration.
The pump stops and returns to
Ready Mode
360-0166-01 - (Rev A)
14
CC / MIN
Gilian 800i Air Sampling Pump
Note On Field Calibration
The above display calibration procedure serves to make internal pump adjustments and
improve the accuracy of the flow display. It does not replace field calibration as described by
OSHA and NIOSH. A flow verification using the Gilibrator and the exact field sampling train
should be conducted before and after each field sample. Procedures for field calibration
may be referenced in the NIOSH Manual of Analytical Methods at www.cdc.gov/niosh or in
the OSHA Technical Manual at www.osha.gov.
15
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
SECTION THREE
Program
The program capability allows a time based sampling program to be set and executed.
From the time it is initiated until it completes the program or is canceled, the pump will use
the programmed time sequence to turn the pump on and off at specified intervals. All
programs specify the flow rate, and a delay time before starting. If option 4 is enabled, the
program will consist of a single run time. If option 4 is disabled (the default setting) the
program consists of four sequential program segments each specifying an on period and off
period, and a cycle count that will repeat the four on/off segments the designated number of
times.
Each on/off interval is set to a number of minutes. If any interval is specified as zero it
terminates that cycle of the program, even if there are non zero intervals at later points in
the program. After each cycle ends, the cycle count is evaluated and the program
terminates or the next cycle starts. Setting a cycle count of zero is not significant and one
cycle will be executed, exactly as if the cycle count were set to one.
At each on time there is an SCAL (Self-calibration), which takes ten seconds to establish a
zero reference for the flow control system. The pump does not run during the SCAL and
this time is not counted as part of the program.
If a flow fault occurs and the pump enters “HOLD”, the time that elapses while the pump is
halted is not counted in the sample runtime, but is counted by the program timer. During this
Fault Activated HOLD time, the unit will attempt to restart every 3 minutes. If the “on” time
for a segment has not expired, Hold restarts are enabled, and the fault condition has been
corrected, the pump will restart and continue within the on time segment. If the on segment
expires before the three minutes elapses, the pump will not restart as it has entered the next
off segment. The pump will restart at the next on interval if one occurs. Up to 10 restarts
will be attempted before the program is terminated.
PROG
3.1 Programming
At Ready Mode, press SET/CAL button three
times. “Pro0”, program disabled or “Pro1”,
program enabled, is displayed.
360-0166-01 - (Rev A)
16
Gilian 800i Air Sampling Pump
Disabled
3.2 Program enable/disable
PROG
Enabled
Press the ▲ button to enable Program Mode,
“Pro1” and “PROG” are displayed.
Press ▼ button to disable Program Mode.
“Pro0” is displayed.
NOTE: If program is not to be edited, press the
enter button repeatedly to advance through the
programming options and return pump to ready
mode.
3.3 Program Editing
The program capability is controlled by the Option 4 setting. If the option is disabled
(default) there are eleven program parameters. If Option 4 is enabled ("single run"), there
are three parameters. Program editing proceeds through each separate numeric
parameters. Press ENTER button to begin setting of parameters.
Press the ▲ button to increase each parameter or ▼ button to decrease each parameter.
Press the ENTER button to accept each setting and move to next parameter.
The parameters are as follows:
Parameter Name
Flow rate
Delay Time before start
On Time, Segment 1
Off Time, Segment 1
On Time, Segment 2
Off Time, Segment 2
On Time, Segment 3
Off Time, Segment 3
On Time, Segment 4
Off Time, Segment 4
Cycle count
Display
0000
d.000
1.000
.000
2.000
.000
3.000
.000
4.000
.000
C.000
Parameter Range
200-800 cc/min
0-999 minutes
0-999 minutes
0-999 minutes
0-999 minutes
0-999 minutes
0-999 minutes
0-999 minutes
0-999 minutes
0-999 minutes
0-999 number of cycles to run
If a parameter is modified, the program is saved to non volatile memory and preserved.
After the last parameter, the pump returns to Ready mode.
17
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
SECTION FOUR
Options
Options allow the functionality of the pump to be modified. Each option can be set to
On (1) or Off (0). The option settings are stored in nonvolatile memory and preserved over
power down and battery changes.
The options are:
Option
01
Option Title
Fault Hold Lock
02
Auto Lock
03
Auto Start
04
Single Event Program
360-0166-01 - (Rev A)
Option Description
If set to On, and the pump
goes into Fault for any reason,
the pump will be locked in Fault
Hold and will not try to restart.
If set to Off, the pump will try to
restart 3 minutes after entering
Fault Hold.
If set to On, the program will
enter Keyboard Lock when a
manual or programmed sample
is started, The keyboard can
be unlocked as described in
section 5.4.
If set to On, the pump will start
a sample when turned on. If
sample data has not been
cleared, the pump will display
“dCLr” and enter idle mode.
If set to On, the program will
specify flow rate, delay time
and a single on time ("Single
Event"). If set to off, the
program will be full length with
4 on/off segments and a cycle
counter.
18
Default
Off (0)
Off (0)
Off (0)
Off (0)
Gilian 800i Air Sampling Pump
4.1 Option Settings
At Ready Mode, press SET/CAL button four
times. “OP” is displayed.
.
.
Press ENTER button to set indicated option.
Option ON
.
.
.
.
Option OFF
Press ▲ button to turn On indicated Option
(o.1.xx on display),
press ▼ button to turn Off Option
(o.0.xx on display).
19
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
Press ENTER button to accept setting, and
advance to Option 2.
Edit additional Options in the same manner as Option 1.
Press ENTER to advance to remaining Options.
CC / MIN
Press ENTER to continue to Ready mode.
All option changes will be saved through power off.
360-0166-01 - (Rev A)
20
Gilian 800i Air Sampling Pump
SECTION FIVE
Operation
5.1 Starting The Sample Run
NOTE: Total Run Time and Total Volume Sampled are cumulative from one sample
run to the next unless you reset the flow rate, clear the display, or calibrate the
display. If you want to clear the values before starting a sample run, see Section 5.5
for instructions on clearing the run data.
Make sure pump is fully charged, that flow rate has been properly set, and that the pump
has been field calibrated using actual sampling set-up. Make certain all sample tubing and
any sample media have been properly installed.
If programmed operation is desired, enable and configure according to section 3. When
program mode is enabled, “PROG” is displayed and starting the pump will begin the
program. “PROG” will remain displayed until program mode is disabled or pump is turned
off. If “PROG” is not displayed the pump will start in manual mode.
Program Mode:
Press and hold the RUN button until “IN” is
displayed.
“IN” signifies the pump is in a program and will
be displayed until the program is complete or
the pump is stopped.
“Pro1” will appear momentarily and will be
added to the rotation of displays shown, while in
a program. Once the program is complete,
“Pro1” will not be displayed.
IN
PROG
If the “delay” program setting is zero, the pump
will start immediately; if a delay has been
programmed, the delay will start.
At each program on time, “SCAL” (as described
in Manual Mode) will be displayed.
When program is complete, pump will return to
ready mode, with program enabled (“PROG”
displayed). To run program again, Press and
Hold the RUN button. If Program mode is no
longer desired, disable according to section 3.
After the program starts, the controls may be
locked if desired. (described in section 5.3)
21
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
Manual Mode:
Press and hold the RUN button until “SCAL” is
displayed, then release button. Pump will start
10 seconds later. Note: “SCAL” indicates pump
is doing an internal Self Adjustment. This self
adjustment may occur during the course of a
sample if the temperature changes by more
than 3ºC. The pump is not operating and the
clock does not count the time while pump is in
SCAL mode.
After the program starts, the controls may be
locked if desired. (described in section 5.3)
During sampling, pump alternately displays
following screens:
CC / MIN
Live Flow Rate
MIN
Total Run Time
VOL
Total Volume Sampled (liters)
.
Program Mode
(Only displayed while a program is active)
L
IN
PROG
360-0166-01 - (Rev A)
22
Gilian 800i Air Sampling Pump
5.2 Stopping The Sample Run
Press and hold the STOP button until pump
motor stops.
Pump alternately displays following screens:
CC / MIN
Set Flow Rate
MIN
Total Run Time
VOL
Total Volume Sampled (liters)
.
L
NOTE
If the pump motor does not stop, go to Section 5.4 to unlock the keypad.
If pump is in “READY” mode, you may power down by pressing and holding the
POWER button for 4-5 seconds. The display will show “OFF” before shutting down.
Sample data will be retained until the clear run function is performed (section 5.5).
23
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
5.3 Locking The Keypad
The keypad can be locked to prevent
tampering.
SET
To lock, press and hold both SET/CAL and
▲/CLR buttons for 5 seconds until “LOCK” is
displayed.
Note: Keypad cannot be locked during “SCAL”.
5.4 Unlocking The Keypad
SET
To unlock, press and hold both SET/CAL and
▲/CLR buttons for 5 seconds until “UnLK” is
displayed.
• Note: The word “LOCK” will replace the TOTAL VOLUME SAMPLED on display.
360-0166-01 - (Rev A)
24
Gilian 800i Air Sampling Pump
5.5 Clearing The Run Data
In Ready Mode, press and hold ▲/CLR button.
“CLr” will be displayed and flash for a total of 8
seconds. When Flow Rate is displayed, release
button. Data will be cleared and pump will return
to Ready mode.
Releasing the CLEAR button during the 8
seconds (before flow rate is displayed), will
retain data.
MIN
Total Run Time (Cleared)
.
Total Volume Sampled (Cleared)
VOL
L
NOTE
If you remove the battery pack before powering down the pump, all stored data will be lost.
Changing the flow rate will also clear previous run data.
5.6 Flow Calculation
The Gilian 800i calculates the total air volume sampled using the following formula:
Total Air Volume (Liters) =
Air Flow Rate (cc/min) x Sample Time (minutes) / 1000 cc/Liter
OR
AFR x ST
1000
25
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
SECTION SIX
Maintenance
6.1 Battery Maintenance
NOTE
Do not charge or replace battery pack while in an explosive atmosphere. Use only
Sensidyne charger PN 298-0013-01 or other charger designated for Gilian 800i.
The Gilian 800i pump uses rechargeable Nickel-Metal-Hydride batteries that must be fully
charged and properly maintained for maximum run time. The battery pack has a charge
time under 4 hours using Fast Charger (PN 298-013-01). Battery pack may be charged
separately or while on the pump.
Make certain charger plug is fully inserted into jack on battery pack (see #3 in Components
of Section 1, for charger jack location).
See Appendix E for more information on charger operation.
After charging is complete, make certain the rubber jack cover is plugged back into the
charging jack to protect the jack during operation.
CAUTIONS & NOTES
Both charger and battery pack become warm during charging.
Charger switches automatically to trickle mode when battery is fully charged.
DO NOT operate pump while charger is attached.
Do not short battery terminals. Shorting will blow internal fuse.
All NiMH batteries lose charge when not in use. If battery pack has not been charged for
3-4 days, recharge battery before use. This ensures that batteries are fully charged just
prior to sampling. NiMH batteries stored for extended time periods should be recharged
every 1-2 months to avoid complete discharge.
Battery pack has an estimated life of 300–500 charge/discharge cycles, depending on use.
Table below shows estimated battery life based on usage level.
Pump Usage Weekly Use
Est. Battery Life
High ................40-60 hrs ........1.0-1.5 yrs
Medium...........20-39 hrs ........1.5-2.5 yrs
Low .................< 20 hrs ..........2.5 yrs
360-0166-01 - (Rev A)
26
Gilian 800i Air Sampling Pump
6.2 Pump Filter Maintenance
Change internal pump filter PN 811-0012-01 when it is dirty or damaged. Reuse o-ring and
ensure it is properly seated when reinstalling.
27
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
APPENDIX A
Troubleshooting Guide
Symptom
Possible Cause
Corrective Action
Pump will not turn
on
Low battery charge
Blown fuse in battery
Dead Cells in battery
Control board problem
Dead cells in battery
Charge battery
Replace battery
Replace battery
Return for service
Replace battery
Pump shows Fault
in display/Enters
HOLD
Inlet filter clogged
Intake obstructed
Replace Filter
Examine sample holder and
remove obstruction or run at
lower flow rate
Return for service
Charge battery
Correct the flow rate per
sample method guidelines
Correct tubing kink
Control board problem
Low battery
Flow rate is set too high for sample
media
Sample media tubing pinched shut
Pump runs flat out
Pump will not make
flow specs
Pump runs erratic &
faults
360-0166-01 - (Rev A)
Internal flow transducer
disconnected
Control board problem
Bad Calibration
Return for service
Recalibrate
Valve dirty or torn
Return for service
Torn diaphragm on Yoke assembly
Leak in pump
Battery not sufficiently charged
Return for service
Input manifold screws may
not be tight after replacing
input filter. Return for service
if tightening screws does not
solve leakage issue
Charge battery
Faulty bearing
Return for service
Faulty motor
Liquid in pump
Charger connected
Return for service
Return for service
Don’t run pump with charger
connected
28
Return for service
Gilian 800i Air Sampling Pump
APPENDIX A
Troubleshooting Guide
Pump surges
Display calibration adjusted out of
range
Charger connected
Reset display calibrations to
factory setting (see
procedure at end of table)
Don’t run pump with charger
connected
Pump will not run
program; Pro1
flashes briefly
Program time is set to zero
Enter non zero program
duration
Keyboard
inoperative
Keyboard is locked
Unlock keyboard (keyboard
may lock automatically if
AutoLock option is turned
on)
Wait for program to complete
or stop program
Pump in off phase of program
Pump stops
occasionally and
restarts after 10
seconds
Normal operation. Flow control is
being rezeroed.
Displayed flow rate
does not match
calibration
Flow display is out of calibration
Pump will not run at Wrong sample tube holder selected
for constant pressure control
desired flow rate
adapter
with low flow
adapter in place
Calibrate
Select tube holder that
incorporates a needle valve
Pump starts when
power is turned on
Auto Start option turned on
Turn AutoStart option off
“dCLr” displayed
when power turned
on
Auto Start option turned on and data
in pump not cleared at the end of
run
IF AutoStart desired, clear
data to allow pump to start at
power on. If Auto Start not
desired, turn option off.
To reset display calibration, Turn off pump power, when Off appears in display press
and hold Clr key, Clr will appear and flash, continue holding key until display blanks.
29
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
APPENDIX B
Parts List
Spare Parts & Accessories
Part Number
Description
783-0007-01
800143
811-0012-01
Battery Pack
Filter Cassette Kit
10 Filters
360-0166-01
Operation Manual
811-9914-02
Diagnostic Panel
800149
Tube Holder Kit, Single Tube Holder Kit (No Manifold), 6 x
70 mm
Tube Holder Kit, Single Tube Holder Kit (No Manifold), 710 x 110 mm
Tube Holder Kit, Dual Manifold (Holders/Ends/Tubing), 6 x
70 mm
Tube Holder Kit, Dual Manifold (Holders/Ends/Tubing), 10
x 110 mm
800259
800148
801407
200484
800159
200505
360-0166-01 - (Rev A)
Tubing, 36”, 1/4” ID
Tubing, 36”, 1/8” ID (with 1/4” ID adapter)
Tubing, 36”, 1/8” ID
30
Gilian 800i Air Sampling Pump
APPENDIX C
Specifications
Performance
Operating High Flow Range ........... 200–800 cc/min
Accuracy ....................................... ± 5%
Constant Flow control..................... < ± 5% of set flow (after calibration)
Constant Flow Compensation ........ 200 to 800 cc/min up to 80” water back pressure
Flow Fault....................................... If flow changes exceed 5%, fault icon appears.
If fault exceeds 30 seconds, pump shuts down.
If Enabled: Pump attempts to restart every 3
minutes for up to 30 minutes.
General
Controls.......................................... Power/Enter, Set/Cal, ▲/Clear, ▼/Run/Stop
Indicators........................................ Flashing Green LED (“Normal Operation”)
Icons (LCD) .................................... Battery Indicator, Hold, Fault, Set
Dimensions .................................... 3.2” (W) x 5.4” (H) x 2.3” (D)
Weight ............................................ 20.5 oz.
Display (Normal Operation)............ Live Flow, Elapsed Time & Volume Sampled
Electrical
Battery Pack................................... Removable, Sealed,
Rechargeable Nickel-Metal-Hydride (6 cells)
Battery Level Indicator.................... Icon displays Full, Mid, & Low charge levels
Interface Connectors ...................... Charging Jack
Charge Time .................................. < 4 hours
31
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
APPENDIX C
Specifications
Intrinsic Safety Equipment Ratings:
(Refer to Certifications and Approvals Section)
US and Canada ..............................Class I, II, III Div 1, Gps A, B, C, D, E, F, G T4
Class I, Zone 0 Group IIC T4
Europe ............................................ATEX Ex II 1 G, Ex ia IIC T4
CE Compliance:
EMC Emissions ..............................EN 55011:1998/A1:1999 Group 1 Class B
EMC Immunity ................................EN 61326:1997/A1:1998/A2:2001
IEC 1000-4-2:1995/EN 61000-4-2:1995
IEC 1000-4-3:2002/EN 61000-4-3:2002/A1:2002
EN1232:1997..................................Type G Compliant
Environmental
Temperature
Operating .......................................0°C to 45°C (32°F to 113°F)
Storage ..........................................-20°C to 45°C (-4°F to 113°F)
Charging (max) ...............................5°C to 40°C (41°F to 104°F)
Charging (for best charge and life) .5°C to 30°C (41°F to 90°F)
Humidity
Operating .......................................0–85 %RH, non condensing
Storage ..........................................0–98 %RH, non-condensing
360-0166-01 - (Rev A)
32
Gilian 800i Air Sampling Pump
APPENDIX D
Fast Charger PN 298-0013-01
The fast charger is available as a single unit and with a five unit power station.
The fast charger is a universal input (100 - 240 VAC, 50-60Hz) charger with the capacity to
rapidly charge NiMH battery packs. It delivers 1 amp in fast charge mode and monitors the
battery for dV/dt changes to terminate charge when the battery reaches full charge. After
the completion of fast charge, charging current is reduced and the battery is topped off for a
fixed time. After completion of top off, the battery enters a trickle charge mode that
automatically maintains full charge. Before charging starts, the charger makes sure the
battery is able to be fast charged by measuring the cell voltage; if the cell voltage is too low,
the battery is trickle charged until the cells are conditioned for fast charge.
Warning: Before charging battery, check to be sure that the charger is idle (LED indicator
orange). The charger cycle will initiate correctly only if started from Idle mode. The
charger will change to Idle mode after being disconnected from the battery for about 20
seconds. Plugging charger into battery pack while charger is not in idle mode will result in
an incomplete charge.
Indicator
Orange
Idle; No connection or bad battery;
Initialization of charge cycle
Red
Fast charge
Green/Orange flash
Top off charge
Green
Trickle charge
Orange/Green flash
Charge cycle error (typically battery fault)
Red flashing
Internal charger fault
33
360-0166-01 - (Rev A)
Gilian 800i Air Sampling Pump
APPENDIX E
Factory Calibration & Service
USA
Sensidyne, LP
1000 112th Circle N, Suite 100
St. Petersburg, Florida 33716 USA
800-451-9444
727-530-3602
727-539-0550 [Main fax]
727-538-0671 [Service fax]
e-mail: info@sensidyne.com
web: www.Sensidyne.com
360-0166-01 - (Rev A)
34
THIS PAGE INTENTIONALLY LEFT BLANK
THIS PAGE INTENTIONALLY LEFT BLANK
Manufactured by:
Sensidyne, LP
1000 112th Circle N, Suite 100
St. Petersburg, Florida 33716
USA
800-451-9444 • 727-530-3602 • 727-539-0550 [fax]
www.Sensidyne.com • info@sensidyne.com
Authorized EU Representative
Schauenburg Electronic Technologies GmbH
Weseler Str. 35 · 45478
Mülheim-Ruhr Germany
+49 (0) 208 9 99 10 • +49 (0) 208 5 41 10 [fax]
www.schauenburg.com • international@schauenburg.com