Introduction to On-Wafer Characterization at Microwave Frequencies

Transcription

Introduction to On-Wafer Characterization at Microwave Frequencies
Innovating Test
Technologies
for better
measurements
faster
Introduction to
On-Wafer Characterization
at Microwave Frequencies
Chinh Doan
Graduate Student
University of California, Berkeley
Innovating Test
Technologies
for better
measurements
faster
Introduction to
On-Wafer Characterization
at Microwave Frequencies
Dr. Tariq Alam
Senior Applications Engineer
Cascade Microtech, Inc.
E-Mail: tariqa@cmicro.com
presented by: Chinh Doan,
University of California, Berkeley
Innovating Test
Technologies
for better
measurements
faster
Presentation Outline
• Microwave Probing Technology
– Air CoplanarTM Probes
• On-Wafer Calibration Methods
– SOLT, TRL/LRM, SOLR, LRRM
• On-Wafer Verification Methods
• Layout Rules
• Calibration and Measurement Software
– WinCal TM
What Are My Measurement Objectives ?
Innovating Test
Technologies
for better
measurements
faster
• Now
– Determine S-parameters of on-wafer
active devices between 500 MHz to 5 or
50 GHz or beyond
• Future?
– ↑ Wafer size (6 to 8 to 12 inch)
– ↑ Frequency range of interest
– ↑ Need for thermal measurements
– ↑ Test automation for throughput
Innovating Test
Technologies
for better
measurements
faster
What Equipment Do I Need ?
•
•
•
•
•
•
•
•
Vector Network Analyzer
Cables
Probes
Probe positioners
Probe station
Controller
Contact Substrate
Impedance Standard Substrate (ISS)
Only Cascade Microtech provides the Total Measurement Solution
from the Test Ports of the VNA down to the wafer level
Innovating Test
Technologies
for better
measurements
faster
Microwave Probing
• On-wafer fixturing needs:
• Electrically
– wide BW transmission
lines
– low contact R
• Mechanically
– consistent probe shape
– placement
– durability
Optimize for loss, impedance match, power and current handling
capability, contact force, tip visibility...
Innovating Test
Technologies
Microwave Probe Transmission Line Contacts
for better
measurements
faster
Poor
Variable loop inductance
prevents calibration
(Requires repeatable transition)
Better
Long path to single ground
contact limits bandwidth
Best
Precise line impedance right
to the ground-signal-ground
contacts
ACP Probe Technology
Innovating Test
Technologies
for better
measurements
faster
K-Connector
Block
Hard
Absorber
Absorber
Air Coplanar
Waveguide Tip
Soft
Absorber
Low-Loss
Cable
Low-loss, low density teflon dielectric coax
Microwave absorber
- consistent attenuation
- termination of coaxial shield energy
- provides rigidity
15 W CW at 10 GHz
5 A DC current
Innovating Test
Technologies
for better
measurements
faster
Low-loss ACP
• Low-loss and
standard ACP
Probes
• Application
– Noise
measurements
– Load Pull
Air-Coplanar Tips
Innovating Test
Technologies
for better
measurements
faster
• Precision tip fabrication
for tight impedance control
l
BeCu tips for Au or Cu pads
Tungsten tips for Al pads
High tip visibility for
consistent placement
on pads
Clear view of
contact point
Preferred 23°
contact angle
Wide contact
area
Innovating Test
Technologies
Installing and Using ACP Probes
for better
measurements
faster
φ
Use Cable strain relief on
positioners
Use Contact Substrate to
planarize probes
*definition: planarization the ability to insure all
contacts are at the same
height
Innovating Test
Technologies
Using ISS Alignment Marks
for better
measurements
faster
Internal Apex
Initial contact
Full skate and overtravel
• Used to set ‘skate’ and probe separation
Maintaining ACP Probes
Innovating Test
Technologies
for better
measurements
faster
• Keep tips clean of dirt and debris
• Inspect and clean connectors
• Electrical verification -- Probe Test
Innovating Test
Technologies
for better
measurements
faster
Calibration
ao
for better
measurements
faster
°°
bo
Port 1
Forward
•
vv v
°
°
•
Perfect
Reflectometer
Switch
Error
Adapter
DUT
[S]
Reverse
Port 2
a3
°°
Innovating Test
Technologies
How Do I Calibrate My VNA ?
b3
Microwave Errors (Forward)
Directivity
Port-1 Match
Reflection Tracking
Port-2 Match
Transmission Tracking
Crosstalk
Calibration Standards
Open
Short
Load
Thru
Line
Etc.
Innovating Test
Technologies
One-port Network Analyzer Model
for better
measurements
faster
ER
aO
ED
ES
ΓDUT
bO
1
ER = frequency response of measurement channel
ED = Directivity of coupler
ES = Port match
Calibrate with three known reflection coefficients
Short - open - load (SOL)
Short - offset short 1 - offset short2
l
l
Two-port Network Analyzer Model
(Forward Model)
Innovating Test
Technologies
for better
measurements
faster
EX
S21
ER
aO
b3
DUT
ED
bO
ET
ES
1
S22
S11
S12
ET = Models imperfections in transmission response
EL = Models signal reflected back into DUT from P2
Calibrate with short-open-load on each port
plus a Thru (SOLT) (uses 10 knowns)
EL
VNA Calibration
Innovating Test
Technologies
bO
for better
measurements
faster
a3
Ideal
Network
Analyzer
aO
b3
DUT
e00
e01
e10
S21
e11
S11
S12
S22
e22
e23
N1
One port or
two port
calibration standards
N2
N3
• Switch modelled and measured separately
– signals simultaneously measured
– two two-port error boxes
e32
e33
Innovating Test
Technologies
for better
measurements
faster
Calibrating the Probe Tips with Coplanar
Waveguide Impedance Standards
ISS
GROUND
GROUND
GROUND
GROUND
GROUND
GROUND
SIGNAL
SIGNAL
SIGNAL
SIGNAL
SIGNAL
SIGNAL
GROUND
GROUND
GROUND
GROUND
GROUND
GROUND
Thru
Short
Loads
• Electrical behavior of standards
– standard dimensions
– probe pitch
– probe placement (use alignment
marks)
– cal coefficients supplied with probe
Innovating Test
Technologies
for better
measurements
faster
SOLT Calibration
Short
Open
(probes in air)
Lshort
Copen
Load
Thru
Lterm
DelayThru
• All standards must be
perfectly known
– available on virtually every
vector network analyzer (CalKit
required)
– open has capacitance (often
negative)
– short and load have
inductance
– sensitive to probe placement
– mathematically
overdetermined
– unpredictable behavior
Innovating Test
Technologies
for better
measurements
faster
TRL/LRM Calibration
Thru
Reflect
Line(s)
OR
Match
• Thru-Reflect-Line
–
–
–
–
–
–
requires least info about standards
S-parameters referenced to line Zo
reference plane at centre of Thru
requires multiple probe spacings
Design rules
Zo is inherently complex at low
frequencies
– not suitable for fixed spacing probes
(e.g., probe card)
• Line-Reflect-Match
– referenced to Zmatch
Innovating Test
Technologies
for better
measurements
faster
Multi-line TRL Benchmark Cal
Method
Thru
Reflect
Line(s)
• Modified TRL algorithm developed by
the U.S. National Institute of
Standards and Technology
• Benchmark on-wafer calibration
method
• Takes an optimal weighted average
of all the line measurements
• On-wafer standards (with DUT)
preferred
• Renormalizes the S-parameter
impedance to 50 Ohms
Innovating Test
Technologies
for better
measurements
faster
SOLR Calibration
Short
Load
Open
(probes in air)
Reciprocal
OR
• Short-Open-LoadReciprocal Thru
– reciprocal Thru requires only S12 =
S21
– tolerant to lossy or highly reactive
insertion standard
– convenient for use with fixed probe
spacing in probe cards
– Does not require a custom
Thru
– convenient for use when DUT
terminals are orientated at 90°
– available in WinCal
SOLR for Right Angle Measurements
Innovating Test
Technologies
for better
measurements
faster
• Carefully constructed right
angle ‘Thru’ standard
• Thru is non-ideal, large dip
at 20 GHz
• Errors in standard cal’s
• SOLR immune to Thru
errors
1.0
0.5
Orthogonal SOLT
|S21|
Orthogonal LRRM
[dB]
0.0
Orthogonal SOLR
-0.5
Straight LRRM
-1.0
0
5
10
15
20
25
30
[GHz]
35
40
45
50
LRRM
Innovating Test
Technologies
for better
measurements
faster
Line
• Cascade Microtech Calibration Research
– Line-Reflect-Reflect-Match Calibration
– like TRL, only Match acts as infinitely high loss line
Reflect
Reflect
(probes in air)
Match
– one transmission line standard only allows fixed probe
spacing calibration
– Thru (line) delay, Match resistance must be known
– measurements referenced to laser trimmed resistor
– required measurement of only one load standard
– load inductance compensation
– uses off-wafer standards (ISS)
– same standards as SOLT only - no need for cal kit
– available in WinCal
Innovating Test
Technologies
for better
measurements
faster
LRRM Calibration
l
l
System drift baseline
LRRM compares with
system drift limit
–
l
best fixed probe
position calibration
SOLT /LRM
–
–
–
growing error w/freq
possible CalKit error
possible ref plane
error
Innovating Test
Technologies
for better
measurements
faster
SOLT
TRL
Popular Calibration Methods
for Wafer Probing
Z0
Reference
Trimmed
Resistor
Inherently
Consistent
No
Probe Card
Support
Absolute
Accuracy
Fair
Fair
Best (if
Corrected)
Transmission
Lines
Yes
Poor
LRM/
LRRM
Trimmed
Resistor
Yes
Fair
Good
SOLR
Trimmed
Resistor
Yes
Best
Good
Innovating Test
Technologies
for better
measurements
faster
Calibration Verification
How Do I Know If My Calibration is Successful ?
Innovating Test
Technologies
for better
measurements
faster
GROUND
GROUND
GROUND
SIGNAL
SIGNAL
SIGNAL
GROUND
GROUND
GROUND
Pad
Open stub
GROUND
GROUND
SIGNAL
SIGNAL
GROUND
GROUND
Line (delay)
Calibration Verification Standards
Innovating Test
Technologies
for better
measurements
faster
Thru
0.1 GHz
0.5
1.0
40.0 GHz
0.10
0.05
[dB]
•Unity Gain
0.00
•1 pS Line
-0.05
•Phase Lag
-0.10
0
5
10
15
20
[GHz]
25
30
35
40
for better
measurements
faster
Reflect
0.5
(probes in air)
0.4
0.3
0.2
0.1
40.0
10
GHz
[dB]
Innovating Test
Technologies
Calibration Verification Standards
0.0
-0.1
-0.2
20
50
0.1
-0.3
GHz
-0.4
10 2050
-0.5
-50
-20
0
5
10
15
20
[GHz]
-10
•Unity Gain
•Negative Capacitance
25
30
35
40
Innovating Test
Technologies
for better
measurements
faster
Calibration Verification Standards
Match
S21 of Standard Loads
0
-10
-20
-30
-40
[dB]
-50
-60
HPC40 on
HPC ISS
-70
-80
0
5
10
15
20
[GHz]
25
30
35
40
• Measured S21 when probes placed on 50ohm loads
for better
measurements
faster
1
0.8
1.5
0.6
2
G
0.4
Open
stub
S
G
3
4
0.2
5
10
20
0.2
0.4
0.6
0.8
1
1.5
2
3
4 5
10 2050
50
0.1
-50
-20
-10
0.0
-0.2
-5
-4
-0.5
40.0 GHz-3
-0.4
[dB]
Innovating Test
Technologies
Independent Calibration Verification
Standards
-2
-1.0
-0.6
-1.5
-0.8
-1
-1.5
0
5
10
15
20
[GHz]
25
30
35
40
• Linear Phase Lag
GHz
Innovating Test
Technologies
for better
measurements
faster
Common Calibration Errors
• Poor Calibration
Verification/
Repeatability
•
•
•
•
Inaccurate Probe Placement
Inaccurate Probe Definitions
Poor Probe Contact
LRM Sensitivity to Differences
in Load Standards
• Incorrect Models for Lines
• Calibration Drift
• Poor Phase Stability of Cable
• Narrow Band ‘Suck out’ • Dirty Connectors
Innovating Test
Technologies
Beware of an Unconnected Substrate
for better
measurements
faster
GSG pads shield like CPW
Fields terminate on
backside of wafer
on one side
GS pads fringe to the ground plane or chuck
• Parasitic couplings to the conductors near DUT
• GSG shields magnetic and electric fields better than GS
Innovating Test
Technologies
for better
measurements
faster
Layout Guidelines
Innovating Test
Technologies
Device Characterization with Coplanar
Waveguide Microwave Probes
for better
measurements
faster
GROUND
SOURCE
DRAIN
GATE
SIGNAL
SIGNAL
GROUND
GROUND
SOURCE
GROUND
Innovating Test
Technologies
for better
measurements
faster
Pad Size (Passivation Window)
• Recommended minimum pad size is 80um x 80um for
ACP Probes when performing automated
measurements
• Smaller pad dimensions can be used for manual
probing
• HPC Probe allows 40um x 70um manual probing
• Passivation height must be considered
• Pad height variation must not exceed 25um
Innovating Test
Technologies
for better
measurements
faster
Process Control Monitor Device
Layout
G
G
S
S
G
G
GSG Test Device
Open Pads & Metal
(Remove Ypad)
Shorted Metal
(Remove Zmetal)
Innovating Test
Technologies
for better
measurements
faster
PPR Corrected H21 Measurement
0.25 µm CMOS Transistor
H21
60
40
Corrected for pad parasitics
FT = 33 GHz
As measured
FT = 25 GHz
20
0
-20
.1
1
10
FREQUENCY (GHz)
100
Innovating Test
Technologies
WinCal
for better
measurements
faster
• VNA Calibration and Measurement Tool
Innovating Test
Technologies
Calibration Features
for better
measurements
faster
• Calibration
– SOL, SOLT, SOLR,
LRM, LRRM, TRL
– Stability tests
Innovating Test
Technologies
Setting Up a CalKit
for better
measurements
faster
CalKit can be easily defined and downloaded to VNA
– Removes one of the most common sources of error
Innovating Test
Technologies
Measurement Features
for better
measurements
faster
• Tools
– Read/Save SParameters
– Pad Parasitic
Removal
– Probe Test
Innovating Test
Technologies
Monitoring the Calibration
for better
measurements
faster
• Stability check checks
system drift
Innovating Test
Technologies
Pad Parasitic Removal
for better
measurements
faster
• Measures intrinsic devices
Innovating Test
Technologies
Calibration Repeatability
for better
measurements
faster
LRRM automatic calibration is
very repeatable
Operator dependent probe
placement errors
– manual cal’s are not
as repeatable
Innovating Test
Technologies
for better
measurements
faster
Summary
• Microwave measurements require careful
calibration, verifications and attention to detail
• Many new applications to accommodate
the varied needs of growing wireless and
high-speed digital needs
• Let Cascade Microtech help you keep up with
future innovations
Innovating Test
Technologies
for better
measurements
faster
E-Mail: tariqa@cmicro.com