Introduction to On-Wafer Characterization at Microwave Frequencies
Transcription
Introduction to On-Wafer Characterization at Microwave Frequencies
Innovating Test Technologies for better measurements faster Introduction to On-Wafer Characterization at Microwave Frequencies Chinh Doan Graduate Student University of California, Berkeley Innovating Test Technologies for better measurements faster Introduction to On-Wafer Characterization at Microwave Frequencies Dr. Tariq Alam Senior Applications Engineer Cascade Microtech, Inc. E-Mail: tariqa@cmicro.com presented by: Chinh Doan, University of California, Berkeley Innovating Test Technologies for better measurements faster Presentation Outline • Microwave Probing Technology – Air CoplanarTM Probes • On-Wafer Calibration Methods – SOLT, TRL/LRM, SOLR, LRRM • On-Wafer Verification Methods • Layout Rules • Calibration and Measurement Software – WinCal TM What Are My Measurement Objectives ? Innovating Test Technologies for better measurements faster • Now – Determine S-parameters of on-wafer active devices between 500 MHz to 5 or 50 GHz or beyond • Future? – ↑ Wafer size (6 to 8 to 12 inch) – ↑ Frequency range of interest – ↑ Need for thermal measurements – ↑ Test automation for throughput Innovating Test Technologies for better measurements faster What Equipment Do I Need ? • • • • • • • • Vector Network Analyzer Cables Probes Probe positioners Probe station Controller Contact Substrate Impedance Standard Substrate (ISS) Only Cascade Microtech provides the Total Measurement Solution from the Test Ports of the VNA down to the wafer level Innovating Test Technologies for better measurements faster Microwave Probing • On-wafer fixturing needs: • Electrically – wide BW transmission lines – low contact R • Mechanically – consistent probe shape – placement – durability Optimize for loss, impedance match, power and current handling capability, contact force, tip visibility... Innovating Test Technologies Microwave Probe Transmission Line Contacts for better measurements faster Poor Variable loop inductance prevents calibration (Requires repeatable transition) Better Long path to single ground contact limits bandwidth Best Precise line impedance right to the ground-signal-ground contacts ACP Probe Technology Innovating Test Technologies for better measurements faster K-Connector Block Hard Absorber Absorber Air Coplanar Waveguide Tip Soft Absorber Low-Loss Cable Low-loss, low density teflon dielectric coax Microwave absorber - consistent attenuation - termination of coaxial shield energy - provides rigidity 15 W CW at 10 GHz 5 A DC current Innovating Test Technologies for better measurements faster Low-loss ACP • Low-loss and standard ACP Probes • Application – Noise measurements – Load Pull Air-Coplanar Tips Innovating Test Technologies for better measurements faster • Precision tip fabrication for tight impedance control l BeCu tips for Au or Cu pads Tungsten tips for Al pads High tip visibility for consistent placement on pads Clear view of contact point Preferred 23° contact angle Wide contact area Innovating Test Technologies Installing and Using ACP Probes for better measurements faster φ Use Cable strain relief on positioners Use Contact Substrate to planarize probes *definition: planarization the ability to insure all contacts are at the same height Innovating Test Technologies Using ISS Alignment Marks for better measurements faster Internal Apex Initial contact Full skate and overtravel • Used to set ‘skate’ and probe separation Maintaining ACP Probes Innovating Test Technologies for better measurements faster • Keep tips clean of dirt and debris • Inspect and clean connectors • Electrical verification -- Probe Test Innovating Test Technologies for better measurements faster Calibration ao for better measurements faster °° bo Port 1 Forward • vv v ° ° • Perfect Reflectometer Switch Error Adapter DUT [S] Reverse Port 2 a3 °° Innovating Test Technologies How Do I Calibrate My VNA ? b3 Microwave Errors (Forward) Directivity Port-1 Match Reflection Tracking Port-2 Match Transmission Tracking Crosstalk Calibration Standards Open Short Load Thru Line Etc. Innovating Test Technologies One-port Network Analyzer Model for better measurements faster ER aO ED ES ΓDUT bO 1 ER = frequency response of measurement channel ED = Directivity of coupler ES = Port match Calibrate with three known reflection coefficients Short - open - load (SOL) Short - offset short 1 - offset short2 l l Two-port Network Analyzer Model (Forward Model) Innovating Test Technologies for better measurements faster EX S21 ER aO b3 DUT ED bO ET ES 1 S22 S11 S12 ET = Models imperfections in transmission response EL = Models signal reflected back into DUT from P2 Calibrate with short-open-load on each port plus a Thru (SOLT) (uses 10 knowns) EL VNA Calibration Innovating Test Technologies bO for better measurements faster a3 Ideal Network Analyzer aO b3 DUT e00 e01 e10 S21 e11 S11 S12 S22 e22 e23 N1 One port or two port calibration standards N2 N3 • Switch modelled and measured separately – signals simultaneously measured – two two-port error boxes e32 e33 Innovating Test Technologies for better measurements faster Calibrating the Probe Tips with Coplanar Waveguide Impedance Standards ISS GROUND GROUND GROUND GROUND GROUND GROUND SIGNAL SIGNAL SIGNAL SIGNAL SIGNAL SIGNAL GROUND GROUND GROUND GROUND GROUND GROUND Thru Short Loads • Electrical behavior of standards – standard dimensions – probe pitch – probe placement (use alignment marks) – cal coefficients supplied with probe Innovating Test Technologies for better measurements faster SOLT Calibration Short Open (probes in air) Lshort Copen Load Thru Lterm DelayThru • All standards must be perfectly known – available on virtually every vector network analyzer (CalKit required) – open has capacitance (often negative) – short and load have inductance – sensitive to probe placement – mathematically overdetermined – unpredictable behavior Innovating Test Technologies for better measurements faster TRL/LRM Calibration Thru Reflect Line(s) OR Match • Thru-Reflect-Line – – – – – – requires least info about standards S-parameters referenced to line Zo reference plane at centre of Thru requires multiple probe spacings Design rules Zo is inherently complex at low frequencies – not suitable for fixed spacing probes (e.g., probe card) • Line-Reflect-Match – referenced to Zmatch Innovating Test Technologies for better measurements faster Multi-line TRL Benchmark Cal Method Thru Reflect Line(s) • Modified TRL algorithm developed by the U.S. National Institute of Standards and Technology • Benchmark on-wafer calibration method • Takes an optimal weighted average of all the line measurements • On-wafer standards (with DUT) preferred • Renormalizes the S-parameter impedance to 50 Ohms Innovating Test Technologies for better measurements faster SOLR Calibration Short Load Open (probes in air) Reciprocal OR • Short-Open-LoadReciprocal Thru – reciprocal Thru requires only S12 = S21 – tolerant to lossy or highly reactive insertion standard – convenient for use with fixed probe spacing in probe cards – Does not require a custom Thru – convenient for use when DUT terminals are orientated at 90° – available in WinCal SOLR for Right Angle Measurements Innovating Test Technologies for better measurements faster • Carefully constructed right angle ‘Thru’ standard • Thru is non-ideal, large dip at 20 GHz • Errors in standard cal’s • SOLR immune to Thru errors 1.0 0.5 Orthogonal SOLT |S21| Orthogonal LRRM [dB] 0.0 Orthogonal SOLR -0.5 Straight LRRM -1.0 0 5 10 15 20 25 30 [GHz] 35 40 45 50 LRRM Innovating Test Technologies for better measurements faster Line • Cascade Microtech Calibration Research – Line-Reflect-Reflect-Match Calibration – like TRL, only Match acts as infinitely high loss line Reflect Reflect (probes in air) Match – one transmission line standard only allows fixed probe spacing calibration – Thru (line) delay, Match resistance must be known – measurements referenced to laser trimmed resistor – required measurement of only one load standard – load inductance compensation – uses off-wafer standards (ISS) – same standards as SOLT only - no need for cal kit – available in WinCal Innovating Test Technologies for better measurements faster LRRM Calibration l l System drift baseline LRRM compares with system drift limit – l best fixed probe position calibration SOLT /LRM – – – growing error w/freq possible CalKit error possible ref plane error Innovating Test Technologies for better measurements faster SOLT TRL Popular Calibration Methods for Wafer Probing Z0 Reference Trimmed Resistor Inherently Consistent No Probe Card Support Absolute Accuracy Fair Fair Best (if Corrected) Transmission Lines Yes Poor LRM/ LRRM Trimmed Resistor Yes Fair Good SOLR Trimmed Resistor Yes Best Good Innovating Test Technologies for better measurements faster Calibration Verification How Do I Know If My Calibration is Successful ? Innovating Test Technologies for better measurements faster GROUND GROUND GROUND SIGNAL SIGNAL SIGNAL GROUND GROUND GROUND Pad Open stub GROUND GROUND SIGNAL SIGNAL GROUND GROUND Line (delay) Calibration Verification Standards Innovating Test Technologies for better measurements faster Thru 0.1 GHz 0.5 1.0 40.0 GHz 0.10 0.05 [dB] •Unity Gain 0.00 •1 pS Line -0.05 •Phase Lag -0.10 0 5 10 15 20 [GHz] 25 30 35 40 for better measurements faster Reflect 0.5 (probes in air) 0.4 0.3 0.2 0.1 40.0 10 GHz [dB] Innovating Test Technologies Calibration Verification Standards 0.0 -0.1 -0.2 20 50 0.1 -0.3 GHz -0.4 10 2050 -0.5 -50 -20 0 5 10 15 20 [GHz] -10 •Unity Gain •Negative Capacitance 25 30 35 40 Innovating Test Technologies for better measurements faster Calibration Verification Standards Match S21 of Standard Loads 0 -10 -20 -30 -40 [dB] -50 -60 HPC40 on HPC ISS -70 -80 0 5 10 15 20 [GHz] 25 30 35 40 • Measured S21 when probes placed on 50ohm loads for better measurements faster 1 0.8 1.5 0.6 2 G 0.4 Open stub S G 3 4 0.2 5 10 20 0.2 0.4 0.6 0.8 1 1.5 2 3 4 5 10 2050 50 0.1 -50 -20 -10 0.0 -0.2 -5 -4 -0.5 40.0 GHz-3 -0.4 [dB] Innovating Test Technologies Independent Calibration Verification Standards -2 -1.0 -0.6 -1.5 -0.8 -1 -1.5 0 5 10 15 20 [GHz] 25 30 35 40 • Linear Phase Lag GHz Innovating Test Technologies for better measurements faster Common Calibration Errors • Poor Calibration Verification/ Repeatability • • • • Inaccurate Probe Placement Inaccurate Probe Definitions Poor Probe Contact LRM Sensitivity to Differences in Load Standards • Incorrect Models for Lines • Calibration Drift • Poor Phase Stability of Cable • Narrow Band ‘Suck out’ • Dirty Connectors Innovating Test Technologies Beware of an Unconnected Substrate for better measurements faster GSG pads shield like CPW Fields terminate on backside of wafer on one side GS pads fringe to the ground plane or chuck • Parasitic couplings to the conductors near DUT • GSG shields magnetic and electric fields better than GS Innovating Test Technologies for better measurements faster Layout Guidelines Innovating Test Technologies Device Characterization with Coplanar Waveguide Microwave Probes for better measurements faster GROUND SOURCE DRAIN GATE SIGNAL SIGNAL GROUND GROUND SOURCE GROUND Innovating Test Technologies for better measurements faster Pad Size (Passivation Window) • Recommended minimum pad size is 80um x 80um for ACP Probes when performing automated measurements • Smaller pad dimensions can be used for manual probing • HPC Probe allows 40um x 70um manual probing • Passivation height must be considered • Pad height variation must not exceed 25um Innovating Test Technologies for better measurements faster Process Control Monitor Device Layout G G S S G G GSG Test Device Open Pads & Metal (Remove Ypad) Shorted Metal (Remove Zmetal) Innovating Test Technologies for better measurements faster PPR Corrected H21 Measurement 0.25 µm CMOS Transistor H21 60 40 Corrected for pad parasitics FT = 33 GHz As measured FT = 25 GHz 20 0 -20 .1 1 10 FREQUENCY (GHz) 100 Innovating Test Technologies WinCal for better measurements faster • VNA Calibration and Measurement Tool Innovating Test Technologies Calibration Features for better measurements faster • Calibration – SOL, SOLT, SOLR, LRM, LRRM, TRL – Stability tests Innovating Test Technologies Setting Up a CalKit for better measurements faster CalKit can be easily defined and downloaded to VNA – Removes one of the most common sources of error Innovating Test Technologies Measurement Features for better measurements faster • Tools – Read/Save SParameters – Pad Parasitic Removal – Probe Test Innovating Test Technologies Monitoring the Calibration for better measurements faster • Stability check checks system drift Innovating Test Technologies Pad Parasitic Removal for better measurements faster • Measures intrinsic devices Innovating Test Technologies Calibration Repeatability for better measurements faster LRRM automatic calibration is very repeatable Operator dependent probe placement errors – manual cal’s are not as repeatable Innovating Test Technologies for better measurements faster Summary • Microwave measurements require careful calibration, verifications and attention to detail • Many new applications to accommodate the varied needs of growing wireless and high-speed digital needs • Let Cascade Microtech help you keep up with future innovations Innovating Test Technologies for better measurements faster E-Mail: tariqa@cmicro.com