Avalon FM20/FM30 Fetal/Maternal Monitor
Transcription
Avalon FM20/FM30 Fetal/Maternal Monitor
Avalon FM20/FM30 Fetal/Maternal Monitor Philips M2702A/M2703A Technical Data Sheet - Release J.3 The Philips Avalon FM20/FM30 fetal monitor offers a reliable, cost-effective, high-performance solution for fetal monitoring needs in clinics and hospitals, doctors’ practices and private households, from prenatal check-ups to antepartum monitoring of highrisk pregnancies, and during labor and delivery. With a cutting-edge feature set, including advance transducers, triplets monitoring, high-quality signal processing, and a touchscreen interface, the Avalon FM20/FM30 sets new standards for performance, flexibility, convenience, and ease of use. The unique Smart Pulse measurement integrated in the Toco transducer simplifies monitoring. Without an additional sensor, the maternal pulse rate is monitored and continuously compared with the heart rate from the fetal sensor (cross-channel verification). In combination with an Avalon CL Transducer System all fetal and maternal measurements can be monitored cableless, including twins, triplets, and Smart Pulse. The battery option gives monitoring flexibility e.g. for in-house transport or bridging mains power outage. Features (FM20/FM30) Transducers and Measurements • Smart transducers are comfortable to wear, waterproof, and easy to clean • Ultrasound fetal heart rate monitoring of a singleton, twins, and even triplets • Fetal Movement Profile (FMP) automatically detects fetal body movements for antepartum applications using ultrasound transducer • External uterine activity (Toco) monitoring • Smart Pulse integrated in the Toco MP transducer detects the maternal pulse rate without the need for an additional sensor (SpO2 or MECG), and reassures correct fetal heart rate monitoring • Cross-Channel Verification (CCV) detects coincidence between all fetal and maternal heart/pulse rates • NST timer for antepartum applications • Optional Non-Stress Test (NST) report software allows interpretation of fetal heart rate and Toco traces, and generates a printed report automatically for an NST, using NICHD guidelines from 2008 for electronic fetal heart rate monitoring • Optional Maternal non-invasive blood pressure including pulse rate • Heart rate with maternal ECG (MECG) • Cableless monitoring in connection with an Avalon CL or Avalon CTS Fetal Transducer System • Monitoring of maternal temperature in connection with the Tympanic Temperature Module • Entering of manually taken maternal temperature FM30 ONLY • Direct fetal heart rate monitoring with fetal ECG (DECG) including waveform display • Intrauterine (IUP) pressure monitoring • Optional Maternal pulse oximetry (SpO2) and continuous maternal pulse rate Monitor • High-resolution color TFT touchscreen display with a wide viewing angle, displaying fetal and maternal measurements are as large, easyto-read numerics 2 • Intuitive user functionality via the touchscreen • Quick and simple cleaning due to the absence of mechanical controls • Optimized size and presentation of measurements with automatic screen layouts • Displaying information such as: Patient identification (including the bed label when connected to a surveillance system such as OB TraceVue/IntelliSpace Perinatal), date and time, alarm information, recorder status, and prompt and error messages • Fast access to functions e.g. starting a recording or resetting the Toco baseline via Configurable SmartKeys • Adjustment of the heart rate and alarm volume with an indication of the current setting • Automatic channel management allowing free selectable connection of all wired fetal transducers, patient module, or remote event marker to any of the fetal sensor sockets at the left side of the monitor • Identification of the relation between transducer and numeric through a blue transducer Finder LED, while the position indicator on the screen shows to which fetal sensor socket a transducer is connected • Comprehensive patient data presentation, including at-a-glance patient identification, and the choice to add user-configurable notes (Roman-8 character set) • Optional battery operation based on a built-in Smart Battery (M4605A) and on an external AC/DC power supply (M8023A #E25 option); the monitor can operate using battery power for at least four hours Recorder (FM20/FM30) • Built-in six-inch high-resolution thermal array recorder with paper type detection and paper-out warning • Trace output of fetal and maternal heart rates, fetal movements, and uterine activity and annotation of time, date, trace identification symbols, patient information, and software versions • Maternal noninvasive blood pressure values are annotated at the end of the measurement • Internal backup memory for at least 7 hours of trace data (>3.5 hours with previous hardware revision <A 00.18): – Stored data recording to print a specific patient episode from a selected patient at high speed, even multiple times – Bridge paper-out for printing the remaining trace data after new paper has been loaded – Data recovery after an interruption of the online data transmission to an OB TraceVue/IntelliSpace Perinatal system over a LAN connection. • Paper Save Mode for maternal measurements FM30 ONLY • Maternal heart rate, SpO2 saturation, and pulse rate values are printed every five minutes • Recording of six-second MECG and/or DECG wave snapshots with 25 mm/sec scaling in parallel or separate to the fetal traces Alarms When an alarm condition is fulfilled, it is signaled by the monitor in the following ways: • An alarm tone sounds • An alarm message is shown on the screen • The numeric of the alarming measurement flashes on the screen Flexible Nurse Call Interface • Means for alarms generated on the monitor to be signaled on an external device such as a nurse call system, a beeper, or a light • Providing three general alarm relays • Connection of the external device to the alarm relay of the monitor, and configurable alarm criteria • Active open and closed contacts and a user-definable delay time ALARM CATEGORIES MIB/RS-232 Interface • MIB, Medical Information Bus (IEEE P1073), is a standard for interfacing medical devices, used for the Data Export functionality to export measurement values from the monitor to other devices, or to connect the Tympanic Temperature Module • Physiological Alarms for fetal and maternal parameters (to indicate vital sign alarm limit violations, for instance) • Technical Alarms (INOPs) are triggered by signal quality problems, equipment malfunction, or equipment disconnect Device Interface (USB Interfaces) • Allowing connection of USB devices (mouse, keyboard, bar code reader) to the monitor ALARM MODES • INOP only: INOPs are enabled, with audible and visual indication active (default setting) • All: patient alarms and INOPs are enabled, with audible and visual indicators active ALARM SUSPENSION • Alarms Off/Pause Alarms function • Alarm pause period: indefinite, or one, two, or three minutes ALARM LATCHING/NON-LATCHING • Visual • Audible • Visual and audible Standard Interfaces • Cableless monitoring using the Avalon CL Transducer System, or the Avalon CTS Cordless Transducer System with connection at the fetal sensor socket Optional Interfaces (FM20/FM30) LAN/RS-232 System Interface The LAN/RS-232 interface has two fully isolated ports: • The LAN port is for connecting the monitor to an OB TraceVue/ IntelliSpace Perinatal obstetrical information and surveillance system (revision E.0 or later) on a network. OB TraceVue/IntelliSpace Perinatal controls all patient admit/discharge events. • The RS-232 port can be used for connecting the monitor to an obstetrical information and surveillance system, such as OB TraceVue/IntelliSpace Perinatal. • Dual system interface support. If an obstetrical information and surveillance system is connected via LAN, another information system (EMR) can be connected via RS232 in read-only mode. Service Features (FM20/FM30) • The Configuration Mode is password-protected and allows trained users to customize the monitor configuration • The Demo Mode is password-protected and is intended for training and educational purposes • The Service Mode is password-protected and ensures that only trained staff can access service tests and tasks • The Support Tool uses the LAN interface of the fetal monitor • The Support Tool helps technical personnel to – carry out configuration, upgrades, and troubleshooting on an individual monitor – share configuration settings between monitors – back up the monitor settings Related Products (FM20/FM30) Avalon CL and Avalon CTS Transducer System The monitor is compatible with the Avalon CL and CTS, allowing expectant mothers to be continuously monitored without wires, whether in the bath or shower, in the bed, or while ambulating. Refer to the Avalon CL and the Avalon CTS Data Sheets for further information. Avalon CL Transducer System Avalon CTS Transducer System 3 Tympanic Temperature Module The monitor is compatible with the Tympanic Temperature Module allowing the monitoring of maternal temperature. The temperature is taken in the ear, but result of this measurement can be automatically adjusted to correspond to a different body reference site. The result is shown on the display of the thermometer and transmitted to the monitor. Refer to the Tympanic Temperature Module Data Sheet for further information. The monitor is classified as electromagnetic emissions Class B, except whenever used with the IUP/ECG patient module M2738A, when it is classified as Class A. This ISM device complies with Canadian ICES-003. Cet appareil ISM est conforme à la norme NMB-001 du Canada. Environmental Specifications (FM20/FM30) The monitor may not meet the given performance specifications if stored and used outside the specified temperature and humidity ranges. FM20/FM30 Monitor Temperature Range Operating without battery: 0°C – 45°C (32°F – 113°F) with fully charged battery: 0°C – 40°C (32°F – 104°F) with battery charging: 0°C – 35°C (32°F – 95°F) Tympanic Temperature Module Regulatory and Standards Compliance (FM20/FM30) The monitor is in conformity with the requirements of the European Medical Devices Directive 93/42/EEC and the following major international standards: Safety, Performance • IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+AC:2010 • IEC 60601-1-6:2010 / EN 60601-1-6:2010 • IEC 60601-1-8:2006+A1:2012 / EN 60601-1-8:2007+AC:2010 • IEC 60601-2-49:2011 / EN 60601-2-49:2001 • ANSI/AAMI ES60601-1:2005+A1:2012+C1:2009+A2:2010 • CAN/CSA C22.2#60601-1-08 • JIS T 1303 2005 • AS/NZS 3200.1.0-1998 The possibility of hazards arising from hardware and software errors was minimized in compliance with ISO 14971:2012, IEC 60601-1:2005+A1:2012/EN 60601-1:2006+AC:2010. Alarm sounds are compliant with Standard IEC 60601-1-8:2006+A1:2012 / EN 60601-1-8:2007+AC:2010. Electromagnetic Compatibility • IEC 60601-1-2:2007/EN 60601-1-2:2007+AC:2010 • ICES-003:2012 4 Storage -20°C – 60°C (-4°F – 140°F) (except recorder paper) Humidity Range Operating <95% relative humidity @ 40°C (104°F) Storage <90% relative humidity @ 60°C (140°F) (except recorder paper) Altitude Range Operating -500 – 3.000 m (-1.640 – 9.840 ft.) Storage -500 – 13.100 m (-1.640 – 43.000 ft.) Transducers (M2734B/M2735A/M2736A/M2738A) Temperature Range Operating 0°C – 40°C (32°F – 104°F) Storage -20°C – 60°C (-4°F – 140°F) Humidity Range Operating <95% relative humidity @ 40°C (104°F) Storage <90% relative humidity @ 60°C (140°F) Altitude Range Operating -500 – 3.000 m (-1.640 – 9.840 ft.) Storage -500 – 13.100 m (-1.640 – 43.000 ft.) FOR FM30 ONLY Philips SpO2 Sensors Operating Temperature Range Reusable sensors (M119xX) 10°C – 37°C (50°F – 98.6°F) Disposable sensors (M113xA) 0°C – 55°C (32°F – 131°F) Physical Specifications (FM20/FM30) FM20/FM30 Monitor Power (without battery option) Supply Voltages 100 VAC – 240 VAC ± 10% Supply Frequency Range 50 Hz – 60 Hz Power consumption (current) 0.7 – 0.4 A Dimensions and Weight Size mm/(in) ±1%: W x H x D 286 x 133 x 335 mm ± 1% (without options) (11.3 x 5.2 x 13.2 in ± 1%) Weight without battery 5.1 kg (11.2 lbs) Weight with battery 5.3 kg (11.7 lbs) Classifications Degree of Protection Against Type CF Electrical Shock Electrical Class Transducers (M2734B/M2735A/M2736A/M2738A) Shock Resistance Withstands a 1 m drop to concrete surface with possible cosmetic damage only Water Ingress Protection Code M2734B/M2735A/ IP 68 (immersion up to 1 m water depth M2736A for 5 hours) M2738A IP 67 (immersion up to 0.5 m water depth for 30 minutes) Dimensions and Weight M2734B/M2735A/ Size (diameter) M2736A Weight (without cable) Maximum size mm/(in.): WxHxD Cable length Weight 83 mm (3.27 in.) < 220 g (7.8 oz.) 50 x 28 x 135 mm (2.0 x 1.1 x 5.3 in.) 2.5 m < 150 g (5.3 oz.) Degree of Protection Against Electrical Shock Type CF Fetal/Maternal Performance Specifications Ultrasound Measurement Method Ultrasound Pulsed Doppler Measurement US 50 – 240 bpm Range Resolution Display 1 bpm Printer Class II equipment Mode of Operation Continuous operation Startup Time (Time taken < 30 seconds from switching on the monitor to seeing the first parameter labels) Ingress Protection IP X1 (provided recorder drawer is shut) M2738A Performance Specifications (FM20/FM30) Complies with IEC 60601-2-37:2007/EN 60601-2-37:2008. ECG measurement follows IEC 60601-2-27:2011+AC:2012/EN 60601-227:2006+AC:2006 except clauses listed below: • 201.6.2, 201.8.5.5 • 201.12.1.101 • 202.6.2.101 Jitter @ 200 bpm US Intensity Average output power Peak rarefactional pressure Output beam intensity (Iob) (= spatial average temporal average intensity) Spatial-peak temporal average intensity Effective radiating area @ -12 dB 1/4 bpm ≤ 3 bpm P = (7.4 ± 0.4) mW pr = (40.4 ± 4.3) kPa Iob= Isata = (2.38 ± 0.59) mW/cm2 Ispta = (15.0 ± 3.2) mW/cm2 (3.11 ± 0.74) cm2 US Frequency 1 MHz ± 100 Hz US Signal range Repetition Rate Duration Good 3.5 μVpp – 350 μVpp @ 200 Hz 3.0 kHz ≤ 100 μs Full Acceptable Half-full US Burst Signal Quality Indication Poor Beat to Beat Change (max.) for Ultrasound Toco Measurement Method Sensitivity Resolution Measurement Range Display Printer Empty 28 bpm Strain Gauge Sensor Element 1 unit = 2.5 g 1 unit 1/4 unit 400 units 5 Fetal/Maternal Performance Specifications Signal Range 0 – 127 Units Maximum Offset Range -300 units Baseline Setting 20 units Auto Offset Correction 3 seconds after connecting the transducer, the Toco value is set to 20 units Auto Zero Adjust Toco value is set to zero following a negative measurement value for 5 seconds Pulse from Toco MP Range 40 – 240 bpm Accuracy ±2% or 1 bpm, whichever is greater Resolution Wavelength Range Emitted light energy Maternal Heart Rate Measurement Range Resolution Display Recorder Accuracy Filter Bandwidth Inop Auxiliary Current (Leads Off Detection) Input Signal Range MECG Defibrillator Protection ESU Protection 1 bpm 780 – 1100 nm ≤ 15 mW 30 – 240 bpm 1 bpm 1/4 bpm ± 1 bpm or 1%, whichever is greater 0.8 – 80 Hz < 100 μA 150 μVpp to 6 mVpp None None FM30 ONLY Direct and Maternal ECG DECG Single Lead ECG (derived from Fetal Scalp Electrode) MECG Single Lead ECG (derived from RA and LA electrodes) DECG 20 μVpp to 6 mVpp Input Signal IUP Measurement Method Measurement Range Signal Range Resolution Sensitivity Offset Compensation 6 Display Printer Passive Resistive Strain Gauge Elements -100 – 300 mmHg -99 – 127 mmHg 1 mmHg 1/4 mmHg 5 μ/V/mmHg +100 – 200 mmHg IUP Baseline Setting Accuracy (not including sensor accuracy) Auto Offset Correction 0 mmHg ± 0.5% per 100 mmHg 3 seconds after connecting the transducer, the IUP value is set to 0 mmHg Alarm Specifications (FM20/FM30) Fetal Heart Rate (Ultrasound Alarm Specifications FHR/DECG Alarm Limits Bradycardia (low limit) 60 – 200 bpm adjustable in 10 bpm steps Default: 110 bpm Tachycardia 70 – 210 bpm adjustable in 10 bpm steps (high limit) Default: 170 bpm FHR/DECG Alarm Delay Bradycardia 10 – 300 seconds in steps of 10 s (low limit) Default: 240 s Tachycardia 10 – 300 seconds in steps of 10 s (high limit) Default: 300 s Signal Loss 10 – 300 seconds in steps of 10 s Delay MHR/MECG Alarm Specifications High Range 31 – 240 bpm adjustable in 1 bpm steps (30 – 40 bpm), and 5 bpm steps 40 – 240 bpm) Default: 120 bpm Low Range 30 – 235 bpm adjustable in 1 bpm steps (30 – 40 bpm), and 5 bpm steps 40 – 240 bpm) Default: 50 bpm Tachycardia Difference to high limit: 0 – 50 bpm adjustable in 5 bpm steps, Default: 20 bpm Clamping at: 150 – 240 bpm adjustable in 5 bpm steps, Default: 200 bpm Bradycardia Difference to low limit: 0 – 50 bpm adjustable in 5 bpm steps, Default: 20 bpm Clamping at: 30 – 100 bpm adjustable in 5 bpm steps, Default: 40 bpm Noninvasive Blood Pressure (FM20/FM30) Measurement Validation: In adult mode, the blood pressure measurements determined with this device comply with the American National Standard for Electronic or Automated Sphygmomanometers (ANSI/AAMI SP10 - 1992) in relation to mean error and standard deviation, when compared to intra-arterial or auscultatory measurements (depending on the configuration) in a representative patient population. For the auscultatory reference the 5th Korotkoff sound was used to determine the diastolic pressure. Complies with IEC 80601-2-30:2010/EN 80601-2-30:2010. Noninvasive Blood Pressure Performance Specifications Measurement Systolic 30 – 270 mmHg (4 – 36 kPa) Ranges Diastolic 10 – 245 mmHg (1.5 – 32 kPa) Mean 20 – 255 mmHg (2.5 – 34 kPa) Accuracy Max. Std. Deviation: 8 mmHg (1.1 kPa) Max. Mean Error: ± 5 mmHg (± 0.7 kPa) Pulse Rate Range 40 – 300 bpm Accuracy 40 – 100 bpm: ± 5 bpm (average 101 – 200 bpm: ± 5% of reading over non201 – 300 bpm: ± 10% of reading invasive blood pressure measurement cycle) Measurement Time Cuff Inflation Time Initial Cuff Inflation Pressure Auto Mode Repetition Times Venipuncture Mode Inflation Inflation Pressure Automatic deflation after Typical at HR > 60 bpm Auto/manual: 30 seconds maximum time: 180 seconds Typical for normal cuff: Less than 10 seconds 165 ± 15 mmHg 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60, or 120 minutes SpO2 (FM30 only) Maternal pulse oximetry provides SpO2, pulse numerics, and a pulse trace for oxygen saturation (SpO2) of maternal aterial hemoglobin. Pulse rate is derived from SpO2 if ECG is not connected. Complies with ISO 80601-2-61:2011/EN 80601-2-61:2011 (except alarm system; alarm system complies with IEC 60601-2-49:2011/EN 60601-2-49:2001). Measurement Validation: The SpO2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, only about two-thirds of the measurements can be expected to fall within the specified accuracy compared to COoximeter measurements. Display Update Period: Typical: 2 seconds, Maximum: 30 seconds. Maximum with noninvasive blood pressure INOP suppression on: 60 seconds. The specified accuracy is the rootmean-square (RMS) difference between the measured values and the reference values. SpO2 Performance Specifications Range 0 – 100% Accuracy Philips Reusable Sensors: M1191A/B, M1191AL/BL, M119 1ANL, M1192A, M1192AN = 2% (70% to 100%) M1191T, M1192T, M1194A, M1194AN, M1196A, M1196T = 3% (70% to 100%) Philips Disposable Sensors with M1943A(L): M1131A, M1901B, M1903B, M1904B = 3% (70% – 100%) M1133A, M1134A = ±2% (70% – 100%) NellcorPB® Sensors with M1943A(L): 20 – 120 mmHg (3 – 16 kPa) 170 seconds MAX-A, MAX-AL, MAX-P, MAX-N, D-25, D-20, N-25, OxiCliq A, P, N = 3% (70% – 100%) Accuracy ALARM SPECIFICATIONS NBP (FM20/FM30) Alarm Specifications Systolic Diastolic Mean LNC MP10: LNOP DC-I, LNOP DC-IP, LNOP YI, Range Adjustment 30 – 270 mmHg (4 – 36 kPa) 10 – 245 mmHg (1.5 – 32 kPa) 10 – 30 mmHg: 2 mmHg (0.5 kPa) > 30 mmHg: 5 mmHg (1 kPa) 20 – 255 mmHg (2.5 – 34 kPa) Overpressure Settings > 300 mmHg (40 kPa) > 2 sec Masimo Reusable Sensors® with LNOP MP12 or Adjustment not user adjustable LNCS DC-I, LNCS DC-IP: 2% (70% – 100%) LNOP TC-I, LNCS TC-I: 3.5% (70% – 100%) Masimo Disposable Sensors® with LNOP MP12 or LNC MP10: LNOP Adt, LNOP Adtx, LNOP Pdt, Resolution Pulse Range Accuracy Resolution LNOP Pdtx, LNCS Adtx, LNCS Pdtx: 2% (70% – 100%) LNOP Neo-L, LNCS Neo-L: 3% (70% to 100%) 1% 30 – 300 bpm ± 2% or 1 bpm, whichever is greater 1 bpm 7 SpO2 Performance Specifications Sensors Wavelength 500 – 1000 nm information about the wavelength range range can be especially useful to clinicians (for instance, when photodynamic therapy is performed) Emitted ≤ 15 mW Light Energy Pulse 70% – 100% Oximeter Calibration Range ALARM SPECIFICATIONS SPO2 SpO2 Alarm Specifications SpO2 Desat Pulse Tachycardia Bradycardia (FM30 ONLY) Range Adjustment Delay 50 – 100% 50 – low alarm limit 30 – 300 bpm 1% steps 1% steps (0, 1, 2, 3,... 30) + 4 sec max. 14 sec Difference to high limit 0 – 50 bpm Clamping at 150 – 300 bpm Difference to low limit 0 – 50 bpm Clamping at 30 – 100 bpm 1 bpm steps (30 – 40 bpm) 5 bpm steps (40 – 300 bpm) 5 bpm steps max. 14 sec max. 14 sec 5 bpm steps External Power Supply Specifications (FM20/FM30) M8023A External Power Supply Specifications Power Specifications Line Voltage 100 – 240 V approximately Current 1.3 – 0.7 A Frequency 50 – 60 Hz approximately Indicators AC Power green LED Ingress Protection for External Power Supply (M8023A) IP31 (protected against the ingress of solid foreign objects 2.5 mm in diameter or larger, and the ingress of water when water is dripping vertically) when rested on its rubber feet on a flat, level surface 8 Battery Specifications (FM20/FM30) M4605A Battery Specifications Performance Specifications Avalon FM20/30 Battery Operating Time (with new, fully Option #E25 charged battery) Basic monitoring configuration: >2 hours (Display Brightness: 70%, Recorder: "On" at 3 cm/min, NBP: Auto Mode at 15 min, 2 US Transducers, 1 Toco+ with MECG, 1 Patient Module with DECG) Charge Time When monitor is off: approx. 6 hours When monitor is in use: more than 10 hours (depending on monitor configuration) Recorder Specifications (FM20/FM30) 5 bpm steps 5 bpm steps M8023A External Power Supply Specifications IP32 when mounted with the connectors facing downwards Built-in Thermal Array Fetal Trace Recorder Mechanism Thermal Array Recorder Paper and Type Standard Z-fold Printing paper Standard Speeds (real- 3 cm/min, 2 cm/min, time traces) 1 cm/min ECG Wave Recording Emulated 25 mm/s (FM30 only. Not realPrint speed is time.) variable up to 20 mm/s and depends on the print load Stored Data Print speed is Recording/Trace Recovery Printout Paper Advance Sensing Accuracy @ 3 cm/min, 2 cm/min, 1 cm/min ± 5 mm/page variable up to 20 mm/s and depends on the print load Up to 20 mm/s Optical Reflex Sensor for black page marks Built-in Thermal Array Fetal Trace Recorder Usable Print 128 mm Width Resolution 8 dots/mm (200 dpi) Time Delay to < 30s @ 1 cm/min see trace on paper Trace Separation Twin Standard FHR2 +20 bpm Offset for FHR Classic FHR1 +20 bpm (Ultrasound/ Triplet Standard FHR2 +20 bpm DECG) FHR3 -20 bpm Classic FHR1 +20 bpm FHR3 -20 bpm Ordering Guide and Accessory Options (FM20/FM30) You can order the Avalon FM20/FM30 fetal monitor under the product number M2702A/M2703A. To order, precede the required option number with M2702A (for example, M2702A#B71 adds the noninvasive blood pressure measurement). “K” options let you modify your order. The following tables give an overview of the monitor’s standard capabilities, and what is available as options. Refer to “Standard Accessories Included (FM30 only)” on page 11 for a list of items shipped as part of the standard configuration. Start-up quantities of fetal supplies are available in five convenient kits. Features listed as options can be added at a later time (see “Upgrade Options (FM20/FM30)” on page 17). Standard Measurements (FM20/FM30) Description FM20 Recommended Accessory Option FM30 FM30 Standard Fetal Measurements Toco MP K10 DFHR via Direct ECG (DECG) – including DECG waveform Intrauterine Pressure (IUP) K03 Standard Maternal Measurements Maternal ECG and Heart Rate via K03 MECG electrodes, including MECG waveform Optional Measurements (FM20/FM30) Description – K01 Fetal Movement Profile (FMP) – Dual PS/2 Interface for keyboard and mouse connection System Interface (1 x RS232 port and 1 x LAN port) Dual MIB/RS232 interface board (for Data Export functionality and for maternal infrared temperature measurement) Quad. USB Interface (Keyboard, Mouse, Bar Code Reader) You can measure MHR using the Toco+ transducer Order M2735A separately (option K03) Option Recommended Accessory Option(s) n/a K02 K31 K31 + K91 Optional Interfaces (FM20/FM30) Fetal Heart Rate (FHR) via ultrasound Twins monitoring via ultrasound Toco MP – Standard Maternal Measurements Maternal Heart Rate via MECG K03 electrodes Required NST Report Capability Non-Stress Test reporting C71 Optional Fetal Measurements Triple FHR via ultrasound C73 Optional Maternal Measurements Noninvasive Blood Pressure B71 (NIBP) with Pulse Rate Pulse Oximetry (maternal B73 SpO2) together with NBP with Pulse Rate (FM30 only) Description labor. Recommended Accessory Option Description Standard Fetal Measurements We recommend switching off FMP during FM20 ADDITIONAL STANDARDS MEASUREMENTS (FM30 ONLY) J22 Recommended Accessory Option M8024A #A01 J70 – J13 – J25 – Required Option 9 Required Option Description Flexible Nurse Call Relay Interface J30 Recommended Accessory Option – K30 K31 Avalon CL Transducer System Avalon CL Transducer System Product Number CL Base Station 866074 CL Toco+ MP Transducer 866075 CL Ultrasound Transducer 866076 CL ECG/IUP Transducer 866077 Mounting brackets order directly from GCX Tympanic Temperature Module Tympanic Temperature Module Tympanic Temperature Module, including: • a base station for mounting on a table, wall, or a roll stand • A starter pack of 96 probe covers Adding a 1.5 m, RS232 connection cable for use with the fetal monitors Adding a 3.0 m, RS232 connection cable for use with the fetal monitors Product Number 866149 866149 #A01 K02 K03 10 Cable and Transducer Options Description The monitor comes with one ultrasound transducer. To add a second M2736A ultrasound transducer, order option K01 The monitor comes with one ultrasound transducer. To add two additional M2736A ultrasound transducers, order option K02 (only in conjunction with C73) Order ECG Patient Module M2738A, using option K03 (includes MECG adapter cable M1363A) Cable and Transducer Options Description Order the Remote Event Marker (989803143411) using option K04 Avalon CTS Interface Cable M2731-60001 Order Antimicrobial NBP Cuff assortment kit with 3 m (10 ft) connection tubing (M1599B) using option K31: M1573A; M1574A; M1575A ACCESSORIES (FM30 ONLY) Option Number K10 K11 Cable and Transducer Options Description Toco MP transducer and Patient Module (instead of Toco+ transducer) as standard Toco+ Transducer (replaces K10) Paper Scaling Options (FM20/FM30) Option Number P01 P02 Description 50 – 210 bpm paper scaling 30 – 240 bpm paper scaling Transport Options (Battery and External Power Supply FM20/ FM30) Option Number E25 866149 #A02 Accessories (FM20/FM30) Option Number K01 Option Number K04 Description Battery (M4605A), external power supply (M8023A #E25) and MSL cable (1 m) Fetal Supplies Starter Kits (FM20/FM30) Kit Number 862214 862215 862216 Description Contents Standard Antepartum Kit Paper (30 – 240), (North America) belts, gel, belts buttons, belt clips Paper (50 – 210), belts, gel, belts buttons, belt clips Paper (50 – 210), belts, gel, belts buttons, belt clips Standard Antepartum Kit (Europe, Latin America, Asia) Standard Antepartum Kit (Japan) Fetal Accessories (FM20/FM30) FETAL SUPPLIES STARTER KITS (FM30 ONLY) Kit Number 862217 862218 Description Contents Intrapartum DECG Kit (single spiral) Fetal scalp electrodes, fetal adhesive pad electrodes Fetal scalp electrodes, fetal adhesive pad electrodes Intrapartum DECG Kit (double spiral) Europe only, not for USA Standard Accessories Included (FM20/FM30) Description Ultrasound transducer Toco MP transducer Butterfly Belt Clips Fetal paper pack (country-specific, installed) Power cable Instructions for Use Documentation DVD (includes Service Guide and Instructions for Use) Quantity 1 1 1 per Transducer 1 1 1 1 Quantity 1 1 1 Transducers (FM30/FM20) Supported Transducers Ultrasound transducer Toco MP transducer for monitoring external Toco and maternal pulse Part Number M2736A M2734B TRANSDUCER (FM30 ONLY) Supported Transducers Toco+ transducer for monitoring Toco, DECG, MECG, or IUP (K11) Patient Module for monitoring maternal heart rate, DECG or IUP Part Number Belt buttons, pack of 10 Butterfly belt clip (pack of 6) Telemetry Interface Cable Telemetry Avalon CTS Cable M2731interface 60001 cable External Marker M1569A 989803143401 M4601A1 M4602A M4603A M1562B M2208A 40483A 40483B 453564203941 989803143411 1 Note: 32 mm belts for use with M2738A Patient Module only FETAL ACCESSORIES (FM30 ONLY) STANDARD ACCESSORIES INCLUDED (FM30 ONLY) Description Patient module for DECG, MECG, or IUP monitoring (K10) DECG adapter cable 989803137651 MECG adapter cable M1363A Fetal Accessories Belt (reusable, gray, water resistant) 32 mm wide, 15 m roll 60 mm wide, 5 belts 60 mm wide, 15 m roll 50 mm wide, 5 belts Belt (disposable, yellow, water resistant) 60 mm wide, pack of 100 Ultrasound Gel 12 Bottles 5 liter refill (with dispenser) for 40483A Shelf life: 24 months maximum Belt Buttons and Belt Clip Part Number M2735A M2738A Fetal Accessories Part Number DECG Accessories: New Philips DECG Solution DECG reusable leg plate adapter cable (with 989803137651 flushing port) DECG leg attachment electrode for DECG leg 989803139771 plate adapter cable DECG fetal scalp electrode: single spiral, 989803137631 worldwide availability DECG fetal scalp electrode: double spiral, 989803137641 Europe only. Not for USA DECG Accessories: QwikConnect Plus™ Solution DECG reusable leg plate adapter cable M1362B (QwikConnect Plus™) DECG leg attachment electrode for DECG leg M1349A plate adapter cable DECG fetal scalp electrode: single spiral, 15133E worldwide availability DECG fetal scalp electrode: double spiral, 15133D Europe only. Not for USA Koala IUP Accessories Disposable Koala IUP catheter M1333A Reusable Koala IUP adapter cable 989803143931 11 MECG ACCESSORIES (FM30 ONLY) Maternal Heart Rate Accessories MECG adapter cable Foam MECG electrodes, snap-fit, for MECG adapter cable Part Number M1363A 40493D/E Recorder Paper (FM20/FM30) Supplied in cases of 40 packs. Each pack has 150 numbered pages Product No. Geography M1910A USA/Canada and Asia M1911A Europe/Japan M1913A Japan M1913J Japan *Bradycardia and tachycardia alarm ranges are shaded. FHR Scale Grid Color Scale Units 30 – 240 50 – 210 50 – 210 50 – 210 Red/Orange Green Green Green* mmHg mmHg and kPa mmHg mmHg High-lighted 3cm Lines? Yes No Yes Yes Noninvasive Blood Pressure Accessories (FM20/FM30) Multi-Patient Comfort Cuffs and Disposable Cuffs Maternal Patient Category Adult (Thigh) Large Adult Adult Small Adult Cuff kit of 4 adult sizes 12 Limb Circumference 42.0—54.0 cm 34.0—43.0 cm 27.0—35.0 cm 20.5—28.0 cm Bladder Width 20.0 cm 16.0 cm 13.0 cm 10.5 cm Disposable cuff Part No. M1879A M1878A M1877A M1876A Reusable cuff Part No. M1576A M1575A M1574A M1573A M1578A Tubing M1598B (1.5 m) or M1599B (3.0 m) Antimicrobial Coated Reusable Cuffs Maternal Patient Category (color) Limb Circumference Adult Thigh (gray) Adult Thigh (gray) pack of 5 cuffs Large Adult X-Long (burgundy) Large Adult X-Long (burgundy) pack of 5 cuffs Large Adult (burgundy) Large Adult (burgundy) pack of 5 cuffs Adult X-Long (navy blue) Adult X-Long (navy blue) pack of 5 cuffs 45.0—56.5 cm 45.0—56.5 cm Bladder Width 21.0 cm 21.0 cm 35.5—46.0 cm 35.5—46.0 cm 17.0 cm 17.0 cm M4558B M4558B5 35.5—46.0 cm 35.5—46.0 cm 17.0 cm 17.0 cm M4557B M4557B5 27.5—36.5 cm 27.5—36.5 cm 13.5 cm 13.5 cm M4556B M4556B5 Adult (navy blue) Adult (navy blue) pack of 5 cuffs Small Adult (royal blue) Small Adult (royal blue) pack of 5 cuffs Cuff kits containing one small adult, one adult, one large adult and one thigh cuff Cuff kits containing one small adult, one adult, one adult X-long, one large adult, one large adult X-long and one thigh cuff 27.5—36.5 cm 27.5—36.5 cm 13.5 cm 13.5 cm M4555B M4555B5 20.5—28.5 cm 20.5—28.5 cm 10.6 cm 10.6 cm M4554B M4554B5 Part No. Tubing M4559B M4559B5 M1598B (1.5 m) or M1599B (3.0 m) 864288 864291 Soft Single Patient Single-Hose Disposable Cuffs Patient Category Adult (Thigh) Large Adult X-Long Large Adult Adult X-Long Adult Small Adult Limb Circumference (cm) 45.0 – 56.5 35.5 – 46.0 35.5 – 46.0 27.5 – 36.5 27.5 – 36.5 20.5 – 28.5 Bladder Width (cm) 20.4 16.4 16.4 13.1 13.1 10.4 Part No. M4579B M4578B M4577B M4576B M4575B M4574B Tubing M1598B (1.5 m) or M1599B (3.0 m) 13 SpO2 Accessories (FM30 only) All listed sensors operate without risk of exceeding 41°C on the skin if the initial skin temperature is below 37°C. Philips Reusable Sensors Description Adult sensor (2.0 m cable), for patients over 50 kg. Any finger, except thumb. M1191A with longer cable (3.0 m) Small adult, pediatric sensor (1.5 m cable) for patients between 15 kg and 50 kg. Any finger except thumb. Use only on adult patients with FM30/40/50. Ear sensor (1.5 m cable) for patients more than 40 kg. Use only on adult patients with FM30/40/50. Adult clip sensor (3 m cable) for patients over 40 kg. Any finger except thumb. Adult clip sensor (2 m cable) for patients over 40 kg. Any finger except thumb. Adult sensor (0.45 m cable), for patients over 50 kg. Any finger except thumb. Small adult, pediatric sensor (0.45 m cable) for patients between 15 kg and 50 kg. Any finger except thumb. Use only on adult patients with FM30/40/50. Adult clip sensor (0.9 m cable) for patients over 40 kg. Any finger except thumb. Special Edition (SE) Adult sensor (3 m cable), for patients over 50 kg. Any finger except thumb. Special Edition (SE) Small adult, pediatric sensor (1.5 m cable) for patients between 15 kg and 50 kg. Any finger except thumb. Use only on adult patients with FM30/40/50. Special Edition (SE) Ear sensor (1.5 m cable) for patients more than 40 kg. Comments No adapter cable required. Product Number M1191B M1191BL M1192A M1194A M1196A M1196S Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable. M1191T M1192T M1196T No adapter cable required. M1191ANL SE sensors work with FM30/40/50, as well as with OxiMax-compatible SpO2 versions of other Philips monitors. M1192AN M1194AN Philips Disposable Sensors Not available in the USA: Description Identical to OxiMax MAX-A Identical to OxiMax MAX-P Identical to OxiMax MAX-N 14 Comments Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable Product Number M1904B M1903B M1901B Available worldwide: Description Adult/Pediatric finger sensor (0.45 m cable) Use only on adult patients with FM30/40/50. Adult/Pediatric finger sensor (0.9 m cable) for patients >40 kg. Any finger except thumb. Use only on adult patients with FM30/40/50. Adult/Pediatric finger sensor (0.9 m cable) for patients >40 kg. Any finger except thumb. Adhesive-free Use only on adult patients with FM30/40/50. Comments Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable Product Number M1131A M1133A M1134A Nellcor Sensors Nellcor sensors must be ordered from Nellcor/Covidien. Product Number OxiMax MAX-A OxiMax MAX-AL OxiMax MAX-P OxiMax MAX-N Oxisensor II D-25 Oxisensor II D-20 Oxisensor II N-25 OxiCliq A OxiCliq P OxiCliq N Description Adult finger sensor (patient size >30 kg) OxiMax MAX-A with long cable Pediatric foot/hand sensor (patient size 10—50 kg) Use only on adult patients with FM30/40/50. Adult finger or neonatal foot/hand sensor (patient size >40 kg or <3 kg) Use only on adult patients with FM30/40/50. Adult sensor (patient size >30 kg) Pediatric sensor (patient size 10—50 kg) Use only on adult patients with FM30/40/50. Neonatal/Adult sensor (patient size <3 kg or >40 kg) Use only on adult patients with FM30/40/50. See OxiMax MAX-A See OxiMax MAX-P Use only on adult patients with FM30/40/50. See OxiMax MAX-N Use only on adult patients with FM30/40/50 Comments Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable. Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable. Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable together with OC-3 adapter cable. Masimo LNOP Reusable Sensors Product Number LNOP DC-I LNOP DC-IP Description Adult Finger Sensor (>30 kg) Pediatric Finger Sensor (10—50 kg) Use only on adult patients with FM30/40/50 Philips Part Number 989803140321 989803140331 LNOP YI Multi-Site Sensor (>1 kg) Use only on adult patients with FM30/40/50 Ear Sensor (>30 kg) Use only on adult patients with FM30/40/50 n/a LNOP TC-I 989803140341 For use with this sensor the adapter cable LNOP MP12 (M1020-61102) is needed. 15 Masimo LNCS Reusable Sensors Product Number LNCS DC-I LNCS DC-IP Description Adult Finger Sensor (>30 kg) Pediatric Finger Sensor (10—50 kg) Use only on adult patients with FM30/40/50 Philips Part Number 989803148281 989803148291 LNCS TC-I Ear Sensor (>30 kg) Use only on adult patients with FM30/40/50 989803148301 For use with this sensor the adapter cable LNC MP10 (989803148221) is needed. Masimo LNOP Disposable Adhesive Sensors Product Number LNOP Adt Description Adult Sensor (>30 kg) Philips Part Number 989803140231 LNOP Adtx LNOP Pdt Adult Sensor (>30 kg) Pediatric Sensor (10—50 kg) Use only on adult patients with FM30/40/50 Pediatric Sensor (10—50 kg) Use only on adult patients with FM30/40/50 Neonatal (<3 kg) or Adult adhesive Sensor (>40 kg) Use only on adult patients with FM30/40/50 n/a 989803140261 LNOP Pdtx LNOP Neo-L n/a 989803140291 Masimo LNCS Disposable Adhesive Sensors Product Name LNCS Adtx LNCS Pdtx LNCS Neo-L Description Adult Sensor (>30 kg) Pediatric Finger Sensor (10—50 kg) Use only on adult patients with FM30/40/50 Neonatal Foot Sensor (<3 kg) or Adult Finger Sensor (>40 kg) Use only on adult patients with FM30/40/50 Philips Part Number 989803148231 989803148241 989803148271 Input Devices (FM20/FM30) Description Slimline keyboard with integrated trackball (includes spill cover) Optical mouse 16 Option M8024A #A01 M8024A #B01 Description Trackball Wireless trackball Hand-Track Option M8024A #C01 M8024A #C02 M8024A #C03 Mounting Hardware (FM20/FM30) Mounts, Carts, and Rollstands M2740A #A01 M2740A #A05 M2740A #B01 M2740A #C02 Cart with fixed angle mount and two drawers Flush Wall Mount for a flat wall mounting M2740A #R01 Wall Mount (Arm) Bed Hanger M2740A #U01 M2740A #W01 Rollstand with tray Mounting Kit for Avalon CTS for use with M2740A #C02 Wall Channel required for wall mounts A01 and A05 Upgrade Options (FM20/FM30) Upgrade options are prefixed with M2702AU/ M270203AU. For example, to add a system interface upgrade, order M2702AU Option J70. Upgrade Options Option Option Adds Number B71 Noninvasive Blood Pressure, no supplies included B72 Pulse oximetry (SpO2), no supplies included B73 Connector cover for NBP and SpO2 connectors C71 NST Report Capability C73 Triplets monitoring capability CL2 Support for cableless maternal measurements (IntelliVue CL NBP and CL SpO2). Not available with E25 Upgrade Options Option Option Adds Number J13 Dual MIB/RS232 Interface for Data Export functionality, and for maternal infrared temperature measurement J22 Dual PS/2 Interface for connecting a keyboard and mouse J25 Quad. USB Interface (Keyboard, Mouse, Bar Code Reader) J30 Flexible Nurse Call Relay Interface J70 System Interface: 1 x RS232 port and 1 x LAN port K30 Avalon CTS Interface Cable M2731-60001 SJ3 Rel. J.3 Software Upgrade option E25. Battery upgrade for monitors with Rev. G software or higher 17 18 19 Philips Healthcare is part of Royal Philips How to reach us www.philips.com/healthcare healthcare@philips.com fax: +31 40 27 64 887 0366 Avalon FM20/M2702A and FM30/M2703A comply with the requirements of the Council Directive 93/42/EEC of 14 June 1993 (Medical Device Directive). Asia +852 2821 5888 Europe, Middle East, Africa +49 7031 463 2254 Latin America +55 11 2125 0744 North America +1 425 487 7000 800 285 5585 (toll free, US only) Please visit www.philips.com/ © 2014 Koninklijke Philips N.V. All rights are reserved. Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. Printed in The Netherlands. 4522 962 99821 * APR 2014