Standard Specifications for Laboratory Weights p y g and Precision
Transcription
Standard Specifications for Laboratory Weights p y g and Precision
p y Weights g Standard Specifications for Laboratory and Precision Mass Standards (2013 Revision) NCSLI 2013 Nashville At the end of this technical session, you will be able to: ◦ Name the three major weight specifications used in the U.S.; US; ◦ Identify the new ASTM weight classifications; ◦ Visualize the new microgram weights; ◦ Summarize the significant changes in ASTM E617; ASTM E617: Standard Specification for Laboratory Weights and Precision Mass Standards. Industry & Laboratory Standard. OIML R111: Metrological & Technical Requirements for Classes E1 to M3 used for verification of weighing instruments and calibration of weights of a lower accuracy class. L Legal lM Metrology t l (I (Industry d t &L Laboratory) b t ) NIST Handbook 105-1: Specification and T l ffor R f S d d and d Tolerances Reference Standards Field Standard Weights & Measures. Legal metrology & commercial weighing instruments. instruments NBS Circular 3 (1918) NBS S Circular C l 547 (1954) ( 9 ) ◦ NBS Classes J, M, S, S-1, P, Q, & T abo ato y C Classes asses A,, B,, & C ◦ State Laboratory ◦ Commercial weights referred to HB 44 ASTM E617 (1981) Industry Standard ◦ ASTM Classes 1-6 1 6 ◦ Adopted by NBS/NIST ASTM E617 (1997) ◦ Metric Only, Added Classes 0 & 7 ASTM E617 (2013) Uniformity ◦ ◦ ◦ ◦ & Consistency Users of ASTM classified mass standards Mass calibration laboratories ISO/IEC 17025:2005 Assessors Accrediting Bodies Weight Tolerance Classes: ◦ Add an additional weight classification with metrological properties and tolerances similar to OIML E1. Designated g ASTM Class 00. ◦ Add an additional weight classification with metrological properties similar to OIML E1 and tolerances approximately one one-half half of Class 00. 00 Designated ASTM Class 000. Add sub-milligram weights of nominal values 500, 300, 200, 100 & 50 micrograms (µg). Support the new ASTM E617 weight classes. Redefinition of the kilogram. Uncertainties to support the new weight classes l b before/after f / f the h redefinition d fi i i off the h kilogram. Pharma, Forensic, Biotechnology, Silicon Valley, Valley etc. Users NIST Mass & Force Manufacturers Group NIST Calibration OWM Laboratories Major alignment with OIML R111 with the exception of MPE/Classes 0-7. 0 7 Move towards international harmonization while meeting the needs of U.S. industry/commerce. Sub-milligram weight specifications. Mandatory and prescriptive uncertainty components & estimation. Prescriptive minimum comparisons (method). Prescriptive environmental ranges. P i i stabilization bili i i Prescriptive times. Prescriptive manufacturer verification and subsequent q calibrations tests to p perform. Weight manufacturers must be able to provide evidence that all new weights comply with specifications in this standard (e.g., material density material, density, magnetism magnetism, surface finish finish, mass values, uncertainties). Calibration laboratories must meet all the requirements of E617 to make a compliance statement to an ASTM class. BIPM/SI Prototype Kilogram NMI Prototype Kilograms Primary Mass Laboratory Accredited Mass Calibration Lab Mass Used On Weighing Instrument Measurand On g g Instrument Weighing Measurement Process: A. Reference Standard B. Measurand C. Measurement Result D. Measurement Traceability: 1. Unbroken Chain of Comparisons 2 Measurement Uncertainty 2. 3. Documented Procedure 4. Technical Competence 5. Realization of SI Units 6. Periodic Recalibration 7. Measurement Assurance Commonly heard in the USA: “NIST-traceable” traceable” Correct: “SI SI-traceable NIST HB 105-1 (1990) OIML R111 ASTM E617 Metric M t i (SI units) it ) Avoirdupois Apothecary Grains SI Units U it Only O l (metric) ( t i ) Metric M t i (SI units) it ) Avoirdupois Apothecary Grains Troy ounces Pennyweight Carats (2013) ASTM Class 000: ◦ New ASTM class tolerance N l t l ◦ ≥ 1 gram is approximately one-half of ASTM Class 00 ◦ Uncertainty must be less than 1/3 of MPE ASTM Class 00: ◦ New ASTM class tolerance ◦ Alignment to OIML Class E1 ≥ 10 gram ◦ Uncertainty must be less than 1/3 of MPE ASTM Class 0 – 7: ◦ No changes to maximum permissible errors ◦ Uncertainty must be less than 1/3 of MPE The expanded uncertainty of the conventional mass shall be less than or equal to one-third of the MPE. The conventional mass shall not differ by more than the difference: MPE minus expanded uncertainty, from the nominal value of the weight. Reported in SI units, typically milligrams. 000 00 0 1 1 kg k 0.25 0 25 mg 0.5 0 5 mg 1.3 1 3 mg 2.5 2 5 mg 50 µg 0.002 mg 0.003 mg 0.005 mg --- SI/Metric (kilogram, gram, & milligram) Reestablished MPE for the following measurement units (restored from E617 revision 1981): ◦ ◦ ◦ ◦ ◦ ◦ Avoirdupois (pounds, ounces, & fractions) y ounces Troy Pennyweights Grains C t Carats Apothecary (Ounces, Drams, Scruples) X X X X x 500 µg, 300 µg, 200 µg, 100 µg, & 50 µg S Specifications ifi ti MPE for portions of ASTM Classes 000 - 6 Manufacturer Specifications: ◦ ◦ ◦ ◦ ◦ ◦ ◦ ◦ ◦ Construction C i Design Surface Area Magnetism* Density Finish Finish* Adjustment* Marking Cases *Considerations in the subsequent recalibrations of the weight. weight Laboratories issuing calibration reports for weights and weight sets shall have evidence of metrological traceability to the International System of Units (SI). (SI) Calibration certificates shall be issued only by laboratories having g a quality q y system y complying with the requirements of ISO/IEC 17025, which has preferably been verified by thi d party t assessmentt ((accreditation). dit ti ) third Conventional mass of each weight Indication if adjusted Expanded uncertainty Value of the coverage factor, k Meet the requirements of ISO/IEC 17025, Section 5.10 5 10 These represent the minimum requirements for a Certificate of Calibration for an ASTM E617 class weight. Weight Class Alcohol Al h l Distilled H2O 000, 00, & 0 1 2 3-7 7-10 7 10 days d 3-6 3 6 days d 1-2 1 2 days d 1 hour h 4-6 days 2-3 days 1 day 1 hour Weight Class Range ± Per Hour ± Per 12 Hours 000, 000 00, 00 & 0 17 to t 23.5 23 5 ºC 0.3 0 3 ºC 0.5 0 5 ºC 1 17 to 23.5 ºC 0.7 ºC 1 ºC 2 1.5 ºC 2 ºC 3 2 ºC 3.5 ºC 4-7 3 ºC 5 ºC ASTM E617 (2013) T Table bl 11 If the air density deviates from 1.2 kg/m3 by more than 10 % %, (true) mass values shall be used in calculations and the conventional mass shall be calculated from the (true) mass. Two methods for the determination of the conventional mass of weights: ◦ Direct comparison ◦ Subdivision/multiplication (commonly known as dissemination) Class 000, 00, 0 1 2&3 4&5 6&7 ABBA 3 2 1 1 1 ABA 5 3 2 1 1 AB….BA 5 3 2 1 1 ASTM E617 (2013) Table 12 •ABBA & ABA appropriate for weight classes 000, 000 00, 00 0 0, 1 1, 2 2, & 3 •AB…BA may be used for weight classes 4, 5, 6, and & 7 •AB…BA may be used for weight classes 000, 00, 0, 1, 2, & 3 if a mass comparator with an automatic weight exchange mechanism is used and if the system is installed in a protecting housing. Based on the GUM & supplements Based on repeatability of the intercomparisons Based on historical data (check standards and controll charts) h ) NIST Mass Code software being updated to meet ASTM E617 & OIML R111 uncertainty requirements and to better apply air buoyancy corrections to measurement results Standard uncertainty of the weighing process Uncertainty off the standard U i h reference f d d ◦ Covariance considerations must be applied ◦ Drift & instability Uncertainty of the air buoyancy correction (ABC) ◦ ABC equation and/or estimations y of the Reference Standard & Unknown ◦ Density (measured or assumed) ◦ Environmental measuring equipment Uncertainty comparator y of the balance/mass / p ◦ Sensitivity ◦ Display Resolution (zero & load) y ◦ Eccentricity HB 105-1 (DRAFT) OIML R111 ASTM E617 Mandatory & M d t prescriptive for components & calculations,, including g air density (based on OIML R111 recommendations) Mandatory & M d t prescriptive for components & calculations,, including g air density. Mandatory & M d t prescriptive for components & calculations,, including g air density (based on OIML R111 recommendations) (2013) Required: U ≤ 1/3 MPE Required: U ≤ 1/3 MPE Required: U ≤ 1/3 MPE Conventional mass and ABC on all lib ti calibrations exceptt M class. Conventional mass ABC Air ABC: Ai Buoyancy B C Correction i Conventional mass and ABC Dr. George Rodrigues (Artel), ASTM E41 Committee Chair Tracey Hill (Troemner), ASTM E617 Secretary NIST (M (Mass & F Force G Group, Offi Office off Weights W i h & Measures, Adminstration) Dr Richard Davis, Dr. Davis BIPM Errata update to E617 (2013) ASTM E898 ASTM E1154 NIST Seminars & Tutorials ◦ F Fundamentals d t l off M Metrology t l S Seminars i (1 week) k) ◦ Mass Seminars (2 weeks) weeks)) ◦ Advanced Mass Seminars (1-2 ( Mark Ruefenacht Heusser Neweigh 925 798 8900 925-798-8900 mark@neweigh.com