MACO Breeze IIc User Screens Manual
Transcription
MACO Breeze IIc User Screens Manual
MACO Breeze II C User Screen Manual Table of Contents Contents 1. Introduction ................................................................................................. 6 1.1 Configurations ........................................................................................... 7 1.2 Main Menu ................................................................................................ 7 1.3 Recipe Storage ......................................................................................... 7 1.4 Parison Profile ........................................................................................... 8 1.5 Machine Monitor ........................................................................................ 8 1.6 Statisical Process Control (SPC) ............................................................... 8 2. Powerup ........................................................................................................ 9 2.1 Security........................................................................................................ 9 3. Recipe Transfer......................................................................................... 11 4. The Parison Profile Screen ........................................................................ 13 4.1 Setpoints ................................................................................................. 13 4.1.1 Profile Mode ......................................................................................... 13 4.1.2 Profile Graph Limits .............................................................................. 14 4.1.3 Profile Number ..................................................................................... 14 4.1.4 Weight (Head 1/Head 2) ....................................................................... 14 4.2 Values ..................................................................................................... 14 4.2.1 Current Point Size ................................................................................ 14 4.2.2 Active Cursor Position .......................................................................... 14 4.3 Commands.............................................................................................. 14 4.3.1 Read Profile ......................................................................................... 14 4.3.2 Send to Parison .................................................................................... 16 4.3.3 Toggle Lock Pt ..................................................................................... 16 4.3.4 Interpolate ............................................................................................ 16 4.3.5 Toggle Master ...................................................................................... 16 4.3.6 Cursor Mode ........................................................................................ 16 4.3.7 Shift Up (Increment Points) ................................................................... 16 4.3.8 Shift Down (Decrement Points) ............................................................. 16 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 1 of 62 MACO Breeze II C User Screen Manual 4.3.9 Shift Profile Left .................................................................................... 16 4.3.10 Shift Profile Right ................................................................................ 17 4.3.11 Spike Profile ....................................................................................... 17 4.3.12 Clear Profile ....................................................................................... 17 5. Machine Monitors ...................................................................................... 18 5.1.1 Shotsize Time Setpoint ......................................................................... 18 5.1.2 Time Sync Delay Setpoint..................................................................... 19 5.1.3 Profile Delay Time Setpoint .................................................................. 19 5.1.4 Time Correction Limit Setpoint/Time Correction Enable ........................ 19 5.1.5 Actual Position Value............................................................................ 19 5.1.6 Percent Out Value ................................................................................ 19 5.1.7 Head Target Values.............................................................................. 19 5.1.8 Profile Point Values .............................................................................. 19 5.1.9 Cycle Counter Value............................................................................. 20 5.1.10 Sync Time Values............................................................................... 20 5.1.11 Machine Monitor Messages ................................................................ 20 5.2 Position Based Monitor ............................................................................ 22 5.2.1 Shotsize Setpoint ................................................................................. 22 5.2.2 Cushion Setpoint .................................................................................. 23 5.2.3 Actual Position Value............................................................................ 23 5.2.4 Percent Out Value ................................................................................ 23 5.2.5 Target Values ....................................................................................... 23 5.2.6 Accumulator Position Value .................................................................. 23 5.2.7 Cycle Time ........................................................................................... 23 5.2.8 Total Cycles ......................................................................................... 23 5.2.9 Machine Monitor Messages .................................................................. 23 5.3 Accumulator (A1) Calibration (Position Based only) ................................. 24 6. Parison Setup ........................................................................................... 27 6.1 Time Based Parison Setup ...................................................................... 27 6.1.1 Control Span and Span 1-4 .................................................................. 28 6.2 Gain & Reset ........................................................................................... 28 6.3 Loop Config............................................................................................. 29 6.4 Position Based Parison Setup ................................................................. 30 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 2 of 62 MACO Breeze II C User Screen Manual 7. Calibration................................................................................................. 31 7.1 Selection of Head Tooling Type (converging or diverging)........................ 31 7.2 Loops 1 – 4 Config. Setpoints .................................................................. 32 7.3 Open Loop Automatic Head Calibration ................................................... 32 7.4 Accumulator (A1) Calibration (Position Based only) ................................. 32 8. Output Linearization Tables ....................................................................... 33 9. Option Setup Time Based ........................................................................... 35 9.1 Head Tooling Setup.................................................................................... 35 9.2 Time Correction.......................................................................................... 36 9.3 Comparator Setup ...................................................................................... 36 10. Option Setup 1- 2 Head Position Based, 1 Accumulator.............................. 37 10.1 Head Tooling Setup.................................................................................. 37 10.2 Parison Shot Start Type ........................................................................... 37 11. Option Setup 1- 2 Head Position Based, 2 Accumulator.............................. 38 11.1 Accumulator Selection .............................................................................. 38 11.2 Comparator Setup .................................................................................... 39 12. Option Setup 1- 4 Head Position Based, 1 Accumulator.............................. 40 12. 1 Head Tooling Selection ........................................................................... 40 13. Line graphs ................................................................................................ 42 14. Security Change......................................................................................... 45 14.1 Security Change Screen ........................................................................ 45 14.2 Setpoints ............................................................................................... 46 14.2.1 Password Change Pointer .................................................................. 46 14.2.2 New Password Change ...................................................................... 46 14.3 14.3.1 14.4 14.4.1 14.5 Values ................................................................................................... 46 Security Level Password Pointer......................................................... 46 System Commands ............................................................................... 46 Accept New Password ........................................................................ 46 New Module ASCII Strings .................................................................... 47 14.5.1 Current Password ............................................................................... 47 14.5.2 New Password ................................................................................... 47 14.6 14.6.1 Module Displayable Messages .............................................................. 47 Password Index Level 1-4................................................................... 47 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 3 of 62 MACO Breeze II C User Screen Manual 14.6.2 15 15.1 16 Password Index-complete................................................................... 47 SPC Introduction ...................................................................................... 49 Definitions ............................................................................................. 49 SPC Setup Setpoints ................................................................................ 50 16.1 Trig ....................................................................................................... 50 16.2 Sample Size .......................................................................................... 51 16.3 Sample Time ......................................................................................... 51 16.4 Group Period ......................................................................................... 51 16.5 LSL Spec limit ....................................................................................... 51 16.6 USL Spec limit ....................................................................................... 52 16.7 Recalc Limits ......................................................................................... 52 17 17.1 SPC Screens............................................................................................ 53 SPC X-BAR Values ............................................................................... 53 17.2 SPC Range Values .................................................................................. 54 17.3 SPC Graphs ............................................................................................ 55 17.4 SPC Table Data ...................................................................................... 56 17.4 SPC Distribution Curve............................................................................ 57 18 Parison Head 100 Setpoint Screens ......................................................... 58 19 Parison Head 100 Setpoint Screens .......................................................... 60 19.1 Flash Reload Count / Scroll Rates .......................................................... 60 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 4 of 62 MACO Breeze II C User Screen Manual Figures Figure 1 Security & Menu ........................................................................................................................ 9 Figure 2 Recipe Manager ...................................................................................................................... 11 Figure 3 Parison Profile Screen (Time Based)........................................................................................ 13 Figure 5 Interpolation Modes ................................................................................................................. 15 Figure 6 Machine Monitor (2 Head Time Based) .................................................................................... 18 Figure 7 Machine Monitor (4 Head Time Based) .................................................................................... 21 Figure 8 Machine Monitor (2 Head Position Based)................................................................................ 22 Figure 9 Machine Monitor (2 Head Position Based 2 Accum.) ................................................................ 25 Figure 10 Machine Monitor (4 Head Position Based) .............................................................................. 26 Figure 11 Profiler & StateSetup (Time Based) ........................................................................................ 27 Figure 12 Parison Setup (Position Based 2 Head 2 Accum.) .................................................................. 30 Figure 13 Tuning & Calibration .............................................................................................................. 31 Figure 14 Output Linearization ............................................................................................................... 33 Figure 15 Linearization example ............................................................................................................ 34 Figure 16 Option Setup (Time Based) .................................................................................................... 35 Figure 17 Option Setup (Position Based 1 Accum.) ................................................................................ 37 Figure 18 Option Setup (Position Based 2 Accum.) ................................................................................ 38 Figure 19 Option Setup (Position Based 4 Head) ................................................................................... 40 Figure 20 Parison Comp Help ................................................................................................................ 41 Figure 21 Line Graphs ........................................................................................................................... 42 Figure 22 Line Graph Help..................................................................................................................... 44 Figure 23 Security Change .................................................................................................................... 45 Figure 24 SPC Setup ............................................................................................................................. 50 Figure 25 Group Period diagram ............................................................................................................ 51 Figure 26 SPC X-Bar Data..................................................................................................................... 53 Figure 27 SPC Range Data ................................................................................................................... 54 Figure 28 SPC Graphs .......................................................................................................................... 55 Figure 29 SPC Table Data ..................................................................................................................... 56 Figure 30 SPC Distribution Curve .......................................................................................................... 57 Figure 31 Head 1 Setpoints ................................................................................................................... 58 Figure 32 Head 2 Setpoints ................................................................................................................... 59 Figure 33 Display Config........................................................................................................................ 60 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 5 of 62 MACO Breeze II C User Screen Manual 1. Introduction The Breeze IIc parison-only controller is specifically designed for simple installation and ease of use. It can be retrofitted on a wide variety of blow molding machines to provide consistent wall thickness of blow-molded products for improved quality, less material consumption and reduced scrap. It can also be specified on new OEM blow molding machinery. The Breeze IIc is factory configured for most blow molding applications. Specify the machine type and the Breeze IIc will be factory programmed and wired for that machine. Preconfigured applications are: Continuous Extrusion, 1 Head Continuous Extrusion, 2 Heads Continuous Extrusion, 3 Heads Continuous Extrusion, 4 Heads Position Based 1 Accumulator, 1 Head Solenoid Valve Position Based 1 Accumulator, 2 Heads Solenoid Valve Position Based 2 Accumulators, 2 Heads Solenoid Valve Position Based 1 Accumulator, 3 Heads Solenoid Valve Position Based 1 Accumulator, 4 Heads Solenoid Valve Wheel (Encoder), 1 Head, Incremental or Quadrature Encoders Wheel (Encoder), 2 Heads, Incremental or Quadrature Encoders The all-in-one package comes standard with a 120 VAC or 230 VAC Input 24 VDC, 5 amp output and a 4 Channel Servo Amplifier card. External machine wiring is connected to connectors on the back of the unit. After connecting the cables, the installer follows the labels diagram shipped with the Breeze IIc to connect wires to the proper points on the machine. Once the unit is installed, simply calibrate the head/accumulator and enter the desired parison profile. The 100 point parison profile is displayed graphically on a large color display. This enables the user to quickly modify the profile and compare it to the actual process value. The entered profiles can then be stored either internally on Compact Flash or in the internal NAND directory. It can also be saved externally to USB cartridges. The Breeze IIc uses the same cards as Barber-Colman’s MACO ® Compact total machine controller. This allows users to stock fewer spare parts. It also provides users with highly precise parison control. The Breeze IIc updates parison tooling position every 0.1 milliseconds to minimize material usage and improve part quality. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 6 of 62 MACO Breeze II C User Screen Manual 1.1 Configurations Based on the model number ordered, the Breeze IIc is shipped with either a time based or position based (accumulator or reciprocating screw) or wheel (encoder) configuration. Each configuration requires both a different program downloaded into the Breeze and different internal/external wiring. All configurations share the same internal cards and system screens. Model Number MCJx-11x21-1C0-0-L0 MCJx-12x21-1C0-0-L0 MCJx-17x21-1C0-0-M0 MCJx-18x21-1C0-0-M0 MCJx-14x21-1C0-0-L0 MCJx-15x21-1C0-0-L0 MCJx-16x21-1C0-A-L0 MCJx-1?x21-1C0-A-M0 MCJx-1?x21-1C0-0-L0 MCJx-1?x21-1CH-0-L0 1.2 Description Continuous Extrusion, 1 Head Continuous Extrusion, 2 Heads Continuous Extrusion, 3 Heads Continuous Extrusion, 4 Heads Position Based 1 Accumulator, 1 Head Position Based 1 Accumulator, 2 Heads Position Based 2 Accumulator, 2 Heads Position Based 1 Accumulator, 3-4 Heads Wheel Based, 1-2 Heads Incremental Encoder Wheel Based, 1-2 Heads Quadrature Encoder Main Menu The power up security code allows the Breeze IIc to be operated with only 4 screens. The main menu screen allows the user to choose one of four security levels. Setup personnel can be given security codes that allow complete access to setup information as well as to the normal operating screens. Operators can be given security codes that provide access only to the operating screens. The user can move through the screens either by pressing the path keys on the bottom of each screen or by pressing the left most path key to return to the main menu and then selecting the desired screen from the main menu. Only screens that are at or below the entered security code will be accessible. The screens are grouped according to the level of access: Level 1 Level 2 Level 3 Level 4 1.3 Normal Operation without SPC access Normal Operation with SPC access Field Configuration and Tuning Factory Configuration User User Setup personnel Factory personnel Recipe Storage Recipe storage can reduce startup time and allow easy job transfer between different machines. All entered setpoints can be saved to either internal (NAND) memory or to USB memory sticks. Approximately 200 different recipes can be saved internally or to a single usb memory stick. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 7 of 62 MACO Breeze II C User Screen Manual 1.4 Parison Profile The parison profile screen is used to enter and view the parison profile. The entire profile can be viewed on this screen. The user can enter any number of master setpoints (two or more), and select any one of five modes of interpolation (flat, linear, and three types of parabolic) to automatically interpolate the remainder of the 100 profile points. To ease the entry of new parison profile you can now specify a desired number of master points and auto populate the master points. The generation of the profile setpoints and closed loop update of the analog output takes place at a 1 millisecond rate for very precise control of the tooling. Separate profiles can be entered for Head 1, 2, 3, and 4. The parison process value can be viewed on the parison profile screen. In multiple head applications one profile can be sent to all heads as a base starting position. Then each profile for up to 4 heads can then be adjusted individually for each mold. An orientation mark can be placed on a part, with the parison “skip” point feature. Specify any one of the 100 parison points as a “lock” point and interpolation will not change the point. The time based profile screen is identical to the position based screen except that several of the die-gap setpoints have been removed. If the operator requires that more than one profile point be “skipped” then the spike profile feature can be used to spike a group of profile points above and below the active cursor. 1.5 Machine Monitor The machine monitor screen provides an overview of the entire process. Graphical sprites move with either one, two, three, or four heads and a bar graph moves with the pushout motion. Both cycle time and total number of cycles are tracked. A timer for Pushout and Fill are used to abort auto cycle if cushion or shot-size are not reached before the timer expires. The time based machine monitor screen functions identically to the position based screen except that the setpoints and timers have been changed to reflect time based operation. 1.6 Statisical Process Control (SPC) SPC provides for the simultaneous calculation of different process parameters. Xbar, R and histogram charts can be displayed for each of the values. Calculated X-bar, R, Cr, and CpK values are also available. The last 100 calculated values for each of the parameters remain in memory and are available for the operator to view and print on demand or automatically after 100 points have been collected. SPC alarms based on industry accepted standards are available to make machine decisions based on part quality. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 8 of 62 MACO Breeze II C User Screen Manual Figure 1 Security & Menu 2. Powerup As delivered, the controller will power up displaying a screen similar to Figure 1 (only security level 1 screens will be displayed). 2.1 Security The setpoint entry area for the security code is located on the upper right of the Header in the Security & Main Menu screen (in the user screen set). If it is necessary to change the security level to go to one of those screens, use the arrow keys to select the security setpoint entry area, and type in the code for the security level desired (see appendixes). Observe the lower right of the screen. An asterisk will appear for each character entered. Use the plus/minus key for any dashes contained in the security code. Pressing Enter will cause the security to change to the level indicated. Note that it will be necessary to refresh the Main Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 9 of 62 MACO Breeze II C User Screen Manual Menu screen in order to display the available paths for the newly entered security level (use the up and down page keys). Level 1 is the lowest security level. Operating at Level 1 allows the least access to the controller. Level 4 is the highest user security level. Operating at Level 4 allows the most access to the controller. Only screens with a security level equal to or less than the operating security level will be accessible. If operating at Security Level 1, only Security Level 1 screens (or Paths) appear. If at Security Level 2, only Security Level 1 & 2 screens (or Paths) appear. If at Security Level 3, only Security Level 1, 2 & 3 screens (or Paths) appear. If at Security Level 4, Security Level 1, 2, 3 & 4 screens (or Paths) appear. Note that once a screen is displayed, only those setpoints can be changed which have an assigned security level less than or equal to the operating security level. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 10 of 62 MACO Breeze II C User Screen Manual Figure 2 Recipe Manager 3. Recipe Transfer The Powerup recipe is a special recipe located in recipe location ”0”. As the name implies, this recipe is loaded into the system upon system power up. Any recipe can be stored as the power up recipe. All other recipes are stored in the locations indicated, 10 recipes per page. (Recipes 1-10 are showing in this example.). A unique alpha-numeric name, up to a maximum of 29 characters in length, can be entered for each recipe. All setpoints within the system are stored or loaded to and from a recipe. To save a recipe, enter an alpha-numeric name using the keyboard, enter the desired recipe location setpoint and press “Save Recipe”. To load a recipe, enter the desired recipe location setpoint and press “Load Recipe”. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 11 of 62 MACO Breeze II C User Screen Manual To delete a recipe, enter the desired recipe location setpoint and press “Delete Recipe”. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 12 of 62 MACO Breeze II C User Screen Manual Figure 3 Parison Profile Screen (Time Based) 4. The Parison Profile Screen The parison profile screen for a time based system is shown in figure 3. The screen will vary slightly depending on whether you have a time or position based system. If changes are made to setpoints on this screen, be sure to do a "save recipe" (on the Recipe Storage screen). 4.1 Setpoints 4.1.1 Profile Mode This setpoint determines how the profile will be interpolated (refer to Figure 5). Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 13 of 62 MACO Breeze II C User Screen Manual 4.1.2 Profile Graph Limits These setpoints are shown beneath the profile graph on the left and right sides. Typical entries are 00.00 (on the far left) to 100.00 (on the far right) for full display. Note that these setpoints require an entry (and save) on initial powerup. 4.1.3 Profile Number This setpoint determines which profile the system displays. The range is 1, 2, 3 or 4. 4.1.4 Weight (Head 1/Head 2) This setpoint determines the minimum weight of all parison profile points. Weight is an offset to the profile. The means that the actual Target = Weight + Profile point. 4.2 Values 4.2.1 Current Point Size This is the value of the profile point for the position (1-100) that the active cursor is on. 4.2.2 Active Cursor Position This value shows the position of the active cursor on the graph. To move the active cursor, press the Up or Down Arrows that are above and below the Yellow Cursor Key near the middle of the Parsion Profile screen. The cursor will then move either “Point to Point” or “Master to Master” depending on which cursor mode you are in. 4.3 Commands 4.3.1 Read Profile This command causes the profile (as selected by the Profile Number Setpoint) to be read from the module. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 14 of 62 MACO Breeze II C User Screen Manual Figure 4 Interpolation Modes Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 15 of 62 MACO Breeze II C User Screen Manual 4.3.2 Send to Parison This command sends the displayed profile (as selected by the Profile Number setpoint) to the parison module. 4.3.3 Toggle Lock Pt Turns lock points on or off. A lock point is one which is not included in interpolation. A lock point can be inserted by moving the active cursor to the desired point on the profile and pressing "Toggle Lock Pt." Pressing "Toggle Lock Pt" a second time will remove the lock point. 4.3.4 Interpolate This command causes the profile to be interpolated in the selected mode. 4.3.5 Toggle Master This command allows the insertion or removal of master points (1 to 100 master points are available) at the position of the active cursor. A master point can be inserted by moving the active cursor to the desired point on the profile and pressing "Toggle Master." Pressing "Toggle Master" a second time removes the point. 4.3.6 Cursor Mode This changes how the cursor moves. It can either be “Point to Point” which moves the cursor 1 profile point at a time or “Master to Master” in which the cursor moves just between Master points. Master to Master is a faster way to adjust the profile. 4.3.7 Shift Up (Increment Points) Increases all profile points by the Step Size setpoint. 4.3.8 Shift Down (Decrement Points) Decreases all profile points by the Step Size setpoint. 4.3.9 Shift Profile Left Moves the profile toward the origin (point 1). Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 16 of 62 MACO Breeze II C User Screen Manual 4.3.10 Shift Profile Right Moves the profile away from the origin (point 1). 4.3.11 Spike Profile This is a feature that is similar to Skip (Lock) Points that allows a marking of the profile either above or below the active cursor by a selected amount for a given time period. Then after the time expires, the “mark” is automatically removed. So, what is needed to make this work is an amount (of Spike) entered, the number of points to Spike above and below the active cursor, and a Spike Time. Then you can energize this command and the Profile will Spike. NOTE: Remember to set the Spike timer longer than your Cycle Time. Otherwise you may never see the Spike appear in the bottle. 4.3.12 Clear Profile Clears the presently displayed profile from the screen. If the profile had been sent to the module, it is not lost (perform a "read profile to recall it). Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 17 of 62 MACO Breeze II C User Screen Manual Figure 5 Machine Monitor (2 Head Time Based) 5. Machine Monitors The machine monitor screens will vary slightly depending on whether a time based or position based system has been ordered Time Based Monitor Refer to Figure 6. The time based machine monitor screen displays monitor information for two die heads. If only one is used, it should be Head 1. If any changes are made to the setpoints on this screen, be sure to do a "save recipe" (on the Recipe Storage screen). 5.1.1 Shotsize Time Setpoint This setpoint determines the total time over which profiling will be performed. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 18 of 62 MACO Breeze II C User Screen Manual 5.1.2 Time Sync Delay Setpoint This setpoint determines the time at which profiling is to begin after receiving a sync pulse (from DC Input 1). 5.1.3 Profile Delay Time Setpoint This setpoint determines the amount of time that the profiled value is held at the segment 1 target. This time is subtracted from the shotsize time setpoint. 5.1.4 Time Correction Limit Setpoint/Time Correction Enable This setpoint imposes a limit on the amount of correction that can be applied to shotsize time after a machine cycle. It is used only in time based profiling and both Time Correction and Alternating Time Correction (if desired) must be enabled on the “Option Setup Screen”. “Time correction” is used in application in which there is variation in the Sync to Sync time. Time Correction examines the variation and adjust the shotsize time to balance the times from shot to shot. If “Time Correction” is enabled a message of “T” will appear next to the “Correction Limit” setpoint. Alternating Time Correction is typically used for Two-Sided shuttle continuous extrusion machines and attempts to balance the variation in Sync Times from Side 1 to Side 2 of the machine. If “Alternating Time Correction” is enabled a message of “A” will appear next to the “Correction Limit” setpoint. 5.1.5 Actual Position Value The head position values (Head 1 and Head 2) are the current process value for each respective loop. 5.1.6 Percent Out Value The head percent out values (Head 1 and Head 2) are the current output demand value for each respective loop. 5.1.7 Head Target Values The head target values (Head 1 and Head 2) are the current target value for each respective loop. 5.1.8 Profile Point Values The profile points values represent what profile point is being generate when in “Run Mode.” Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 19 of 62 MACO Breeze II C User Screen Manual 5.1.9 Cycle Counter Value In the Run mode, the total cycle value increments by one each time the profile segment reaches profile point 1. setpoint (on the Parison Setup screen). 5.1.10 Sync Time Values The sync time values are the last 5 sync-to-sync time values. 5.1.11 Machine Monitor Messages Run: OFF Machine function key 1 is toggled Off (the LED will be off) and the Breeze is in the idle mode. The diegap opening is determined by the respective Head 1/Head 2 idle diegap setpoint. Run: ON Machine function key 1 is toggled On (the LED will be on) and the Breeze is in the profile mode. Sensor out of Range the "Input V" is more than 0.500 volts outside of the calibration range. On a time based machine, the outputs are disabled. To recover, follow this procedure: Verify the wire connections and sensor integrity. Enter the calibration mode to regain loop outputs and verify calibration. Adjust gain and reset for quick response with minimal overshoot of zero and span (view "Input V" and compare it to Low V/High V to determine overshoot/ undershoot at the calibration voltage). Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 20 of 62 MACO Breeze II C User Screen Manual Figure 6 Machine Monitor (4 Head Time Based) Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 21 of 62 MACO Breeze II C User Screen Manual Figure 7 Machine Monitor (2 Head Position Based) 5.2 Position Based Monitor Note: The position based machine monitor screen displays monitor information for two die heads. If only one is used, it should be Head 1. If any changes are made to the setpoints on this screen, be sure to do a "Save Setpoint" . 5.2.1 Shotsize Setpoint This setpoint determines the shotsize (in engineering units) and is based on the Loop 5 Accum. 1 Range setpoint (0 to 100) on the Tuning/Calibrate screen (Figure 11). If other than 0-100% engineering units are desired, change the Loop 5 Accum 1 Range setpoint. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 22 of 62 MACO Breeze II C User Screen Manual 5.2.2 Cushion Setpoint This setpoint determines the cushion or accumulator empty (in engineering units) and is based on the Loop 5 Accum 1 Range setpoint (0 to 100) on the Tuning/Calibrate screen (Figure 11). If other than 0-100% engineering units are desired, change the Loop 5 Accum 1 Range setpoint. 5.2.3 Actual Position Value The head position values (Heads 1-4) are the current process value for each respective loop. 5.2.4 Percent Out Value The head percent out values (Heads 1-4) are the current percent output value for each respective loop. 5.2.5 Target Values The head target values (Heads 1-4) are the current target value for each respective loop. 5.2.6 Accumulator Position Value The accumulator position value is the accumulator position based on the Loop 5 Accum 1 Range setpoint (0-100) on the Tuning/Calibrate 5.2.7 Cycle Time The cycle time value is the elapsed time between Run Profile signals. The Run Profile signal is generated by either the accumulator reaching shotsize or the accumulator reaching shotsize in conjunction with Logic Input 1 going high ( Shoot ) 5.2.8 Total Cycles The total cycles value is the total number of Run Profile signals during the Auto (Run) mode. 5.2.9 Machine Monitor Messages Run: Off (Manual Mode) Machine function key 1 ("Run") is toggled Off and the Breeze IIc is in the manual mode. Manual “Fill” and manual “Purge” momentary screen buttons are functional. Note that the Run mode being off is required in order to perform calibration of the heads. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 23 of 62 MACO Breeze II C User Screen Manual Run: On (Auto Mode) Machine function key 1 ("Run") is toggled On (the LED will be on) and the Breeze is in the automatic mode. Run Mode is required in order to recognize the initiate shot (DC In 1). Shoot: The accumulator is moving toward cushion. Fill: The fill delay timer has expired. The fill speed DC Out 1 are energized. Timeout: Auto cycle has aborted due to shot or fill timeout (investigate the physical condition of the machine and recover via manual [Run = Off] mode). Sensor out of Range The "Input V" is more than 0.500 volts outside of the calibration range. On a position based machine, the outputs are disabled and the cycle in progress is aborted. To recover, follow this procedure: Verify the wire connections and sensor integrity. Enter the calibration mode to regain loop outputs and verify calibration. Adjust gain and reset for quick response with minimal overshoot of zero and span (view "Input V" and compare it to Low V/High V to determine overshoot/ undershoot at the calibration voltage).be 10 or 15% for the “OL” setpoints and 10.00 or 15.00 seconds for the calibrate timer. First, make sure that “Run” is turned off so that calibration can take placed. Second, make sure that the proper tooling is selected, either converging or diverging. Third, Select the Input to calibrate. Fourth, Select Zero. The calibration timer should time down to Zero and then Calibration Enable should turn On for a few seconds and then turn off. After turning off you should see the Lw Volt number automatically change to what the voltage was at Zero. Fifth, repeat the above procedure for Span. Repeat for any Input calibration that is desired. Last, remember to “Save Setpoint” on button 9. 5.3 Accumulator (A1) Calibration (Position Based only) Note that the accumulator sensor must be wired so that the sensor signal DECREASES as the accumulator moves from shotsize to cushion. Move the accumulator by hand to zero and span and note the voltage at each position. Enter the lesser of the two voltages (watch "Input V") as the A1 "Low V" setpoint and the greater of the two as the A1 "High V" setpoint. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 24 of 62 MACO Breeze II C User Screen Manual Figure 8 Machine Monitor (2 Head Position Based 2 Accum.) Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 25 of 62 MACO Breeze II C User Screen Manual Figure 9 Machine Monitor (4 Head Position Based) Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 26 of 62 MACO Breeze II C User Screen Manual Figure 10 Profiler & StateSetup (Time Based) 6. Parison Setup The parison setup screen will vary slightly depending on whether a time based or position based system has been ordered (see Figures 2 and 3). Note: The 24 vdc Input #3 on this unit must be wired to a contact that denotes that the machine has its hydraulic pumps running and that an Emergency Stop is not active so that the machine is ready to safely operate. If Input #3 is wired correctly a “Ready” message will appear indicating that you are ready for safe Breeze IIC automatic operation. 6.1 Time Based Parison Setup Refer to Figure 11. Profiler Setup, Allocation State Setup, Tuning, Control Span, and Span (analog output) must be left as set at the factory. The tuning settings Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 27 of 62 MACO Breeze II C User Screen Manual for gain and reset may require changing. If changes are made, be sure to do a "Save Setpoints” using the button 9 on the right of the screen. 6.1.1 Control Span and Span 1-4 These are locked at 100.00% from the factory and determine the control scale and output span. In this case both are set to 100.00% 6.2 Gain & Reset After calibration is complete, gain and reset can be adjusted (slowly) in order to optimize performance. Good starting numbers are generally in the range of: Gain = 0.5 and Reset = 5. On the “Machine Monitor” screen with run off you should be able to enter tooling target numbers into the idle setpoint and observe how the tooling responds when moving from entries of 10% to 90%. If the tooling responds quickly and is able to hold position at target with little error, you are probably close enough to start running in automatic by turning run on and putting the machine in automatic. Final tuning adjustments can be made using the Linegraph screen to see how closely the actual value follows the target for the parison tooling. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 28 of 62 MACO Breeze II C User Screen Manual 6.3 Loop Config The loop config setpoints determine the action or function of the loops inputs and outputs. See the table and example below. Example of Parison Loop Configuration If Converging = 9, which is 1 + 8: Repeats/Min + Bipolar Output; then diverging should be 1 + 2 + 8 + 16 + 16384 = 16411 which is: +1 = +2 = +8 = +16 = +16384 = Repeats/Min Invert Output Bipolar Swap Zero/Span Use Tune Set 1 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 29 of 62 MACO Breeze II C User Screen Manual Figure 11 Parison Setup (Position Based 2 Head 2 Accum.) 6.4 Position Based Parison Setup Refer to Figure 12. Profiler Setup, Allocation State Setup, Tuning, Control Span, and Span (analog output) must be left as set at the factory. The tuning settings for gain and reset may require changing. If changes are made, be sure to do a "Save Setpoints” using the button 9 on the right of the screen. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 30 of 62 MACO Breeze II C User Screen Manual Figure 12 Tuning & Calibration 7. Calibration The parison tuning and calibration screen will vary slightly depending on whether a time based or position based system has been ordered (see Figure 4 - note that time based systems do not have accumulator entries). "Lw Volt" and "HI Volt" represents the measured voltages at either 0 or 100% of diegap, depending on whether the head is converging or diverging. The sensor should be wired so that for CONVERGING tooling voltage ("Input V" above) DECREASES as the head approaches 0% diegap. Conversely, wire the sensor so that for DIVERGING tooling voltage INCREASES as the head approaches 0% diegap. 7.1 Selection of Head Tooling Type (converging or diverging) Touching the Select button is used to toggle between converging and diverging (Note: The message to the right of the “Select” button will toggle between Converge or Diverge). All heads will be of the same head tooling type. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 31 of 62 MACO Breeze II C User Screen Manual Remember to press the “Save Setpoints” button on the middle right of the screen to save the Converging or Diverging Selection. 7.2 Loops 1 – 4 Config. Setpoints Converging Head A converging head Loop Config 1 setpoint will be either "9" or "11." Entering the wrong number will result in a process readout 180 degrees out of phase with the actual head position. Entries of “11” invert the output versus entries of “9” Normally, 9 (converge) is used in conjunction with 16411 (diverge) else it would be 11 (converge) or 16409 (diverge). Diverging Head A diverging head Loop Config 2 setpoint will be either "16411" or "16409." Entering the wrong number will result in a process readout 180 degrees out of phase with the actual head position. Entries of “16409” invert the output versus entries of “16411”. Normally, 9 (converge) is used in conjunction with 16411 (diverge) else it would be 11 (converge) or 16409 (diverge). NOTE: See Table 1 for a full definition of the Loop Config entry. 7.3 Open Loop Automatic Head Calibration The calibration procedure is Open Loop in its nature for Analog Inputs 1 – 4 which are wired to the “LVDT’s” for head tooling 1 – 4. Thus, entries for “OpenOL” and “CloseOL” are required to get output to move the head tooling. These are out of 100% of output span. Thus, an entry of 10% would equal 1 DC volt outputted to the servo valve. Also, there needs to be a non-zero entry present for the calibrate time setpoint. This setpoint should be set long enough to allow time for the head tooling to move full stroke. Good starting numbers would be 10 or 15% for the “OL” setpoints and 10.00 or 15.00 seconds for the calibrate timer. First, make sure that “Run” is turned off so that calibration can take placed. Second, make sure that the proper tooling is selected, either converging or diverging. Third, Select the Input to calibrate. Fourth, Select Zero. The calibration timer should time down to Zero and then Calibration Enable should turn On for a few seconds and then turn off. After turning off you should see the Lw Volt number automatically change to what the voltage was at Zero. Fifth, repeat the above procedure for Span. Repeat for any Input calibration that is desired. Last, remember to “Save Setpoint” on button 9. 7.4 Accumulator (A1) Calibration (Position Based only) Note that the accumulator sensor must be wired so that the sensor signal DECREASES as the accumulator moves from shotsize to cushion. Move the accumulator by hand to zero and span and note the voltage at each position. Enter the lesser of the two voltages (watch "Input V") as the A1 "Low V" setpoint and the greater of the two as the A1 "High V" setpoint. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 32 of 62 MACO Breeze II C User Screen Manual Figure 13 Output Linearization 8. Output Linearization Tables Note that these setpoints MUST be entered BEFORE pumps are turned ON. Note also that before adding any segments to these setpoints that pumps MUST be turned OFF. These setpoints determine the output voltage associated with a percent output. Each output can have up to 8 linear regions defined using "% Demand" and "Out Volts". These setpoints allow the use to effectively eliminate and null in the valves. Refer to the table below. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 33 of 62 MACO Breeze II C User Screen Manual % Demand Output Volts Figure 14 Linearization example Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 34 of 62 MACO Breeze II C User Screen Manual Figure 15 Option Setup (Time Based) 9. Option Setup Time Based 9.1 Head Tooling Setup These push buttons determine the head toolings that the system is configured for. The “<”beside the button indicates which selection the system is currently configured for. NOTE that changes made to these selects will need to be “saved” by pressing the Save Setpoints button. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 35 of 62 MACO Breeze II C User Screen Manual 9.2 Time Correction These push buttons turn on and off Time Correction and well as Alternating Time Correction. For Alternating Time Correction to function BOTH push buttons must be set to the ON position. (For an explanation of the function refer to section 5.1.4) 9.3 Comparator Setup These Setpoints configure the Comparators 1 and 2. These Comparators determine the functionality of DC Output 1 and 2 ( Time Based) or DC Output 3 and 4 ( Position Based). Each Comparator requires and IN A, IN B, Type and CompSP setpoints to be entered. Refer to Figure 20 for a list of commonly used Modbus ID numbers to be entered into the IN A Sel and IN B Sel setpoints as well the numbers to be entered into the type setpoint for the different types of comparators. Note that the if either of the IN A Sel or IN B Sel setpoints are set to “19817” or “19818” then the CompSP will be utilized by the comparator and must have a number entered into it that is scaled the same as the value it is being compared to. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 36 of 62 MACO Breeze II C User Screen Manual Figure 16 Option Setup (Position Based 1 Accum.) 10. Option Setup 1- 2 Head Position Based, 1 Accumulator 10.1 Head Tooling Setup Pressing the # of Heads Select button will toggle between 1 head and 2 head configurations. NOTE that any changes should be saved by pressing the Save Setpoints Button. 10.2 Parison Shot Start Type This button Enables and Disables the use of DC Input 1 as an OK to shoot input. When this button is Enabled the system will wait to turn on the Run Profile until the Accumulator has reached Shotsize AND DC Input 1 is turned ON. This can be used to ensure the machine is ready to accept the Parison. If this input is not necessary to the cycle of the machine it can be Disabled and the system will turn on the Run Profile once the accumulator reaches Shotsize. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 37 of 62 MACO Breeze II C User Screen Manual Figure 17 Option Setup (Position Based 2 Accum.) 11. Option Setup 1- 2 Head Position Based, 2 Accumulator 11.1 Accumulator Selection Pressing the # of Accumulators will change the number of Accumulators that the system is configured for. The number of Accumulators required is determined by the physical machine. Note that changes to this selection should be saved by pressing the Saved Setpoints button. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 38 of 62 MACO Breeze II C User Screen Manual 11.2 Comparator Setup In addition to comparator 1 and 2 there are comparators 3 and 4 which configure outputs 3 and 4 on the Option slot B DC I/O installed only when the unit was ordered as a 2 Accumulator type system. These comparators configure the same as those described in section 9.3 (refer to Figure 20) Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 39 of 62 MACO Breeze II C User Screen Manual Figure 18 Option Setup (Position Based 4 Head) 12. Option Setup 1- 4 Head Position Based, 1 Accumulator 12. 1 Head Tooling Selection These push buttons determine the head toolings that the system is configured for. The “<”beside the button indicates which selection the system is currently configured for. NOTE that changes made to these selects will need to be “saved” by pressing the Save Setpoints button Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 40 of 62 MACO Breeze II C User Screen Manual Figure 19 Parison Comp Help Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 41 of 62 MACO Breeze II C User Screen Manual Figure 20 Line Graphs 13. Line graphs Line graphs can be used to display setpoints, values, and output signals. The line graph display can plot 4 separate graphs. Refer to Figures 22 for available variables. Note that the Line Graph Channel 1 assignment can be displayed on the parison profile graph screen (but will not include the weight setting). GRAPHS SETUP Changing this setpoint ( 1- 4 ) will setup the graph to display the graphs for heads 1- 4 respectively. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 42 of 62 MACO Breeze II C User Screen Manual HOLD GRAPHS – This command toggles the display between a Run and a Hold state. In the Run state the display updates with new information on each machine cycle or time period. In the Hold state the display is not updated. The Hold state is typically used to study the displayed graphs. To toggle between the two, use the cursor keys to select the command and then press the Enter key. A message will display the status. SAMPLE – This setpoint indicates the time period for the 100 points displayed on the line graph. Entering a setpoint of 10.0 seconds will cause sampling to occur every 0.1 seconds. Entry range is 0 to 65.0 seconds. PHASE – This setpoint allows selection of either segmenter based sampling or time based sampling. The following setpoints can be entered: 1 = Time Based This mode allows sampling based on a time entry. The sampling starts after the Graph Trigger is energized. In this mode, a Sample Time setpoint entry must be made. 3 = Segmenter Based This mode samples 100 points within the shotsize period (position or time based). The linegraph will update at the end of shotsize. FILL – When Fill is selected the area under the first line graph will be filled in (rather than displayed as a line graph it is displayed as a solid – useful when plotting values against setpoints). DISPLAY MIN/MAX SETPOlNTS – The Minimum and Maximum Display setpoints scale the "Y" axis (vertical) values. To separate data, different offsets can be achieved by using different minimum setpoints. CURSOR – The cursor is indicated by a vertical bar. The current position for the cursor is indicated in the lower right position on the screen. The values of the line graphs at the current cursor position are shown below the line graphs. Cursor position is changed by using the Cursor Right/Cursor Left soft keys. The values for all channels are updated as the cursor is moved. The current cursor position ( 1-100) as well as the time is reflected in the values just below the graph. Pos – is a setpoint (1-100) that will cause the cursor to move to the position Mult – a non zero entry into this setpoint tells the graph to enter into a Multi graph mode. In this mode the display will draw multiple traces for each channel 1 for each trigger up to the value of this setpoint. Once this setpoint is reached it will redraw to graph on the next trigger signal and start over. Each cycle of multi graphs will generate a Multigraph.csv file in the Compact flash that can be used to document the consistency of the profiling. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 43 of 62 MACO Breeze II C User Screen Manual Figure 21 Line Graph Help Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 44 of 62 MACO Breeze II C User Screen Manual Figure 22 Security Change 14. Security Change 14.1 Security Change Screen The security change screen allows passwords to be changed at the operator station (an operation that formerly required screen editing). Up to 40 different passwords per security level can be programmed from this screen. Enter the number of the setpoint to be reviewed or changed. Once the existing password is displayed, enter a new password. To accept the change, cursor to the active area and press return. The passwords will be stored in the display. On power-up (or Display Reset) the display checks to see if new user screens have been downloaded. If new screens have been downloaded, the display will automatically update the local passwords to the passwords used in the new screen set. This update creates 40 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 45 of 62 MACO Breeze II C User Screen Manual password positions per security level. If a screen set had more than 40 per level only the first 40 would be utilized. If there is less than 40 (more likely), then blanks will be loaded in the remaining spaces. This automatic update means that if the local passwords were different from those used in the newly downloaded screen set the local passwords will have to be re-entered. There are several methods of securing the password display and limiting changes to passwords for a given level. This can be accomplished by security level (protection of the screen or individual elements) or by limiting the security pointer range to less than its maximum of 200. For example, if the security pointer range is limited from 0 to 120, then only levels 1, 2 and 3 (not 4 or 5) could be changed. Note that level 5 security does not allow any additional changes beyond level 4, so that setting a range of 0 to 160 can prevent a user from changing an OEM’s highest level of security codes. 14.2 Setpoints 14.2.1 Password Change Pointer The password change pointer setpoint is used to move to a specific position in the password file. 14.2.2 New Password Change The new password change setpoint is the setpoint used to enter a new password. Once the new password is entered here, pressing enter will cause it to appear as the “proposed password.” 14.3 Values 14.3.1 Security Level Password Pointer This value is used to display the security level of the current password change pointer setpoint. The value will correspond to the levels shown in figure 1 above. Password 1-40 are level 1; 41-80 are level 2; 81-120 are level 3; 121-160 are level 4; and 161-200 are level 5. 14.4 System Commands 14.4.1 Accept New Password Once a “proposed password” is displayed, this system command is used to make the “proposed password” the “existing password.” Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 46 of 62 MACO Breeze II C User Screen Manual 14.5 New Module ASCII Strings 14.5.1 Current Password This module ASCII string is used to display the password currently pointed at by the password change pointer setpoint. Note that the content of a module ASCII string cannot be programmed (internal firmware determines the character string that appears within the area). If programming this to a screen, be certain to assign enough space for the entire string of 13 characters. 14.5.2 New Password This module ASCII string is used to display the new (proposed) password. Once the new password has been accepted, it will appear in both the current and new message blocks. Note that the content of a module ASCII string cannot be programmed (internal firmware determines the character string that appears within the area). If programming this to a screen, be certain to assign enough space for the entire string of 13 characters. 14.6 Module Displayable Messages 14.6.1 Password Index Level 1-4 These 4 module displayable messages can be used to display a message (for instance, an operator’s name) with a specific password. In order to use this function, you need to know how many passwords will exist for each security level. Use the Message Editor (in the screen editor program) to create a series of messages relating to each security level. Each series should start with a blank message. For example, say that 3 operators require a Level 2 security password. You will need 4 contiguous locations in the message file (a blank message, Tom, Dick and Jane). Once the messages are created, use the screen editor to lay down the “Password index level 2” message block and point to the blank message as the message source. Whenever the password assigned to position 41 is being used on the machine, the word “Tom” will appear as the message. 14.6.2 Password Index-complete This module displayable message works in exactly the same manner as the “Password Index Level x” messages described above except that this message requires 201 positions in the message file (a blank message followed by a message location for every potential password). This message has been created in order to provide for the original version of the Security Change function. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 47 of 62 MACO Breeze II C User Screen Manual Starting Factory Security Codes Level 1 2 3 4 Code 1 222 5966 123 8118 1397 6800 159753 Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 48 of 62 MACO Breeze II C User Screen Manual 15 SPC Introduction Statistical Process Control (SPC) is a software option to the MACO control system. SPC provides for the simultaneous calculation of 10 different user selected parameters. Any process value from the system can be selected as an SPC parameter. Each of the 10 parameters allows selection of sample size, time or event triggering, time between readings, time between sample groups, upper and lower specification limits. The software will calculate and display each sample group’s X-Bar, R, Cr, Cpk and Cp. The Iast 100 calculated values for each parameter will remain in the SPC buffer memory. SPC X-Bar and R charts, along with the last 100 samples can be printed (via the optional printer port) on demand, event based, or automatically after 100 points have been gathered. SPC can also be performed on setpoints provided for that purpose in the Data Handler modfile (Spare Setpoint #1 thru #8). These setpoints allow the calculation of statistical data on as many as eight parameters which are not directly measured by the system, such as part weights or dimensions. The data is measured by the machine operator and then manually entered as the setpoint. Titles for the parameters as well as the menus for the parameter choices are entered when screen programming is performed. When programming screens, up to thirty machine parameters may be chosen. Any value or setpoint from any controller can be used. Ten of the thirty can then be selected for SPC data collection. Screens are also made available for X-Bar and R chart display and distribution curves. 15.1 Definitions Sample Size: Number of readings in a group (2 to 25) Sample Time: Time between individual samples (1 to 999 seconds) Trigger Mode: Time based or event based through sequence logic Group Period: Time between beginning of sample groups (0 to 65535 seconds) Upper Specification Limit (USL): The highest acceptable limit Lower Specification Limit (LSL): The lowest acceptable limit X-Bar: Average of all samples in the sample group X-BarBar: Average of all X-Bars on the X-Bar chart R: Range: The difference between the largest sample of the subgroup and the smallest sample R-Bar: Average of all the ranges on the Range chart STDV: Standard Deviation UCLR: Upper control limit for Range values LCLR: Lower control limit for Range values UCLX: Upper control limit for X-Bar values LCLX: Lower control limit for X-Bar values Cpk: Process Capability Index Cr: Capability Ratio Cp: Inverse of Cr Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 49 of 62 MACO Breeze II C User Screen Manual Figure 23 SPC Setup 16 SPC Setup Setpoints 16.1 Trig SPC can be set up to sample data at a specific event (through the logic program) or at specified time intervals. Enter "1" for event based initiation, or enter "0" for time based initiation. Control relays CR 6633 through CR 6640 are used to start the sampling in event based operation. The trigger CR must be energized at least 1 second (when deenergized it must also remain deenergized for 1 second). Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 50 of 62 MACO Breeze II C User Screen Manual 16.2 Sample Size Enter the number of measurement readings, from 2 to 25, that are to constitute the sample. SPC data that has accumulated up to this point will be cleared when sample size is changed. 16.3 Sample Time This is the time between samples when the variable has been set to the time based mode. Units are in seconds and can range from 0 to 999 (16-2/ 3 minutes). The minimum entry is 1. An entry of 0 will cause the sampling to cease. 16.4 Group Period This is the time between the beginning of each sample group. The units are in seconds and can range from 0 to 65535 (18-1/5 hours). If the setpoint is 0 or is less than the time required to collect a sample group, then the channel will be sampled continuously at the sample rate with no time lapses between the end of one sample and the beginning of another. This setpoint is also active when in the CR Triggered mode. Figure 24 Group Period diagram 16.5 LSL Spec limit This is usually set to the lower limit of acceptability for the parameter being monitored. When a sampled sub-group average is less than this entry, then the specification alarm control relay for this channel will be activated. This entry is also used in the calculations for Cr, Cp, and Cpk. The number of decimal places for this entry is the same as the sampled parameter (and is determined in the SPCSetup.csv file in the Nand Flash). Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 51 of 62 MACO Breeze II C User Screen Manual 16.6 USL Spec limit This is usually set to the Upper limit of acceptability for the parameter being monitored. When a sampled sub-group average is less than this entry, then the specification alarm control relay for this channel will be activated. This entry is also used in the calculations for Cr, Cp, and Cpk. The number of decimal places for this entry is the same as the sampled parameter (and is determined in the SPCSetup.csv file in the Nand Flash). 16.7 Recalc Limits The upper and lower control limits are automatically calculated after the first twenty-five sample groups have been collected (an "F" will appear on the setup screen, indicating the limits have been calculated). These limits will remain in effect until the SPC data is cleared or until Recalc limits are entered. Recalc limits are used to recalculate control limits on a specific group of SPC data after the limits have been frozen. For example, a change in material may cause the transfer pressure to increase for a given part. For the lower recalc limit, enter the lowest numbered sample group to be used in the limits calculation. For the upper recalc limit, enter the highest numbered sample group to be used in the limits calculation. Most recent sample groups have the highest sample group number. The oldest data will have the lowest sample group number. There must be a difference of five between the group numbers and the lower recalc limit must always be lower than the upper recalc limit. In addition, a statistically significant R-Bar must exist. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 52 of 62 MACO Breeze II C User Screen Manual Figure 25 SPC X-Bar Data 17 SPC Screens 17.1 SPC X-BAR Values The screen in Figure 26 shows current values for X-Bar, X-BarBar, Upper and Lower Control Limits for X-Bar values, and the Standard Deviations for all 10 variables. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 53 of 62 MACO Breeze II C User Screen Manual Figure 26 SPC Range Data 17.2 SPC Range Values This screen in Figure 27 is similar to Figure 26 except that it is for the Range values. Shown are the current values for R, R-Bar, Upper and Lower Control Limits, Cr (Capability Ratio), and Cpk (Process Capability Index). Figure 26 and Figure 27 both contain messages such as S, X, R and F. These messages are controlled by data handler control relays (CR 785-814). When one or more of these control relays energizes one the following messages will appear: S – Specification Alarm X – X-Bar Trend Alarm R – R-Bar Trend Alarm F – Upper and Lower Control Limits are Frozen Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 54 of 62 MACO Breeze II C User Screen Manual Figure 27 SPC Graphs 17.3 SPC Graphs Figure 28 is a graphic screens. The screent is a plot of both the X-Bar and Range values on the same screen. The up and down arrows as well as the setpoint below allow the displayed graph to be changed to display a graph of any of the configured selections. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 55 of 62 MACO Breeze II C User Screen Manual Figure 28 SPC Table Data 17.4 SPC Table Data The SPC Tabular Data screen (Figure 29) is a listing of all 100 data points. Included on the screen are the X-Bar and Range values as well as the sample time. Scroll Up and Down functions allow you to access the other points as only 30 can be displayed on a screen. The screen also has provisions for removing a data point if desired. To remove a data point simply enter the number of the point to be removed. Lower numbered data points will then be shifted up one position. An up arrow will be displayed to the right of each line which has an X-Bar data point greater than the current upper spec limit (USL). A down arrow will be displayed to the right of each line which has an X-Bar data point less than the current lower spec limit (LSL). If the X-Bar is within the spec limits (or none have been entered) a dash will be displayed. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 56 of 62 MACO Breeze II C User Screen Manual Figure 29 SPC Distribution Curve 17.4 SPC Distribution Curve Each sampled channel has 2 distribution curves (Figure 30). One is the distribution of the last 100 subgroups sampled. The other is the total history distribution since the channel has been setup and the control limits have been calculated. The curves are not be displayed until control limits have been calculated. Each distribution curve is formed with a series of twenty vertical bar graphs. There are six equally sized divisions on both sides of the centerline. One group extends to the LCL and the other to the UCL. There are also four more divisions above and below the control limits. The distribution curves are saved in battery-backed RAM and so are available on powerup. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 57 of 62 MACO Breeze II C User Screen Manual Figure 30 Head 1 Setpoints 18 Parison Head 100 Setpoint Screens There are up to 4 screens that display the current setpoint value of all 100 Profile setpoints for the displayed. Each setpoint can be changed from this screen regardless of whether it is defined as a master point or not. Changes to these setpoints will automatically change the profile. To view changes made to the profile on the Parison Graph screen a Read Profile will have to initiated. Interpellation of non master points will not function on this screen. Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 58 of 62 MACO Breeze II C User Screen Manual Figure 31 Head 2 Setpoints Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 59 of 62 MACO Breeze II C User Screen Manual Figure 32 Display Config 19 Parison Head 100 Setpoint Screens This screen contains functions that allow for changes to how display functions. Changes can be made to the color pallet, security timeouts, how the highlighted cursor is displayed, as well as touch screen calibration. 19.1 Flash Reload Count / Scroll Rates Flash Reload count- The flash reload count determines the rate at which the cursor flashes. This setpoint is shipped from the factory set at approximately 0.30 seconds and should be left somewhere close to that setting. Using a much faster or much slower setting makes determining the location of the cursor difficult (leave it alone). Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 60 of 62 MACO Breeze II C User Screen Manual Scroll Rates- The scroll rate setpoints determine the rate at which the flashing cursor moves from one reverse video area to the next when a cursor key is pressed. The scroll rate mode setpoints provide three “gears” for the scroll rate. When a cursor key is initially held depressed, the cursor will move at the rate determined by scroll rate mode 1. Releasing the key and then pressing that key or any other cursor key will cause the cursor to move at the rate determined by scroll rate mode 2. Releasing the key again and then again pressing that key or any other cursor key will cause the cursor to move at the rate determined by scroll rate mode 3. These setpoints are initially shipped from the factory set at approximately 0.20 seconds. Try settings of 0.20; 0.15; and 0.10 for Scroll Rate Mode 1, 2 and 3, respectively. Scroll Dead Time- The scroll dead time setpoint determines how soon the cursor moves after a cursor key is pressed. The scroll dead time setpoint is initially shipped from the factory set at approximately 0.40 seconds (leave as is). BlnkTmr- The blank reload count setpoint determines the time between keystrokes before the display “blanks out.” The minimum setting is 60.00 (seconds). Setting this entry to zero disables screen blanking (which may cause damage to the display). This setpoint is initially shipped from the factory set at 600.00 seconds (10 minutes). Settings on the order of 300.00 seconds (5 minutes) might be desirable. Anything less than 5 minutes could be a nuisance (Don’t mess with fractional second settings). Vectr 1- 4 – These setpoints determine the screen number the display will vector when the corresponding Alarm Cr is energized in RLD. Note that these are NOT utilized in Version 2.1 of the Breeze IIC Key1CLM – Ket3CLM- These determine which column in the Bitmap_Keyword_Setup_Assignments.csv will utilize to the functionality of Bitmap keys on the left and right of the display. These setpoints will change automatically if the configuration of the Unit is changed ( DO NOT make manual changes to these setpoints.) Save SP Time - This is the amount of time in seconds before the system will automatically save all setpoints. This timer will only time when there is no display activity. Setting this timer to 0 will cause the system not to function as the system will constantly be saving setpoints and will not allow for any other functions. ( NOTE for best results leave at the default of 600.00 sec.) Operator 17-32/ Selector SW 1-4- These setpoints are the system save the state of control relays. They are populated by the system and no entry should be made. I/O Slot #- Leave at factory setting of 3( Note pertinent to the Breeze IIC ) Scrn Capture Loc- This determines where the system will store the generated .BPM files when a screen capture is preformed. 0 = Nand Flash ( DO NOT USE) 1= USB 1 2= USB 2 3= Compact Flash Card ( Factory default and should be left at 3) Sec Time Out- The minimum security reload setpoint is a timer setpoint (in seconds) used to determine the starting point of a countdown to zero. The minimum setting is 60 seconds; the maximum setting is 655.35 seconds. Each keystroke resets the count to the setpoint. Once this timer counts to zero, the security level will be set Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 61 of 62 MACO Breeze II C User Screen Manual to the "Timeout Security Level" (see below). Sec Time Out Level- The timeout security level setpoint determines the security level of the system after the minimum security reload timer setting times out. This setting, along with the timeout setting above, can be used to prevent an unattended machine from inadvertently being left at a high security level. The minimum setting is 1 (Level 1) and the maximum setting is 4 (Level 4). Touch Screen Calibration- Pressing this button will initiate the calibration routine. Once the calibration is complete the display will return to this screen. Sys Mess as Numbers- enabling this function will display any generated system messages to be displayed in HEX code number for factory troubleshooting ( NOTE do not enable) Enable Limit Display- Should be enable by default. This displays the setpoint limits for the selected setpoint. ( Note that must be disabled if the Enable ID display is enabled ) Enable ID Display- When enabled allows the display to Modbus ID along with the HEX ID for any setpoint or value highlighted by the cursor. ( Note must be disabled if the Enable Limit display is enabled) Copy all files to USB- This function will copy all files stored in the Nand flash to the USB memory stick. Capture All Screens- separate functions for user screens and system screen. Initiating the user screen this function will automatically start at screen #1 and generate and store .BMP files for each screen in the user screen set. Once completed the display will return to this display config screen. Initiating the system screen function will start at screen at # 257 and generate and store .BMP files for each system screen in the system. Once completed the display will return to these display config screen. Color Pallet Setpoints – each color in the pallet has separate RGB settings. Each color has a reset push button to return settings to factory defaults. Color settings should be left at “as shipped” settings as these have been set to match colors utilized in the bitmaps. The above figure 33 shows the correct settings. (Note if changes are made a save setpoints should be preformed ) Cursor Mode – changes the appearance of the selected element when highlighted. 0= default 1= 2 boxes outside of selection 2= one box outside of selection. Box Color – numbers entered match the color pallet number 0- 15 Color for customer descriptions- numbers entered match the color pallet number 0- 15. (Note not utilized in ver. 2.1 of the Breeze IIC ) Store Modfile Desc. In Recipes- defines if the modfile descriptions of the setpoints are included within the recipe file. 0= No ( Default) 1= Yes 2= store in Unicode format Location for Customer Descriptions- 0= messages will display on the mini screen 1= messages will display on the mini screen area.( NOTE not utilized in ver. 2.1 of the Breeze IIC) Eurotherm USA Ashburn Virginia, 20147 1670-IN-038-0-00 Page 62 of 62