PDF - PENTAX Medical
Transcription
PDF - PENTAX Medical
INSTRUCTIONS FOR USE (REPROCESSING/MAINTENANCE) PENTAX VIDEO GI SCOPES 90i SERIES 90K SERIES This instruction for use describes the recommended procedures for the reprocessing and maintenance of the equipment after its use. For the inspection and direction for use, please refer to the separate operation instruction for use. Intended Use (Gastroscopes) These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum and Small Bowel. These instruments are introduced via the mouth or the nose as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations. Never use these endoscopes for any purpose other than that for which they have been designed. The EG-2790i, 2990i, and 2990Zi can only be used with PENTAX video processor, model EPK-i series. The EG-1690K, 2490K, 2790K, 2990K, 3490K, and 3890TK can be used with PENTAX video processor, model EPK-i series or EPK-1000. Intended Use (Duodenoscopes) The model ED-3490TK, the Video Duodenoscopes are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the Billiary Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Common Bile, Hepatic and Cystic Ducts. These instruments are introduced via the mouth when indications consistent with the need for the procedure are observed in adult and pediatric patient populations. Never use these endoscopes for any purpose other than that for which they have been designed. The ED-3490TK can only be used with PENTAX video processors, model EPK-i series or EPK-1000. Intended Use (Colonoscopes) These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Large Bowel. These instruments are introduced via the rectum when indications consistent with the need for the procedure are observed in adult and pediatric patient populations. Never use these endoscopes for any purpose other than that for which they have been designed. The EC-2990Li/Fi/Mi, 3490Fi/Mi/Li, 3490TLi/TFi/TMi, 3890Fi/Fi2/Mi/Mi2/Li, and 3890LZi/FZi/MZi can only be used with PENTAX video processor, model EPK-i series. The EC-3490FK/MK/LK, 3890FK/FK2/MK/MK2/LK, and 3890TLK/TFK/TMK can be used with PENTAX video processor, model EPK-i series or EPK-1000. Intended Use (Small Bowel Enteroscopes) The model VSB-2990i, the Video Small Bowel Enteroscopes are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum and Small Bowel. These instruments are introduced via the mouth or the rectum as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations. Never use these endoscopes for any purpose other than that for which they have been designed. The VSB-2990i can only be used with PENTAX video processor, model EPK-i series. Sterility Statement These endoscopes identified in this instruction for use (IFU) are reusable semi-critical medical devices. Since they are packaged non-sterile, they must be cleaned and either high-level disinfected or sterilized BEFORE initial use. Prior to each subsequent procedure, they must be subjected to an appropriate cleaning and either high-level disinfection or sterilization process. After servicing, all endoscopes and their components must be reprocessed prior to clinical use. Contraindication Please consult regional and national health authority recommendations and requirements regarding protocols to follow in order to reprocess and/or destroy endoscopes that will be used or have been determined to have been used (post procedure) on patients afflicted with Creutzfeldt-Jacob Disease (CJD or vCJD). Notes This IFU has been written having regard to ISO 17664 and national guidelines on reprocessing of medical products (e.g., RKI, DGS/DHOS). Read this IFU before operating, and save this book for future reference. Failure to read and thoroughly understand the information presented in this IFU, as well as those developed for ancillary endoscopic equipment and accessories, may result in serious injury including infection by cross contamination to the patient and/or user. Furthermore, failure to follow the instructions in this IFU and the companion IFU may result in damage to, and/or malfunction of, this equipment. This IFU describes the recommended procedures for the reprocessing and maintenance of the equipment after its use. It is the responsibility of each medical facility to ensure that only well educated and appropriately trained personnel, who are competent and knowledgeable about the endoscopic equipment, antimicrobial agents/processes and hospital infection control protocol be involved in the reprocessing of these medical devices. Current infection control guidelines require that G.I. endoscopes and other semi-critical medical devices, that normally come into contact with intact mucous membranes, such as in the gastrointestinal tract, must at least be cleaned and high-level disinfected before patient use. Only the user can determine if an instrument has undergone appropriate infection control procedures prior to each clinical use. It must be recognized that infection control practices involve many complex and often controversial issues which are constantly evolving. PENTAX strongly recommends that user remain informed of the latest federal and local regulations, and encourages users to follow infection control guidelines developed by various organizations for health care professionals. The text contained in this IFU is common for various types/models of PENTAX endoscopes and users must carefully follow only those sections and instructions pertaining to the specific instrument models identified on the separate IFU. If you have any questions regarding any of the information in this IFU or concerns pertaining to the safety and/or use of this equipment, please contact your local PENTAX representative. TABLE OF CONTENTS 1. INTERNAL CHANNELS OF PENTAX ENDOSCOPES ...............................................................................1 1-1. Internal Channels ..........................................................................................................3 1-1-1. Video Colonoscopes/Gastroscopes ............................................................................3 1-1-2. Video Gastroscopes with a Single/Common Channel for Air/Water Delivery ...........................4 1-1-3. Video Duodenoscopes ...........................................................................................5 1-1-4. Two-Channel (Biopsy/Instrument) Video Endoscopes ......................................................6 1-2. Reprocessing System .....................................................................................................7 1-2-1. Video Colonoscopes/Gastroscopes ............................................................................7 1-2-2. Video Gastroscopes with a Single/Common Channel for Air/Water Delivery ...........................8 1-2-3. Video Duodenoscopes ......................................................................................... 10 1-2-4. Two-Channel (Biopsy/Instrument) Video Endoscopes .................................................... 11 2. CARE AFTER USE ................................................................................................................. 13 2-1. Endoscopes ............................................................................................................... 15 2-1-1. Pre-Cleaning .................................................................................................... 16 2-1-2. Leakage Testing ................................................................................................ 17 2-1-3. Cleaning ......................................................................................................... 17 2-1-4. High-Level Disinfection ....................................................................................... 36 2-1-5. Sterilization ..................................................................................................... 42 2-2. Endoscopic Accessory Instruments (EAIs) and Endoscope Components ....................................... 45 2-2-1. Cleaning ......................................................................................................... 45 2-2-2. High-Level Disinfection ....................................................................................... 47 2-2-3. Sterilization ..................................................................................................... 49 2-3. Water Bottle Assembly .................................................................................................. 51 3. POST REPROCESSING ............................................................................................................ 53 3-1. Servicing .................................................................................................................. 54 3-2. Care and Maintenance Tips ............................................................................................ 56 4. NOMENCLATURE AND FUNCTION ............................................................................................... 59 4-1. Video Endoscope ........................................................................................................ 59 Conventions Throughout this IFU, the following conventions will be used to indicate a potentially hazardous situation which, if not avoided; WARNING : could result in death or serious injury. CAUTION : may result in minor or moderate injury or property-damage. NOTE : may result in property-damage. Also, advises owner/operator about important information on the use of this equipment. Prescription Statement Federal (U.S.A) law restricts this device to sale by or on the order of a physician or other appropriately licensed medical professional. Symbols on Marking Symboles distinctifs Symbol for “MANUFACTURER” Attention, consult instructions for use Attention, consulter le manuel d’utilisation Symbol for “DATE OF MANUFACTURE” Type BF applied part (Safety degree specified by IEC 60601-1) Partie appliquée de type BF (niveau de sécurité spécifié par la norme CEI 60601-1) Symbol for “Authorised Representative in the European Union” このCEマーキングはEC指令への適合宣言マークです。 The CE marking assures that this product complies with the requirements of the EC directive for safety. Das CE Zeichen garantiert, daß dieses Produkt die in der EU erforderlichen Sicherheitsbestimmungen erfüllt. Le logo CE certifie que ce produit est conforme aux normes de sécurité prévues par la Communauté Européenne. II marchio CE assicura che questo prodotto è conforme alle direttive CE relative alla sicurezza. La marca CE asegura que este producto cumple todas las directivas de seguridad de la CE. 1. INTERNAL CHANNELS OF PENTAX ENDOSCOPES The following internal schematics have been provided as a service to help users better understand the intricate construction of PENTAX endoscopes. Knowledge of the various internal channels and tubes within an instrument and their relation to each other allows one to care for and reprocess the endoscope more easily and with greater confidence. Much time and effort has been expended into designing endoscopes and their cleaning/disinfecting components so that reprocessing of the instruments before each patient use can be effectively and efficiently performed by either manual methods or automated processes. Connectors on most PENTAX cleaning/disinfecting adapters and instrument channel inlet incorporate standard size luer-lock and/or luerslip fittings to easily accommodate reprocessing devices or systems available from other manufacturers. As can be seen from these internal schematics, the PENTAX cleaning system promotes efficient unidirectional flow of solution beginning from connections at the light guide plug, traveling up tubings in the umbilical cord to the valve cylinders in the control body, passing through the channels in the insertion tube and finally exiting nozzles or channel openings at the distal tip of the endoscope. The elimination of multiple branching of channels, combined with a direct and straightforward pathway for solutions to travel maximizes flow efficiency and ensures contact of detergent/disinfectant/sterilant with all internally exposed channel surfaces. WARNING: (in the U.S.A. or other countries adhering to FDA regulations) It is imperative that flexible endoscopes and other semi-critical devices be reprocessed using at least high-level disinfection with a legally marketed disinfectant. Only legally marketed endoscope automated reprocessing devices/systems whose device specific claims have been validated by the Automated Endoscope Reprocessor(AERs)/Washer-Disinfectors(WD) manufacturers and/or anti-microbial agents which have been tested and found to be compatible by PENTAX should be used with PENTAX products. It should be noted that only legally marketed AERs and reprocessing devices/systems, or cleared antimicrobial agents which have been tested by PENTAX and found to be compatible with the materials in PENTAX endoscopes should be used with PENTAX products. All AER reprocessing claims, reprocessing instructions, validation studies, and compliance with local regulations and/or professional guidelines are the responsibility of the AER manufacturer. –1– Items required for reprocessing: • Syringes: various sizes can be used including 10/20cc, 50/60cc, etc. • Tray/Reprocessing basin (at least 40 cm by 40 cm/16 in × 16 in) • Soft Sponge • Personal protective equipment (PPE) including gloves, face mask, etc. • Soft sterile gauze • PVE soaking cap OE-C9 • PENTAX cleaning adapter OF-B153 • PENTAX cleaning adapter OF-G17 • PENTAX water jet check valve adapter OE-C12 • PENTAX irrigation tube OF-B113 • PENTAX suction channel selector OF-B161 • PENTAX two-channel cleaning adapter OF-B162 Model Instrument Channel Single Two • PENTAX inlet seal • PENTAX channel cleaning brush CS6021T • PENTAX cylinder cleaning brush CS-C9S • PENTAX channel cleaning brush CS-6030SN • Detergent • Disinfectant • 70-90% medical grade ethyl or isopropyl alcohol • Source of compressed/forced air Common Direction of View A/W channel Forward Oblique Water Jet EG-2790i X X EG-2990i X X EG-1690K X EG-2490K X X EG-2790K X X EG-2990K X X X EG-2990Zi X X X EG-3490K X X X X X EG-3890TK X X X X ED-3490TK X X EC-2990Li/Fi/Mi X X X X* EC-3490Fi/Mi/Li X X X EC-3490FK/MK/LK X X X EC-3490TLi/TFi/TMi X X X EC-3890Fi/Fi2/Mi/ Mi2/Li X X X EC-3890FK/FK2/MK/ MK2/LK X X X EC-3890LZi/FZi/MZi X X X X X X EC-3890TLK/TFK/TMK VSB-2990i X X * See p.5 for drawing of elevator mechanism –2– Elevator mechanism 1-1. Internal Channels 1-1-1. Video Colonoscopes/Gastroscopes EG-2490K/2790K/2990K/3490K, EG-2790i/2990i, EG-2990Zi, EC-3490FK/MK/LK, EC-3890FK/FK2/MK/MK2/LK, EC-2990Li/Fi/Mi, EC-3490Fi/Mi/Li, EC-3490TLi/TFi/TMi, EC-3890Fi/Fi2/Mi/Mi2/Li, EC-3890LZi/FZi/MZi, VSB-2990i Inlet Seal Water Feeding Channel Instrument Channel Inlet Air/Water Feeding Valve Suction Control Valve Air Feeding Channel Water Nozzle Air Nozzle Instrument Channel *Water Jet Channel EG-2990i/2990Zi Water Feeding Channel Air Feeding Channel Instrument Channel Air/Water Nozzle *Water Jet Channel *Scopes with Water Jet System *Water jet check valve adapter OE-C12 with Check-valve OE-C14 Water Feeding Tube Air Feeding Tube *Water Jet Port Water Bottle Assembly Air Pump Suction Tube Suction Source Video Processor The illustration above shows the actual routes taken by air, water, suction and water jet through “single biopsy/instrument channel” PENTAX Video endoscopes. During clinical use a check-valve, OE-C14 should be attached to the check valve unit, OE-C12 within the PENTAX water jet channel system. –3– 1-1-2. Video Gastroscopes with a Single/Common Channel for Air/Water Delivery EG-1690K Air/Water Feeding Valve Suction Control Valve Instrument Channel Inlet Inlet Seal Air/Water Feeding Channel Air/Water Nozzle Instrument Channel Air Feeding Tube Water Feeding Tube Suction Tube Air Pump Water Bottle Assembly Suction Source Video Processor The illustration above shows the actual routes taken by air, water and suction through PENTAX video endoscopes with a single/common channel for air/water delivery and suction. –4– 1-1-3. Video Duodenoscopes ED-3490TK Air/Water Feeding Valve Suction Control Valve Instrument Channel Inlet Water Feeding Channel Inlet Seal Air Feeding Channel Air/Water Nozzle Instrument Channel * Elevator mechanism in ED-series distal tip Air Feeding Tube Water Feeding Tube Suction Tube Air Pump Water Bottle Assembly Suction Source Video Processor The illustration above shows the actual routes taken by air, water and suction through PENTAX video duodenoscopes. NOTE: The elevator control wire/channel is completely sealed and therefore not exposed to potential contaminants and/or patient debris. –5– 1-1-4. Two-Channel (Biopsy/Instrument) Video Endoscopes EG-3890TK, EC-3890TLK/TFK/TMK Instrument Channel Inlet Air Feeding Channel Air Nozzle Inlet Seal Air/Water Feeding Valve Suction Control Valve Instrument Channel Selecter Water Feeding Channel Water Nozzle Instrument Channel *Water Jet Channel Suction Tube * Scopes with Water Jet System * Water jet check valve adapter OE-C12 with Check-valve OE-C14 Air Feeding Tube * Water Jet Port Air Pump Water Feeding Tube Water Bottle Assembly Suction Source Video Processor The illustration above shows the actual routes taken by air, water, suction and water jet through PENTAX Two-Channel (biopsy/instrument) video endoscopes. During clinical use a check-valve, OE-C14 should be attached to the check valve, OE-C12 within the PENTAX water jet channel system. –6– 1-2. Reprocessing System 1-2-1. Video Colonoscopes/Gastroscopes EG-2490K/2790K/2990K/3490K, EG-2790i/2990i, EG-2990Zi, EC-3490FK/MK/LK, EC-3890FK/FK2/MK/MK2/LK, EC-2990Li/Fi/Mi, EC-3490Fi/Mi/Li, EC-3490TLi/TFi/TMi, EC-3890Fi/Fi2/Mi/Mi2/Li, EC-3890LZi/FZi/MZi, VSB-2990i Cleaning Adapter Inlet Seal Air Feeding Channel Water Nozzle Water Feeding Channel Instrument Channel Air Nozzle EG-2990i/2990Zi Instrument (OF-B153) Channel Inlet *Water Jet Channel Water Feeding Channel Air Feeding Channel *Scopes with Water Jet System Instrument Channel Air/Water Nozzle Water Feeding Tube *Water Jet Channel PVE Soaking Cap (ATTACH) Air Feeding Tube Cleaning Adapter OF-G17 *Irrigation Tube OF-B113 *OE-C12 without check-valve Suction Tube Suction Nipple Ventilation Cap (REMOVE) Luer-Slip Syringe with Cleaning/Disinfecting Solution CAUTION: BEFORE IMMERSING: a) The “Red” ventilation cap must be taken OFF. b) The PVE Soaking cap should be securely ON the electrical contacts. To reprocess a PENTAX video endoscope, first an appropriate cleaning detergent and then a high-level disinfectant or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and endoscope components (air/water feeding valve, suction control valve, etc.). Exposure times, concentration, and temperature of detergent and disinfectant/sterilant must be strictly adhered to. –7– Please note that all solution entrance ports and flow pathways are illustrated above. *Be aware that during reprocessing check-valves, OE-C14 should be removed from the check valve, OE-C12 to allow for unrestricted flow of reprocessing agents through the PENTAX water jet channel system. WARNING: Prior to disinfection or sterilization of all instruments and components must be scrupulously cleaned including manual brushing of all brushable channels. 1-2-2. Video Gastroscopes with a Single/Common Channel for Air/Water Delivery EG-1690K Instrument Channel Inlet Inlet Seal Air/Water Nozzle Cleaning Adapter (OF-B153) Air/Water Feeding Channel Instrument Channel Air Feeding Tube Water Feeding Tube Suction Tube Cleaning Adapter OF-G17 PVE Soaking Cap (ATTACH) Suction Nipple Ventilation Cap (REMOVE) Luer-Slip Syringe with Cleaning/Disinfecting Solution –8– CAUTION: BEFORE IMMERSING: a) The “Red” ventilation cap must be taken OFF. b) The PVE Soaking cap should be securely ON the electrical contacts. To reprocess a PENTAX video endoscope with a single/common channel, first an appropriate cleaning detergent and then a high-level disinfectant or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and endoscope components (air/water feeding valve, suction control valve, etc.). Exposure times, concentration, and temperature of detergent and disinfectant/sterilant must be strictly adhered to. Please note that all solution entrance ports and flow pathways are illustrated above. WARNING: Prior to disinfection or sterilization of all instruments and components must be scrupulously cleaned including manual brushing of all brushable channels. –9– 1-2-3. Video Duodenoscopes ED-3490TK Instrument Channel Inlet Inlet Seal Water Feeding Channel Air/Water Nozzle Cleaning Adapter OF-B153 Air Feeding Channel Instrument Channel Air Feeding Tube Water Feeding Tube Suction Tube Cleaning Adapter OF-G17 PVE Soaking Cap (ATTACH) Suction Nipple Ventilation Cap (REMOVE) Luer-Slip Syringe with Cleaning/Disinfecting Solution CAUTION: BEFORE IMMERSING: a) The “Red” ventilation cap must be taken OFF. b) The PVE Soaking cap should be securely ON the electrical contacts. To reprocess a PENTAX video endoscope, first an appropriate cleaning detergent and then a high-level disinfectant or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and endoscope components (air/water feeding valve, suction control valve, etc.). Exposure times, concentration, and temperature of detergent and disinfectant/sterilant must be strictly adhered to. Please note that all solution entrance ports and flow pathways are illustrated above. – 10 – NOTE: • Since the elevator control wire/channel is fully enclosed and not exposed to patient material, it does not require special reprocessing. However, the actual elevator mechanism (front, back and base) must be cleaned and thoroughly reprocessed. • Be aware that all recessed areas around the elevator mechanism should be thoroughly cleaned with an appropriately sized cylinder cleaning brush (e.g. CS-C9S) and in a cleaning detergent solution. WARNING: Prior to disinfection or sterilization of all instruments and components must be scrupulously cleaned including manual brushing of all brushable channels. 1-2-4. Two-Channel (Biopsy/Instrument) Video Endoscopes EG-3890TK, EC-3890TLK/TFK/TMK Instrument Channel Inlet Air Feeding Channel Water Nozzle Suction Channel Selector OF-B161 or Two channel cleaning adapter OF-B162 Inlet Seal Cleaning Adapter OF-B153 Water Feeding Channel Air Nozzle Instrument Channel Water Jet Channel* *Scopes with Water Jet System Suction Tube Cleaning Adapter OF-G17 *lrrigation Tube *OF-B113 PVE Soaking Cap (ATTACH) Water Feeding Tube OE-C12 without check-valve* Air Feeding Tube Ventilation Cap (REMOVE) Luer-Slip Syringe with Cleaning/Disinfecting Solution – 11 – CAUTION: BEFORE IMMERSING: a) The “Red” ventilation cap must be taken OFF. b) The PVE Soaking cap should be securely ON the electrical contacts. To reprocess a PENTAX Two-Channel Video Endoscope, first a cleaning detergent and then a high-level disinfectant or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and endoscope components (air/water feeding valve, suction control valve, etc.). Exposure times, concentration, and temperature of detergent and disinfectant/ sterilant must be strictly adhered to. Please note that all solution entrance ports and flow pathways are illustrated above. Be aware that during reprocessing check-valves, OE-C14 should be removed from the check valve, OE-C12 to allow for unrestricted flow of reprocessing agents through the PENTAX water jet channel system. WARNING: Prior to disinfection or sterilization of all instruments and components must be scrupulously cleaned including manual brushing of all brushable channels. – 12 – 2. CARE AFTER USE IMPORTANT INSTRUCTIONS Cleaning-Disinfection-Sterilization: PENTAX Endoscopes To maintain maximum performance and a long service life of the endoscope, proper care after each procedure is extremely important. Immediately after (at least within one hour) the completion of a procedure, the endoscope and its components should be thoroughly and carefully cleaned. If the endoscope or its components are left uncleaned for some time after use, dried blood, mucous, contrast material or other patient debris may cause damage to the instrument (component) or may interfere with the ability of the user to properly reprocess the endoscope or component. NOTE: This reprocessing/maintenance instruction for use contains detailed recommendations on the manual reprocessing of PENTAX endoscopes using PENTAX supplied cleaning/disinfecting adapters. Automated endoscope reprocessors(AERs)/WasherDisinfectors (WDs) may provide a means of reprocessing flexible endoscopes, including PENTAX instruments. However, only those Automated endoscope reprocessors(AERs)/Washer-Disinfectors (WDs) should be used whose manufacturers provide devicespecific instructions and have validation data that support each AER/WD claim with respect to PENTAX model instruments. AER/ WD manufacturers should be consulted for their specific claims including but not necessarily limited to a) the ability of the AER/WD to provide a cleaned and high-level disinfected (or sterilized) endoscope and endoscope components (eg., valves), b) the identification of any special feature area (internal channel) or endoscope component that can not be reprocessed and therefore requires manual reprocessing, c) the microbial quality of the rinse water, d) the inclusion of an “automated” alcohol rinse cycle, e) the inclusion of a terminal drying cycle that removes the majority of water/fluid within endoscope channels, f) maintenance procedures for water filter replacement and/or decontamination of the filtration system to ensure the microbial claim of water, etc. g) compliance with local regulations and/or guidelines. – 13 – CAUTION: (Automated Reprocessing) Unless the AER/WD has validated channel monitoring capability for device-specific endoscope models, • prior to "automated reprocessing" check and confirm the patency of any lumens/channels to ensure that all internal channels are unblocked and/or unclogged • prior to use of any AER/WD confirm the integrity of all channel tubes and recommended channel cleaning adapters • once tubing sets are connected to the appropriate channels, check the tubing to confirm that there are no kinks or crimps that could restrict flow of solution throughout the endoscope • (if the unit allows) during reprocessing check and confirm that reprocessing solutions are exiting expected channel ports Failure to adhere to the above recommendations could result in inadequate cleaning, disinfection (or sterilization) of and/or improper removal of residues from all instrument channels/ surfaces. Summary of PENTAX Reprocessing Instructions PENTAX reprocessing instructions can be broken down into six basic steps. 1. 2. 3. 4. Pre-Cleaning (in the Examination Room) Section 2-1-1 Leakage Testing (in a designated Decontamination Area) Section 2-1-2 Cleaning (in a designated Decontamination Area) Section 2-1-3 High-Level Disinfection or Sterilization (in a designated Decontamination Area) Section 2-1-4, 2-1-5 5. Endoscopic Accessory Instruments (EAIs) and Endoscope Components Section 2-2 6. Post Reprocessing Chapter 3 NOTE: The instructions for manual reprocessing of PENTAX endoscope contained in this instruction for use are consistent with reprocessing guidelines, developed by medical professional organizations (eg., SGNA, ASGE, APIC, ESGE) and/or national consensus groups (including ASTM). NOTE: Contact the manufacturer and follow local regulations regarding safe use, appropriate handling and disposal of cleaning and disinfection solution including alcohol. Material Safety Data Sheets (Health and Safety Data Sheets or similar documents depending upon country) available from the cleaning and disinfection solution (including alcohol) manufacturer should provide guidance to end users about composition, hazards, chemical and physical properties, first aid, handling and storage, stability, precautions, disposal, etc. associated with cleaning and disinfection solution including alcohol. – 14 – 2-1. Endoscopes WARNING: The importance of meticulous mechanical cleaning of the endoscope and its removable components cannot be overemphasized. Prior to disinfection or sterilization, all instruments and components must be scrupulously cleaned. Failure to do so could result in incomplete or ineffective disinfection and sterilization. During the reprocessing process, always wear protective garments such as gloves, gowns, face masks, etc. to minimize the risk of cross contamination. CAUTION: To avoid damaging the endoscopes, do NOT twist, rotate or bend excessively any of the strain relief boots (1), (2) during pre-use inspection, clinical use, reprocessing, or any handling activity. Be particularly cautious for the insertion tube strain relief boot(1). When wiping the insertion tube and the cable, use a slow back and forth motion to wipe them along the tube/cord. Never apply excessive force or torque on these strain relief boots or slim tubes/cords. L U L U F F F F D D R R Do NOT Bend D R F F L U Do NOT Twist or Rotate (1) (2) Figure 2.1 – 15 – 2-1-1. Pre-Cleaning Immediately after removing the endoscope from the patient, gently wipe all debris from the insertion tube with gauze or the like moistened with a cleaning detergent solution. F L U 1) D R Figure 2.2 2) Place the distal end of the endoscope into detergent solution and aspirate through the biopsy/suction channel for 5 ~ 10 seconds. Alternate aspiration of solution and air several times to create agitation for better pre-cleaning. When finished, pull the distal tip out of the solution. Figure 2.3 3) 4) Set A/W Drain lever on the water bottle assembly to the drain position. With the air pump of the video processor ON and set to the HIGHEST pressure setting, depress the air/water feeding valve of the endoscope fully until all water has been discharged from the endoscope. Alternate covering of the hole in the valve and depressing the valve to forcefully expel mucous, debris, etc. which may have entered the air and water nozzles. (Only for models with a water jet channel system) Attach the irrigation tube, OF-B113 to the water jet check valve adapter, OE-C12. Connect a syringe filled with detergent solution to the OF-B113, and flush the channel with the detergent. Fill the syringe with air and flush through the channel several times to force any residual solution out of the channel. – 16 – (1) (1) Irrigation Tube (OF-B113) (2) Water Jet Check Valve Adapter (OE-C12) (2) Figure 2.4 5) (For ED-3490TK) Using a 10/20 mL syringe filled with detergent solution, inject the solution onto and surrounding the elevator mechanism so that all recessed areas within the distal endoscope tip are exposed to the detergent. 2-1-2. Leakage Testing Before reprocessing and/or immersion in any fluids, PENTAX endoscopes should be tested for the loss of integrity in their watertight construction by using PENTAX brand leakage testers. For specific details on PENTAX recommended leak detection procedures, please refer to the instructions supplied with PENTAX leakage testers. CAUTION: Various types of endoscope leakage testers exist including manual, electro-mechanical and “automated” versions, some of which are stand alone units and others which may be integrated into AERs/WDs. It must be recognized that PENTAX does not evaluate non-PENTAX leakage tester systems to satisfy their specific product claims, for their effectiveness to accurately detect leaks and/or for their compatibility with PENTAX endoscopes. Insufficient pressures may reduce the likelihood for accurate leak detection, especially if the endoscope’s distal bending section is not flexed during testing. Excessive pressures may adversely affect the endoscope, especially if pressurization occurs during automated reprocessing at elevated temperatures. PENTAX accepts no responsibility for use of non-PENTAX leakage testers. Users should check with the leakage tester manufacturer and confirm their specific product claims, including compatibility with PENTAX endoscopes at various temperatures and their ability to detect leaks with/without fluid immersion and with/without flexing of the endoscope’s distal bending section. 2-1-3. Cleaning 1) Prepare a basin with water and an appropriate cleaning detergent per detergent manufacturer’s instructions. The solutions should either be cleaning detergents or other cleaning agents specially formulated to clean flexible endoscopes. For specific brands of compatible solutions, please contact your local PENTAX service facility or sales representative. – 17 – CAUTION: BEFORE IMMERSING: a) The PVE Soaking cap should be securely ON the electrical contacts. b) The “Red” ventilation cap must be taken OFF. (1) PVE Soaking Cap (OE-C9) (2) Ventilation Cap (OF-C5) (1) (2) Figure 2.5 CAUTION: The use of a cleaning detergent immediately after each procedure to dissolve and remove organic contaminants and proteinaceous debris is essential to the care and maintenance of the endoscope from the standpoints of infection control and functionality. 2) Immerse the endoscope and its components in fresh detergent solution. After removing the removable components, thoroughly (but gently) wash the entire surface of the endoscope and its components. (1) (1) Suction Control Valve (OF-B120) (2) Air/Water Feeding Valve (OF-B188) (3) Inlet Seal(s) (OF-B190) (the second one is for two-channel video endoscopes only) (4) Suction Channel Selector (OF-B161) (two-channel endoscopes only) (5) Water Jet Check Valve Adapter (OE-C12) (endoscopes with jet system only) (6) Water Jet Connector Cap (OF-B118) (endoscopes with jet system only) (2) (4) (3) (5) (6) Figure 2.6 Make sure that the recessed areas such as the channel openings, valve cylinders, elevator mechanism (if present), distal endoscope tip, etc. are brushed clean using the provided brushes. – 18 – NOTE: Cleaning detergent refers to cleaning solution specially formulated for use and compatible with flexible endoscopes. 3) Manual Cleaning of Channels and/or Ports/Cylinders by Brush A variety of special brushes have been provided to mechanically brush clean the entire suction/instrument channels, all cylinders, ports and valves. The entire endoscope should be immersed in detergent solution during the remainder of the cleaning procedure. CAUTION: 1) It is highly recommended that ONLY PENTAX cleaning brushes specified in our instructions for use should be used to clean PENTAX endoscopes. 2) PENTAX cleaning brushes have been specially designed to clean various PENTAX internal channel systems and valves/ ports/cylinders. Validation studies have been performed supporting the use of PENTAX supplied brushes and cleaning adapters for cleaning PENTAX endoscopes following PENTAX manual reprocessing instructions. 3) Over the years some other manufacturers’ cleaning brushes/ devices have been found to damage PENTAX endoscopes and/or create the need for servicing as some cleaning devices can become lodged (“stuck”) inside various lumens of PENTAX endoscopes. The likelihood for endoscope damage or servicing increases if a cleaning device does not have a protective tip (or contains any sharp-edged surface), if its flexible shaft uses a flimsy plastic material that is not firm enough to allow for easy accessory advancement and/or if the proper sequence and/or direction of channel cleaning is not followed as described in PENTAX instructions for use. 4) The cleaning brushes should always be inserted as described in PENTAX reprocessing/maintenance instructions for use. 5) It is recommended that cleaning brushes should be grasped about 5 cm (2 inches) from any channel ports/openings while advancing/withdrawing a brush. To avoid endoscope and brush damage, do not use force. 6) To prevent excess friction between brush and channel, do NOT tightly coil the endoscope while brushing. Keep the insertion tube and umbilical cord as straight as possible and never attempt to pass the cleaning brush through a fully angulated endoscope. Failure to follow these instructions could result in endoscope or brush damage. 7) The channel cleaning brushes CS6021T (disposable) should NOT be inserted from the suction cylinder into suction tube within the umbilical cord. Bristles from the CS6021T could become stuck within the suction cylinder when pulling it out from the suction nipple at the PVE Connector. – 19 – a) Models with a Water Jet channel EG-2990Zi, 2990K, 3490K, 3890TK, EC-3490FK/MK/LK, 3890FK/FK2/MK/MK2/LK, EC-2990Li/Fi/Mi, 3490Li/Fi/Mi, 3490TLi/TFi/TMi, 3890Fi/Fi2/Mi/Mi2/Li, 3890LZi/FZi/MZi EG-2990i/2990Zi (7) Air/Water Feeding Valve Water Feeding Channel (5) Suction Control Valve Air Feeding Channel (8) Air/Water Cylinder Instrument Channel Air/Water Nozzle Water Jet Channel* Inlet Seal (3) Instrument Channel Inlet (4) Suction Cylinder (2) Instrument Channel Air Feeding Tube Water Feeding Tube OE-C12 without check-valve (9) Water Jet Check Valve Adapter Water Jet Connector Cap Air/Water Port (10) Water Jet Cylinder (1) Suction Tube Suction Nipple Fig Place Brush Entrance (1) Suction Tube CS6021T Suction Nipple (2) Instrument Channel CS6021T Suction Cylinder Distal Tip [p.25]4)-c) (3) Instrument Channel Inlet CS-C9S Instrument Channel Inlet — [p.25]4)-a) (4) Suction Cylinder CS-C9S Suction Cylinder — [p.27]4)-e) (5) Suction Control Valve CS-C9S Suction Control Valve — [p.27]4)-f) (6) Elevator Mechanism CS-C9S — — — (7) Air/Water Feeding Valve CS-C9S Air/water feeding valve — [p.28]5)-a) (8) Air/Water Cylinder CS-C9S Air/Water Cylinder — [p.29]5)-b) CS-C9S Water Jet Adapter — [p.29]6)-a) CS-C9S Water Jet Cylinder — [p.30]6)-b) (9) (10) Water Jet Check Valve Adapter Water Jet Cylinder Exit – 20 – Section Suction Cylinder [p.25]4)-b) b) Models without a Water Jet channel system EG-1690K (7) Air/Water Feeding Valve (5) Suction Control Valve (8) Air/Water Cylinder (4) Suction Cylinder Inlet Seal Air/Water Feeding Channel (3) Instrument Channel Inlet Air/Water Nozzle (2) Instrument Channel Water Feeding Tube Air Feeding Tube (1) Suction Tube Air/Water Port Suction Nipple Fig Place Brush Entrance (1) Suction Tube CS6021T Suction Nipple (2) Instrument Channel CS6021T Suction Cylinder Distal Tip [p.25]4)-c) (3) Instrument Channel Inlet CS-C9S Instrument Channel Inlet — [p.25]4)-a) (4) Suction Cylinder CS-C9S Suction Cylinder — [p.27]4)-e) (5) Suction Control Valve CS-C9S Suction Control Valve — [p.27]4)-f) (6) Elevator Mechanism CS-C9S — — — (7) Air/Water Feeding Valve CS-C9S Air/water feeding valve — [p.28]5)-a) (8) Air/Water Cylinder CS-C9S Air/Water Cylinder — [p.29]5)-b) – 21 – Exit Suction Cylinder Section [p.25]4)-b) ED-3490TK (7) Air/Water Feeding Valve (5) Suction Control Valve (8) Air/Water Cylinder Inlet Seal Water Feeding Channel (6) Elevator Mechanism (3) Instrument Channel Inlet (4) Suction Cylinder Air Feeding Channel (2) Instrument Channel Water Feeding Tube Air Feeding Tube Air/Water Port (1) Suction Tube Suction Nipple Fig Place Brush Entrance (1) Suction Tube CS6021T Suction Nipple (2) Instrument Channel CS6021T Suction Cylinder Distal Tip [p.25]4)-c) (3) Instrument Channel Inlet CS-C9S Instrument Channel Inlet — [p.25]4)-a) (4) Suction Cylinder CS-C9S Suction Cylinder — [p.27]4)-e) (5) Suction Control Valve CS-C9S Suction Control Valve — [p.27]4)-f) (6) Elevator Mechanism CS-C9S — [p.27]4)-g) (7) Air/Water Feeding Valve CS-C9S Air/water feeding valve — [p.28]5)-a) (8) Air/Water Cylinder CS-C9S Air/Water Cylinder — [p.29]5)-b) The areas in back of/ surrounding elevator – 22 – Exit Suction Cylinder Section [p.25]4)-b) EG-2490K, EG-2790K, EG-2790i, VSB-2990i (7) Air/Water Feeding Valve (5) Suction Control Valve (8) Air/Water Cylinder (4) Suction Cylinder Inlet Seal (3) Instrument Channel Inlet (2) Instrument Channel Air Feeding Tube Water Feeding Tube (1) Suction Tube Fig Place (1) Suction Tube (2) Instrument Channel (3) Instrument Channel Inlet (4) Brush CS6021T/ CS6030SN* CS6021T/ Entrance Suction Nipple Exit Suction Cylinder Section [p.25]4)-b) Suction Cylinder Distal Tip [p.25]4)-c)/ [p.26]4)-d)* CS-C9S Instrument Channel Inlet — [p.25]4)-a) Suction Cylinder CS-C9S Suction Cylinder — [p.27]4)-e) (5) Suction Control Valve CS-C9S Suction Control Valve — [p.27]4)-f) (6) Elevator Mechanism N/A — — [p.27]4)-g) (7) Air/Water Feeding Valve CS-C9S Air/water feeding valve — [p.28]5)-a) (8) Air/Water Cylinder CS-C9S Air/Water Cylinder — [p.29]5)-b) CS6030SN* *(For VSB-2990i) – 23 – c) Models with Two Blopsy/Instrument Channels EG-3890TK, EC-3890TLK/TFK/TMK (8) Instrument Channel Inlet (9) Air/Water Feeding Valve Inlet Seal (5) Selector (10) Air/Water Cylinder (7) Suction Control Valve (4) Selector Cylinder* (6) Suction Cylinder Water Jet Channel (3) Instrument Channel (2) Suction Channel (11) Water Jet Check Valve Adapter Air Tube (1) Suction Tube (12) Water Jet Cylinder Water Tube Suction Nipple Fig Place Brush Entrance (1) Suction Tube CS6021T Suction Nipple (2) Suction Channel CS6021T Suction Cylinder (3) Instrument Channel CS6021T Instrument Channel Inlet Distal Tip [p.26]4)-c) (4) Selector Cylinder CS-C9S Selector Cylinder — [p.28]4)-h) (5) Selector CS-C9S Selector — [p.28]4)-i) (6) Suction Cylinder CS-C9S Suction Cylinder — [p.27]4)-e) (7) Suction Control Valve CS-C9S Suction Control Valve — [p.27]4)-f) (8) Instrument Channel Inlet CS-C9S Instrument Channel Inlet — [p.25]4)-a) (9) Air/Water Feeding Valve CS-C9S Air/Water Feeding Valve — [p.28]5)-a) (10) Air/Water Cylinder CS-C9S Air/Water Cylinder — [p.29]5)-b) (11) Water Jet Check Valve Adapter CS-C9S Water Jet Adapter — [p.29]6)-a) (12) Water Jet Cylinder Water Jet Cylinder — [p.30]6)-b) CS-C9S – 24 – Exit Suction Cylinder Selector Cylinder Section [p.25]4)-b) [p.26]4)-c) 4) Brush clean the entire instrument/suction channel system: (a) Insert the large bristle of the cylinder cleaning brush (CS-C9S) into the instrument channel inlet until the grip is stopped at the inlet and scrub clean the surface inside the instrument channel inlet and then gently withdraw the brush. Repeat several times ensuring that only a clean brush is introduced into the cylinder each time. (1) (3) (1) Cylinder Cleaning Brush (CS-C9S) (2) Channel Cleaning Brush (CS6021T/ CS6030SN) (3) Channel Cleaning Brush (CS6021T/ CS6030SN) (2) Figure 2.7 (1) (1) Cylinder Cleaning Brush (CS-C9S) (2) Instrument Channel Inlet (2) Figure 2.8 (b) (c) Using the long channel cleaning brush CS6021T, insert back end (blue) of the shaft of brush into the opening of the suction nipple and gently pass the brush until it appears in the suction control valve receptacle (suction cylinder). Grasp the back end of the brush shaft and gently pull the brush from the receptacle. This will clean the suction tube within the umbilical cord. Brush this channel at least 2 times OR until it is visibly clean. (For VSB-2990i) Using the long channel cleaning brush CS6030SN, insert the brush into the opening of the suction nipple and gently pass the brush until it appears in the suction control valve receptable (suction cylinder). (ALL ENDOSCOPES EXCEPT VSB-2990i and TWOCHANNEL VIDEO ENDOSCOPES) Insert the CS6021T (blue coded end) into the opening at the bottom of the suction control valve receptacle (suction cylinder) on the control head and gently advance until it exits the endoscope distal tip. Grasp the back end of the brush shaft and then gently pull the brush from the endoscope distal tip. DO NOT USE EXCESS FORCE. Brush this channel at least 2 times OR until it is visibly clean. – 25 – (TWO-CHANNEL VIDEO ENDOSCOPES ONLY) Insert the CS6021T (blue coded end) into the opening at the bottom of the suction control valve receptacle (suction cylinder) on the control head and gently advance until it exits the selector cylinder. Then, brush clean the selector cylinder and the selector itself. Grasp the back end of the brush shaft and then gently pull the brush from the cylinder of the suction channel selector. DO NOT USE EXCESS FORCE. Brush this channel at least 2 times OR until it is visibly clean. (TWO-CHANNEL VIDEO ENDOSCOPES ONLY) Insert the CS6021T (blue coded end) into the instrument channel inlet and gently advance until it exits the endoscope distal tip. Be sure to clean both channels very carefully. Then, grasp the back end of the brush shaft and then gently pull the brush from the endoscope distal tip. DO NOT USE EXCESS FORCE. Repeat at least 2 times until it is visibly clean. TWO-CHANNEL VIDEO ENDOSCOPES ONLY (1) (2) (2) (3) (1) Cylinder Cleaning Brush (CS-C9S) (2) Channel Cleaning Brush (CS6021T) (3) Channel Cleaning Brush (CS6021T) (4) Channel Cleaning Brush (CS6021T) (4) Figure 2.9 (d) (For VSB-2990i) Insert the brush (CS6030SN) into the opening at the bottom of the suction control valve receptacle (suction cylinder) on the control head and gently advance until it exits the endoscope distal tip. DO NOT USE EXCESS FORCE. Then gently withdrawal the brush. Repeat at least 3 times until it is visibly clean. NOTE: • The CS6021T is provided non-sterile for one time use. Never reuse the brush on more than one instrument. • When using the tri-bristled brush CS6021T, slowly remove it from the endoscope being careful not to fling patient debris/ contaminants into the air as the bristles exit the endoscope. – 26 – (e) Using the large bristle of the cylinder cleaning brush (CS-C9S), scrub clean the surfaces inside the suction control valve receptacle (suction cylinder) on the control head. Do NOT insert the brush excessively. (1) (1) Cylinder Cleaning Brush (CS-C9S) Figure 2.10 NOTE: Be sure to inspect the bottom of the suction control valve receptacle (suction cylinder) on the control head for any debris. (f) Scrub all internal and external surfaces of the suction control valve OF-B120 using the smaller end of the cylinder cleaning brush (CS-C9S). (1) Suction Control Valve (OF-B120) (2) Cylinder Cleaning Brush (CS-C9S) (1) (2) Figure 2.11 (g) (For ED-3490TK) Using the CS-C9S brush clean the distal tip of the endoscope including the elevator mechanism and all area around the elevator itself. (1) Cylinder Cleaning Brush (CS-C9S) (1) Figure 2.12 – 27 – NOTE: • Do NOT squeeze or severely bend the insertion tube. • Do NOT use any abrasive materials. • Be careful to avoid damage to the distal lenses. WARNING: Worn or damaged valves and/or O-rings should be replaced with new ones. The entire valve mechanism should be subjected to a high-level disinfection or sterilization procedure prior to use (O-ring set, model OF-B127, is optionally available). Failure to do so could result in continuous aspiration which in certain clinical situations can suction tissue into the distal channel opening at the scope tip and/or create a loss of insufflated air via the suction system. A compromised valve could also result in the potential for reflux or spit-back of patient fluids that may present infection control risks. (h) Insert the large bristle of the cylinder cleaning brush (CS-C9S) into the opening of the suction channel selector cylinder and scrub clean the surface inside the cylinder of the suction channel selector and then gently withdraw the brush. Repeat several times ensuring that only a clean brush is introduced into the cylinder each time. (1) Cylinder Cleaning Brush (CS-C9S) (1) Figure 2.13 (i) Scrub all internal and external surfaces of the selector using the smaller side of the cylinder cleaning brush (CS-C9S). (1) (1) Suction Channel Selector (OF-B161) (2) Cylinder Cleaning Brush (CS-C9S) (2) Figure 2.14 5) Manual Cleaning of A/W Valve and Cylinder by Brush a) Remove the air/water feeding valve (OF-B188). Scrub all internal and external surfaces of the valve using the smaller end of the cylinder cleaning brush (CS-C9S). Repeat several times OR until the bristle of the brush becomes visibly clean. – 28 – (1) (1) Air/Water Feeding Valve (OF-B188) (2) Cylinder Cleaning Brush (CS-C9S) (2) Figure 2.15 WARNING: Worn or damaged valves (OF-B188 and O-rings in particular) should be replaced with new ones which has already been subjected to a high-level disinfection or sterilization procedure (O-ring set, model OF-B192, is optionally available). Failure to do so could create a risk of cross contamination to the end users due to the potential for reflux or spit-back of patient fluids out of the air/water feeding valve. It could also create continuous air flow or excessive air insufflation and result in the potential patient injury such as pneumatic perforation. b) Using the larger end of the CS-C9S cleaning brush, scrub clean the surfaces inside the air/water cylinder. Repeat brushing at least 3 times OR until it becomes visibly clean. (1) Cylinder Cleaning Brush (CS-C9S) (1) Figure 2.16 6) Manual Cleaning of Water Jet Check Valve Adapter by brush (Endoscopes with Water Jet System) a) The water jet check valve adapter (OE-C12) should already have been removed from the endoscope. Remove the detachable check-valve, model OE-C14 from the adapter unit, OE-C12. Clean all surfaces of the OE-C14 and the OE-C12 using the special cylinder cleaning brush CS-C9S and detergent solution. (1) (2) (3) Figure 2.17 – 29 – (1) Water Jet Check Valve Adapter (OE-C12) (2) Check-Valve (OE-C14) (3) Cylinder Cleaning Brush (CS-C9S) NOTE: The check-valve OE-C14 is a reusable component and it as well as the OE-C12 water jet check valve adapter, irrigation tube and water jet channel should be reprocessed after each use. WARNING: A worn or damaged check valve should be replaced with a new one which has already been subjected to cleaning, a high level disinfection or sterilization procedure. For repeated use, always ensure that the check valve has already been reprocessed. A damaged, worn or missing check valve could create a risk of cross-contamination to the end user due to the potential for reflux (spit-back) of patient fluids through an unsealed path. If the check valve is not attached properly, it could fall off during the endoscopic procedure and results in potential for end user cross-contamination. CAUTION: If there is any doubt as to the suitability for use of the check valve, replace with a fully reprocessed one prior to each procedure. A set of 10 pieces of the OE-C14 check-valve are optionally available in a package as model OE-C15. NOTE: The PENTAX endoscope components, OE-C12 water jet check valve adapter, OE-C14 check-valve and OF-B113 irrigation tube may all be subjected to ultrasonic cleaning and prevacuum steam sterilization. b) Using the larger bristled end of CS-C9S, brush clean inside the surfaces of the water jet cylinder. (1) Cylinder Cleaning Brush (CS-C9S) (1) Figure 2.18 NOTE: Brushing of internal channels and/or ports/cylinders is not a substitute for exposure to an appropriate cleaning solution. Manual cleaning by brush complements and augments the cleaning effectiveness of chemical cleaning. – 30 – 7) Chemical Cleaning by Detergent Solution CAUTION: Current PENTAX manual cleaning instructions have been validated using cleaning detergent solutions. For non-enzymatic solutions always adhere to the specific manual cleaning instructions developed by the manufacturer of the detergent product. When using a non-enzymatic detergent during automated reprocessing refer to the validated cleaning instructions developed by the manufacturer of the Automated Endoscope Reprocessor(AER) or Washer-Disinfector (WD) units/ systems. Channel Adapters/Components to be attached Air and OF-B153 Water OF-G17 Syringe Instument Rinse water 10/20 mL 25 mL or more 80 mL or more OF-B153 Biopsy/ Detergent 50 mL or more Inlet Seal(s) OF-B161/OF-B162 (Two-Channel Endoscopes) 50/60 mL 200 mL or more 100 mL or 400 mL or more more OE-C12 with OE-C14 Water Jet removed 10/20 mL 10 mL or more 30 mL or more OF-B113 a) For Air/Water Cleaning (I) The cleaning adapter OF-B153 should be attached to the air/ water cylinder and suction cylinder. This adapter caps (seals) off the air/water and suction cylinders to allow unidirectional flow of solution through these delivery/aspiration systems. i) Raise the locking cap and slide the side covers (marked with ) downward. ii) Align the cylinder guide over the air/water and suction cylinder ports. iii) Slide the OF-B153 forward. iv) Holding the side covers (marked with ), push down and slide the locking cap under the locking tab to secure. v) Lock the locking cap with the locking tab of the base. i) (2) (1) (1) Locking Cap (2) Rubber Seal (3) Base (3) ii) (2) (1) (3) – 31 – (1) Cylinder Guide (2) Cylinder Ports (3) Cylinder Inlets iii) iv) (1) End of Locking Cap (2) Locking Tab (1) (2) (1) Locking Tab v) (1) (1) Figure 2.19 NOTE: It is imperative that the cleaning adapter OF-B153 be securely attached to their respective valve cylinders. Failure to properly match and secure the cleaning adapter could result in ineffective and incomplete reprocessing. (II) The cleaning adapter OF-G17 has a standard ANSI luer lock connector to which a 10/20 mL syringe or other device should be attached. Connect OF-G17 to the air/water port on the PVE connector/plug. At least 25 mL of fresh detergent should be flushed through this connector and will simultaneously flow through both the air and water feeding channels within the endoscope. Check and confirm that the flushed detergent solution exits the distal Air/Water nozzle(s). If the endoscope has two separate nozzles, solution should flow out of BOTH nozzles. Once proper flow has been confirmed, the OF-G17 should be removed. Please refer to the internal schematics. (2) (1) Cleaning Adapter (OF-G17) (2) Cleaning Adapter (OF-B153) (1) Figure 2.20 Provided the cleaning detergent is allowed to remain in contact with the internal channel surfaces for the recommended exposure time, the cleaning solution should dissolve and clean any remaining debris within these channels. – 32 – NOTE: If blockage of the line is encountered, avoid use of excessive pressure to prevent endoscope damage. b) For Biopsy/Suction Cleaning (I) Install a inlet seal to the instrument channel inlet prior to injecting cleaning solution into the suction system. (TWO-CHANNEL VIDEO ENDOSCOPES) Install inlet seals to each instrument channel inlets. Reattach the suction channel selector (OF-B161). If the two channel cleaning adapter, OF-B162, was used, attach the OF-B162 instead of the OF-B161. (1) (2) (1) Cleaning Adapter (OF-B153) (2) Inlet Seal (OF-B190) or Suction Channel Selecter (OF-B161/ Two Channel Cleaning Adapter (OF-B162) for TWO-CHANNEL VIDEO ENDOSCOPES Figure 2.21 WARNING: (TWO-CHANNEL VIDEO ENDOSCOPES ONLY) When the suction channel selector, OF-B161, is used, each channel should be flushed separately with detergent to ensure contact with all instrument channel surfaces. Use the suction channel selector to direct flow to each, separate channel. Provided the entire endoscope remains immersed under detergent solution during this cleaning process, the suction channel selector mechanism should be removed and soaked separately in detergent. (II) A 50/60 mL syringe or other device may be attached to the suction nipple. At least 50 mL of fresh detergent solution should be flushed through the entire instrument/suction channel system. Check and confirm that the flushed detergent solution exits the distal instrument channel opening(s). Once proper flow has been confirmed, the OF-B153 should be removed. As an alternative, solution can be drawn into the instrument channel by attaching tubing from an aspirator to the suction nipple, as long as the aspirator is turned on, detergent solution can be suctioned through the endoscope. Then, remove OF-B153, inlet seal(s) and OF-B161/162 (TWOCHANNEL VIDEO ENDOSCOPES). – 33 – c) For Water Jet System (1) (1) Irrigation Tube (OF-B113) (2) Water Jet Check Valve Adapter (OE-C12) (2) Figure 2.22 WARNING: Although the water jet system may not be clinically used during each procedure, it still MUST be properly cleaned and subjected to the same disinfection/sterilization processes as other internal channels of the endoscope. (I) (II) (III) The water jet check valve adapter (OE-C12) should first be attached to the water jet port. Do NOT attach a check-valve (OE-C14) during the cleaning process. Next, snap the irrigation tube (OF-B113) into the water jet check valve adapter. Using a 10/20 mL syringe, fresh detergent solution at least 10 mL or more should be injected into the irrigation tube through the entire length of the water jet channel until it exits the endoscope distal tip. Check and confirm that the flushed detergent solution flows out from the distal Water Jet channel opening. Once proper flow has been confirmed, the OE-C12 should be removed. d) (For ED-3490TK) (I) (II) Using a 10/20 mL syringe, inject the cleaning detergent solution to clean the elevator mechanism and recessed areas surrounding it within the endoscope distal tip. While the ENTIRE endoscope remains immersed, inject cleaning detergent via syringe all around the elevator mechanism, including behind it to ensure that entrapped air bubbles do not prevent contact of detergent solution with potentially contaminated surfaces. WARNING: Avoid introduction of air during the that no air bubbles exit from the endoscope distal tip. The presence prevent contact of the detergent with – 34 – flushing process. Confirm channel openings at the of the air bubbles could channel surfaces. WARNING: The cleaning detergent solution should remain in contact with ALL internal channels and external endoscope surfaces for the time period recommended by the manufacturer of the detergent. 8) Prior to rinsing, attach the previously removed adapters and components and purge all internal channels with air (using a syringe) to expel residual detergent solution out from each channel. 9) Using clean water, immerse the entire endoscope as well as all removed components and thoroughly rinse all items. While fully immersed, manipulate valve mechanisms and inject clean water via syringe into/ through removable endoscope components. 10) With all cleaning adapters, including irrigation tube, attached to the endoscope, flush all previously air purged channels with clean water (80 mL for the air water feeding channel, 200 mL for the suction channel, 30 mL for the water jet channel). All internal channels must be thoroughly rinsed to remove residual detergent and debris. WARNING: It is important that ALL internal channels (air, water, instrument, water jet, etc.), external endoscope surfaces and components be thoroughly rinsed with clean water to remove residual detergent solution. 11) Remove the endoscope and its components from the solution. 12) Rinse water remaining within the channels should be purged using air to prevent dilution and/or adulteration of antimicrobial agents to be used in the subsequent disinfection or sterilization process. NOTE: Following a water rinse and air purge 70-90% medical grade ethyl or isopropyl alcohol, (30 mL for the air and water feeding channels, 70 mL for the suction/ instrument channels and 10 mL for the water jet channel) followed by compressed air, not greater than 165kPa (24psi), may be used to facilitate drying. 13) Gently dry all external surfaces of the endoscope with a soft gauze or the like. Do NOT put tension on the insertion tube on the endoscope while drying since the outer cover of the bending section may be excessively stretched. WARNING: Prior to disinfection or sterilization, it is imperative that any solutions previously used in the cleaning process be thoroughly rinsed and dried. Failure to do so could result in ineffective or incomplete disinfection and sterilization. – 35 – CAUTION: NEVER subject the endoscope to ultrasonic cleaning methods employing high-frequency ultrasound. 2-1-4. High-Level Disinfection Before any attempt is made to disinfect the endoscope, the complete cleaning procedure described elsewhere in this instructions for use must have been completed. Prior to high-level disinfection, the end user should confirm the minimum effective concentration (MEC) of reused disinfectant as per the manufacturer’s instructions. Before complete immersion in any disinfecting solution the endoscope should have been “Leak Tested” as described elsewhere in this instruction for use. WARNING: It is imperative that flexible endoscopes and other semi-critical devices be reprocessed using at least high-level disinfection with a legally marketed disinfectant. Only legally marketed endoscope automated reprocessing devices/systems whose device specific claims have been validated by the AER/WD manufacturer and/or anti-microbial agents which have been tested and found to be compatible by PENTAX should be used with PENTAX products. Endoscopes are semi-critical devices that require at least highlevel disinfection. Only use legally marketed solutions and/or automatic endoscope reprocessing machines or washing disinfection machines whose manufacturers have made validation testing with PENTAX products (specific to Pentax endoscope model number). A listing of legally marketed solutions/systems which have been determined to be compatible with PENTAX brand products is available from your local PENTAX dealer/authorized service provider. Be aware of the important note regarding infection control on the inside cover of this instruction for use. CAUTION: BEFORE IMMERSING, confirm that a) The “Red” ventilation cap is taken OFF. b) The PVE Soaking cap is securely ON the electrical contacts. (1) PVE Soaking Cap (2) Ventilation Cap (1) (2) Figure 2.23 – 36 – Adapters/ Channel Components to be Syringe Disinfectant attached Air and Water OF-B153 10/20 OF-G17 mL Inlet Seal(s) Instument OF-B161/OF-B162 (Two-Channel Video OE-C12 with OE-C14 removed OF-B113 1) flush 80 mL or 30 mL or more more more 200 mL or 70 mL or more more more 100 mL or 400 mL or 140 mL or more more more 10/20 10 mL or 30 mL or 10 mL or mL more more more 50/60 mL Endoscopes) Water Jet Alcohol 50 mL or OF-B153 Biopsy/ 25 mL or Rinse water The air/water/instrument channel disinfecting adapters consisting of two separate components OF-B153 and OF-G17 should be installed on the endoscope from the previous cleaning procedure. (I) Air/Water feeding channel cleaning adapter OF-G17 incorporates a standard ANSI luer lock connector to which a 10/ 20 mL syringe or other device should be attached. At least 25 ml of fresh (or reused actively effective) disinfecting solution should be flushed through this connector and will simultaneously flow through both the air and water feeding channels of the endoscope. (Please refer to the internal schematics) (2) (1) Cleaning Adapter (OF-G17) (2) Cleaning Adapter (OF-B153) (1) Figure 2.24 (II) After the entire endoscope is completely immersed, and the air and water feeding channels have been filled with disinfecting solution, cleaning adapter OF-G17 and syringe should be removed. (III) Adhere to proper exposure times for the disinfectant. (IV) Confirm that an inlet seal is already attached to the instrument channel inlet during the next step. – 37 – WARNING: (TWO-CHANNEL VIDEO ENDOSCOPES ONLY) The suction channel selector OF-B161 or the two channel cleaning adapter OF-B162 should already be installed on the endoscope from the previous cleaning procedure. When the suction channel selector OF-B161 is used, each channel should be flushed separately with disinfectant to ensure contact with all instrument channel surfaces. Use the suction channel selector mechanism to direct flow to each separate channel. Providing the entire endoscope remains immersed under disinfectant solution during reprocessing, the suction channel selector mechanism or the two channel cleaning adapter OF-B162 should be removed and soaked separately in disinfectant. 2) A 50/60 mL syringe (or other device) can be attached to the suction nipple. At least 50 mL of fresh (or reused actively effective) disinfecting solution should be flushed through or drawn into the entire suction system. (1) (2) (1) Cleaning Adapter (OF-B153) (2) Inlet Seal(s) (OF-B190) Figure 2.25 3) (For ED-3490TK) While the endoscope tip remains immersed, inject disinfectant via syringe all around the elevator mechanism, including behind it to ensure that entrapped air bubbles do not prevent contact of disinfecting solution with potentially contaminated surfaces. 4) Only for models with a water jet channel system (1) (2) Figure 2.26 – 38 – (1) Irrigation Tube (OF-B113) (2) Water Jet Check Valve Adapter (OE-C12) NOTE: Ensure that the removable check-valve OE-C14 been detached from the adapter unit prior disinfectant into the water jet channel system. This unrestricted flow of reprocessing agents into and entire water jet channel system. (I) (II) (III) has already to injecting will allow for through the The irrigation tube (OF-B113) with standard luer lock fitting should already be attached to the water jet cylinder. Fresh (or actively effective used) disinfecting solution more than 10 mL should be flushed through this connection into the water jet channel and exit from the endoscope distal tip. Please refer to the internal schematics. Avoid introduction of air bubbles during this flushing process. Confirm that no air bubbles are present and the entire water jet channel is filled with disinfectant. Remove the irrigation tube OF-B113, the water jet check valve adapter OE-C12 and syringe. WARNING: Although the water jet system may not be clinically used during each procedure, it still MUST be properly cleaned and subjected to the same disinfection/sterilization processes as other internal channels of the endoscope. WARNING: Avoid introduction of air during the flushing process. Confirm that no air bubbles exit from the channel openings at the endoscope distal tip. The presence of the air bubbles could prevent contact of the disinfectant with channel surfaces. 5) As long as the entire instrument is immersed in disinfecting solution, the cleaning adapters, the syringes used in previous steps, the inlet seal should be removed while the instrument remains fully under the disinfecting solution. The removal of component parts and cleaning adapters from the endoscope will eliminate the risk of mated surfaces not being exposed to the liquid chemical germicide. (1) Never remove the PVE soaking cap. D R F L U DC (1) Figure 2.27 – 39 – WARNING: The disinfecting solution should remain in contact with ALL internal surfaces of the channels and external endoscope surfaces for the time period recommended by the manufacturer of the disinfectant. 6) 7) 8) 9) While fully immersed, manipulate valve mechanisms and inject disinfectant via syringe into/through removable endoscope components. This will remove entrapped air bubbles that could inhibit contact of solution with component surfaces and provide for better exposure of surfaces to germicide. Make sure disinfectant is injected into/through the inlet seal. The endoscope’s component parts, including all valves, should remain in contact with the disinfecting solution for the time period recommended by the manufacturer of the solution. After the endoscope and its component parts have been in contact with the disinfecting solution for an appropriate time, flush all channels with air to purge remaining disinfectant, then remove the endoscope and its components from the solution. Thoroughly rinse the entire endoscope and all its components with sterile water. Typical rinsing recommendations following exposure to an appropriate disinfectant solution containing no surfactants call for three separate complete immersions in fresh sterile water. A syringe filled with sterile water should be attached to adapter OF-G17 to flush at least 80 mL of water from the air and water feeding channels of the endoscope. Then fill a syringe with air and flush through the air and water feeding channels several times to force any residual water out of the tubing and nozzles. Dry thoroughly. With a syringe filled with sterile water, the cleaning adapter OF-B153 and inlet seal attached, rinse the entire suction system, including the instrument channel by flushing a total of at least 200 mL of water. Then flush air through the instrument channel several times to remove residual water. Dry thoroughly. (I) (II) Attach the water jet connector cap, OF-B118, and the water jet check valve adapter OE-C12 to the endoscope. Prior to attaching the adapter to the water jet port, ensure the detachable check-valve has already been removed from the adapter unit. Connect a syringe filled with sterile water to the irrigation tube, Part OF-B113 and flush at least 30 mL of water out of the channel. Then fill the syringe with air and flush through the channel several times to force any residual water out of the channel. Dry thoroughly. The quantity of rinse water described above equates to at least three times the volume per internal channel system. – 40 – CAUTION: Adequately rinse all internal channels and endoscope surfaces with sterile water to remove residual disinfectant solution. Insufficient rinsing may result in mucosal inflammation due to exposure to residual disinfectant solution. NOTE: Ideally, all final rinses should be made with sterile water or bacteria-free water whose microbial quality has been confirmed via monitoring. After water rinsing, 70-90% medical grade ethyl or isopropyl alcohol should be flushed through all channels, (30 mL for the air and water feeding channels, 70 mL for the suction channel and 10 mL for the water jet channel) followed by compressed air, not greater than 165kPa (24 psi) to facilitate drying. Channel cleaning adapters should be used for alcohol flushing/rinsing and forced air drying. The outer surfaces of the endoscopes can be dried by gently wiping with a sterile gauze or lint-free cloth saturated with alcohol. Regardless of the quality of the rinse water used, a dry instrument accomplished by a final alcohol rinse followed by forced air is essential to prevent bacteria colonization and/or infections associated with waterborne microorganisms. Such infections are more likely to occur when wet/contaminated instruments are used on patients whose immune systems are compromised or suppressed or when these devices are used in anatomical areas considered sterile and/or susceptible to these organisms. 10) Following an alcohol rinse, the following steps may be performed to aid in the drying process, if filtered compressed or forced air is not available. (I) Remove the channel adapters OF-B153, channel adapter OFG17, and irrigation tube OF-B113, reinstall the previously reprocessed suction control valve, air/water feeding valve, water jet check valve adapter OE-C12 with the check-valve OEC14, and the inlet seal. (II) Attach the endoscope to an external suction source, and aspirate air through the channel of the endoscope to remove any residual alcohol and to air dry channel surfaces. – 41 – (III) Remove the PVE soaking cap and attach the endoscope to the processor with the air pump turned ON to its HIGHEST pressure setting and the A/W drain lever of the water bottle assembly set in the DRAIN position, depress the air/water feeding valve of the endoscope fully until all alcohol has been discharged from the endoscope. Thoroughly drain all alcohol from the air channel as well by covering the hole in the air/water feeding valve. Repeat until no moisture or alcohol is seen exiting the endoscope or distal tip. (1) A/W (2) DRAIN (1) (2) Figure 2.28 11) Gently dry all external surfaces of the endoscope with a soft gauze or the like. Do NOT put tension on the insertion tube while drying since the outer cover of the bending section may be excessively stretched. Dry the objective lens with a cottontip applicator. WARNING: If the endoscope is to be stored after reprocessing, detach removable valves, components, etc. All channels should be completely dry before storage. CAUTION: PENTAX recommended rinsing parameters described in this section on high-level disinfection apply to the use of a 2.4% alkaline Glutaraldehyde solution containing no surfactants. Always check with the manufacturer of the liquid chemical germicide (or AER/WD for “automated reprocessing”) to confirm that their specific rinsing instructions are based upon validated test data demonstrating that no harmful levels of residues remain on any device or lumen surfaces that may pose a risk to patient and/or user. 2-1-5. Sterilization Before any attempt is made to sterilize the endoscope, the complete cleaning procedure as described elsewhere in this instruction for use must have been completed. NOTE: NEVER place the endoscope in a steam autoclave!! – 42 – A) Ethylene Oxide Gas Sterilization (Recommended) Ethylene Oxide (ETO) Gas Sterilization can be performed on these endoscopes, provided the following special instructions, which may differ from other endoscopes, are followed to ensure the proper performance of the instrument. Adhere to the sterilization manufacturer’s instructions and always use a biological indicator. 1) The endoscope must first have been properly cleaned and thoroughly dried according to the instructions in this instruction for use and each of the component parts should be removed. (1) (1) Suction Control Valve OF-B120 (2) Air/Water Feeding Valve OF-B188 (3) Inlet Seal(s) (the second one is for twochannel endoscopes only) (4) Suction Channel Selector OF-B161 (two-channel video endoscopes only) (5) Water Jet Check Valve Adapter OE-C12 (endoscopes with jet system only) (6) Water Jet Connector Cap OF-B118 (endoscopes with jet system only) (2) (4) (3) (5) (6) Figure 2.29 WARNING: Failure to thoroughly dry all surface areas could result in incomplete or ineffective sterilization. Moisture could prevent contact of the ETO gas with the actual contaminated surfaces. CAUTION: Prior to placing these Endoscopes in a Gas Sterilizer or Aeration Chamber: a) The “Red” ventilation cap MUST be “ON” securely. b) The PVE soaking cap should be “OFF” the electrical contacts. This is opposite of the immersion instructions. (1) PVE Soaking Cap (OE-C9) (2) Ventilation Cap (OFC5) (1) (2) Figure 2.30 – 43 – 2) The following parameters for Ethylene Oxide Gas Sterilization are proposed. Temperature: 55°C (131°F) Relative Humidity: 50%RH EO Concentration : 600-650 mg/L Gas Exposure Time: 5 Hours Aeration: 12 Hours at 55°C (131°F) CAUTION: Adequately aerate the endoscope to remove any remaining ethylene oxide. Insufficient aeration may adversely affect the human body. B) Other Methods of Sterilization Other types of cleaning and/or disinfection/sterilization systems/processes are available for the reprocessing of medical devices. However, due to the heat sensitive nature and/or the specific biocompatible materials used in the construction of flexible endoscopes, some of these marketed systems/ processes/solutions could have detrimental effects on flexible endoscopes. To avoid the potential for instrument damage and/or endoscope failure, check the compatibility of such reprocessing systems/solutions first with the OEM of the reprocessing agent/sterilization process. Prior to using other methods, confirm the specific compatibility and sterilization claims of ANY sterilizing methods/processes with the sterilizer manufacturer and ensure manufacturer that they have performed microbiological validation studies that support their claims of achieving sterilization of device specific flexible endoscopes and endoscope components. CAUTION: Use appropriate chemical indicators and/or biological monitors as recommended by the manufacturer of the sterilizer. WARNING: In addition to efficacy and compatibility claims always check with the manufacturer of the sterilizer/disinfector to confirm their specific product claims including compatibility and confirm that they have test data supporting their claims and demonstrating that no harmful levels of any residues (active/ inert ingredients, their byproducts or derivatives of the processed devices) remain on any instrument or lumened surfaces that may pose a risk to patients and users. – 44 – 2-2. Endoscopic Accessory Instruments (EAIs) and Endoscope Components WARNING: Current infection control guidelines require that biopsy forceps and similar endoscopic accessory instruments which enter sterile tissue or the vascular system or break the mucosal barrier must be sterilized before each patient use. For patient contact endoscopic accessories, follow the specific and detailed reprocessing instructions supplied with each product. 2-2-1. Cleaning CAUTION: Not all manufacturers of Automated Endoscope Reprocessors(AERs)/Washer-Disinfectors (WDs) make specific claims nor provide special instructions for reprocessing all of the removable endoscope components that are integral to the safe and effective operation of flexible endoscopes. Therefore, should the AER/WD manufacturer’s instructions not specifically address reprocessing of any particular endoscope component (air/water feeding valve, suction control valve, inlet seal, irrigation tube, check valve, selector mechanism, etc.) in the AER/WD, then those components must be reprocessed manually as described in PENTAX instructions/labeling. Prior to use, check with each AER/WD manufacturer as to their specific claims with respect to reprocessing individual endoscope components. CAUTION: NEVER use ultrasonic cleaning methods with ultrasound on the endoscope itself. 1) 2) 3) 4) Reusable endoscopic accessory instruments such as biopsy forceps and removable endoscope components such as suction valve should be cleaned immediately (at least within one hour) after each use since dried blood, mucous, or other debris may cause damage to the instrument and render the mechanism inoperable, or may interfere with the ability of the user to reprocess the device or component. Place the EAIs and/or components in a basin with fresh cleaning detergent solution for the time period, concentration, and temperature recommended by the manufacturer of the detergent being careful not to tightly coil or kink the flexible shaft. Clean the handle and flexible shaft by gently wiping with a soft gauze or the like. The biopsy cups and pivot pin area should be carefully and gently cleaned with a soft brush. Removal components such as air/water, suction control valve should be manipulated and detergent injected directly into/ onto components surfaces and then brushed clean. Rinse all residual detergent from the forceps by immersing the entire forceps under clean water and manipulating the handle and cup mechanism. – 45 – 5) Ultrasonic cleaning of forceps and similar accessories is then recommended, provided the manufacturer’s instructions and the parameters below are followed: Heavily soiled and/or difficult-to-manually clean components such as valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior to subsequent high-level disinfection or sterilization. Frequency Range 44 kHz ± 6% Time 5 minutes CAUTION: DO NOT use caustic or abrasive solutions in the ultrasonic cleaner. NOTE: All detergent must be removed from the inner mechanism of the forceps and individual endoscope components. Any detergent that remains after the water evaporates may cause increased friction that may render the mechanism inoperable. Residual detergent may also interfere in the subsequent biocidal process. 6) After cleaning and thorough rinsing, the endoscopic accessory instruments and components should be gently dried using a soft gauze or the like. Avoid tight coiling or kinking and do NOT put tension on the flexible shaft of the forceps and similar endoscopic accessory instruments. NOTE: Other PENTAX reusable accessories (channel cleaning adapters, cleaning brushes, bite block, etc.) and endoscope components (inlet seals, air/water and suction control valves, etc.) not specifically identified previously should be cleaned in a similar manner as above. Ultrasonic cleaning methods are recommended for accessories and endoscope components whose entire surfaces are not easily accessible by manual cleaning. Heavily soiled components such as valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior to subsequent high-level disinfection or sterilization. For these specific items, please refer to their separate operation instructions for uses or instruction sheets. – 46 – NOTE: It is imperative that ultrasonic cleaning of the Endoscopic Accessory Instruments(EAIs) and other components be performed PRIOR to steam sterilization. Only those PENTAX accessories identified by their pink colored handle, labeled as being autoclavable or identified below may be subjected to steam sterilization. • • • • • • • • • • • • • • • • • • PENTAX biopsy forceps PENTAX bite block OF-Z5 PENTAX water bottle OS-H4/OS-H5 PENTAX suction control valve OF-B120 PENTAX A/W feeding valve OF-B188 PENTAX inlet seal OF-B190 PENTAX cleaning Adapter OF-G17 PENTAX cylinder cleaning Brush CS-C9S PENTAX channel cleaning brushes CS6030SN PENTAX cleaning adapter OF-B153 PENTAX Gas/Water Feeding Valve OF-B194 *PENTAX suction channel selector OF-B161 *PENTAX instrument channel cleaning adapter OF-B162 **PENTAX irrigation tube OF-B113 **PENTAX water jet check valve adapeter OE-C12 **PENTAX check-valve OE-C14 **PENTAX water jet connector cap OF-B118 ***PENTAX distal rubber hood OE-A58/OE-A59 *two-channel endoscopes only **endoscopes with a water jet system only ***for Zoom endoscopes only Also, please refer to the instruction for use supplied with each accessory. 2-2-2. High-Level Disinfection Before any attempt is made to disinfect endoscopic accessory instruments and/or scope components such as bite block, air/water and suction control valve, brushes, etc. the complete cleaning procedure as described elsewhere in this instruction for use, must have been completed. Heavily soiled components such as valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior to subsequent high-level disinfection. 1) 2) 3) The entire accessory or component should be immersed in disinfecting solution. Accessory and component surfaces should remain in contact with the disinfecting solution for the time period recommended by the manufacturer of the solution. To ensure better contact, manipulate components such as valves while injecting disinfectant into/onto components surfaces, including all lumens. After the item has been in contact with the disinfecting solution for the appropriate amount of time, remove it from the disinfecting solution. – 47 – 4) 5) Rinse all residual disinfecting solution from the accessory/component by immersing it under sterile water. After thoroughly rinsing, the items should be gently dried using a soft gauze or the like. Compressed air may also be used to facilitate drying. NOTE: Ideally, all final rinses should be made with sterile water or bacteria-free water whose microbial quality has been confirmed via monitoring. After water rinsing, 70-90% medical grade ethyl or isopropyl alcohol should be flushed through lumens of the endoscopic instruments, as well as any removable scope components including valve mechanisms, followed by compressed air, not greater than 165kPa (24psi) to facilitate drying. External instrument and component surfaces can be dried by gently wiping with a sterile gauze or lint-free cloth saturated with alcohol. Regardless of the quality of the rinse water used, a dry instrument accomplished by a final alcohol rinse followed by forced air is essential to prevent bacteria colonization and/or infections associated with waterborne microorganisms. Such infections are more likely to occur when wet/contaminated instruments are used on patients whose immune systems are compromised or suppressed or when these devices are used in anatomical areas considered sterile and/or susceptible to these organisms. WARNING: The addition of defoaming agents to the water supply and/or automated reprocessing system is NOT recommended. Due to their nature, these silicone based agents cling tenaciously to surfaces. Unless they are rinsed very thoroughly, a “barrier” which could reduce the effectiveness of the disinfection/ sterilization process could be created. Additionally, repeated use of such defoamers could eventually lead to residual silicone build up resulting in equipment malfunction such as clogged air and water feeding channels. Similarly, silicone residues can deposit a “film” onto the distal objective lens causing “blurry” endoscopic images. – 48 – 2-2-3. Sterilization WARNING: Current infection control guidelines require that biopsy forceps and similar accessories which enter sterile tissue or vascular system or break the mucosal barrier must be sterilized before each patient use. For patient contact endoscopic accessories, follow the specific and detailed reprocessing instructions supplied with each product. Before any attempt is made to sterilize the accessories and/or individual endoscope components, the complete cleaning procedures as described elsewhere in this instruction for use must have been completed. Heavily soiled components such as valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior to subsequent sterilization. CAUTION: Use only the type of packaging material and package configuration as recommended by the manufacturer of the sterilizer. Use appropriate chemical indicators and/or biological monitors as recommended by the manufacturer of the sterilizer. NOTE: The following sterilization parameters are only valid with sterilization equipment that is properly maintained and calibrated. • Steam Sterilization (Autoclaving) Recommended NOTE: The PENTAX accessories and scope components listed on page 47 may be subjected to Steam Autoclaving. 1) 2) Prior to steam sterilization, all autoclavable endoscopic accessory instruments and endoscopic components previously identified should be thoroughly cleaned using manual and ultrasonic cleaning methods as described elsewhere in this instruction for use. Autoclaving can then be performed under the following conditions: For USA Sterilizer Type Prevacuum Temperature Exposure Time Drying Time 132°C 4 minutes 30 minutes 135°C 3 minutes 16 minutes NOTE: (For USA) Validation testing of these parameters was performed using Kimbely Clark Wrap KC-600. – 49 – For countries except USA Item Condition Sterilizer Type Prevacuum Exposure Condition 134°C - 137 °C x 3-18 minutes CAUTION: NEVER place the endoscope in a steam autoclave nor subject it to high-frequency ultrasonic cleaning methods! • ETO GAS sterilization 1) ETO gas sterilization can be performed on these accessories and/or components, provided they have first been properly cleaned and thoroughly dried. CAUTION: PENTAX Distal Rubber Hood OE-A58/A59 can’t be ETO sterilized. 2) Following ETO gas sterilization, aeration is required (12 Hours at 55). NOTE: For ethylene oxide sterilization of PENTAX accessories and endoscope components, follow the same ETO parameters as for PENTAX endoscopes (see page 47). – 50 – 2-3. Water Bottle Assembly Each water bottle assembly should be cleaned and sterilized at least on a daily basis or more frequently depending upon the patient and/or type of endoscopic procedure. As with all endoscopic accessories, prior to sterilization, water bottle assemblies must be scrupulously cleaned. Failure to do so could result in incomplete or ineffective sterilization. NOTE: Be careful when handling the water bottle assembly. DO NOT carry the bottle by the A/W Connector or the hose. When the bottle cap has been removed from the bottle, be careful in handling the water feeding stem. I. Cleaning of Water Bottle Assembly Reprocessing At-A-Glance Autoclave Manual Pre-Cleaning 1. 2. 3. Ultrasonic Cleaning Rinse and Dry Steam Sterilize Cooling After use, the entire bottle assembly (bottle, cap and tubing) should be washed with fresh detergent solution and a dampened gauze or scrub brush. Complete immersion in a cleaning detergent solution should be performed for soiled items to make contact with all surfaces of the water bottle assembly including the inner surface of the tubing system. Ultrasonic cleaning of the entire water bottle assembly is then recommended to access difficult to reach areas, provided the operating frequency is 44 kHz ± 6% for a period of at least 5 minutes. After washing with a cleaning solution, all surfaces of the water bottle assembly should be thoroughly rinsed with clean water and dried. Gauze or lint-free cloth can be used to wipe dry most surfaces. Compressed air and 70-90% medical grade ethyl or isopropyl alcohol should be used to facilitate drying of hard to reach areas. CAUTION: To avoid disconnection and/or bursting of the internal tubing, always set A/W Drain lever to the Air/Water position (upright) and use less than 165 kPa (24psi) air pressure during forced air drying. II. Sterilization of Water Bottle Assembly Before any attempt is made to sterilize the water bottle assembly, the complete cleaning procedure described above must have been completed and all items are thoroughly dry. Failure to scrupulously clean the water bottle assembly prior to the sterilizing process could result in incomplete or ineffective sterilization. – 51 – CAUTION: Use only the type of packaging material and package configuration as recommended by the manufacturer of the sterilizer. Use appropriate chemical indicators and/or biological monitors as recommended by the manufacturer of the sterilizer. NOTE: The following sterilization parameters are only valid with sterilization equipment that is properly maintained and calibrated. Steam Sterilization (Autoclaving) PENTAX water bottle assembly, Model OS-H4/OS-H5 has been designed to withstand high-pressure steam sterilization procedures providing the following parameters are adhered to: For USA Sterilizer Type Prevacuum Temperature Exposure Time Drying Time 132°C 4 minutes 30 minutes 135°C 3 minutes 16 minutes NOTE: Validation testing of these parameters was performed using Kimberly Clark Wrap KC-600. For countries except USA Item Condition Sterilizer Type Prevacuum Exposure Condition 134°C - 137°C x 3-18 minutes During steam sterilization, ensure that the cap and tubing section have been removed from the water bottle container. Make sure that the drain lever on the water bottle cap has been set on the A/W position (upright). – 52 – 3. POST REPROCESSING 1) Following reprocessing, the endoscope may either be reused or placed in storage. WARNING: If the endoscope is to be stored after reprocessing, detach removable valves, components, etc. All channels should be completely dry before storage. NOTE: When utilizing chemo-thermal processes for reprocessing PENTAX endoscopes, the instruments should be allowed to return to room temperature prior to use and/or further handling. 2) Prior to reuse, ensure that instrument has been properly inspected and fully prepared for the next clinical procedure. NOTE: Refer to the companion PENTAX Instructions for Use(Operation) describing in detail the recommended instructions on the inspection, use and operation of these endoscopes. 3) 4) 5) Prior to storage, ensure that all internal channels, endoscope components, instrument surfaces and accessories are thoroughly dry. A cotton tipped applicator moistened with 70-90% medical grade ethyl or isopropyl alcohol may be used to carefully remove any films or residues left upon the lens surfaces, such as the distal objective lens. The endoscope should be hung in a clean, dry, well-ventilated storage cabinet at room temperature. The insertion tube and light guide cable should be hung and kept as straight as possible during storage. CAUTION: • Make sure that all removable components such as the air/water feeding valve, suction control valve, water jet components, inlet seal are detached from the endoscope. This will allow for better air circulation throughout the internal channels to ensure thorough drying. • NEVER store the endoscope, its components and accessories in the carrying case, as this type of dark, humid and unventilated environment is conducive to bacteria colonization, which increases the risk of cross contamination. These cases are intended for transportation of the instrument, not storage. • NEVER store the endoscope in areas of high humidity, high temperatures or in direct exposure to sunlight or X-rays. • Avoid storage of the endoscope in cabinets, which have any sharp edges, exposed nails/screws, etc. Contact with sharp objects can puncture, scratch or otherwise damage the endoscope. – 53 – 3-1. Servicing Prior to returning any instrument for repair to PENTAX, the instrument should first undergo appropriate reprocessing/decontamination procedures for the purpose of infection control. Check with your local PENTAX service facility for more details. 1) 2) 3) 4) 5) 6) All instruments requiring repair should be shipped in the original carrying case with appropriate packing along with comments describing the instrument damage and complaint. A repair purchase order number, contact name and phone number of the individual responsible for authorizing repairs, as well as shipping address should be included. The “Red” ventilation cap should be attached to the instrument if it will be shipped by air freight. Any accessories and/or endoscope components potentially related to the endoscope damage or complaint should also be returned with the endoscope. PVE soaking caps should also be returned with the endoscope to check/ confirm the integrity of their watertight seal. After servicing, all endoscopes must be reprocessed prior to patient use. – 54 – WARNING: Instrument repairs should only be performed by an authorized PENTAX service facility. PENTAX assumes no liability for any patient/user injury, instrument damage or malfunction, or REPROCESSING FAILURE due to repairs made by unauthorized personnel. Your local PENTAX distributor can provide a list of “compatible” reprocessing agents with PENTAX endoscopes based upon material compatibility and functionality studies performed by PENTAX, Japan. These tests of course apply only to genuine PENTAX parts, components and materials including proprietary adhesives, sealants, lubricants, etc. specifically selected for use in PENTAX endoscopes to satisfy their original design criteria. PENTAX manual reprocessing instructions supplied with each product have been validated for PENTAX endoscopes utilizing exclusive PENTAX parts/materials and assembled based upon proprietary PENTAX manufacturing technologies and/or servicing techniques. It must be recognized that PENTAX does not evaluate nonPENTAX parts, components, materials and/or servicing methods and therefore questions regarding material compatibility and/or functionality of PENTAX instruments built with these unauthorized, untested and unapproved items, materials, repair/assembly methods must be referred to the third party service organization and/or device remanufacturer. It is unknown to PENTAX if serviced or remanufactured instruments (performed by unauthorized PENTAX entities) which still bear a PENTAX label are within PENTAX device specifications and/or if unauthorized activities have significantly changed the instrument’s performance, intended use, safety and/or effectiveness. These companies should confirm the ability for these serviced/ remanufactured devices to be reprocessed safely and effectively with reprocessing agents/systems recognized as compatible by PENTAX for standard PENTAX products. These third party companies and/or remanufacturers should be consulted to confirm if they have performed reprocessing validation studies on instrument models which they have serviced (or remanufactured) that support the cleaning, high-level disinfection and/or sterilization of these endoscopes via the normal endoscope OEM reprocessing recommendations, standard AER device-specific instructions and/or their own unique reprocessing recommendations. Ultimately, owners of these medical devices are responsible for selecting an appropriate service facility or vendor whose activities render an instrument to the same expectations and quality of a finished device supplied by the endoscope OEM. 7) For disposal of instruments, follow local or country regulations. – 55 – 3-2. Care and Maintenance Tips Flexible endoscopes have been an invaluable tool in the medical community’s armamentarium to successfully diagnose and treat a wide variety of illnesses in patients for several decades. Perhaps due to their longevity and progressive design changes over the years which have simplified their use, flexible endoscopes have been somewhat taken for granted and have erroneously not been considered highly technological medical devices. In fact, current generation flexible endoscopes although easier to clinically use, are much more sophisticated than ever. Special reprocessing instructions must be followed to ensure the instruments are patient ready and patient safe. Special care and handling must be exercised and practiced to prevent instrument malfunction and prolong the reliability of the endoscope. The burden of responsibility to ensure safe and reliably functioning instruments is left in the hands of the healthcare professionals who actually care for and reprocess flexible endoscopes. Naturally, equipment manufacturers share in this responsibility and tremendous efforts have been made in designing instruments which could be reprocessed and maintained as easy as possible. However, due to the nature of their use and application, flexible endoscopes must be subjected to special cleaning procedures, followed by a disinfection or sterilization process after each and every patient use. To highlight and simplify, what may appear to some as being complicated maintenance and reprocessing instructions, PENTAX strongly recommends the users review the following suggestions and advice on the care and maintenance of your PENTAX flexible endoscopes. These tips, particularly those involving endoscope reprocessing should not be construed as “shortcuts” and are not intended as substitute directions for complete instructions found elsewhere in the instruction for use. Avoid soaking of the endoscope with accessories (forceps, injection or aspiration needles, etc.) or any sharp edged objects which could inadvertently scratch or cut the distal bending section sheath. (Subsequent flexing back and forth of the rubber sheath could eventually stretch the scratched rubber until a pinhole and leak develops.) Exposure to a compatible cleaning detergent is essential to thorough cleaning of all surfaces of the endoscope. Rinsing and drying after cleaning is imperative to prevent dilution and/or adulteration of the disinfectant/ sterilant. – 56 – Not all manufacturers of automated endoscope reprocessors (AERs) make specific claims nor provide special instructions for reprocessing all of the removable scope components that are integral to the safe and effective operation of flexible endoscopes. Therefore, should the AER manufacturer’s instructions not specifcally address reprocessing of any particular scope component (air/water feeding valve, suction valve, inlet seal, irrigation tube, check valve, etc.) in the AER, then those components must be reprocessed manually as described in PENTAX instructions/labeling. Prior to use, check with each AER manufacturer as to their specific claims with respect to reprocessing individual endoscope components. Do NOT reuse disposable endoscope accessory instruments (EAIs) or endoscope component intended for single patient or one time use. Do NOT expose the endoscope or Endoscope Accessory Instruments (EAIs) or endoscope component to harsh chemical solutions. Strictly adhere to exposure times recommended by the manufacturers of compatible solutions. Avoid contact of any flexible portion of the endoscope with any sharp edge objects (bed frames, table top corners, sink drains, accessories hanging in storage cabinets, etc.) at any time during the handling, reprocessing or storage of the endoscope. Avoid stretching of the bending section rubber sheath at the distal portion of the endoscope. During mechanical cleaning with a dampened gauze, do NOT use excessive force. A gentle back and forth wiping motion should be sufficient to remove gross debris. Subsequent soaking in a cleaning detergent will clean the remainder of debris.+ Disinfectants and sterilants are toxic substances by nature. All residual solution must be thoroughly rinsed and dried prior to each patient use. The key to preventing clogged air, water or water jet channels/nozzles is to immediately flush the channels with either air pressure or fluid/detergent right after removal from the patient. This should be followed by brushing of each accessible channel and then soaking with a cleaning detergent. Avoid attempting to remove or unscrew endoscope components which should not be removed. Parts such as the distal portion of the light guide plug, any inlet ports and any strain relief boots on either the insertion tube or umbilical cable are essential to the watertight integrity of the instrument. Removal or loosening of these components and subsequent immersion could lead to fluid invasion into the endoscope and/or compromise effective reprocessing. Check for any sharp edges on all surfaces of an automated cleaning/ disinfecting unit which may come in contact with an endoscope. Some units may have sharp edged wire mesh filters and baskets or inlet/outlet ports which could damage your endoscope. When utilizing chemo-thermal processes for reprocessing PENTAX endoscopes, the instruments should be allowed to return to room temperature prior to use and/or further handling. – 57 – Do NOT overtighten the cap to the water bottle assembly. The metal pipe at the top of the PENTAX water bottle assembly functions as an inlet port for air from the light source. This inlet pipe should not be used as a leverage tool to tighten the cap to the water container. Overtightening could cause the plastic cap to crack. Make sure that the air/water/instrument channel cleaning adapter is securely attached to the top of the air/water and suction cylinders. Do NOT introduce air bubbles into the endoscope’s internal channels during flushing of cleaning and/or disinfecting/sterilizing solutions as these bubbles could interfere in the effectiveness of the disinfection/sterilization process. Do NOT store the endoscope, its components or accessories in the carrying case as this type of dark, humid and unventilated environment is conducive to bacteria colonization which increases the risk of cross-contamination. Prior to each use, check the condition of all endoscope accessory instruments. Do NOT use any accessories with kinked or bent flexible shafts. Do NOT use forceps with misaligned cups and/or bent needles/spikes. Do NOT use aspiration or injection needles which are not retractable or whose sharp tips can not be protected. Do NOT use cleaning brushes without smooth or rounded distal tips. Do NOT use instruments with exceptionally long distal rigid sections or whose outer diameter restricts passage through the instrument channel/ channel inlet. Use of any of the above accessories could result in channel damage and costly repairs. Verification of the Minimum Effective Concentration (MEC) of active ingredients (via test strips or similar methods) is recommended to ensure potency of the liquid chemical germicide to achieve high-level disinfection/ sterilization. WARNING: Instrument repairs should only be performed by an authorized PENTAX service facility. PENTAX assumes no liability for any patient/user injury, instrument damage or malfunctions, or REPROCESSING FAILURE due to repairs made by unauthorized personnel. WARNING: NEVER drop this equipment or subject it to severe impact as it could compromise the functionality and/or safety of the unit. Should this equipment be mishandled or dropped, do not use it. Return it to an authorized PENTAX service facility for inspection or repair. – 58 – 4. NOMENCLATURE AND FUNCTION 4-1. Video Endoscope AIR/WATER FEEDING VALVE (OF-B188) Covering of hole in the top of the valve delivers pressurized air. Covering of the hole and fully depressing the valve delivers pressurized water. INLET SEAL Allows passage of accessories while preventing escape of fluids and air. SUCTION CONTROL VALVE (OF-B120) Depress to remove fluids or air through the instrument channel. INSTRUMENT CHANNEL INLET For introduction of biopsy forceps and other accessories. BUTTON 1 Push to freeze an image. BUTTON 2 Push to activate the hardcopy system that was selected between “FILE” and “HARD COPY”. CONTROL BODY BUTTON 1 BUTTON 2 BUTTON 4 MODEL DESIGNATION BUTTON 3 RIGHT/LEFT ANGULATION CONTROL KNOB UP/DOWN ANGULATION LOCK LEVER When this lever is in the “F” position, turned clockwise, the bending section moves freely. When turned fully counterclockwise, the bending section becomes progressively more stabilized. UP/DOWN ANGULATION CONTROL KNOB RIGHT/LEFT ANGULATION LOCK KNOB Functions similar to Up/Down lock ELEVATOR MECHANISM CONTROL KNOB (DUODENOSCOPES ONLY) BUTTON 4 Turn clockwise to activate magnification (EG-2990i/2990Zi, EC-3490i/3490Ti/3890i/3890Zi Series). To guide and direct cannula or forceps. BUTTON 3 Push to activate the VCR for recording live procedures or to enhance the endoscopic image (EPK-1000). STRAIN RELIEF BOOT UMBILICAL CORD NOTE: The function of each button can be changed. For more details, refer to the manual supplied with the EPK-i series or the EPK-1000. EG-1690K/2490K/2790K/2990K/3490K, EG-3890TK, ED-3490TK, EC-2990i/3490Ti/3490K/3890K /3890TK series Scope EG-2790i VSB-2990i EG-2990i/2990Zi EC-3490i/3890i/ 3890Zi series Processor EPK-i series EPK-i series EPK-1000 Button 1 Freeze Freeze Freeze Freeze Button 2 Hard Copy Hard Copy Hard Copy Hard Copy Button 3 VCR VCR VCR (for NTSC) Enhance (for PAL) VCR Button 4 Enhance — Enhance (for NTSC) AVE/PEAK (for PAL) Enhance Magnification Control Lever — Magnification — — – 59 – EPK-i series NOTE: To avoid damaging the endoscope, do NOT twist, rotate or bend excessively any of the strain reliefs. DISTAL END (Refer to the inside rear cover of this instruction for use) STRAIN RELIEF BOOT BENDING SECTION INSERTION TUBE WATER JET PORT Allows connection of special irrigation tube (OF-B113) for pressurized source of a spray directed at the endoscopically visualized surface. AIR/WATER PORT To connect feeding tube from water bottle assembly. VENTING CONNECTOR Accepts “RED” Ventilation cap. Also accepts Leakage Tester. VENTILATION CAP OF-C5 RED Provides venting of endoscope interior to equalize internal and external pressures. This cap must be removed before immersion. NOTE: See important separate section regarding the use of this cap! PVE CONNECTOR Can be rotated within a 180 range. ELECTRICAL CONTACTS PVE SOAKING CAP OE-C9 This cap must be securely attached before immersion. Align the black arrow on the soaking cap with the green dot at the base of the silver collar surrounding the electrical contacts on the PENTAX PVE connector. Press the cap down onto the metal collar and turn clockwise to secure. FEED BACK TERMINAL (Not available in EG-1690K) To connect the OL-Z3 cable from the PENTAX video processor, model EPK-i series, EPK-1000. SUCTION NIPPLE For attachment to external suction source. LIGHT GUIDE PLUG Transmits light from light source to distal end of endoscope. NOTE: Ensure that the PVE soaking cap has been securely attached (by properly rotating it) to prevent the cap from coming off during reprocessing. Failure to securely attach the PVE soaking cap can result in scope damage. WARNING Immediately after use, the metal light guide plug and the electrical contacts/pins of the endoscope may be HOT. To avoid burns, do not touch these areas immediately after use. For safer handling after a procedure, grasp the PVE connector housing. – 60 – HOYA Corporation 2-7-5 Naka-Ochiai, Shinjuku-ku, Tokyo 161-8525 Japan PENTAX Medical (Shanghai) Co., Ltd. Room 701, 291 Fumin Road, Shanghai, China 200031 Tel: +86-21-61701555 Fax: +86-21-61701655 PENTAX Medical Singapore Pte. Ltd. 438A Alexandra Road, #08-06 Alexandra Technopark Singapore 119967 Tel: +65-6271-1669 Fax: +65-6271-1691 PENTAX Medical Company A Division of PENTAX of America, inc. 3 Paragon Drive Montvale, New Jersey 07645-1782 USA Tel: +1-201-571-2300 Toll Free: +1-800 431-5880 Fax: +1-201-391-4189 PENTAX Canada Inc. 1770 Argentia Road Mississauga, Ontario L5N 3S7 Canada Tel: +1-905-286-5585 Fax: +1-905-286-5571 PENTAX Europe GmbH Julius Vosseler Strasse 104, 22527 Hamburg, Germany Tel: +49-40-561-920 Fax: +49-40-560-4213 EC REP PENTAX U.K. Ltd. PENTAX House, Heron Drive, Langley SLOUGH SL3 8PN, Great Britain Tel: +44-1753-792792 Fax: +44-1753-792794 Manufacturing Site HOYA Corporation, PENTAX Miyagi Factory 30-2 Okada, Aza-Shimomiyano, Tsukidate, Kurihara-shi, Miyagi 987-2203 Japan Specifications are subject to change without notice and without any obligation on the part of the manufacturer. 88151 2014. 03. 6217001 Z863 R16 printed in JAPAN