Triton TR8 Instruction Manual

Transcription

Triton TR8 Instruction Manual
ELECTRO-CHEMICAL DEVICES, INC.
Triton TR8 Instruction Manual
Turbidity Analyzer/Controller
ELECTRO-CHEMICAL DEVICES, INC.
1681 Kettering, Irvine, CA 92614, USA
Tel: +1-949-336-6060, FAX: +1-949-336-6064
www.ecdi.com
Rev: B - 08/09
PREFACE
Purchasing products from Electro-Chemical Devices, Inc. provides you with the finest liquid analytical
instrumentation available. If this is your first purchase from ECD, please read this manual before
installing and commissioning your new equipment.
If there are any questions concerning this equipment, please contact your local ECD representative, or
the factory directly at:
Electro-Chemical Devices, Inc.
1681 Kettering St.
Irvine, CA 92614 USA
Telephone: +1-949-336-6060
FAX: +1-949-336-6064
Website: www.ecdi.com
Email: sales@ecdi.com
© 2009 Electro-Chemical Devices, Inc. All rights reserved. No part of this manual may be used or reproduced in any form or by any means, or
stored in a database or retrieval system without prior written permission from Electro-Chemical Devices, Inc. Making copies of any part of this
manual for any purpose other than personal use is a violation of United States copyright laws. Document printed in the United States of
America.
Page ii
Triton TR8
TABLE OF CONTENTS
PREFACE ........................................................................................................................................................ ii
TABLE OF CONTENTS.................................................................................................................................... iii
WARRANTY .................................................................................................................................................. vi
IMPORTANT SERVICE INFORMATION .......................................................................................................... vi
UNPACKING THE INSTRUMENT................................................................................................................... vii
1.0 GENERAL DESCRIPTION ........................................................................................................................... 1
1.1 TR8 SENSOR FEATURES ....................................................................................................................... 1
1.1.1 Sensor Options ............................................................................................................................. 2
1.2 TR8 Analyzer Features ........................................................................................................................ 2
1.2.1 Analyzer Options .......................................................................................................................... 2
2.0 INSTALLATION ......................................................................................................................................... 3
2.1 ANALYZER MOUNTING........................................................................................................................ 3
2.1.1 Panel Mounting ............................................................................................................................ 3
2.1.2 Pipe Mounting.............................................................................................................................. 3
2.2 SENSOR MOUNTING ........................................................................................................................... 3
2.2.1 Immersion Assembly .................................................................................................................... 4
2.2.2 Flow-Through Assembly ............................................................................................................... 5
2.2.3 De-Bubbler Flow Assembly .......................................................................................................... 6
2.3 WIRING ................................................................................................................................................ 7
2.3.1 Sensor Wiring ............................................................................................................................... 7
2.3.2 Instrument Wiring ........................................................................................................................ 7
3.0 OPERATION ............................................................................................................................................. 8
3.0.1 General Information .................................................................................................................... 8
3.0.2 Liquid Crystal Display ................................................................................................................... 9
3.1 KEY Assignment ................................................................................................................................. 10
3.1.1 Auto/Manual Mode Switching ................................................................................................... 11
3.2 MENUS .............................................................................................................................................. 11
3.2.1 Access Codes .............................................................................................................................. 12
3.2.2 Menu structure .......................................................................................................................... 13
3.2.3 HOLD Function ........................................................................................................................... 13
Triton TR8
Page iii
3.3 START UP and SETUP FUNCTION GROUP MENUS ............................................................................ 14
3.3.1 SETUP 1: TURBIDITY ................................................................................................................... 14
3.3.2 SETUP 2: TEMPERATURE and Wiper .......................................................................................... 15
3.3.3 CURRENT INPUT ......................................................................................................................... 16
3.3.4 CURRENT OUTPUT ..................................................................................................................... 17
3.3.5 ALARM FUNCTIONS .................................................................................................................... 20
3.3.6 CHECK ......................................................................................................................................... 21
3.3.7 RELAY ......................................................................................................................................... 23
3.3.8 CONCENTRATION MEASUREMENT ............................................................................................ 35
3.3.9 SERVICE ...................................................................................................................................... 36
3.3.10 INTERFACE................................................................................................................................ 38
3.4 CALIBRATION FUNCTION GROUP MENUS ........................................................................................ 38
3.4.1 CALIBRATION.............................................................................................................................. 41
3.4.2 OFFSET ....................................................................................................................................... 46
3.4.3 SLOPE ......................................................................................................................................... 46
4.0 MAINTENANCE ...................................................................................................................................... 48
4.1 CLEANING THE TRANSMITTER .......................................................................................................... 48
4.2 CHECKING THE MEASURING POINT .................................................................................................. 48
4.2.1 Transmitter ................................................................................................................................ 48
4.2.2 Triton TR8 Sensor ....................................................................................................................... 49
4.3 CLEANING THE SENSOR..................................................................................................................... 49
4.4 RECALIBRATION ................................................................................................................................ 49
5.0 ORDERING INFORMATION .................................................................................................................... 50
5.1 ACCESSORIES ..................................................................................................................................... 50
6.0 TROUBLESHOOTING .............................................................................................................................. 51
6.1 TROUBLESHOOTING INSTRUCTIONS ................................................................................................. 51
6.2 SYSTEM ERROR MESSAGES ............................................................................................................... 51
6.3 PROCESS SPECIFIC ERRORS ............................................................................................................... 54
6.4 INSTRUMENT SPECIFIC ERRORS ........................................................................................................ 57
6.5 SENSOR TROUBLESHOOTING ............................................................................................................ 58
6.5.1 Troubleshooting instructions ..................................................................................................... 58
6.5.2 Checking the Sensor ................................................................................................................... 59
Page iv
Triton TR8
7.0 ENGINEERING DOCUMENTATION ......................................................................................................... 59
7.1 SPECIFICATIONS ................................................................................................................................ 59
7.1.1Triton TR8 Sensor ........................................................................................................................ 59
7.1.2 Triton Analyzer Specifications .................................................................................................... 60
7.2 OUTLINE & DIMENSIONAL DRAWING ............................................................................................... 62
7.3 WIRING DIAGRAM............................................................................................................................. 62
Triton TR8
Page v
WARRANTY
Electro-Chemical Devices, Inc. (ECD) warrants all products it manufactures to be free from defect in materials and factory
workmanship, and agrees to repair or replace any product that fails to perform, as specified, within one (1) year after date of
shipment. This warranty shall not apply to any product that has been:
1.
2.
3.
Subjected to misuse, negligence or accident;
Connected, installed, adjusted or otherwise used not in accordance with the instructions furnished by ECD;
Repaired, modified or altered by persons not authorized by ECD, resulting in injury to the performance, stability or
reliability of the product.
This warranty is in lieu of any other warranty, expressed or implied. ECD reserves the right to make changes in the design or
construction of its products at any time, without prior notification, and without incurring any obligation to make any changes in
previously delivered products.
Seller’s sole liabilities and the buyer’s sole remedies under this agreement shall be limited to a refund in the purchase price, or
at ECD’s discretion, to the repair or replacement of any product that proves, upon ECD’s examination, to be defective, when
returned to the factory, transportation prepaid by the buyer, within one (1) year of the product’s original shipment date. Seller
shall not be liable for damages consequential or incidental to defects in any product, for failure of delivery in whole or in part,
for injuries resulting from its use, or for any other cause.
This warranty and the writing attached constitute the full understanding of seller and the buyer, and no terms, conditions,
understanding, or agreement purporting to modify or vary the terms hereof shall be binding unless hereafter made in writing
and signed by an authorized official of Electro-Chemical Devices, Inc.
This warranty does not cover pH, ORP or Specific Ion measurement, reference or combination electrodes or electrode
cartridges that have been commissioned in service.
IMPORTANT SERVICE INFORMATION
Use only factory authorized components for repair. Tampering or unauthorized substitution of components may
adversely affect the operation of this product and may void the warranty.
If service or repair is required, please obtain the serial number(s) or sales order number of the product(s) in
question and contact ECD’s Service Department at:
+1-800-729-1333 (USA/Canada) or +1-949-336-6060
or email Service@ecdi.com
A Return Material Authorization (RMA) number must be obtained from the service department before returning
any material to ECD. All material returned to ECD shall be shipped prepaid to the factory.
Page vi
Triton TR8
UNPACKING THE INSTRUMENT
Your Electro-Chemical Devices instrument has been carefully packaged to protect it from damage during
shipment and dry storage. Upon receipt please follow the procedure outlined below.
1. Before unpacking, inspect the condition of the shipping container to verify proper handling by
the carrier. If damage is noted, save the shipping container as proof of mishandling for the
carrier.
2. Check the contents of the shipping container with the items and quantities shown on the
packing list. Immediately report any discrepancies to ECD.
3. Save the original packing material until you are satisfied with the contents. In the event the
product(s) must be returned to ECD, the packing material will allow you to properly ship it to
ECD.
4. Familiarize yourself with the instrument before installation, and follow proper installation and
wiring procedures.
Triton TR8
Page vii
1.0 GENERAL DESCRIPTION
The Triton TR8 is a nephelometric turbidity analyzer designed for use in water and wastewater.
Turbidity, the cloudiness or haziness of a water sample, is caused by particles suspended in the
water, typically clay and silt. Since bacteria and viruses can be attached to these particles,
turbidity has become a critical indicator of the overall water quality.
The Triton TR8 uses an optical method for determining the turbidity, a light beam is directed
into the sample where it is scattered by suspended particles in the water. The amount of
scattering depends on the amount of material in the water, the wavelength of light used and
the size and composition of the suspended particles. The 90° scattered light detection method
used by the Triton TR8 sensor is the most common sensor design for turbidity.
The Triton TR8 uses a long lived near infrared LED light source (880 nm) and the 90° scattered
light method in accordance with ISO 7027 / EN 27027 to assure accurate turbidity values under
standardized and comparable conditions. The turbidity value is calculated inside the sensor and
digitally transmitted to the Triton TR8 analyzer. The analyzer interprets the signal, displays the
turbidity value and transmits the appropriate 4-20 mA output.
1.1 TR8 SENSOR FEATURES
The TR8 sensor has a high sensitivity at low levels of turbidity, a simple
optical configuration and a balanced sensitivity to all particle sizes. This
smart sensor is factory calibrated and ready to use when received. All of
the signal processing is performed in the sensor and a noise resistant
digital signal is sent to the analyzer/transmitter. A High Range version,
(-2), is available for applications with turbidity higher than 500 FNU.
Figure 1
Three detectors monitor the light beam at an angle of 90°. The first detector, inside the sensor,
(see Figure 1) is the reference detector that compensates for changes in the LED light source
(#1) caused by aging or other variations. The second detector (#2) measures a short path length
which is best for high concentration measurements. The third detector (#3) measures the
longer path length which is best for lower concentrations. The turbidity signal is constantly
adjusted versus the reference detector, digital filter functions suppress interfering signals and
self monitoring diagnostics assure a highly reliable measurement.
An optional mechanical wiper assembly is available to keep the scratch resistant sapphire
optical windows clean in dirty applications. The cleaning cycle and duration times are user
defined parameters.
Triton TR8
Page 1
1.1.1 Sensor Options
Part#
Description
Triton TR8 Turbidity Sensor with 7 meter cable
1398000-1
Triton TR8 Turbidity Sensor High Range with 7 meter cable
1398000-2
Triton TR8 Turbidity Sensor with 15 meter cable
1398001-1
Triton TR8 Turbidity Sensor High Range with 15 meter cable
1398001-2
Triton TR8 Turbidity Sensor with 7 meter cable & Automatic Wiper
1398010-1
Triton TR8 Turbidity Sensor High Range with 7 meter cable & Automatic Wiper
1398010-2
Triton TR8 Turbidity Sensor with 15 meter cable & Automatic Wiper
1398011-1
Triton TR8 Turbidity Sensor High Range with 15 meter cable & Automatic Wiper
1398011-2
Triton TR8 Turbidity Sensor with 7 meter cable & Flow Through Assembly
1398100-1
Triton TR8 Turbidity Sensor with 15 meter cable & Flow Through Assembly
1398101-1
Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & Flow Through Assembly
1398110-1
Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & Flow Through Assembly
1398111-1
Triton TR8 Turbidity Sensor with 7 meter cable & De-Bubbler Assembly
1398200-1
Triton TR8 Turbidity Sensor with 15 meter cable & De-Bubbler Assembly
1398201-1
Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & De-Bubbler Assembly
1398210-1
Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & De-Bubbler Assembly
1398211-1
(-1) Low Range Sensor < 500 FNU, (-2) High Range Sensor > 500 FNU
1.2 TR8 Analyzer Features
The Triton TR8 Analyzer/Transmitter is a microprocessor based, menu driven, turbidity/suspended solids
transmitter. The NEMA 4X TR8 Analyzer can be Panel mounted, Wall mounted or Pipe mounted. The
base unit is available as a 110 VAC, 220 VAC or 24 VDC powered instrument. The standard configuration
has a 250VAC/30VDC 2A contact relay that can be used as an alarm. The single 4-20 mA output is
configured for 0-10 FNU, although it can be easily configured for NTU, ppm, mg/l or g/l. Optional
configurations with (2) 4-20 mA and/or (4) relays/contacts are available upon request to the factory.
1.2.1 Analyzer Options
Part#
1290100-1
1290100-2
1290100-3
1290100-*
Page 2
Description
Triton TR8 Turbidity Analyzer, 115 VAC, (1) 0/4-20 mA output, (1) Failure Alarm Relay*
Triton TR8 Turbidity Analyzer, 230 VAC, (1) 0/4-20 mA output, (1) Failure Alarm Relay*
Triton TR8 Turbidity Analyzer, 24 VDC, (1) 0/4-20 mA output, (1) Failure Alarm Relay*
Consult factory for optional dual 0/4-20 mA outputs and multiple relays up to (4) additional relays
Triton TR8
2.0 INSTALLATION
Mount the TR8 in a location where there is easy access to the analyzer and sensors. Install the system in
an area where vibrations, electromagnetic and radio frequency interference are minimized or absent.
Do not mount in direct sunlight or areas of extreme heat. The TR8 is suitable for outdoor use if mounted
with a protective cover or sunshield.
2.1 ANALYZER MOUNTING
The Universal Mounting allows the Analyzer to be Wall mounted, Panel Mounted or Pipe/Rail Mounted.
The drawings below show the Universal Mounting used with the C-22 analyzer for panel mounting and
the Universal Mounting with the T-23 transmitter that uses the same mounting as the C-22 for Pipe/Rail
Mounting. The Triton TR8 Analyzer has the same dimensions as the Panel Mounted C-22 Analyzer.
2.1.1 Panel Mounting
Figure 2
2.1.2 Pipe Mounting
Figure 3
2.2 SENSOR MOUNTING
The Triton TR8 turbidity sensor can be installed as an immersion sensor into a tank with the optional
Immersion assembly PN 1000223, or in-line with a Flow Through fitting, PN 1000219, or the special Debubbler Flow through fitting designed especially for drinking water applications.
Care must be taken when installing the sensor in a pipe or near a tank wall to avoid the back scattering
of light which leads to a higher sensor response than expected. The lower the turbidity of the measured
Triton TR8
Page 3
solution is the greater the possibility of back scattering. The following graph can be used as an
installation guideline.
Figure 4
2.2.1 Immersion Assembly
The immersion assembly consists of a 1 meter down tube, ¾” FNPT on the process end and a cap with
cable feed through on the other. Verify that the sensor is at least 6” from the tank bottom and side
walls. The cable is not water proof and under no circumstances should the sensor be suspended in the
application by the cable.
Figure 5
Page 4
Triton TR8
2.2.2 Flow-Through Assembly
Conventional turbidity measurements are carried out in an unpressurized sample. When the pressure on
a sample (which was pressurized beforehand) is released, fine bubbles are produced which distort the
turbidity measurement. The pressure must remain constant or rise as the water moves from a pipe and
into the flow cell in order to minimize the formation of air bubbles. The outfall from the flow cell must
not create a low pressure in the flow cell by siphoning the sample to drain, see examples below.
If air bubbles cannot be eliminated then the mechanical wiper assembly can help remove the bubbles
that form on the measurement windows, otherwise use the De-Bubbler Assembly below.
Figure 6: Proper Installation examples to avoid bubble formation.
Figure 7: Sensor alignment in flow cell with optional spray cleaner and without.
Triton TR8
Page 5
Figure 8: Dimensional Drawing of the Flow-Through Assembly with optional union adapters
2.2.3 De-Bubbler Flow Assembly
When air bubbles are present in the sample the De-Bubbler Flow assembly effectively removes them
from the sample. The lower half of the incoming sample flow is diverted down the central tube (1) to the
“Ring Channel” where air bubbles can escape through the outlets, (2), upwards to the exiting sample
flow. Bubble free water is pushed through the tubes, (3), into the measuring Chamber, (5). The design
facilitates a high flow rate by the TR8 sensor, (6), that optimizes response time and minimizes
sedimentation forming in the chamber. A large drain plug, (4), aids in flushing any sediment that does
deposit in the chamber.
Figure 9: De-Bubbler Flow Cell Diagram
Page 6
Figure 10: Dimensional Drawing, De-Bubbler Assembly
Triton TR8
2.3 WIRING
The various cables, measurement signal, outputs and power, enter the Triton TR8 instrument through
the three cable glands on the bottom of the instrument. Only properly trained personnel should attempt
to wire the instrument.
2.3.1 Sensor Wiring
The Triton TR8 sensor uses a digital communication protocol that allows the sensor to be placed up to
200 meters from the analyzer. The TR8 Sensor is supplied with either a 7 meter or 15 meter cable.
Installation at distances greater than 15 meters requires the use of an extension cable, PN
9640004.cond and Junction Box, PN 1000222.
Figure 11: Sensor wiring Color Code
2.3.2 Instrument Wiring
Optional Relay Card
(4-20 mA)
Figure 12: Terminal Designations, shown with optional Relay Card
Triton TR8
Page 7
3.0 OPERATION
3.0.1 General Information
Figure 13: Operator Interface
1. LC display for displaying the measured values and configuration data
2. Field for user labeling
3. 4 main operating keys for calibration and device configuration
4. Changeover switch for automatic/manual control mode of the relays
LED (Green) Automatic Mode
LED (Yellow) Manual Mode
5. LEDs for status of working relays (switch status)
LED (Green) indicates the measured value is within the allowed limits, relay inactive
LED (Red) indicates the measured value is outside the allowed limits, relay active
6. LED for alarm function
LED (Red) indicates Alarm condition, e.g. continuous limit violation, temperature sensor failure
or system error/failure
7. Display of the active contact and key for relay changeover in manual mode
LED (Red) indicates active relay
Page 8
Triton TR8
3.0.2 Liquid Crystal Display
Figure 14: Display Features
1. Indicator for measuring mode (normal operation)
2. Indicator for calibration mode
3. Indicator for setup mode (configuration)
4. Indicator for "Hold" mode (current outputs remain at last current state)
5. Indicator for receipt of a message for devices with communication, e.g. Hart, Profibus
6. Indicator of working status of relays 3 and 4: inactive, active
7. Function code display
8. In measuring mode: The measured variable
In setup mode: The configured variable
9. In measuring mode: The secondary measured value
In setup or calibration mode: e.g. setting value
10. Auto Temperature Compensation (ATC)
11. Manual Temperature Compensation (MTC)
12. "Error": error display
13. Sensor symbol
Triton TR8
Page 9
3.1 KEY Assignment
CAL key
When you press the CAL key, the device first prompts you for the calibration access code:
• Code 22 for calibration
• Code 0 or any other code for reading the last calibration data
Use the CAL key to accept the calibration data or to switch from field to field within the
calibration menu.
ENTER key
When you press the ENTER key, the device first prompts you for the setup mode access code:
• Code 22 for setup and configuration
• Code 0 or any other code for reading all configuration data.
The ENTER key has several functions:
• Calls up the Setup menu from the measuring mode.
• Saves (confirms) data entered in the setup mode.
• Moves on within function groups.
PLUS key and MINUS key
In the setup mode, the PLUS and MINUS keys have the following functions:
• Selection of function groups.
! Note!
Press the MINUS key to select the function groups in the order given in the "System
configuration" section.
• Configuration of parameters and numerical values
• Operation of the relay in manual mode
In the measuring mode, you get the following sequence of functions by repeatedly pressing
the PLUS key:
1. Temperature display in F
2. Temperature display hidden
3. Current input signal in %
4. Current input signal in mA
5. Measured value display in FNU or NTU (uncompensated value without reflection
compensation offset and slope, referred to data set 1)
In the measuring mode, the following is displayed in sequence by repeatedly pressing the
MINUS key:
1. Current errors are displayed in rotation (max. 10).
2. Once all the errors have been displayed, the standard measurement display appears. In
the function group F, an alarm can be defined separately for each error code.
REL key
In the manual mode, you can use the REL key to switch between the relay and the manual start
of cleaning.
In the automatic mode, you can use the REL key to read out the switch-on points (for limit
contactor) or set points (for PID controller) assigned to the relay in question.
Press the PLUS key to jump to the settings of the next relay. Use the REL key to get back to the
display mode (automatic return after 30 s).
AUTO key
You can use the AUTO key to switch between automatic mode and manual mode.
Escape function
If you press the PLUS and MINUS key simultaneously, you return to the main menu or are
taken to the end of calibration if calibrating. If you press the PLUS and MINUS key again, you
return to the measuring mode.
Page 10
Triton TR8
Locking the keyboard
Press the PLUS and ENTER key for at least 3 seconds to lock the keyboard against any
unauthorized data entry. All the settings can continue to be read.
The code prompt displays the code 9999.
Unlocking the keyboard
Press the CAL and MINUS key for at least 3 seconds to unlock the keyboard.
The code prompt displays the code 0.
3.1.1 Auto/Manual Mode Switching
The analyzer is normally in Automatic mode and the relay/contacts are controlled by the instrument.
The relay contacts can be actuated manually by switching from the Auto Mode to the Manual Mode as
described below. This allows the user to actuate the wiper assembly, cleaning systems or any items
actuated by the relays.
1. The transmitter is in Automatic mode. The top LED beside the AUTO key is lit.
2. Press the AUTO key to switch to Manual Mode.
3. To enable the manual mode, enter the code 22 via the PLUS and
MINUS keys. The bottom LED beside the AUTO key lights up.
4. Select the relay or the function.
You can use the REL key to switch between the relays. The relay selected and the switch status
(ON/OFF) is displayed on the second line of the display.
In the manual mode, the measured value is displayed continuously (e.g. for measured value
monitoring for dosing functions).
5. Switch the relay. It is switched on with PLUS and switched off with MINUS.
The relay remains in its switched state until it is switched over again.
6. Press the AUTO key to return to the measuring mode, i.e. to the automatic mode. All the
relays are triggered again by the transmitter.
3.2 MENUS
The menu selections are addressed by pressing either the CAL key for Calibration Functions or the Enter
key for the Set-Up Functions. The analyzer will then prompt for an Access Code. Once the access code
has been entered the top level function of the chosen menu will be displayed. Pressing the Enter key will
Triton TR8
Page 11
scroll though the available functions. Pressing the + or – keys will scroll through the menus available in
that function group.
Setup mode (Menu Code)
Calibration Mode (Menu Code)
• SETUP 1 Turbidity (A)
• CALIBRATION (C)
• SETUP 2 Temperature (B)
• OFFSET (V)
• CURRENT INPUT (Z)
• SLOPE (N)
• CURRENT OUTPUT (O)
• ALARM (F)
• CHECK (P)
• RELAY (R)
• CONCENTRATION MEASUREMENT (K)
• SERVICE (S)
• INTERFACE (I)
3.2.1 Access Codes
• Press CAL, then using the + and – keys enter the Access Code 22, press Enter to access to Calibration
and Offset menu
• Press ENTER, then using the + and – keys enter the Access Code 22, press Enter to access to the setup
menus
• Simultaneously Press the PLUS and ENTER keys, enter Code 9999 to lock the keyboard
• Simultaneously Press the MINUS and CAL keys, enter Code 0 to unlock the keyboard
• Press CAL or ENTER + any code: access to read mode, i.e. all the settings can be read but not modified.
Page 12
Triton TR8
3.2.2 Menu structure
Figure 15: Menu Structure Diagram
1. Functions (parameters selected, numbers entered)
2. Function Groups, scroll backwards and forwards with the PLUS and MINUS keys
3. Switch from function to function with the ENTER key
The configuration and calibration functions are arranged in function groups.
• In setup mode, select a function group with the PLUS and MINUS keys.
• In the function group itself, switch from function to function with the ENTER key.
• Within the function, select the desired option with the PLUS and MINUS key or edit the settings with
these keys. Then confirm with the ENTER key and continue.
• Press the PLUS and MINUS keys simultaneously (Escape function) to exit programming (return to the
main menu).
• Press the PLUS and MINUS keys simultaneously again to switch to the measuring mode.
3.2.3 HOLD Function
During setup and calibration functions, the current output is "frozen"; it retains its last value. "HOLD"
appears on the display. If a controller actuating variable is output via current output 2, it is set to 0/4 mA
while in Hold.
The automatic Hold function is accessed in the SERVICE MENU “S2”.
Triton TR8
Page 13
3.3 START UP and SETUP FUNCTION GROUP MENUS
After powering the instrument, Set the language in the (S) menu, then adjust all parameters in the (A)
menu and finally adjust the settings in the (B) menu. The instrument is now operational.
3.3.1 SETUP 1: TURBIDITY
Coding
Field
A
Function Group
Setup 1
Selection or Range
Default in Bold
Display
Information
Initial Display in SETUP 1
A1
Operation Mode
NTU
FNU
Ppm
Mg/l
g/l
%
spec.
A2
Display Unit
kg/l
%
t/m
A2 is only accessible, if A1 =
spec.
A3
Display Format
XX.xx
X.xxx
XXX.x
XXXX
A3 is only accessible, if A1 =
spec.
A4
Connected Sensor
CUS31(-1)
CUS41(-2)
The transmitter
automatically detects which
sensor is connected.
10
1 to 60
Measured value damping
causes averaging over the
specified number of
individual measured values.
It is used, for example, to
stabilize the display with
applications that fluctuate a
great deal.
There is no damping if “1” is
entered.
A5
Page 14
Measurement
Damping
Any change in the operation
mode causes an automatic
reset of the user defined
settings. The offsets for
turbidity and temperature
are reset to zero.
Triton TR8
3.3.2 SETUP 2: TEMPERATURE and Wiper
Selection or Range
Default in Bold
Coding
Field
B
Function Group
Setup 2
Initial Display in SETUP 2
B1
Switch wiper
controller on or off
off
on
auto
If “auto” is selected, the
wiper is operated in
combination with a cleaning
function via
Timer/Chemoclean (“wipe
and clean”). In this case B2
and B3 are not applicable.
B2
Enter operating
period
of wiper
30 s
3 to 999 s
B3
Enter pause time
between the two
wiping cycles
30 min
1 to 7200 min
Select calibration
data
set to be used
3
1 to 3
B5
Copy data sets
no
1 →2
1 →3
2 →3
3 →2
B6
Display measured
value with reflection
compensation?
yes
no
B7
Enter corrected
temperature of the
temperature sensor
Current measured
value
-5.0 to 100.0 °C
B4
Triton TR8
Display
Information
There are 3 calibration data
sets stored in every
operation mode (A1). Data
set 1 cannot be changed.
A hold is active during the
load of a new selected data
set (independent of the
settings in S2).
Data set 1 cannot be
changed (factory
setting).However, it can be
used as a basis for a
customer calibration data
set. To work with the copy of
a data set, select the data set
in field B4.
Displays the measured value
with or without reflection
compensation. Only effective
in NTU, FNU, ppm, mg/l.
This entry can be used to
calibrate the temperature
sensor to an external
measurement.
Page 15
Coding
Field
Selection or Range
Default in Bold
B8
Displays the
temperature
difference
(offset)
Current offset
-5.0 to 5.0 °C
The difference between the
temperatures measured and
entered is displayed.
3.0 %
0.1 to 100 %
Compensates gas bubble
formation, which may arise
from small amounts of
dissolved gas in the medium.
0.1 % = no formation of gas
bubbles. 100 % = strong gas
bubble formation. For clear
media (measured value
below 1000 NTU) always set
the gas bubble barrier to
100 %.
B9
Enter gas bubble
barrier
Display
Information
3.3.3 CURRENT INPUT
To use the "Current input" function group, you need a relay board with current input which is Not Part
of the Basic Version. With this function group you can monitor process parameters and use these for
feed forward control. For this purpose, you must connect the current output of an external measured
variable (e.g. flow meter) to the 4 to 20 mA input of the transmitter. The following assignment applies:
Flow in main stream
Flow meter start of measuring range
Flow meter end of measuring range
Current signal in mA
4
20
Current input signal in %
0
100
Monitoring of flow in main stream
This arrangement is particularly practical if the sample flow through a flow assembly in an open outlet is
completely independent of the flow in the main stream. This permits signaling of an alarm condition in
the main stream (flow too low or has completely stopped) and triggers dosing switch-off even if the
medium flow is retained due to the method of installation.
**The Basic Version Does Not Include Functions in Italics**
Selection or Range
Default in Bold
Coding
Field
Z
CURRENT INPUT
function group
Current input settings.
Z1
Select flow
monitoring of main
stream (with
controller switchoff)
Flow monitoring may only be
switched on if the flowmeter
is connected in the main
stream. If Z1 = off, fields Z2
to Z5 are not available.
Page 16
Off
On
Display
Information
Triton TR8
Coding
Field
Selection or Range
Default in Bold
Z2
Enter the delay for
controller switch-off
through current
input
0s
0 to 2000 s
Z3
Enter the delay for
controller switch-on
through current
input
0s
0 to 2000 s
Z4
Enter the switch-off
limit value for the
current input
50%
0 to 100%
Z5
Enter the switch-off
direction for the
current input
Low
High
Z6
Select feed forward
control to PID
controller
Off
Lin = linear
Basic
Z7
Enter value for
Feed forward
control at
which gain = 1
applies
50%
0 to 100%
Display
Information
Brief flow shortfalls can be
suppressed by a delay and
do not result in controller
switch-off.
In the case of a controller, a
delay until a representative
measured value is received is
useful if the flow fails for an
extended period.
0 to 100% corresponds to 4
to 20 mA at the current
input. Observe measured
value assignment to the
current output of the flow
meter.
The controller is switched off
if the value entered in Z4 is
undershot or overshot.
If Z6 = off, the field Z7 is not
available. Z6 = basic:
disturbance variable only
affects the basic load
(alternatively dosing in
proportion to quantity can
be used if usual PID
controlling is not possible,
due to a defective sensor, for
example).
When the value is set, the
controller actuating variable
is the same size when feed
forward control is switched
on as when feed forward
control is switched off.
3.3.4 CURRENT OUTPUT
Use the "Current output" function group to configure the individual outputs. You can enter either a
linear characteristic (O3 (1)) or you can simulate a current output value (O3 (2)) to check the current
outputs. In addition, a user-defined current output Table (O3 (3)) is available with the Plus Package. If a
second current output is present, you can output the controller actuating variable in accordance with
field R 237 via the current output. Exception: if you have chosen a "continuous controller" for current
output 2, you cannot enter a user-defined current output Table on current output 1.
In a User Defined Output Table, the variable must steadily increase across the table or decrease, the
variable cannot change direction. The distance per mA (SV/mA) between two table values must be
greater than: 0.005 FNU, 0.005 NTU, 0.005 ppm, 0.005 mg/l, 0.005 %, 0.05 g/l or 0.25 C°.
Triton TR8
Page 17
The values for the sample characteristic are entered in the following table. The distance per mA can be
calculated from Δ signal / Δ mA.
Value Pair
Signal Value (SV)
Current Value, mA
SV/mA
1
2
3
0
3000
9000
4
16
20
250
1500
First enter the desired current output configuration into the following blank table with a pencil.
Calculate the resulting signal distance per mA to observe the necessary minimum slope. Then enter the
values in the device.
Value Pair
Signal Value (SV)
Current Value, mA
SV/mA
1
2
3
4
5
6
7
8
9
10
Selection or Range
Default in Bold
Coding
Field
O
CURRENT OUTPUT
function group
O1
Select current
output
Out1
Out 2
O2
Select measured
variable for 2nd
current output
°C
mg/l
Contr
O3 (1)
Enter or output
linear
characteristic
lin = linear (1)
sim = simulation(2)
tab = table (3)
Page 18
Display
Information
Configuration of the current
output (does not apply for
PROFIBUS).
Output 2 not available on all
versions. A characteristic can
be selected for every output.
R237 = curr (current output
2) can only be selected if O2
= Contr is selected (relay
board required).
The characteristic can have a
positive or negative slope for
the measured value output.
In the case of actuating
variable output (O2 = Contr),
an increasing current
corresponds to an increasing
actuating variable.
Triton TR8
Coding
O311
O312
O313
O3 (2)
Field
Selection or Range
Default in Bold
Select current range
4 to 20 mA
0 to 20 mA
0/4 mA value:
Enter corresponding
turbidity or
temperature value
20 mA value:
Enter corresponding
turbidity or
temperature value
Simulate current
output
0.000 FNU
0.000 NTU
0.000 ppm
0.000 mg/l
0.000 g/l
0.000 kg/l
0.000 t/m
0.000 %
0.000 °C
10.00 FNU
10.00 NTU
10.00 ppm
10.00 mg/l
300 g/l / 3.00 g/l
99.99 kg/l
99.99 t/m
10.0 %
100 °C
Lin = linear (1)
Sim =
simulation(2)
Tab = table (3)
Enter simulation
value
Current value
0.00 to 22.00 mA
Enter current output
table (only for Plus
Package)
lin = linear (1)
sim = simulation(2)
tab = table (3)
O331
Select table options
read
edit
O332
Enter number of
table
value pairs
1
1 to 10
O321
O3 (3)
Triton TR8
Display
Information
Here you can enter the
turbidity or temperature
value at which the min.
current value (0/4 mA) is
applied at the transmitter
output. Display format from
A3
Here you can enter the
turbidity or temperature
value at which the maximum
current value (20 mA) is
applied at the transmitter
output. Display format from
A3
Simulation is not ended until
(1) or (3) is selected.
Entering a current value
results in this value being
directly output at the current
output.
Values can also be added or
altered at a later stage. The
values entered are
automatically sorted by
increasing current value.
Enter the number of pairs
from the x and y value
(measured value and current
value) here.
Page 19
Coding
O333
O334
Field
Selection or Range
Default in Bold
Display
Information
Select table value
pair
1
1to No. elements
Assigned
The function chain O333 ...
O335 will run through as
many times as correspond to
the value in O332. "Assign"
appears as the last step.
After confirmation the
system jumps to O336.
Enter x value
0.000 FNU
0.000 NTU
0.000 ppm
0.000 mg/l
0.000 g/l
0.000 kg/l
0.000 t/m
0.000 %
0.000 °C
x value = measured value
specified by user.
O335
Enter y value
4.00 mA
0.00 to 20.00 mA
O336
Message as to
whether table
status is
OK
yes
no
y value = current value
belonging to O334
specified by user.
Return to O333 until all
values are entered.
Back to O3. If status = no,
correct table (all settings
made up until now are
retained) or back to
measuring mode (table is
deleted).
3.3.5 ALARM FUNCTIONS
The monitoring functions are used to define various alarms and configure output contacts. Each
individual error can be defined to be effective or not (at the contact or as an error current). An alarm
condition can be defined to activate a cleaning function (F8)
Coding
Field
F
ALARM function
group
Selection or Range
Default in Bold
Information
Alarm function settings.
F1
Select contact type
Latch = latching
contact Momen
=momentary
contact
F2
Select time unit
s
min
Page 20
Display
The contact type selected
only applies to the alarm
contact.
Triton TR8
Coding
Field
Selection or Range
Default in Bold
F3
Enter alarm delay
0 s (min)
0 ... 2000 s (min)
F4
Select error current
22 mA
2.4 mA
F5
Select error
1
1 ... 255
F6
Set alarm contact to
be effective for the
selected error
yes
no
F7
Set error current to
be effective for the
selected error
no
yes
F8
Automatic cleaning
function start
no
yes
F9
Select return to
menu or next error
next = next error
←R
Display
Information
Depending on the option
selected in F2, the alarm
delay is entered in s or min.
This selection must be made
even if all error reporting is
switched off in F5. Caution!
If "0-20 mA" was selected in
O311, "2.4 mA" may not be
used.
Here you can select all the
errors which should trigger
an alarm. The errors are
selected via the error
numbers. Please refer to the
table in section 6.2 "System
error messages" for the
meaning of the individual
error numbers. The factory
settings remain in effect for
all errors not edited.
If “no” is selected, all the
other alarm settings are
deactivated (e.g. alarm
delay). The settings are
retained. This setting only
applies to the error selected
in F5.
The option selected in F4 is
effective or ineffective in the
event of an error. This
setting only applies to the
error selected in F5.
This field is not available for
certain errors, see "Troubleshooting and fault
elimination" section.
If ←R is selected, you return
to F, if next is selected, you
go to F5.
3.3.6 CHECK
The CHECK function group is only available for devices with a Plus Package.
In the CHECK function group, you can select different monitoring functions for the measurement.
Triton TR8
Page 21
All monitoring functions are off by default. To adapt the Sensor Check System to the current application
conditions, add and set the suitable functions.
Alarm threshold monitoring (fields P1 to P4)
You can use this function to monitor the measured value for permissible upper and lower limits and
trigger an alarm (E154, E155).
PCS alarm (Process Check System), (fields P5 to P8)
AC (Alternating Check): The function AC (field P5) is used to check measuring signals for deviations. If
the measuring signal does not change within an hour an alarm (E152) is triggered. The reason for such
sensor behavior can be contamination, cable rupture or similar.
CC (Controller Check): You can monitor the controller activity with the function CC. This function is
mainly used for batch processes and single-sided limit switches. A malfunction of the controller is
detected and reported thanks to freely adjustable monitoring times (E156 - E157).
The CHECK Function group is an optional package; the BASIC package does not include items listed in
Italics.
Coding
Field
P
CHECK function
group
Selection or Range
Default in Bold
Information
Settings for sensor and
process monitoring
P1
Select alarm
threshold
monitoring
Off
Low
High
Lo+Hi
Low!
High!
Lo+Hi!
P2
Enter alarm delay
0 s (min)
0 to 2000 s (min)
P3
Enter lower alarm
threshold
0.000 FNU
0 to 9999 FNU
P4
Enter upper alarm
threshold
10.00 FNU
0 to 9999 FNU
Page 22
Display
Alarm signaling optionally
with or without
simultaneous controller
switch-off. XXXX = without
controller switch-off
XXXX! = with control switchoff (Errors: E154, E155)
Depending on your selection
in F2, you can enter the error
delay in min or s. Only after
this delay, a high or low limit
violation causes an alarm as
per field P3/P4.
Triton TR8
Coding
Field
Select process
monitoring
(PCS alarm)
P5
Enter maximum
permissible duration
for lower CC set
point limit violation
(field P8)
Enter maximum
permissible duration
for upper CC set
point limit violation
(field P8)
P6
P7
Enter CC set point
(for P6/P7)
P8
Selection or Range
Default in Bold
Off
AC
CC
AC+CC
AC!
CC!
AC+CC!
Display
Information
AC = sensor activity check
(E152) CC = controller check
(E156, E157) XXXX = without
controller switch-off
XXXX! = with controller
switch-off
60 min
0 to 2000 min
Only when P5 = CC or AC+CC
120 min
0 to 2000 min
Only when P5 = CC or AC+CC
1.000 FNU
0 to 9999 FNU
Selected value is an absolute
value. This function is mainly
used for batch processes and
single-sided limit switches.
3.3.7 RELAY
To use the RELAY function group, you need a relay board which is not part of the basic version.
The following relay contacts can be selected and configured as desired (max. four contacts, depending
on options installed):
• Limit contactor for measured turbidity value: R2 (1)
• Limit contactor for temperature: R2 (2)
• PID controller: R2 (3)
• Timer for cleaning function: R2 (4)
• Chemoclean function: R2 (5)
3.3.7.1 Limit contactor for measured turbidity value and temperature
The transmitter has different ways of assigning a relay contact.
Switch-on and switch-off points and pick-up and drop-out delays can be assigned to the limit contactor.
In addition, you can configure an alarm threshold to output an error message and to start a cleaning
function in conjunction with this.
These functions can be used both for turbidity measurement and for temperature measurement.
Triton TR8
Page 23
Please refer to Fig. 28 for a clear illustration of the relay contact states.
• When the measured values increase (maximum function), the relay contact is closed as of t2
after the switch-on point (t1) has been overshot and the pick-up delay has elapsed (t2 - t1). The
alarm contact switches if the alarm threshold (t3) is reached and the alarm delay (t4 - t3) has
also elapsed.
• When the measured values decrease, the alarm contact is reset when the alarm threshold (t5)
is undershot as is the relay contact (t7) after the drop-out delay (t7 - t6).
• If the pick-up and drop-out delays are set to 0 s, the switch-on and switch-off points are also
switch points of the contacts.
Settings can also be made for a minimum function in the same way as for a maximum function.
Figure 16: Illustration of the alarm and limit value functions
A. Switch-on point > switch-off point: Max. function
B. Switch-on point < switch-off point: Min. function
1. Alarm Threshold
2. Switch On Point
3. Switch Off Point
5. Alarm On
6. Alarm Off
7. Contact Off
4. Contact On
3.3.7.2 PID Control
You can define various controller functions for the transmitter. On the basis of the PID controller, P, PI,
PD and PID controllers can be implemented. For an optimum control system, use the controller that best
Page 24
Triton TR8
suits your application. Depending on the option selected in the R 237/R 266 field, the actuating signal
can be output via relays or via current output 2 (if available).
• P controller: Used for simple linear control purposes with small system deviations. Where major
changes are to be controlled, overshooting may occur. In addition, a lasting control deviation is to be
expected.
• PI controller: Is used for control systems where overshooting is to be avoided and no lasting control
deviation should occur.
• PD controller: Is used for processes that require quick changes and where peaks are to be corrected.
• PID controller: Is used for processes where a P, PI or PD controller does not control sufficiently.
Configuration options of the PID controller
The following configuration options are available for a PID controller:
• Change control gain Kp (P influence)
• Set integral action time Tn (I influence)
• Set derivative action time Tv (D influence)
Basic load dosing (Basic)
The basic load dosing (field R231) is used to set a constant dosage (field R2311)
PID controlling plus basic load dosing
If you select this function (PID + Basic) in field R231 the PID controlled dosage will not be lower than the
basic load value entered in field R2311.
Commissioning
If you do not yet have any experience for setting the control parameters, set the values that yield the
greatest possible stability in the control circuit. Proceed as follows to optimize the control circuit further:
• Increase the control gain Kp until the controlled variable just starts to overshoot.
• Reduce Kp slightly and then reduce the integral action time Tn so that the shortest possible correction
time without overshooting is achieved.
• To reduce the response time of the controller, also set the derivative action time Tv.
Triton TR8
Page 25
Fig17: Optimization of settings Tn and Kp
A. Actual value
B. Time
1. Tn too small 2. Tn too large 3. Kp too large 4. Kp too small 5. Optimum setting
Actuating signal outputs (R237 ... R2310)
Each control contact outputs a cyclical signal whose intensity corresponds to the controller's
manipulated variable. A distinction is made according to the type of signal cycle:
• Pulse length modulation
The bigger the calculated manipulated variable is, the longer the contact affected remains picked up.
The period T can be adjusted between 0.5 and 99 s (field R238). Outputs with pulse length modulation
are used to activate solenoid valves.
• Pulse frequency modulation
The bigger the calculated manipulated variable is, the higher the switching frequency of the contact
affected. The maximum switching frequency 1/T can be set between 60 and 180 min-1.
Page 26
Triton TR8
The on-time tON is constant. It depends on the set maximum frequency and is approx. 0.5 s for 60 min-1
and approx. 170 ms for 180 min-1. Outputs with pulse frequency modulation are used to activate
directly controlled solenoid dosing pumps.
Constant controller
Via the current output 2, the minimum actuating variable (0 %) of the controller is output with 0/4 mA
and the maximum actuating variable (100%) of the controller is output with 20 mA.
Control characteristic for direct and inverse control action
You can choose between two control characteristics in the R236 field:
• Direct control action = maximum function
• Inverse control action = minimum function
3.3.7.3 Timer for cleaning function
This function includes a simple cleaning option. You can set the time interval after which cleaning should
start. So you can only select a constant interval sequence. Other cleaning functions are available for
selection in conjunction with the Chemoclean function (version with four contacts, see "Chemoclean
function" section).
Note!
Timer and Chemoclean do not work independently of one another. While one of the two functions is
active, the other cannot be started.
Figure 18: Correlation of cleaning time, pause time and hold dwell period
A. Wiper Function
B. Hold Function
0. Inactive
1. Active
t0 Normal Mode
t1 Start Cleaning
t2-t1 Cleaning Time
t3-t2 Hold Dwell Time
t4-t3 Time between Cleaning cycles
Triton TR8
Page 27
3.3.7.4 Chemoclean function
Just like the timer function, Chemoclean can also be used to start a cleaning cycle. However,
Chemoclean also gives you the added option of defining different cleaning and rinsing intervals. As a
result, it is possible to clean irregularly with different repeat cycles and to separately set the cleaning
times with post rinse times.
Note!
• To use the Chemoclean function the transmitter has to be equipped with a designated relay board (see
product structure or chapter "accessories").
• Timer and Chemoclean do not work independently of one another. While one of the two functions is
active, the other cannot be started.
• For the Chemoclean function, the relays 3 (water) and 4 (cleaner) are used.
• If cleaning is prematurely aborted, a post rinse time always follows.
• If the setting is "Economy", cleaning only takes place with water.
Figure 19: Cleaning Cycle Sequence
A. Hold
B. Water
t0 Normal Mode
t4 - t3 Post Rinse Time
C. Cleaner
t1 Cleaning Start
0. Contact on
1. Contact off
t2 - t1 Pre Rinse Time
t3 - t2 Cleaning Time
t5 - t4 Hold Dwell Time
RELAY FUNCTION
The basic unit does not include items in italics.
Page 28
Triton TR8
Coding
Field
R
RELAY function
group
Selection or Range
Default in Bold
Rel1
Rel2
Rel3
Rel4
R2 (1)
Configure limit
contactor for
turbidity
measurement
LC PV = limit
contactor TU (1)
LC °C = limit
contactor T (2)
PID controller (3)
Timer (4)
Clean =
Chemoclean (5)
R211
Switch function of
R2
(1) off or on
Off
On
Enter the switch-on
point of the contact
9999 FNU
9999 NTU
9999 ppm / 3000
ppm
9999 mg/l / 3000
mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m
200.0 %
Enter the switch-off
point of the contact
9999 FNU
9999 NTU
9999 ppm / 3000
ppm
9999 mg/l / 3000
mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m
200.0 %
R212
R213
Triton TR8
Information
Relay Contact Settings
Select contact to be
configured
R1
Display
Rel3 (water) and Rel4
(cleaner) are only available
with the relevant version of
the transmitter. If
Chemoclean is used as the
cleaning method, Rel4 is not
available.
PV = process value If Rel4 is
selected in the R1 field,
Clean = Chemoclean cannot
be selected. By confirming
with ENTER, another relay
function already switched on
is switched off and its
settings are reset to the
factory settings.
All the settings are retained.
Never set the switch-on
point and the switch-off
point to the same value!
(Only the operating mode
selected in A1 is displayed.)
If two factory settings are
displayed the one on the left
side stands for CUS41 the
one on the right side stands
for CUS31.
Entering a switch-off point
selects either a Max contact
(switch-off point < switch-on
point) or a Min contact
(switch-off point > switch-on
point), thereby
implementing a hysteresis
that is constantly required
(see "Illustration of the alarm
and limit functions "figure). If
two factory settings are
displayed the one on the left
side stands for CUS41 the
one on the right side stands
for CUS31.
Page 29
Coding
Field
Selection or Range
Default in Bold
R214
Enter pick-up delay
0s
0 to 2000 s
R215
Enter drop-out
delay
0s
0 to 2000 s
Display
Information
If the alarm threshold is
undershot / overshot, this
triggers an alarm with the
error message and error
current at the transmitter
(note alarm delay in field F3).
If defined as a Min contact,
the alarm threshold must be
< switch-off point. If two
factory settings are displayed
the one on the left side
stands for CUS41 the one on
the right side stands for
CUS31.
R216
Enter alarm
threshold
9999 FNU
9999 NTU
9999 ppm / 3000
ppm
9999 mg/l / 3000
mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m
200.0 %
R217
Display status for
limit
contactor
MAX
MIN
Display only.
R2 (2)
Configure limit
contactor for
temperature
measurement
LC PV = limit
contactor TU (1)
LC °C = limit
contactor T (2)
PID controller (3)
Timer (4)
Clean =
Chemoclean (5)
By confirming with ENTER,
another relay function
already switched on is
switched off and its settings
are reset to the factory
settings.
R221
Switch function of
R2
(2) off or on
Off
On
Settings made for the limit
contactor are not deleted by
switching the function off.
R222
Enter switch-on
temperature
100.0 °C (212 °F)
-5.0 to 100.0 °C
(23 to 212 °F)
Never set the switch-on
point and the switch-off
point to the same value!
Page 30
Triton TR8
Coding
Field
Selection or Range
Default in Bold
R223
Enter switch-off
temperature
100.0 °C (212 °F)
-5.0 to 100.0 °C
(23 to 212 °F)
R224
Enter pick-up delay
0s
0 to 2000 s
R225
Enter drop-out
delay
0s
0 to 2000 s
Display
Information
Entering a switch-off point
selects either a Max contact
(switch-off point < switch-on
point) or a Min contact
(switch-off point > switch-on
point), thereby
implementing a hysteresis
that is constantly required.
R226
Enter alarm
threshold
(as absolute value)
100.0 °C (212 °F)
-5.0 to 100.0 °C
(23 to 212 °F)
If the alarm threshold is
undershot/overshot, this
triggers an alarm with the
error message and error
current at the transmitter
(note alarm delay in field F3).
If defined as a Min contact,
the alarm threshold must be
< switch-off point.
R227
Display status for
limit
contactor
MAX
MIN
Display only.
R2 (3)
Configure P(ID)
controller
LC PV = limit
contactor TU (1)
LC °C = limit
contactor T (2)
PID controller (3)
Timer (4)
Clean =
Chemoclean (5)
By confirming with ENTER,
another relay function
already switched on is
switched off and its settings
are reset to the factory
settings.
R231
Switch function of
R2
(3) off or on
Off
On
Basic
PID+B
On = PID controller
Basic = basic load dosing
PID+B = PID controller +
basic load dosing
Triton TR8
Page 31
Coding
Field
R232
Enter set point
R233
Enter control gain
Kp
Selection or Range
Default in Bold
0.000 FNU
0.000 NTU
0.000 ppm
0.000 mg/l
0.000 g/l
0.000 kg/l
0.000 t/m
0.000 %
Display
Information
The set point is the value to
be maintained by the control
system. Using this control
process, this value is
restored upwards or
downwards when a
deviation occurs.
1.00
0.01 to 20.00
See "P(ID) controller"
section.
R234
Enter integral action
time Tn (0.0 = no
I-component)
0.0 min
0.0 to 999.9 min
With every Hold, the Icomponent is set to zero.
Although Hold can be deactivated in field S2, this
does not apply for
Chemoclean
and timer!
R235
Enter derivative
action time Tv
(0.0 = no
D-component)
0.0 min
0.0 to 999.9 min
See "P(ID) controller"
section.
R236
Select controller
characteristic
inv = inverse
dir = direct
R237
Select pulse length
or
pulse frequency
len = pulse length
freq = pulse
frequency
curr = current
output 2
R238
Enter pulse interval
10.0 s
0.5 to 999.9 s
R239
Enter maximum
pulse
frequency of the
adjuster
120 min-1
60 to 180 min-1
Page 32
dir = Max. function
inv = Min. function
The setting is required
depending on the
control deviation (upward or
downward deviation, see
"P(ID) controller" section).
Pulse length e.g. for solenoid
valve, pulse frequency e.g.
for solenoid dosing pump,
see "Actuating signal
outputs" section. Curr =
current output 2 can only be
selected if O2 = Contr.
This field only appears if
pulse length is selected in
R237. If pulse frequency is
selected, R238 is skipped and
entries continue with R239.
This field only appears if
pulse frequency is selected in
R237. If pulse length is
selected, R239 is skipped and
entries continue with R2310.
Triton TR8
Coding
Field
Selection or Range
Default in Bold
R2310
Enter minimum
switch-on time tON
0.3 s
0.1 to 5.0 s
Display
Information
This field only appears if
pulse length is selected in
R237.
When you select the basic
load, you enter the desired
dosing quantity. 100% basic
load would correspond to:
– Constantly on for R237 =
len
– Fmax at R237 = freq (field
R239)
– 20 mA at R237 = curr
Cleaning only takes place
with one cleaning agent
(usually water). By
confirming with ENTER,
another relay function
already switched on is
switched off and its settings
are reset to the factory
settings.
Enter basic load
0%
0 to 40 %
R2 (4)
Configure cleaning
function (timer)
LC PV = limit
contactor TU (1)
LC °C = limit
contactor T (2)
PID controller (3)
Timer (4)
Clean =
Chemoclean (5)
R241
Switch function of
R2
(4) off or on
Off
On
Settings made for the timer
are not deleted by switching
the function off.
R242
Enter
rinsing/cleaning
time
30 s
0 to 999 s
Settings for Hold and relay
are active for this time.
R243
Enter pause time
360 min
1 to 7200 min
The pause time is the time
between two cleaning cycles
(see "Timer for cleaning
function" section).
R244
Enter minimum
pause
time
120 min
1 to R243 min
The minimum pause time
prevents constant cleaning if
a cleaning trigger is present.
R2 (5)
Configure cleaning
with Chemoclean
(for version with
four contacts ,
Chemoclean option
and contacts 3 and
4 assigned)
LC PV = limit
contactor TU(1)
LC °C = limit
contactor T (2)
PID controller (3)
Timer (4)
Clean =
Chemoclean (5)
See "Chemoclean function"
section. By confirming with
ENTER, another relay
function already switched on
is switched off and its
settings are reset to the
factory settings.
R2311
Triton TR8
Page 33
Coding
Field
Selection or Range
Default in Bold
R251
Switch function of
R2
(5) off or on
Off
On
R252
Select type of start
pulse
Int = internal
(time-controlled)
Ext = external
(digital input 2)
I+ext = internal +
external
I+stp = internal,
suppressed
by external
The cycle for the "int"
function is started by the end
of the pause time (R257). No
real time clock is available.
External suppression is
required for irregular time
intervals (e.g. weekends).
R253
Enter pre-rinse time
20 s
0 to 999 s
Rinsing with water takes
place.
R254
Enter cleaning time
10 s
0 to 999 s
Cleaning with cleaning agent
and water takes place.
R255
Enter post rinse
time
20 s
0 to 999 s
Rinsing with water takes
place.
R256
Enter number of
repeat cycles
0
0 to 5
R253 to R255 is repeated.
R257
Enter pause time
360 min
1 to 7200 min
The pause time is the time
between two cleaning cycles
(see "Timer function
section).
R258
Enter minimum
pause time
120 min
1 to R257 min
The minimum pause time
prevents constant cleaning if
an external cleaning start is
present.
R259
Enter number of
cleaning cycles
without cleaning
agent (economy
function)
0
0 to 9
After cleaning with cleaner,
up to 9 cleaning sessions can
be carried out with water
only until the next cleaning
session with cleaner takes
place.
Page 34
Display
Information
Triton TR8
3.3.8 CONCENTRATION MEASUREMENT
The basic unit does not include items in italics.
Coding
Field
K
Function group
CONCENTRATION
K1
Selection of
concentration curve,
to be used to
calculate
the display value
Selection or Range
Default in Bold
Display
Information
Four different concentration
curves can be entered in this
function group.
1
1 to 4
The curves are independent
of each other. Therefore,
four different curves can be
defined.
K2
Selection of table to
be edited
2
1 to 4
The modification of a table
(curve) is independent from
the selected curve in K1. The
selected table (curve) will
not be overwritten until the
input of all table values is
complete.
K3
Select table option
read
edit
This selection applies to the
concentration curve selected
in K2.
K4
Enter number of
reference elements
1
1 to 10
Each element consists of two
numeric values.
Select element
1
1 to number of
element in K4
assign
Any element can be edited.
K5
K6
Enter turbidity value
0.00 %
entire measuring
range
The function chain K5 ... K7
will be repeated
automatically as many times
as corresponds to the value
in K4. Then the system jumps
to K8.
K7
Enter concentration
value
entire measuring
range
Measuring unit as selected in
A2.
K8
Message whether or
not the table status
is
ok
yes
no
Only display If not, then set
table correctly (all previous
settings are kept) or back to
measurement mode (this
makes the table invalid).
Triton TR8
Page 35
3.3.9 SERVICE
Coding
Field
S
SERVICE function
group
Selection or Range
Default in Bold
Information
Service function settings.
ENG = English
GER = German
FRA = French
ITA = Italian
NL = Dutch
ESP = Spanish
S+C = Hold during
configuration and
calibration
Cal = Hold during
calibration
Setup = Hold
during
configuration
None = no Hold
S1
Select language
S2
Configure Hold
S3
Manual Hold
off
on
S4
Enter Hold dwell
period
10 s
0 to 999 s
S5
Enter SW upgrade
release code (Plus
Package)
0
0000 to 9999
S6
Enter SW upgrade
release code
Chemoclean
0
0000 to 9999
Page 36
Display
S = setup
C = calibration
The setting is retained even
in the event of a power
failure.
The code is on the name
plate. If an incorrect code is
entered, you are taken back
to the measurement menu.
The number is edited with
the PLUS or MINUS key and
confirmed with the ENTER
key. "1" is displayed if the
code is active.
You can find the code on the
nameplate. If an incorrect
code is entered, you are
taken back to the
measurement menu. "1" is
displayed if the code is
active.
Triton TR8
Coding
Field
Selection or Range
Default in Bold
S7
Order number is
displayed
CUM253-T
S8
Serial number is
displayed
880CB405G
S9
Reset the device to
the
basic settings
no
Sens = sensor data
Facty = factory
settings
S10
Perform device test
no
Displ = display test
Coding
Field
Selection or Range
Default in Bold
E
SERVICE 2
function group
Display
Information
If the device is upgraded, the
order code is not
automatically adjusted.
Sens = last calibration is
deleted and is reset to
factory setting.
Facty = all data (apart from
A1 a. S1) are deleted and
reset to the factory setting!
Display
Information
Information on the device
version
Contr = controller
(1)
Trans = transmitter
(2)
Main = power unit
(3)
Rel = relay module
(4)
E1
Select module
E111
E121
E131
E141
Software version is
displayed
If E1 = contr: instrument
software
If E1 = trans, main, rel:
module firmware
E112
E122
E132
E142
Hardware version is
displayed
Only display function
Triton TR8
Page 37
Selection or Range
Default in Bold
Coding
Field
Display
Information
E113
E123
E133
E143
Serial number is
displayed
Only display function
E114
E124
E134
E144
Module ID is
displayed
Only display function
3.3.10 INTERFACE
Selection or Range
Default in Bold
Coding
Field
Display
Information
I
INTERFACE
function group
Communication settings
(only for device version HART
or PROFIBUS).
I1
Enter bus address
Each address may only be
used once in a network.
If a device address ≠ 0 is
selected, the current output
is automatically set to 4 mA
and the device is set to multidrop operation.
I2
Display of
measuring
point
Address
HART: 0 to 15
or
PROFIBUS: 0 to 126
3.4 CALIBRATION FUNCTION GROUP MENUS
Measuring chain calibration is performed in this function group. The calibration data are saved in an
EEPROM directly in the sensor. For this reason:
• Recalibration is not required in the event of a power failure
• Recalibration is not required when the transmitter is replaced
• Customer-specific recalibration is required, however, when the sensor is replaced
Three calibration data records are saved in the sensor for each of the four main operating modes.
Factory data record no.
1 not changeable
User data record no. 2
changeable
User data record no. 3
changeable
Page 38
FNU/NTU
ppm or mg/l
g/l
Formazine
SiO2
Activated sludge
Formazine
Kaolin
Activated sludge
Formazine
SiO2
Activated sludge
%
Residual concrete
water
Residual concrete
water
Residual concrete
water
Triton TR8
Select the desired data record in the SETUP 2 function group in the B4 field.
• In the FNU operating mode, the sensor is factory calibrated with formazine traceable to ISO 7027.
• In the ppm operating mode, the calibration data records for Kaolin and SiO2 are derived from the FNU
data records.
• In the % operating mode, the calibration data records are set to the average of various residual
concrete waters. They are preset in such a way that correct values are displayed for average clarity.
However the settings do not follow a standard currently applicable.
• In the g/l operating mode also, the sensor is not calibrated to a fixed value as no standard is directly
applicable. You must carry out a calibration because the media of the various applications differ too
greatly here.
Three-point sensor calibration is the standard calibration. It is absolutely essential:
• When commissioning the sensor in sludge applications
• When changing to another sludge type
Three-point sensor calibration is not necessary:
• When commissioning the sensor in the drinking water area (sensor has been calibrated for drinking
water applications in the factory).
• For residual concrete water. Density measurement for determining the concentration of residual
concrete water is based on %–data records. They are preset in such a way that correct values are
displayed for average clarity. One-point calibration is often sufficient to adjust the system in the event of
deviating values.
• When recalibrating with the same sludge type. One-point calibration suffices here if the degrees of
lightness and clarity, for example, do not differ too greatly.
Note!
• Sludge samples tend sediment. Mix the sample well, even during the calibration process, but not
to the extent that gas bubbles are formed.
• The sensor has to be far enough away from the floor and the wall of the calibration vessel during
calibration. The immersion depth must be at least 40 mm.
• The characteristic determined during the calibration is stored in the selected data record (Setup 2, B4
field).
• Calibration is not possible if data record 1 is selected with the factory setting.
• If the calibration data deviate from the standard values by a factor of two or more, a warning (E084) is
output. The calibration results are accepted.
• If the calibration results are outside the permitted range, a calibration error (E045) is indicated. The
calibration results are not accepted.
• For every type of calibration, the installation adjustment and the offset are reset to zero and the slope
to 1.0.
Calibration Menu Options
3–Pt
Three-point calibration
Corr
Three-point correction
Edit
Edit calibration
Refl
Installation adjustment
1–Pt
One-point calibration
Triton TR8
Page 39
Data
Calibration data
Three-point calibration (3–Pt)
You should perform the calibration in the turbidity/solids concentration range in which you plan to
measure. The overall calibration characteristic of the measuring chain is determined using three samples
of known turbidity or known solids content.
Calibration with a very dark, high-absorption medium returns a small slope while light, clear media
return big slopes.
You can create the requisite probes by diluting a medium sample. In general, very good calibration
results are achieved with a concentration gradation of 10 %, 33 % and 100 %. The following condition
must be met for the calibration: Sample A > 1.1 x sample B > 1.1 x sample C
Figure 20: Preparing 3 point calibration samples
A
Original Sample
B
1 part A + 2 parts water
C
1 part A + 9 parts water
A different dilution ratio should be selected for very high-absorption media to achieve accurate
calibration results. Instead of 100 % – 33 % – 10 %, you can use dilution ratios of 100 % – 20 % –
5 % or 100% – 10 % – 3.3 %.
If the scattered light of a light sample hits a sensor that is calibrated for a dark medium, it can happen
that the signal value is so high that it is above the calibration curve.
Three-point correction (Corr)
If the calibration was performed with an unknown sample concentration but with a defined dilution, the
correct value determined afterwards in the laboratory is entered here.
Edit calibration (Edit)
If each of the samples has been determined afterwards in the laboratory, the correction for the
calibration is entered here.
Installation adjustment (Refl)
In installation adjustment, backscatter from the immediate sensor environment is compensated.
Installation adjustment must be performed with a medium lower than 2 FNU or 5 ppm.
Page 40
Triton TR8
One-point calibration (1–Pt)
In the area of TS/concentration measurement, one-point calibration is used to change the conversion
factor (field C166). The slopes are not changed.
In the FNU range, the two slope values are adjusted by a one-point calibration. This is possible because
the conversion factor in the FNU range is always 1 and the editing range is limited to 4000 FNU. In this
way, the two curves are always increasing and the calibration remains clear.
In the ppm range, the slope values are adjusted up to 500 ppm. For higher values, the conversion factor
is changed.
Calibration data (Data)
Here, you can display the calibration points 1 to 3, slope 1 and 2 and the conversion factor.
With three-point calibration, the curves used in the algorithm are adjusted as precisely as possible using
the data points acquired. The difference between the ideal curve of the algorithm and the three actual
calibration points can be found as a correction factor in the C161, C162 and C163 fields of the data
function. The correction values are indicated in %. The values should be as close to 100 % as possible.
Values from 70 to 80 % are acceptable. 50 % in one or two calibration points clearly indicates a
problematic calibration. A warning (E084) is output here for this reason. This can mean that significant
deviations can occur between the calibration points. The calibration points themselves are always
retrieved correctly.
3.4.1 CALIBRATION
Coding
Field
C
Function group
CALIBRATION
Selection or Range
Default in Bold
Display
Information
Calibration settings.
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
Refl = Fitting with
reflection
compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration
data (6)
For data set 1 (B4), only the
“Data” function is accessible.
The offset is reset with 3 Pt
and Edit.
Immerse sensor in the calibration solution (sample 1).
Immerse the sensor so that
there is sufficient distance to
the tank wall (no reflection).
C1 (1)
Triton TR8
Select calibration
Page 41
Coding
Field
Selection or Range
Default in Bold
C111
Enter concentration
of the first
calibration
solution
Value from last
calibration
Immerse sensor in the calibration solution (sample 2).
C112
Enter concentration
of the second
calibration
solution
Value from last
calibration
Immerse sensor in the calibration solution (sample 3).
C113
Enter concentration
of the third
calibration
solution
Value from last
calibration
C114
Calibration status is
displayed
o. k.
E. xxx
Store calibration
results
yes
no
new
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
Refl = Fitting with
reflection
compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration
data (6)
C115
C1 (2)
Page 42
Display
Information
Immerse the sensor so that
there is sufficient distance to
the tank wall (no reflection).
C112 ≥ 1.1 x C111
Immerse the sensor so that
there is sufficient distance to
the tank wall (no reflection).
C113 ≥ 1.1 x C112
If C114 = E xxx, then only no
or new (Exception:
calibration warning E84). If
new, return to C. If yes / no,
return to “Measurement”.
Triton TR8
Coding
Field
Selection or Range
Default in Bold
C121
Enter correct
concentration of the
third calibration
solution
Current value from
C113
entire measuring
range
C122
Calibration status is
displayed
o. k.
Exxx
Store calibration
results
yes
no
new
C1 (3)
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
Refl = Fitting with
reflection
compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration
data (6)
C131
Enter concentration
of the first
calibration
solution
Current value from
C111
entire measuring
range
C132
Enter concentration
of the second
calibration
solution
Current value from
C112
C132 ≥ 1.1 x C131
C133
Enter concentration
of the third
calibration
solution
Current value from
C113
C133 ≥ 1.1 x C132
C134
Calibration status is
displayed
o.k.
Exxx
C123
Triton TR8
Display
Information
If the calibration is
performed with an unknown
sample concentration, but
with a definite dilution (1/10;
1/3;1), the laboratory value is
to be entered.
If C122 = E xxx, then only no
or new (Exception:
calibration warning E84). If
new, return to C. If yes / no,
return to “Measurement”.
Page 43
Coding
C135
Field
Selection or Range
Default in Bold
Store calibration
results?
yes
no
new
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
Refl = Fitting with
reflection
compensation
(4)
1-Pt = Single-point
calibration (5)
Data = Calibration
data (6)
0.0 NTU
0.0 to 2.0 NTU
0.0 FNU
0.0 to 2.0 FNU
0.0 ppm
0.0 to 5.0 ppm
0.0 mg/l
0.0 to 5.0 mg/l
Display
Information
If C134 = E xxx, then only no
or new (Exception:
calibration warning E84). If
new, return to C. If yes / no
return to “Measurement”.
Only for solutions = 2 FNU /
5 ppm!
Backscatter from the
immediate sensor
environment is compensated
for clear media.
C1 (4)
Select calibration
C141
Enter correct
measured value
C142
Calibration status is
displayed
o.k.
Exxx
Store calibration
results?
yes
no
new
If C142 = E xxx, then only no
or new (Exception:
calibration warning E84). If
new, return to C. If yes / no,
return to “Measurement”.
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
Refl = Fitting with
reflection
compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration
data (6)
For FNU: Adaption C164,
C165
For ppm, mg/l:
up to 500 - adaption C164,
C165 above 500 - adaption
C166
For g/l, %: Adaption C166. An
existing basic calibration
(three-point) is corrected by
the single-point calibration.
C143
C1 (5)
Page 44
Only for the ranges NTU,
FNU, ppm, mg/l
Triton TR8
Coding
Field
Selection or Range
Default in Bold
C151
Enter current
calibration value
Current measured
value
Entire measuring
range
C152
Calibration status is
displayed
o.k.
Exxx
Store calibration
results?
yes
no
new
C153
C1 (6)
Select calibration
Display
Information
If C152 = E xxx, then only no
or new (Exception:
calibration warning E84). If
new, return to C. If yes / no,
return to “Measurement”.
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
Refl = Fitting with
reflection
compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration
data (6)
C161
Calibration point 1
is
displayed
Comparison value
Deviation relative to
standard sensor
(= 100 %)
C162
Calibration point 2
is
displayed
Comparison value
Deviation relative to
standard sensor
(= 100 %)
C163
Calibration point 3
is
displayed
Comparison value
Deviation relative to
standard sensor
(= 100 %)
C164
Slope 1 is displayed
Current value
Slope of characteristic 1 of
the sensor
C165
Slope 2 is displayed
Current value
Slope of characteristic 2 of
the sensor
Triton TR8
Page 45
Coding
Field
Selection or Range
Default in Bold
C166
Conversion factor is
displayed
Current value
Display
Information
Conversion factor of internal
turbidity units
into displayed unit
3.4.2 OFFSET
The settings in the OFFSET function group can be used to calibrate the measurement to a reference
measurement. This requires a linear shift of all the measured values, i.e. the adjustment is determined
for one measured value, and all others are calculated using the same adjustment.
Note! Following a calibration, the offset is automatically set to zero.
Selection or Range
Default in Bold
Coding
Field
V
Function group
OFFSET
V1
Enter absolute value
Current measured
value
V2
Enter offset
Current offset
V3
Calibration status is
displayed
o.k.
E xxx
V4
Store calibration
result?
yes
no
new
Display
Information
If V3 = E xxx, then only no or
new. If new, return to V.
If yes/no, return to
"Measurement".
3.4.3 SLOPE
With the settings in the function group SLOPE, a measured value can be adapted to a reference value.
All measured values are proportionally adapted over the entire measuring range according to this
change.
Page 46
Triton TR8
Example:
The displayed measured value is 2.5 g/l. It is adapted to the reference value of 2.0 g/l by means of the
slope function. The change is 20 %, i.e. all measured values are reduced by 20 % over the entire
measuring range.
Note! An OFFSET having been edited before is reset to the factory setting. In contrast to the one-point
calibration, the edited slope can be reset by setting the slope factor to 1.0.
Selection or Range
Default in Bold
Coding
Field
N
Function group
SLOPE
N1
Enter absolute value
Current measured
value
N2
Enter slope
Current slope
N3
Status is displayed
o.k.
E xxx
Store slope?
yes
no
new
N4
Triton TR8
Display
Information
Slope is displayed, can be
edited.
Page 47
4.0 MAINTENANCE
You must carry out maintenance tasks at regular intervals. For this, specify the maintenance times in an
operations logbook or operations calendar.
The following tasks must be carried out:
• Cleaning the transmitter
• Checking the measuring function
• Cleaning the sensor
• Recalibration
4.1 CLEANING THE TRANSMITTER
Clean the front of the housing with usual commercial cleaning agents. In accordance with DIN 42 115,
the front is resistant to:
• Isopropanol • Diluted acids (max. 3%) • Diluted alkalis (max. 5%)
• Esters • Hydrocarbons • Ketones • Household cleaners
Caution! For cleaning purposes, never use:
• Concentrated mineral acids or alkalis • Benzyl alcohol • Methylene chloride • High-pressure steam
4.2 CHECKING THE MEASURING POINT
4.2.1 Transmitter
The TR8 Turbidity sensors cannot be simulated as they contain the complete data processing and all the
measured values are transmitted to TR8 Analyzer using the digital interface RS 485. Therefore a
functional sensor is required for the measuring point test.
Method for testing a measuring point:
• Check that device is operable and that the display reacts appropriately, e.g. by pressing the PLUS key.
• Check the current outputs by carrying out a current simulation (Field O3 (2)).
• Measure the sensor operating voltage: approx. 10 to 16 V at terminals 87 (+) and 88 (–).
• The cause for an incorrect voltage may be present either at the device or at the sensor.
– Replace the sensor.
– If the sensor operating voltage is still too low replace the power supply module (The analyzer
must be returned to the factory for power supply replacement).
Page 48
Triton TR8
• Sensor operating voltage is o.k. but no measured turbidity value even with a new sensor. Replace the
transmitter module (The analyzer must be returned to the factory for power supply replacement).
4.2.2 Triton TR8 Sensor
1. Remove the sensor from the medium.
2. Clean and dry the sensor.
3. Check the slope using CUY22 check unit. A stable measured value (between 2 and 6 FNU) must
register. The exact value is not of importance but must be repeatable for the same sensor.
4.3 CLEANING THE SENSOR
Contamination of the sensor can impact the measurement up to the point that measurement no longer
functions, e.g. by:
• Thick film build-up on the sensor optics. Film build-ups can cause bad measurement results.
• Wiper clogging. This causes bad measurement results.
The sensor must be cleaned regularly to ensure reliable measurement. The frequency and intensity of
the cleaning depend on the medium.
The sensor must be cleaned:
• Prior to every calibration • Regularly during the operation • Prior to returning the sensor for repair.
4.4 RECALIBRATION
After cleaning check the Triton TR8 in the Calibration Cap the reading should be between 2FNU and 6
FNU. If not then you have the following options in the ''Calibration'' menu of the transmitter:
• One-point calibration
• Three-point calibration
• Installation adjustment
• Correction function
• Adjust calibration data
If you are unable to recalibrate the sensor then the sensor must be replaced. When replacing a sensor
pay attention to the following:
• Replacing the Triton TR8 sensor
All factory calibration data are saved in the sensor. The sensor and the assembly are calibrated together.
No additional calibration is required for applications with pure or ultrapure water, if the sensor and the
Triton TR8
Page 49
assembly are replaced together. The calibration data of the sensor are automatically transferred to the
measuring instrument.
5.0 ORDERING INFORMATION
Part#
Description
Triton TR8 Turbidity Sensor with 7 meter cable
1398000-1
Triton TR8 Turbidity Sensor High Range with 7 meter cable
1398000-2
Triton TR8 Turbidity Sensor with 15 meter cable
1398001-1
Triton TR8 Turbidity Sensor High Range with 15 meter cable
1398001-2
Triton TR8 Turbidity Sensor with 7 meter cable & Automatic Wiper
1398010-1
Triton TR8 Turbidity Sensor High Range with 7 meter cable & Automatic Wiper
1398010-2
Triton TR8 Turbidity Sensor with 15 meter cable & Automatic Wiper
1398011-1
Triton TR8 Turbidity Sensor High Range with 15 meter cable & Automatic Wiper
1398011-2
Triton TR8 Turbidity Sensor with 7 meter cable & Flow Through Assembly
1398100-1
Triton TR8 Turbidity Sensor with 15 meter cable & Flow Through Assembly
1398101-1
Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & Flow Through Assembly
1398110-1
Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & Flow Through Assembly
1398111-1
Triton TR8 Turbidity Sensor with 7 meter cable & De-Bubbler Assembly
1398200-1
Triton TR8 Turbidity Sensor with 15 meter cable & De-Bubbler Assembly
1398201-1
Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & De-Bubbler Assembly
1398210-1
Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & De-Bubbler Assembly
1398211-1
(-1) Low Range Sensor < 500 FNU, (-2) High Range Sensor > 500 FNU
5.1 ACCESSORIES
Part #
1000223-1
2000268
3200266
1000219
9640004.cond
1000222-2
1000228-1
1000227
Page 50
Model and Description
Immersion Assembly ( Cap/Cable feed-through, 1 meter down pipe)
Rail Moun3ng Brackets, (2) Quick Release “U” clamps for 2” Guard Rail moun3ng
Wall Moun3ng Bracket
Flow Through Housing
TR8 5 Conductor Cable, per ft.
NEMA 4X Junc3on Box, (2) cable glands, terminal strip, PVC box, 6”x3”x2”, LWD
Service Kit, for Wiper Arm, rubber wipers and mounting material (3 sets)
TR8 Check Unit, used to check Stability of the sensor, (calibration cup)
Triton TR8
6.0 TROUBLESHOOTING
6.1 TROUBLESHOOTING INSTRUCTIONS
The transmitter constantly monitors its functions itself. If an error occurs which the device recognizes,
this is indicated on the display. The error number is shown below the display of the main measured
value. If more than one error occurs, you can call these up with the MINUS key. Refer to the "System
error messages" table for the possible error numbers and remedial measures. Should a malfunction
occur without any transmitter error message, please refer to the "Process-specific errors" or the
"Device-specific errors" tables to localize and rectify the error. These tables provide you with additional
information on any spare parts required.
6.2 SYSTEM ERROR MESSAGES
The system error messages can be called up and selected with the MINUS key.
Error
No.
Display
Test and Remedial Action
E001
EEPROM memory error
E002
Instrument not calibrated,
calibration data invalid, no
user data, user data invalid
(EEPROM error), instrument
software not suitable to
hardware (controller)
E003
Download Error
E004
Instrument software version
not compatible with module
hardware version
1. Switch device off and then
on again.
2. Load device software
compatible with the
Hardware.
3. Load measurementparameter specific device
software.
4. If the error persists, send
in the device for repair
to your local service
organization or replace the
device.
Invalid configuration. Repeat
download.
Load software compatible
with hardware. Load
measurement-parameter
specific device software.
E007
Transmitter malfunction,
instrument software not
compatible with transmitter
version
E008
Sensor or sensor connection
faulty
Triton TR8
Alarm
Contact
Facty
Yes
Error
Current
User
Facty
No
User
Auto
Cleaning
Start
Facty User
_
1
Yes
No
_
Yes
No
No
Yes
No
No
Contact ECD Service.
Yes
No
_
Check sensor and sensor
connection (Service).
Check initialization of wiper.
Yes
No
No
1
1
Page 51
Error
No.
Display
Test and Remedial Action
Alarm
Contact
Facty
E026
Wiper error
E045
Calibration aborted
Below main parameter
measuring
range
Main parameter measuring
range exceeded
Below temperature
measuring range
Temperature measuring
range exceeded
Below current output range
1
Current output range 1
exceeded
Below current output range
2
Current output range 2
exceeded
Set point exceeded limit
contactor 1
Set point exceeded limit
contactor 2
Set point exceeded limit
contactor 3
Set point exceeded limit
contactor 4
Measuring value outside
concentration table
Current output 1 range too
small
Current output 2 range too
small
E055
E057
E059
E061
E063
E064
E065
E066
E067
E068
E069
E070
E079
E080
E081
E084
Calibration warning
E085
Incorrect setting for error
current
E100
Current simulation active
Page 52
Check wiper and test
function using manual
control if necessary.
Repeat calibration
Check measurement and
connections; check device
and measuring cable.
Check measured value and
current assignment.
Error
Current
User
Facty
User
Auto
Cleaning
Start
Facty User
Yes
No
No
No
No
_
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
No
No
_
1
No
No
_
1
No
No
No
No
No
_
1
No
No
_
1
1
Check configuration.
Clean sensor, check table.
Decrease current output
spreading
Calibration data are within
limits but deviate from
standard values by a factor
of more than two.
If the current range "0 to 20
mA" was selected in field
O311, the error current "2.4
mA" may not be set.
Triton TR8
Error
No.
Display
Test and Remedial Action
E101
Service function active
Switch off service function or
switch device off and then
on again.
E102
E106
E116
Manual mode active
Download yes
Download error
E152
PCS alarm
E153
Offset
Below lower alarm threshold
for period exceeding alarm
delay
Above upper alarm
threshold for period
exceeding alarm delay
Current value undershoots
alarm threshold (CC set
point) for longer than the set
permissible maximum
period
Current value exceeds alarm
threshold (CC set point) for
longer than the set
permissible maximum
period
Alarm
Contact
Facty
E154
E155
E156
E157
Wait for download to finish.
Repeat download.
Check sensor and
connection.
Adjustment range exceeded
Perform manual comparison
measurement if necessary.
Service sensor and
recalibrate.
Error
Current
User
Facty
User
Auto
Cleaning
Start
Facty User
No
No
_
1
No
No
No
No
No
No
_
_
_
1
1
1
No
No
No
No
No
No
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
No
E162
Dosage stop
Check settings in the
CURRENT INPUT and CHECK
function groups.
Yes
No
No
E171
Flow in main stream too low
or zero
Restore flow.
Yes
No
No
Check process variables at
sending measuring
E172
Yes
No
No
instrument. Change range
assignment if necessary.
Check process variables at
E173 Current input < 4 mA
sending measuring
Yes
No
No
instrument.
Check process variables at
sending measuring
E174 Current input > 20 mA
Yes
No
No
instrument. Change range
assignment if necessary.
_ 1 If this error occurs, there is no possibility of starting a cleaning session (field F8 not applicable with this error).
Switch-off limit for current
input
exceeded
Triton TR8
Page 53
6.3 PROCESS SPECIFIC ERRORS
Error
Possible Cause
Sensor / sensor cable
defective
Value indicated
0.0
Sensor extension line
interrupted
Check junction boxes and
line.
Incorrect sensor connection
Check connection.
Replace module MKT1 for
testing.
Replace module LSGA (AC
power supply unit) or LSGD
(DC power supply unit) for
testing.
Switch instrument off and
back on, wiper has to wipe
once.
Clean optics
Instrument input defective
Data transfer error
Wiper blocked
Display value 0.0
Sensor completely blocked
Fixed incorrect
measured value
Impermissible instrument
operating state (no response
to key actuation)
Measuring cable
interference
Measured value
fluctuates
Signal output line
interference
Irregular flow rate /
turbulence / air bubbles / big
solids particles
No or incorrect sensor
calibration
Display value
implausible / no
or creeping
change of display
Test and/or Remedial
Measure
Test with new or with
different functional
sensor.
Switch instrument off and
back on
Connect cable screen acc. to
connection diagram (do not
ground)
Check line routing Try
separate line routing,
grounding screen to
PLC/PCS.
Choose better place of
installation or eliminate
turbulences. Possibly use
large measured value
damping factor. Set gas
bubble barrier to 100 %.
Calibration with original
sample required for
concentration or solids
concentration
Clean sensor.
Sensor soiled
Use spray cleaning.
Use wiper version.
Page 54
Equipment, spare parts,
personnel
Triton TR8 (either type is
suitable for rough function
test)
Sensor simulation see
chapter "Maintenance of the
entire measuring point".
See chapter "Wiring".
Repair at manufacturer only.
Repair at manufacturer only.
Repair at manufacturer only.
Use spray cleaning or wiper.
EMC problem: check line
routing if problem persists,
check for possible sources of
interference.
See chapter "Wiring".
Separate signal output,
meas. input and
supply lines.
Measured value damping
see field A5.
See chapter "Calibration".
Remove coarse coats with
brush. Remove carbonates
and similar coats with 3 %
hydrochloric acid. Remove
organic coats and grease
with oxidation agents and /
or grease solvents.
See instructions of assembly
used for spray cleaning.
Wiper upgrade at factory.
Triton TR8
Error
Display value
implausible / no
or creeping
change of display
Possible Cause
Test and/or Remedial
Measure
Wiper rubber defective
Replace wiper arm.
Sensor installed in “dead”
zone or air cushion in
assembly or flange
Incorrect sensor orientation
Incorrect
temperature
value
Controller or
timer cannot be
activated
If temperature display is
required: replace sensor.
Turbidity measurement itself
does not require
temperature measurement.
No relay module installed
Install LSR1-2 or LSR1-4
module.
Repair at manufacturer only.
Pickup delay setting too long
"Hold" function active
Controller/limit
contact works
continuously
Controller in "Manual/On"
mode
Dropout delay setting too
long
Control loop interruption
Line open or short-circuited
No turbidity
current output
signal
Fixed current
output signal
Incorrect current
output signal
Triton TR8
Frontal “bombardment” of
measuring surface with
highly viscous solids may
result in an adhering coating.
Temperature sensor
defective
Controller switched off
Controller in "Manual/Off"
mode
Controller/limit
contact
does not work
Check installation
conditions, move sensor to
area with optimum flow
conditions. Caution if
installed in horizontal lines!
Orient sensor:
• Measuring surface should
face flow in normal media.
• Orient meas. surface at 90°
to flow in media with high
solids concentration
Equipment, spare parts,
personnel
Wiper arm service kit
1000228-1
Activate controller.
Choose "Auto" or
"Manual/On" mode.
Disable or shorten pickup
delay.
“Automatic Hold” during
calibration,
“Hold” input activated;
“Hold” via keyboard active.
Set controller to
"Manual/Off" or "Auto".
See fields R2xx.
Shorten dropout delay.
See field R2xx.
Check measured value,
current output, actuators,
chemical supply.
Disconnect line and measure
directly on instrument.
Total load in current loop
excessive
(>500Ω)
Instrument with PROFIBUS
PA/DP
Current simulation active
PA/DP instruments have no
current output.
Switch off simulation.
Impermissible operating
state of processor system
Switch instrument off and
back on.
Incorrect current assignment
Check current assignment:
0–20 mA or 4–20 mA?
Disconnect line and measure
line.
Keyboard, REL-key
See fields R2xx.
See fields S2 to S4.
Keyboard, REL and AUTO
keys
mA meter 0–20 mA
Ohmmeter
See field O3 (2).
EMC problem: check
installation, screen,
grounding if problem
persists.
Field O311
Page 55
Error
Possible Cause
Incorrect current
output signal
Current output
table not
accepted
Total load in current loop
excessive (> 500 Ω.)
Test and/or Remedial
Measure
Disconnect line and measure
line.
Value interval too small
Select practical intervals.
No central HART module
Verify by looking at
nameplate: HART = -xxx5xx
and -xxx6xx.
No or wrong DD (device
description)
HART interface missing
Instrument not registered
with HART server
Load too low (load > 230 Ω
required)
HART receiver (e.g. FXA 191)
not connected via load but
via power supply
Incorrect device address
(address = 0 for single
operation, address > 0 for
multi-drop operation)
Line capacitance too high
Line interferences
For further information see
“HART® - Field
communication Manual”
Several devices set to same
address
Set addresses correctly.
Communication not possible
with several devices set to
same address.
No central PA/DP module
Verify by looking at
nameplate: PA = -xxx3xx /DP
= xxx4xx.
Upgrade to LSCP module,
see chapter "Spare parts".
Incorrect instrument
software version
Commuwin (CW) II:
Incompatible CW II and
instrument software
versions
No or incorrect DD/DLL
Incorrect baud rate setting
for segment coupler in DPV1 server
Incorrect station (master)
addressed or duplicate
address
Incorrect station (slaves)
address
Bus line not terminated
Line problems (too long,
cross section too small; not
shielded, screen not
grounded, wires not twisted)
For further information, see
"PROFIBUS PA/DP – Field
communication Manual”
No HART
communication
No PROFIBUS®
communication
Page 56
Equipment, spare parts,
personnel
Ohmmeter
Upgrade to LSCH-H1 / -H2.
Triton TR8
Error
Possible Cause
No PROFIBUS®
communication
Bus voltage too low (bus
supply voltage typ. 24 V DC
for non-Ex)
Test and/or Remedial
Measure
Voltage at instrument’s
PA/DP connector must be at
least 9 V.
Equipment, spare parts,
personnel
6.4 INSTRUMENT SPECIFIC ERRORS
The following table helps you during the diagnosis and points to any spare parts required. Depending on
the degree of difficulty and the measuring equipment present, diagnosis is carried out by:
• Trained operator personnel
• The user’s trained electrical technicians
• Company responsible for system installation/operation
Information on the exact spare part designations and on how to install these parts can be found in the
"Accessories" section.
Error
Possible Cause
Device cannot be
operated, display
value 9999
Operation locked
No line voltage
Supply voltage wrong/too
low
Connection faulty
Display dark, no
light-emitting
diode active
Device fuse defective
Power unit defective
Central module defective
Ribbon cable loose
or defective
Device gets hot
Incorrect
measurement
turbidity and/or
temperature
Triton TR8
Voltage wrong/too high
Power unit defective
Transmitter module
defective. Verify the error is
not in the cabling or in the
sensor, See Process Errors
Test and/or Remedial
Measure
Equipment, spare parts,
personnel
Press CAL and MINUS keys
simultaneously.
See "Function of keys"
section.
Check whether line voltage is
present.
Compare actual line voltage
and nameplate data.
Terminal not tightened;
insulation jammed; wrong
terminals used.
Compare line voltage and
the nameplate data and
replace fuse.
Replace power unit, note
variant.
Replace central module,
note variant.
Check ribbon cable, replace
if necessary.
Compare line voltage and
nameplate data.
Replace power unit.
Electrical technician/e.g.
multimeter
Electrical technician/e.g.
multimeter
Simulation of the sensor is
not possible. Test the input
with new or different sensor.
Electrical technician
Electrical technician/suitable
fuse
Consult Factory
Consult Factory
Consult Factory
User, electrical technician
Consult Factory
If test negative: replace
module, Consult Factory
Page 57
Error
Incorrect meas.
Turbidity and/or
temperature
Current output,
current value
incorrect
Possible Cause
Data transfer disturbed
(EMC)
Wrong sensor cable / cable
too long
Adjustment not correct
Load too big
Shunt/short to ground in
current loop
Incorrect mode of operation
No current output
signal
Current output stage
defective
No function of
additional relay
defective or loose ribbon
cable
Test and/or Remedial
Measure
Check cable routing.
Separate sensor cable from
supply cables.
Maximum cable length with
extension is 200 m (656 ft.)
Check with installed current
simulation, connect mA
meter directly to current
output.
Check whether 0–20 mA or
4–20 mA is selected.
Check with installed current
simulation, connect mA
meter directly to current
output.
Check ribbon cable seating,
renew cable if required.
Equipment, spare parts,
personnel
Connect sensor cable screen
to "S" terminal , do not
ground.
If simulation value incorrect:
adjustment in factory or new
module required.
If simulation values correct:
check current loop for load
and shunts.
If simulation value incorrect:
new module required.
6.5 SENSOR TROUBLESHOOTING
6.5.1 Troubleshooting instructions
Troubleshooting must take account of the whole measuring system:
• Transmitter • Electrical leads and connectors • Assembly • Sensor
The possible causes of failure listed in the following table primarily refer to the sensor.
Problem
No display, no sensor
reaction
Reading too high or
too low
Reading greatly
fluctuating
Measured value jump to
9999 FNU
Error 8
Wiper does not find end
position
Page 58
Check
Mains voltage at transmitter?
Sensor connected correctly?
Medium flow present?
Film formation on optical windows?
Film formation on optical windows?
Gas bubbles present?
Sensor calibrated?
Check data record.
Check with check unit.
Gas bubbles present?
Check mounting location.
Gas bubble formation on the optical
windows?
Water in sensor?
Cable breakage?
Incorrect cable connection?
Visual inspection: Initial position not in
end position?
Remedy
Connect mains voltage.
Set up correct connection.
Create flow.
Clean the sensor.
Clean the sensor.
Remove gas bubbles.
Calibrate.
Change, if necessary.
Check in factory.
Remove gas bubbles.
Select other mounting location.
Change orientation. Adjust wiper
interval. Increase gas bubble factor in
the B9 menu.
Send sensor to service.
Check cable.
Check wiring.
Send sensor to service.
Triton TR8
6.5.2 Checking the Sensor
The sensor may only be checked by trained personnel. You will require a voltmeter.
Problem
Voltage check
Slope check
Zero point check
Check
Check supply voltage at transmitter
(sensor connected)
Check slope using 1000227-1, check unit
Produce zero water by filtration (0.2 μm)
Remedy
10 to 16 V between terminals 87 and 88
Stable measured value (1 to 6 FNU)
< 0.1 FNU
Note!
If the values deviate from the set points, carry out trouble-shooting as per the trouble-shooting
instructions or contact your sales center.
7.0 ENGINEERING DOCUMENTATION
7.1 SPECIFICATIONS
7.1.1Triton TR8 Sensor
Measuring Principle:
Measuring Range:
Maximum Measured Error:
Repeatability:
Wavelength:
Optical Reference Compensation:
Factory Calibration:
Temperature/Pressure:
Connecting Cable:
Maximum Cable Length:
Temperature Sensor:
Operating Range:
Storage Range:
Leakage Protection:
Wetted Materials:
Process Connections:
Triton TR8
Nephlelometric 90° NIR Scattered light according to DIN/ISO 7027
(-1) 0.000 to 9999 FNU
(-2) 0.00 to 9999 FNU
0.00 to 3000 ppm
0.00 to 9999 ppm
0.0 to 3.0 g/l
0.0 to 300 g/l
0.0 to 20.0 %
0.0 to 200%
Less Than 5% of the measured value
Less Than 1 % of measured value
880 nm
By reference Diodes
Formazine standard and SiO2
25 °C/ 6 bar to 50 °C/ 1bar
4 cores shielded
200 m
30 k-ohm NTC
-5 – 50 °C
-20 – 60 °C
IP68
Body: PVC, PPS GF40
Window: Sapphire
Wiper: Rubber
Cable: Polyolefin
Flow Assemblies: PVC, PE
G1 and 3/4” NPT
Page 59
7.1.2 Triton Analyzer Specifications
Inputs
Measurements
Measurement principle
Measured Variables
Measuring range (Turbidity)
Measuring range (Temp)
Binary inputs 1 and 2
Current input
Load
Outputs
Output signal
Signal on alarm
Load
Transmission range
Signal resolution
Insulation voltage
Overvoltage protection
Auxiliary voltage output
Contact outputs (floating
changeover contacts)
Limit contactor
Alarm
Turbidity and temperature
The TR8 sensor communicates with the Analyzer through a digital
interface. The analyzer receives a standardized turbidity and
temperature signal from the TR8 sensor.
Turbidity, Solids Content, Temperature
0.000 to 9999 FNU/NTU
0.00 to 9999 ppm
0.0 to 300 g/l
0.0 to 200.0 %
-5.0 to +70.0 °C (+23 to +158 °F)
Voltage: 10 ... 50 V, Current consumption: max. 10 mA
4 to 20 mA, galvanically isolated
Load: 260 Ω for 20 mA (voltage drop 5.2 V)
0/4 to 20 mA, galvanically isolated, active, Terminals 31 & 32
2.4 or 22 mA in case of an error
max. 500 Ω
adjustable, min. Δ 0.1 FNU, Δ 0.1 (Temp) adjustable, Δ 10 to Δ 100
ppm, Δ 0.1 g/l, Δ 0.1 %
% of measuring range
max. 700 Digits/mA
max. 350 Veff / 500 V DC
according to EN 61000-4-5
Output voltage:
15 V ± 0,6 V
Output current:
max. 10 mA
Switching current with ohmic
Max. 2 A
load (cos ϕ = 1):
Switching current with inductive Max. 2 A
load (cos ϕ = 0.4):
Switching voltage:
Max. 250 V AC, 30 V DC
Switching capacity with ohmic
Max. 500 VA , 60 W DC
load (cos ϕ = 1):
Switching capacity with inductive Max. 500 VA
load (cos ϕ = 0.4):
Pick-up/drop-out delay
0 to 2000 s
Function (switchable):
Latching/momentary contact
Alarm delay:
0 to 2000 s (min)
Power Supply
Supply voltage
Power consumption
Mains fuse
Page 60
Depending on ordered version: 100/115/230 V AC +10/-15 %, 48 ...
62 Hz 24 V AC/DC +20/-15 %
Max. 7.5 VA
Fine-wire fuse, semi-delay 250 V/3.15 A
Triton TR8
Performance Characteristics
Measured value resolution
Maximum measured error
Repeatability
Offset range
TR8 Turbidity
0.001 FNU, 0.01 ppm, 0.01 g/l, 0.01 %
Temperature
0.1 °C
Display TR8
± 2 % of measured value (min. 0.02 FNU)
Display (Temp)
max. 1.0 % of measuring range
Signal Output TR8
1 % of current output range (min. 0.02 FNU)
Signal Output (Temp) max. 1.25 % of current output range
max. 1 % of measured value
TR8
± 99.99 FNU, ± 99.99 ppm, ± 99.9 g/l, ± 99.9
%
Temperature
± 5 °C
Environment
Ambient temperature range
Ambient temperature limits
Storage temperature
Electromagnetic compatibility
Overvoltage protection
Degree of protection
Relative humidity
Triton TR8
-10 to +55 C (+14 to +131 °F)
–20 to +60 C (-4 to +140 °F)
–25 to +65 C (-13 to +149 °F)
Interference emission and interference immunity as per EN 61326:
1997 / A1:1998
As per EN
For outputs, binary inputs and current input
61000-4-5:1995
Panel-mounted
IP 54 (front), IP 30 (housing)
instrument:
Field device:
IP 65
10 to 95%, not condensing
Page 61
7.2 OUTLINE & DIMENSIONAL DRAWING
7.3 WIRING DIAGRAM
Optional Relay Card
4-20 mA
Page 62
Triton TR8
Triton TR8
Page 63