Triton TR8 Instruction Manual
Transcription
Triton TR8 Instruction Manual
ELECTRO-CHEMICAL DEVICES, INC. Triton TR8 Instruction Manual Turbidity Analyzer/Controller ELECTRO-CHEMICAL DEVICES, INC. 1681 Kettering, Irvine, CA 92614, USA Tel: +1-949-336-6060, FAX: +1-949-336-6064 www.ecdi.com Rev: B - 08/09 PREFACE Purchasing products from Electro-Chemical Devices, Inc. provides you with the finest liquid analytical instrumentation available. If this is your first purchase from ECD, please read this manual before installing and commissioning your new equipment. If there are any questions concerning this equipment, please contact your local ECD representative, or the factory directly at: Electro-Chemical Devices, Inc. 1681 Kettering St. Irvine, CA 92614 USA Telephone: +1-949-336-6060 FAX: +1-949-336-6064 Website: www.ecdi.com Email: sales@ecdi.com © 2009 Electro-Chemical Devices, Inc. All rights reserved. No part of this manual may be used or reproduced in any form or by any means, or stored in a database or retrieval system without prior written permission from Electro-Chemical Devices, Inc. Making copies of any part of this manual for any purpose other than personal use is a violation of United States copyright laws. Document printed in the United States of America. Page ii Triton TR8 TABLE OF CONTENTS PREFACE ........................................................................................................................................................ ii TABLE OF CONTENTS.................................................................................................................................... iii WARRANTY .................................................................................................................................................. vi IMPORTANT SERVICE INFORMATION .......................................................................................................... vi UNPACKING THE INSTRUMENT................................................................................................................... vii 1.0 GENERAL DESCRIPTION ........................................................................................................................... 1 1.1 TR8 SENSOR FEATURES ....................................................................................................................... 1 1.1.1 Sensor Options ............................................................................................................................. 2 1.2 TR8 Analyzer Features ........................................................................................................................ 2 1.2.1 Analyzer Options .......................................................................................................................... 2 2.0 INSTALLATION ......................................................................................................................................... 3 2.1 ANALYZER MOUNTING........................................................................................................................ 3 2.1.1 Panel Mounting ............................................................................................................................ 3 2.1.2 Pipe Mounting.............................................................................................................................. 3 2.2 SENSOR MOUNTING ........................................................................................................................... 3 2.2.1 Immersion Assembly .................................................................................................................... 4 2.2.2 Flow-Through Assembly ............................................................................................................... 5 2.2.3 De-Bubbler Flow Assembly .......................................................................................................... 6 2.3 WIRING ................................................................................................................................................ 7 2.3.1 Sensor Wiring ............................................................................................................................... 7 2.3.2 Instrument Wiring ........................................................................................................................ 7 3.0 OPERATION ............................................................................................................................................. 8 3.0.1 General Information .................................................................................................................... 8 3.0.2 Liquid Crystal Display ................................................................................................................... 9 3.1 KEY Assignment ................................................................................................................................. 10 3.1.1 Auto/Manual Mode Switching ................................................................................................... 11 3.2 MENUS .............................................................................................................................................. 11 3.2.1 Access Codes .............................................................................................................................. 12 3.2.2 Menu structure .......................................................................................................................... 13 3.2.3 HOLD Function ........................................................................................................................... 13 Triton TR8 Page iii 3.3 START UP and SETUP FUNCTION GROUP MENUS ............................................................................ 14 3.3.1 SETUP 1: TURBIDITY ................................................................................................................... 14 3.3.2 SETUP 2: TEMPERATURE and Wiper .......................................................................................... 15 3.3.3 CURRENT INPUT ......................................................................................................................... 16 3.3.4 CURRENT OUTPUT ..................................................................................................................... 17 3.3.5 ALARM FUNCTIONS .................................................................................................................... 20 3.3.6 CHECK ......................................................................................................................................... 21 3.3.7 RELAY ......................................................................................................................................... 23 3.3.8 CONCENTRATION MEASUREMENT ............................................................................................ 35 3.3.9 SERVICE ...................................................................................................................................... 36 3.3.10 INTERFACE................................................................................................................................ 38 3.4 CALIBRATION FUNCTION GROUP MENUS ........................................................................................ 38 3.4.1 CALIBRATION.............................................................................................................................. 41 3.4.2 OFFSET ....................................................................................................................................... 46 3.4.3 SLOPE ......................................................................................................................................... 46 4.0 MAINTENANCE ...................................................................................................................................... 48 4.1 CLEANING THE TRANSMITTER .......................................................................................................... 48 4.2 CHECKING THE MEASURING POINT .................................................................................................. 48 4.2.1 Transmitter ................................................................................................................................ 48 4.2.2 Triton TR8 Sensor ....................................................................................................................... 49 4.3 CLEANING THE SENSOR..................................................................................................................... 49 4.4 RECALIBRATION ................................................................................................................................ 49 5.0 ORDERING INFORMATION .................................................................................................................... 50 5.1 ACCESSORIES ..................................................................................................................................... 50 6.0 TROUBLESHOOTING .............................................................................................................................. 51 6.1 TROUBLESHOOTING INSTRUCTIONS ................................................................................................. 51 6.2 SYSTEM ERROR MESSAGES ............................................................................................................... 51 6.3 PROCESS SPECIFIC ERRORS ............................................................................................................... 54 6.4 INSTRUMENT SPECIFIC ERRORS ........................................................................................................ 57 6.5 SENSOR TROUBLESHOOTING ............................................................................................................ 58 6.5.1 Troubleshooting instructions ..................................................................................................... 58 6.5.2 Checking the Sensor ................................................................................................................... 59 Page iv Triton TR8 7.0 ENGINEERING DOCUMENTATION ......................................................................................................... 59 7.1 SPECIFICATIONS ................................................................................................................................ 59 7.1.1Triton TR8 Sensor ........................................................................................................................ 59 7.1.2 Triton Analyzer Specifications .................................................................................................... 60 7.2 OUTLINE & DIMENSIONAL DRAWING ............................................................................................... 62 7.3 WIRING DIAGRAM............................................................................................................................. 62 Triton TR8 Page v WARRANTY Electro-Chemical Devices, Inc. (ECD) warrants all products it manufactures to be free from defect in materials and factory workmanship, and agrees to repair or replace any product that fails to perform, as specified, within one (1) year after date of shipment. This warranty shall not apply to any product that has been: 1. 2. 3. Subjected to misuse, negligence or accident; Connected, installed, adjusted or otherwise used not in accordance with the instructions furnished by ECD; Repaired, modified or altered by persons not authorized by ECD, resulting in injury to the performance, stability or reliability of the product. This warranty is in lieu of any other warranty, expressed or implied. ECD reserves the right to make changes in the design or construction of its products at any time, without prior notification, and without incurring any obligation to make any changes in previously delivered products. Seller’s sole liabilities and the buyer’s sole remedies under this agreement shall be limited to a refund in the purchase price, or at ECD’s discretion, to the repair or replacement of any product that proves, upon ECD’s examination, to be defective, when returned to the factory, transportation prepaid by the buyer, within one (1) year of the product’s original shipment date. Seller shall not be liable for damages consequential or incidental to defects in any product, for failure of delivery in whole or in part, for injuries resulting from its use, or for any other cause. This warranty and the writing attached constitute the full understanding of seller and the buyer, and no terms, conditions, understanding, or agreement purporting to modify or vary the terms hereof shall be binding unless hereafter made in writing and signed by an authorized official of Electro-Chemical Devices, Inc. This warranty does not cover pH, ORP or Specific Ion measurement, reference or combination electrodes or electrode cartridges that have been commissioned in service. IMPORTANT SERVICE INFORMATION Use only factory authorized components for repair. Tampering or unauthorized substitution of components may adversely affect the operation of this product and may void the warranty. If service or repair is required, please obtain the serial number(s) or sales order number of the product(s) in question and contact ECD’s Service Department at: +1-800-729-1333 (USA/Canada) or +1-949-336-6060 or email Service@ecdi.com A Return Material Authorization (RMA) number must be obtained from the service department before returning any material to ECD. All material returned to ECD shall be shipped prepaid to the factory. Page vi Triton TR8 UNPACKING THE INSTRUMENT Your Electro-Chemical Devices instrument has been carefully packaged to protect it from damage during shipment and dry storage. Upon receipt please follow the procedure outlined below. 1. Before unpacking, inspect the condition of the shipping container to verify proper handling by the carrier. If damage is noted, save the shipping container as proof of mishandling for the carrier. 2. Check the contents of the shipping container with the items and quantities shown on the packing list. Immediately report any discrepancies to ECD. 3. Save the original packing material until you are satisfied with the contents. In the event the product(s) must be returned to ECD, the packing material will allow you to properly ship it to ECD. 4. Familiarize yourself with the instrument before installation, and follow proper installation and wiring procedures. Triton TR8 Page vii 1.0 GENERAL DESCRIPTION The Triton TR8 is a nephelometric turbidity analyzer designed for use in water and wastewater. Turbidity, the cloudiness or haziness of a water sample, is caused by particles suspended in the water, typically clay and silt. Since bacteria and viruses can be attached to these particles, turbidity has become a critical indicator of the overall water quality. The Triton TR8 uses an optical method for determining the turbidity, a light beam is directed into the sample where it is scattered by suspended particles in the water. The amount of scattering depends on the amount of material in the water, the wavelength of light used and the size and composition of the suspended particles. The 90° scattered light detection method used by the Triton TR8 sensor is the most common sensor design for turbidity. The Triton TR8 uses a long lived near infrared LED light source (880 nm) and the 90° scattered light method in accordance with ISO 7027 / EN 27027 to assure accurate turbidity values under standardized and comparable conditions. The turbidity value is calculated inside the sensor and digitally transmitted to the Triton TR8 analyzer. The analyzer interprets the signal, displays the turbidity value and transmits the appropriate 4-20 mA output. 1.1 TR8 SENSOR FEATURES The TR8 sensor has a high sensitivity at low levels of turbidity, a simple optical configuration and a balanced sensitivity to all particle sizes. This smart sensor is factory calibrated and ready to use when received. All of the signal processing is performed in the sensor and a noise resistant digital signal is sent to the analyzer/transmitter. A High Range version, (-2), is available for applications with turbidity higher than 500 FNU. Figure 1 Three detectors monitor the light beam at an angle of 90°. The first detector, inside the sensor, (see Figure 1) is the reference detector that compensates for changes in the LED light source (#1) caused by aging or other variations. The second detector (#2) measures a short path length which is best for high concentration measurements. The third detector (#3) measures the longer path length which is best for lower concentrations. The turbidity signal is constantly adjusted versus the reference detector, digital filter functions suppress interfering signals and self monitoring diagnostics assure a highly reliable measurement. An optional mechanical wiper assembly is available to keep the scratch resistant sapphire optical windows clean in dirty applications. The cleaning cycle and duration times are user defined parameters. Triton TR8 Page 1 1.1.1 Sensor Options Part# Description Triton TR8 Turbidity Sensor with 7 meter cable 1398000-1 Triton TR8 Turbidity Sensor High Range with 7 meter cable 1398000-2 Triton TR8 Turbidity Sensor with 15 meter cable 1398001-1 Triton TR8 Turbidity Sensor High Range with 15 meter cable 1398001-2 Triton TR8 Turbidity Sensor with 7 meter cable & Automatic Wiper 1398010-1 Triton TR8 Turbidity Sensor High Range with 7 meter cable & Automatic Wiper 1398010-2 Triton TR8 Turbidity Sensor with 15 meter cable & Automatic Wiper 1398011-1 Triton TR8 Turbidity Sensor High Range with 15 meter cable & Automatic Wiper 1398011-2 Triton TR8 Turbidity Sensor with 7 meter cable & Flow Through Assembly 1398100-1 Triton TR8 Turbidity Sensor with 15 meter cable & Flow Through Assembly 1398101-1 Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & Flow Through Assembly 1398110-1 Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & Flow Through Assembly 1398111-1 Triton TR8 Turbidity Sensor with 7 meter cable & De-Bubbler Assembly 1398200-1 Triton TR8 Turbidity Sensor with 15 meter cable & De-Bubbler Assembly 1398201-1 Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & De-Bubbler Assembly 1398210-1 Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & De-Bubbler Assembly 1398211-1 (-1) Low Range Sensor < 500 FNU, (-2) High Range Sensor > 500 FNU 1.2 TR8 Analyzer Features The Triton TR8 Analyzer/Transmitter is a microprocessor based, menu driven, turbidity/suspended solids transmitter. The NEMA 4X TR8 Analyzer can be Panel mounted, Wall mounted or Pipe mounted. The base unit is available as a 110 VAC, 220 VAC or 24 VDC powered instrument. The standard configuration has a 250VAC/30VDC 2A contact relay that can be used as an alarm. The single 4-20 mA output is configured for 0-10 FNU, although it can be easily configured for NTU, ppm, mg/l or g/l. Optional configurations with (2) 4-20 mA and/or (4) relays/contacts are available upon request to the factory. 1.2.1 Analyzer Options Part# 1290100-1 1290100-2 1290100-3 1290100-* Page 2 Description Triton TR8 Turbidity Analyzer, 115 VAC, (1) 0/4-20 mA output, (1) Failure Alarm Relay* Triton TR8 Turbidity Analyzer, 230 VAC, (1) 0/4-20 mA output, (1) Failure Alarm Relay* Triton TR8 Turbidity Analyzer, 24 VDC, (1) 0/4-20 mA output, (1) Failure Alarm Relay* Consult factory for optional dual 0/4-20 mA outputs and multiple relays up to (4) additional relays Triton TR8 2.0 INSTALLATION Mount the TR8 in a location where there is easy access to the analyzer and sensors. Install the system in an area where vibrations, electromagnetic and radio frequency interference are minimized or absent. Do not mount in direct sunlight or areas of extreme heat. The TR8 is suitable for outdoor use if mounted with a protective cover or sunshield. 2.1 ANALYZER MOUNTING The Universal Mounting allows the Analyzer to be Wall mounted, Panel Mounted or Pipe/Rail Mounted. The drawings below show the Universal Mounting used with the C-22 analyzer for panel mounting and the Universal Mounting with the T-23 transmitter that uses the same mounting as the C-22 for Pipe/Rail Mounting. The Triton TR8 Analyzer has the same dimensions as the Panel Mounted C-22 Analyzer. 2.1.1 Panel Mounting Figure 2 2.1.2 Pipe Mounting Figure 3 2.2 SENSOR MOUNTING The Triton TR8 turbidity sensor can be installed as an immersion sensor into a tank with the optional Immersion assembly PN 1000223, or in-line with a Flow Through fitting, PN 1000219, or the special Debubbler Flow through fitting designed especially for drinking water applications. Care must be taken when installing the sensor in a pipe or near a tank wall to avoid the back scattering of light which leads to a higher sensor response than expected. The lower the turbidity of the measured Triton TR8 Page 3 solution is the greater the possibility of back scattering. The following graph can be used as an installation guideline. Figure 4 2.2.1 Immersion Assembly The immersion assembly consists of a 1 meter down tube, ¾” FNPT on the process end and a cap with cable feed through on the other. Verify that the sensor is at least 6” from the tank bottom and side walls. The cable is not water proof and under no circumstances should the sensor be suspended in the application by the cable. Figure 5 Page 4 Triton TR8 2.2.2 Flow-Through Assembly Conventional turbidity measurements are carried out in an unpressurized sample. When the pressure on a sample (which was pressurized beforehand) is released, fine bubbles are produced which distort the turbidity measurement. The pressure must remain constant or rise as the water moves from a pipe and into the flow cell in order to minimize the formation of air bubbles. The outfall from the flow cell must not create a low pressure in the flow cell by siphoning the sample to drain, see examples below. If air bubbles cannot be eliminated then the mechanical wiper assembly can help remove the bubbles that form on the measurement windows, otherwise use the De-Bubbler Assembly below. Figure 6: Proper Installation examples to avoid bubble formation. Figure 7: Sensor alignment in flow cell with optional spray cleaner and without. Triton TR8 Page 5 Figure 8: Dimensional Drawing of the Flow-Through Assembly with optional union adapters 2.2.3 De-Bubbler Flow Assembly When air bubbles are present in the sample the De-Bubbler Flow assembly effectively removes them from the sample. The lower half of the incoming sample flow is diverted down the central tube (1) to the “Ring Channel” where air bubbles can escape through the outlets, (2), upwards to the exiting sample flow. Bubble free water is pushed through the tubes, (3), into the measuring Chamber, (5). The design facilitates a high flow rate by the TR8 sensor, (6), that optimizes response time and minimizes sedimentation forming in the chamber. A large drain plug, (4), aids in flushing any sediment that does deposit in the chamber. Figure 9: De-Bubbler Flow Cell Diagram Page 6 Figure 10: Dimensional Drawing, De-Bubbler Assembly Triton TR8 2.3 WIRING The various cables, measurement signal, outputs and power, enter the Triton TR8 instrument through the three cable glands on the bottom of the instrument. Only properly trained personnel should attempt to wire the instrument. 2.3.1 Sensor Wiring The Triton TR8 sensor uses a digital communication protocol that allows the sensor to be placed up to 200 meters from the analyzer. The TR8 Sensor is supplied with either a 7 meter or 15 meter cable. Installation at distances greater than 15 meters requires the use of an extension cable, PN 9640004.cond and Junction Box, PN 1000222. Figure 11: Sensor wiring Color Code 2.3.2 Instrument Wiring Optional Relay Card (4-20 mA) Figure 12: Terminal Designations, shown with optional Relay Card Triton TR8 Page 7 3.0 OPERATION 3.0.1 General Information Figure 13: Operator Interface 1. LC display for displaying the measured values and configuration data 2. Field for user labeling 3. 4 main operating keys for calibration and device configuration 4. Changeover switch for automatic/manual control mode of the relays LED (Green) Automatic Mode LED (Yellow) Manual Mode 5. LEDs for status of working relays (switch status) LED (Green) indicates the measured value is within the allowed limits, relay inactive LED (Red) indicates the measured value is outside the allowed limits, relay active 6. LED for alarm function LED (Red) indicates Alarm condition, e.g. continuous limit violation, temperature sensor failure or system error/failure 7. Display of the active contact and key for relay changeover in manual mode LED (Red) indicates active relay Page 8 Triton TR8 3.0.2 Liquid Crystal Display Figure 14: Display Features 1. Indicator for measuring mode (normal operation) 2. Indicator for calibration mode 3. Indicator for setup mode (configuration) 4. Indicator for "Hold" mode (current outputs remain at last current state) 5. Indicator for receipt of a message for devices with communication, e.g. Hart, Profibus 6. Indicator of working status of relays 3 and 4: inactive, active 7. Function code display 8. In measuring mode: The measured variable In setup mode: The configured variable 9. In measuring mode: The secondary measured value In setup or calibration mode: e.g. setting value 10. Auto Temperature Compensation (ATC) 11. Manual Temperature Compensation (MTC) 12. "Error": error display 13. Sensor symbol Triton TR8 Page 9 3.1 KEY Assignment CAL key When you press the CAL key, the device first prompts you for the calibration access code: • Code 22 for calibration • Code 0 or any other code for reading the last calibration data Use the CAL key to accept the calibration data or to switch from field to field within the calibration menu. ENTER key When you press the ENTER key, the device first prompts you for the setup mode access code: • Code 22 for setup and configuration • Code 0 or any other code for reading all configuration data. The ENTER key has several functions: • Calls up the Setup menu from the measuring mode. • Saves (confirms) data entered in the setup mode. • Moves on within function groups. PLUS key and MINUS key In the setup mode, the PLUS and MINUS keys have the following functions: • Selection of function groups. ! Note! Press the MINUS key to select the function groups in the order given in the "System configuration" section. • Configuration of parameters and numerical values • Operation of the relay in manual mode In the measuring mode, you get the following sequence of functions by repeatedly pressing the PLUS key: 1. Temperature display in F 2. Temperature display hidden 3. Current input signal in % 4. Current input signal in mA 5. Measured value display in FNU or NTU (uncompensated value without reflection compensation offset and slope, referred to data set 1) In the measuring mode, the following is displayed in sequence by repeatedly pressing the MINUS key: 1. Current errors are displayed in rotation (max. 10). 2. Once all the errors have been displayed, the standard measurement display appears. In the function group F, an alarm can be defined separately for each error code. REL key In the manual mode, you can use the REL key to switch between the relay and the manual start of cleaning. In the automatic mode, you can use the REL key to read out the switch-on points (for limit contactor) or set points (for PID controller) assigned to the relay in question. Press the PLUS key to jump to the settings of the next relay. Use the REL key to get back to the display mode (automatic return after 30 s). AUTO key You can use the AUTO key to switch between automatic mode and manual mode. Escape function If you press the PLUS and MINUS key simultaneously, you return to the main menu or are taken to the end of calibration if calibrating. If you press the PLUS and MINUS key again, you return to the measuring mode. Page 10 Triton TR8 Locking the keyboard Press the PLUS and ENTER key for at least 3 seconds to lock the keyboard against any unauthorized data entry. All the settings can continue to be read. The code prompt displays the code 9999. Unlocking the keyboard Press the CAL and MINUS key for at least 3 seconds to unlock the keyboard. The code prompt displays the code 0. 3.1.1 Auto/Manual Mode Switching The analyzer is normally in Automatic mode and the relay/contacts are controlled by the instrument. The relay contacts can be actuated manually by switching from the Auto Mode to the Manual Mode as described below. This allows the user to actuate the wiper assembly, cleaning systems or any items actuated by the relays. 1. The transmitter is in Automatic mode. The top LED beside the AUTO key is lit. 2. Press the AUTO key to switch to Manual Mode. 3. To enable the manual mode, enter the code 22 via the PLUS and MINUS keys. The bottom LED beside the AUTO key lights up. 4. Select the relay or the function. You can use the REL key to switch between the relays. The relay selected and the switch status (ON/OFF) is displayed on the second line of the display. In the manual mode, the measured value is displayed continuously (e.g. for measured value monitoring for dosing functions). 5. Switch the relay. It is switched on with PLUS and switched off with MINUS. The relay remains in its switched state until it is switched over again. 6. Press the AUTO key to return to the measuring mode, i.e. to the automatic mode. All the relays are triggered again by the transmitter. 3.2 MENUS The menu selections are addressed by pressing either the CAL key for Calibration Functions or the Enter key for the Set-Up Functions. The analyzer will then prompt for an Access Code. Once the access code has been entered the top level function of the chosen menu will be displayed. Pressing the Enter key will Triton TR8 Page 11 scroll though the available functions. Pressing the + or – keys will scroll through the menus available in that function group. Setup mode (Menu Code) Calibration Mode (Menu Code) • SETUP 1 Turbidity (A) • CALIBRATION (C) • SETUP 2 Temperature (B) • OFFSET (V) • CURRENT INPUT (Z) • SLOPE (N) • CURRENT OUTPUT (O) • ALARM (F) • CHECK (P) • RELAY (R) • CONCENTRATION MEASUREMENT (K) • SERVICE (S) • INTERFACE (I) 3.2.1 Access Codes • Press CAL, then using the + and – keys enter the Access Code 22, press Enter to access to Calibration and Offset menu • Press ENTER, then using the + and – keys enter the Access Code 22, press Enter to access to the setup menus • Simultaneously Press the PLUS and ENTER keys, enter Code 9999 to lock the keyboard • Simultaneously Press the MINUS and CAL keys, enter Code 0 to unlock the keyboard • Press CAL or ENTER + any code: access to read mode, i.e. all the settings can be read but not modified. Page 12 Triton TR8 3.2.2 Menu structure Figure 15: Menu Structure Diagram 1. Functions (parameters selected, numbers entered) 2. Function Groups, scroll backwards and forwards with the PLUS and MINUS keys 3. Switch from function to function with the ENTER key The configuration and calibration functions are arranged in function groups. • In setup mode, select a function group with the PLUS and MINUS keys. • In the function group itself, switch from function to function with the ENTER key. • Within the function, select the desired option with the PLUS and MINUS key or edit the settings with these keys. Then confirm with the ENTER key and continue. • Press the PLUS and MINUS keys simultaneously (Escape function) to exit programming (return to the main menu). • Press the PLUS and MINUS keys simultaneously again to switch to the measuring mode. 3.2.3 HOLD Function During setup and calibration functions, the current output is "frozen"; it retains its last value. "HOLD" appears on the display. If a controller actuating variable is output via current output 2, it is set to 0/4 mA while in Hold. The automatic Hold function is accessed in the SERVICE MENU “S2”. Triton TR8 Page 13 3.3 START UP and SETUP FUNCTION GROUP MENUS After powering the instrument, Set the language in the (S) menu, then adjust all parameters in the (A) menu and finally adjust the settings in the (B) menu. The instrument is now operational. 3.3.1 SETUP 1: TURBIDITY Coding Field A Function Group Setup 1 Selection or Range Default in Bold Display Information Initial Display in SETUP 1 A1 Operation Mode NTU FNU Ppm Mg/l g/l % spec. A2 Display Unit kg/l % t/m A2 is only accessible, if A1 = spec. A3 Display Format XX.xx X.xxx XXX.x XXXX A3 is only accessible, if A1 = spec. A4 Connected Sensor CUS31(-1) CUS41(-2) The transmitter automatically detects which sensor is connected. 10 1 to 60 Measured value damping causes averaging over the specified number of individual measured values. It is used, for example, to stabilize the display with applications that fluctuate a great deal. There is no damping if “1” is entered. A5 Page 14 Measurement Damping Any change in the operation mode causes an automatic reset of the user defined settings. The offsets for turbidity and temperature are reset to zero. Triton TR8 3.3.2 SETUP 2: TEMPERATURE and Wiper Selection or Range Default in Bold Coding Field B Function Group Setup 2 Initial Display in SETUP 2 B1 Switch wiper controller on or off off on auto If “auto” is selected, the wiper is operated in combination with a cleaning function via Timer/Chemoclean (“wipe and clean”). In this case B2 and B3 are not applicable. B2 Enter operating period of wiper 30 s 3 to 999 s B3 Enter pause time between the two wiping cycles 30 min 1 to 7200 min Select calibration data set to be used 3 1 to 3 B5 Copy data sets no 1 →2 1 →3 2 →3 3 →2 B6 Display measured value with reflection compensation? yes no B7 Enter corrected temperature of the temperature sensor Current measured value -5.0 to 100.0 °C B4 Triton TR8 Display Information There are 3 calibration data sets stored in every operation mode (A1). Data set 1 cannot be changed. A hold is active during the load of a new selected data set (independent of the settings in S2). Data set 1 cannot be changed (factory setting).However, it can be used as a basis for a customer calibration data set. To work with the copy of a data set, select the data set in field B4. Displays the measured value with or without reflection compensation. Only effective in NTU, FNU, ppm, mg/l. This entry can be used to calibrate the temperature sensor to an external measurement. Page 15 Coding Field Selection or Range Default in Bold B8 Displays the temperature difference (offset) Current offset -5.0 to 5.0 °C The difference between the temperatures measured and entered is displayed. 3.0 % 0.1 to 100 % Compensates gas bubble formation, which may arise from small amounts of dissolved gas in the medium. 0.1 % = no formation of gas bubbles. 100 % = strong gas bubble formation. For clear media (measured value below 1000 NTU) always set the gas bubble barrier to 100 %. B9 Enter gas bubble barrier Display Information 3.3.3 CURRENT INPUT To use the "Current input" function group, you need a relay board with current input which is Not Part of the Basic Version. With this function group you can monitor process parameters and use these for feed forward control. For this purpose, you must connect the current output of an external measured variable (e.g. flow meter) to the 4 to 20 mA input of the transmitter. The following assignment applies: Flow in main stream Flow meter start of measuring range Flow meter end of measuring range Current signal in mA 4 20 Current input signal in % 0 100 Monitoring of flow in main stream This arrangement is particularly practical if the sample flow through a flow assembly in an open outlet is completely independent of the flow in the main stream. This permits signaling of an alarm condition in the main stream (flow too low or has completely stopped) and triggers dosing switch-off even if the medium flow is retained due to the method of installation. **The Basic Version Does Not Include Functions in Italics** Selection or Range Default in Bold Coding Field Z CURRENT INPUT function group Current input settings. Z1 Select flow monitoring of main stream (with controller switchoff) Flow monitoring may only be switched on if the flowmeter is connected in the main stream. If Z1 = off, fields Z2 to Z5 are not available. Page 16 Off On Display Information Triton TR8 Coding Field Selection or Range Default in Bold Z2 Enter the delay for controller switch-off through current input 0s 0 to 2000 s Z3 Enter the delay for controller switch-on through current input 0s 0 to 2000 s Z4 Enter the switch-off limit value for the current input 50% 0 to 100% Z5 Enter the switch-off direction for the current input Low High Z6 Select feed forward control to PID controller Off Lin = linear Basic Z7 Enter value for Feed forward control at which gain = 1 applies 50% 0 to 100% Display Information Brief flow shortfalls can be suppressed by a delay and do not result in controller switch-off. In the case of a controller, a delay until a representative measured value is received is useful if the flow fails for an extended period. 0 to 100% corresponds to 4 to 20 mA at the current input. Observe measured value assignment to the current output of the flow meter. The controller is switched off if the value entered in Z4 is undershot or overshot. If Z6 = off, the field Z7 is not available. Z6 = basic: disturbance variable only affects the basic load (alternatively dosing in proportion to quantity can be used if usual PID controlling is not possible, due to a defective sensor, for example). When the value is set, the controller actuating variable is the same size when feed forward control is switched on as when feed forward control is switched off. 3.3.4 CURRENT OUTPUT Use the "Current output" function group to configure the individual outputs. You can enter either a linear characteristic (O3 (1)) or you can simulate a current output value (O3 (2)) to check the current outputs. In addition, a user-defined current output Table (O3 (3)) is available with the Plus Package. If a second current output is present, you can output the controller actuating variable in accordance with field R 237 via the current output. Exception: if you have chosen a "continuous controller" for current output 2, you cannot enter a user-defined current output Table on current output 1. In a User Defined Output Table, the variable must steadily increase across the table or decrease, the variable cannot change direction. The distance per mA (SV/mA) between two table values must be greater than: 0.005 FNU, 0.005 NTU, 0.005 ppm, 0.005 mg/l, 0.005 %, 0.05 g/l or 0.25 C°. Triton TR8 Page 17 The values for the sample characteristic are entered in the following table. The distance per mA can be calculated from Δ signal / Δ mA. Value Pair Signal Value (SV) Current Value, mA SV/mA 1 2 3 0 3000 9000 4 16 20 250 1500 First enter the desired current output configuration into the following blank table with a pencil. Calculate the resulting signal distance per mA to observe the necessary minimum slope. Then enter the values in the device. Value Pair Signal Value (SV) Current Value, mA SV/mA 1 2 3 4 5 6 7 8 9 10 Selection or Range Default in Bold Coding Field O CURRENT OUTPUT function group O1 Select current output Out1 Out 2 O2 Select measured variable for 2nd current output °C mg/l Contr O3 (1) Enter or output linear characteristic lin = linear (1) sim = simulation(2) tab = table (3) Page 18 Display Information Configuration of the current output (does not apply for PROFIBUS). Output 2 not available on all versions. A characteristic can be selected for every output. R237 = curr (current output 2) can only be selected if O2 = Contr is selected (relay board required). The characteristic can have a positive or negative slope for the measured value output. In the case of actuating variable output (O2 = Contr), an increasing current corresponds to an increasing actuating variable. Triton TR8 Coding O311 O312 O313 O3 (2) Field Selection or Range Default in Bold Select current range 4 to 20 mA 0 to 20 mA 0/4 mA value: Enter corresponding turbidity or temperature value 20 mA value: Enter corresponding turbidity or temperature value Simulate current output 0.000 FNU 0.000 NTU 0.000 ppm 0.000 mg/l 0.000 g/l 0.000 kg/l 0.000 t/m 0.000 % 0.000 °C 10.00 FNU 10.00 NTU 10.00 ppm 10.00 mg/l 300 g/l / 3.00 g/l 99.99 kg/l 99.99 t/m 10.0 % 100 °C Lin = linear (1) Sim = simulation(2) Tab = table (3) Enter simulation value Current value 0.00 to 22.00 mA Enter current output table (only for Plus Package) lin = linear (1) sim = simulation(2) tab = table (3) O331 Select table options read edit O332 Enter number of table value pairs 1 1 to 10 O321 O3 (3) Triton TR8 Display Information Here you can enter the turbidity or temperature value at which the min. current value (0/4 mA) is applied at the transmitter output. Display format from A3 Here you can enter the turbidity or temperature value at which the maximum current value (20 mA) is applied at the transmitter output. Display format from A3 Simulation is not ended until (1) or (3) is selected. Entering a current value results in this value being directly output at the current output. Values can also be added or altered at a later stage. The values entered are automatically sorted by increasing current value. Enter the number of pairs from the x and y value (measured value and current value) here. Page 19 Coding O333 O334 Field Selection or Range Default in Bold Display Information Select table value pair 1 1to No. elements Assigned The function chain O333 ... O335 will run through as many times as correspond to the value in O332. "Assign" appears as the last step. After confirmation the system jumps to O336. Enter x value 0.000 FNU 0.000 NTU 0.000 ppm 0.000 mg/l 0.000 g/l 0.000 kg/l 0.000 t/m 0.000 % 0.000 °C x value = measured value specified by user. O335 Enter y value 4.00 mA 0.00 to 20.00 mA O336 Message as to whether table status is OK yes no y value = current value belonging to O334 specified by user. Return to O333 until all values are entered. Back to O3. If status = no, correct table (all settings made up until now are retained) or back to measuring mode (table is deleted). 3.3.5 ALARM FUNCTIONS The monitoring functions are used to define various alarms and configure output contacts. Each individual error can be defined to be effective or not (at the contact or as an error current). An alarm condition can be defined to activate a cleaning function (F8) Coding Field F ALARM function group Selection or Range Default in Bold Information Alarm function settings. F1 Select contact type Latch = latching contact Momen =momentary contact F2 Select time unit s min Page 20 Display The contact type selected only applies to the alarm contact. Triton TR8 Coding Field Selection or Range Default in Bold F3 Enter alarm delay 0 s (min) 0 ... 2000 s (min) F4 Select error current 22 mA 2.4 mA F5 Select error 1 1 ... 255 F6 Set alarm contact to be effective for the selected error yes no F7 Set error current to be effective for the selected error no yes F8 Automatic cleaning function start no yes F9 Select return to menu or next error next = next error ←R Display Information Depending on the option selected in F2, the alarm delay is entered in s or min. This selection must be made even if all error reporting is switched off in F5. Caution! If "0-20 mA" was selected in O311, "2.4 mA" may not be used. Here you can select all the errors which should trigger an alarm. The errors are selected via the error numbers. Please refer to the table in section 6.2 "System error messages" for the meaning of the individual error numbers. The factory settings remain in effect for all errors not edited. If “no” is selected, all the other alarm settings are deactivated (e.g. alarm delay). The settings are retained. This setting only applies to the error selected in F5. The option selected in F4 is effective or ineffective in the event of an error. This setting only applies to the error selected in F5. This field is not available for certain errors, see "Troubleshooting and fault elimination" section. If ←R is selected, you return to F, if next is selected, you go to F5. 3.3.6 CHECK The CHECK function group is only available for devices with a Plus Package. In the CHECK function group, you can select different monitoring functions for the measurement. Triton TR8 Page 21 All monitoring functions are off by default. To adapt the Sensor Check System to the current application conditions, add and set the suitable functions. Alarm threshold monitoring (fields P1 to P4) You can use this function to monitor the measured value for permissible upper and lower limits and trigger an alarm (E154, E155). PCS alarm (Process Check System), (fields P5 to P8) AC (Alternating Check): The function AC (field P5) is used to check measuring signals for deviations. If the measuring signal does not change within an hour an alarm (E152) is triggered. The reason for such sensor behavior can be contamination, cable rupture or similar. CC (Controller Check): You can monitor the controller activity with the function CC. This function is mainly used for batch processes and single-sided limit switches. A malfunction of the controller is detected and reported thanks to freely adjustable monitoring times (E156 - E157). The CHECK Function group is an optional package; the BASIC package does not include items listed in Italics. Coding Field P CHECK function group Selection or Range Default in Bold Information Settings for sensor and process monitoring P1 Select alarm threshold monitoring Off Low High Lo+Hi Low! High! Lo+Hi! P2 Enter alarm delay 0 s (min) 0 to 2000 s (min) P3 Enter lower alarm threshold 0.000 FNU 0 to 9999 FNU P4 Enter upper alarm threshold 10.00 FNU 0 to 9999 FNU Page 22 Display Alarm signaling optionally with or without simultaneous controller switch-off. XXXX = without controller switch-off XXXX! = with control switchoff (Errors: E154, E155) Depending on your selection in F2, you can enter the error delay in min or s. Only after this delay, a high or low limit violation causes an alarm as per field P3/P4. Triton TR8 Coding Field Select process monitoring (PCS alarm) P5 Enter maximum permissible duration for lower CC set point limit violation (field P8) Enter maximum permissible duration for upper CC set point limit violation (field P8) P6 P7 Enter CC set point (for P6/P7) P8 Selection or Range Default in Bold Off AC CC AC+CC AC! CC! AC+CC! Display Information AC = sensor activity check (E152) CC = controller check (E156, E157) XXXX = without controller switch-off XXXX! = with controller switch-off 60 min 0 to 2000 min Only when P5 = CC or AC+CC 120 min 0 to 2000 min Only when P5 = CC or AC+CC 1.000 FNU 0 to 9999 FNU Selected value is an absolute value. This function is mainly used for batch processes and single-sided limit switches. 3.3.7 RELAY To use the RELAY function group, you need a relay board which is not part of the basic version. The following relay contacts can be selected and configured as desired (max. four contacts, depending on options installed): • Limit contactor for measured turbidity value: R2 (1) • Limit contactor for temperature: R2 (2) • PID controller: R2 (3) • Timer for cleaning function: R2 (4) • Chemoclean function: R2 (5) 3.3.7.1 Limit contactor for measured turbidity value and temperature The transmitter has different ways of assigning a relay contact. Switch-on and switch-off points and pick-up and drop-out delays can be assigned to the limit contactor. In addition, you can configure an alarm threshold to output an error message and to start a cleaning function in conjunction with this. These functions can be used both for turbidity measurement and for temperature measurement. Triton TR8 Page 23 Please refer to Fig. 28 for a clear illustration of the relay contact states. • When the measured values increase (maximum function), the relay contact is closed as of t2 after the switch-on point (t1) has been overshot and the pick-up delay has elapsed (t2 - t1). The alarm contact switches if the alarm threshold (t3) is reached and the alarm delay (t4 - t3) has also elapsed. • When the measured values decrease, the alarm contact is reset when the alarm threshold (t5) is undershot as is the relay contact (t7) after the drop-out delay (t7 - t6). • If the pick-up and drop-out delays are set to 0 s, the switch-on and switch-off points are also switch points of the contacts. Settings can also be made for a minimum function in the same way as for a maximum function. Figure 16: Illustration of the alarm and limit value functions A. Switch-on point > switch-off point: Max. function B. Switch-on point < switch-off point: Min. function 1. Alarm Threshold 2. Switch On Point 3. Switch Off Point 5. Alarm On 6. Alarm Off 7. Contact Off 4. Contact On 3.3.7.2 PID Control You can define various controller functions for the transmitter. On the basis of the PID controller, P, PI, PD and PID controllers can be implemented. For an optimum control system, use the controller that best Page 24 Triton TR8 suits your application. Depending on the option selected in the R 237/R 266 field, the actuating signal can be output via relays or via current output 2 (if available). • P controller: Used for simple linear control purposes with small system deviations. Where major changes are to be controlled, overshooting may occur. In addition, a lasting control deviation is to be expected. • PI controller: Is used for control systems where overshooting is to be avoided and no lasting control deviation should occur. • PD controller: Is used for processes that require quick changes and where peaks are to be corrected. • PID controller: Is used for processes where a P, PI or PD controller does not control sufficiently. Configuration options of the PID controller The following configuration options are available for a PID controller: • Change control gain Kp (P influence) • Set integral action time Tn (I influence) • Set derivative action time Tv (D influence) Basic load dosing (Basic) The basic load dosing (field R231) is used to set a constant dosage (field R2311) PID controlling plus basic load dosing If you select this function (PID + Basic) in field R231 the PID controlled dosage will not be lower than the basic load value entered in field R2311. Commissioning If you do not yet have any experience for setting the control parameters, set the values that yield the greatest possible stability in the control circuit. Proceed as follows to optimize the control circuit further: • Increase the control gain Kp until the controlled variable just starts to overshoot. • Reduce Kp slightly and then reduce the integral action time Tn so that the shortest possible correction time without overshooting is achieved. • To reduce the response time of the controller, also set the derivative action time Tv. Triton TR8 Page 25 Fig17: Optimization of settings Tn and Kp A. Actual value B. Time 1. Tn too small 2. Tn too large 3. Kp too large 4. Kp too small 5. Optimum setting Actuating signal outputs (R237 ... R2310) Each control contact outputs a cyclical signal whose intensity corresponds to the controller's manipulated variable. A distinction is made according to the type of signal cycle: • Pulse length modulation The bigger the calculated manipulated variable is, the longer the contact affected remains picked up. The period T can be adjusted between 0.5 and 99 s (field R238). Outputs with pulse length modulation are used to activate solenoid valves. • Pulse frequency modulation The bigger the calculated manipulated variable is, the higher the switching frequency of the contact affected. The maximum switching frequency 1/T can be set between 60 and 180 min-1. Page 26 Triton TR8 The on-time tON is constant. It depends on the set maximum frequency and is approx. 0.5 s for 60 min-1 and approx. 170 ms for 180 min-1. Outputs with pulse frequency modulation are used to activate directly controlled solenoid dosing pumps. Constant controller Via the current output 2, the minimum actuating variable (0 %) of the controller is output with 0/4 mA and the maximum actuating variable (100%) of the controller is output with 20 mA. Control characteristic for direct and inverse control action You can choose between two control characteristics in the R236 field: • Direct control action = maximum function • Inverse control action = minimum function 3.3.7.3 Timer for cleaning function This function includes a simple cleaning option. You can set the time interval after which cleaning should start. So you can only select a constant interval sequence. Other cleaning functions are available for selection in conjunction with the Chemoclean function (version with four contacts, see "Chemoclean function" section). Note! Timer and Chemoclean do not work independently of one another. While one of the two functions is active, the other cannot be started. Figure 18: Correlation of cleaning time, pause time and hold dwell period A. Wiper Function B. Hold Function 0. Inactive 1. Active t0 Normal Mode t1 Start Cleaning t2-t1 Cleaning Time t3-t2 Hold Dwell Time t4-t3 Time between Cleaning cycles Triton TR8 Page 27 3.3.7.4 Chemoclean function Just like the timer function, Chemoclean can also be used to start a cleaning cycle. However, Chemoclean also gives you the added option of defining different cleaning and rinsing intervals. As a result, it is possible to clean irregularly with different repeat cycles and to separately set the cleaning times with post rinse times. Note! • To use the Chemoclean function the transmitter has to be equipped with a designated relay board (see product structure or chapter "accessories"). • Timer and Chemoclean do not work independently of one another. While one of the two functions is active, the other cannot be started. • For the Chemoclean function, the relays 3 (water) and 4 (cleaner) are used. • If cleaning is prematurely aborted, a post rinse time always follows. • If the setting is "Economy", cleaning only takes place with water. Figure 19: Cleaning Cycle Sequence A. Hold B. Water t0 Normal Mode t4 - t3 Post Rinse Time C. Cleaner t1 Cleaning Start 0. Contact on 1. Contact off t2 - t1 Pre Rinse Time t3 - t2 Cleaning Time t5 - t4 Hold Dwell Time RELAY FUNCTION The basic unit does not include items in italics. Page 28 Triton TR8 Coding Field R RELAY function group Selection or Range Default in Bold Rel1 Rel2 Rel3 Rel4 R2 (1) Configure limit contactor for turbidity measurement LC PV = limit contactor TU (1) LC °C = limit contactor T (2) PID controller (3) Timer (4) Clean = Chemoclean (5) R211 Switch function of R2 (1) off or on Off On Enter the switch-on point of the contact 9999 FNU 9999 NTU 9999 ppm / 3000 ppm 9999 mg/l / 3000 mg/l 300.0 g/l / 3 g/l 99.99 kg/l 99.99 t/m 200.0 % Enter the switch-off point of the contact 9999 FNU 9999 NTU 9999 ppm / 3000 ppm 9999 mg/l / 3000 mg/l 300.0 g/l / 3 g/l 99.99 kg/l 99.99 t/m 200.0 % R212 R213 Triton TR8 Information Relay Contact Settings Select contact to be configured R1 Display Rel3 (water) and Rel4 (cleaner) are only available with the relevant version of the transmitter. If Chemoclean is used as the cleaning method, Rel4 is not available. PV = process value If Rel4 is selected in the R1 field, Clean = Chemoclean cannot be selected. By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings. All the settings are retained. Never set the switch-on point and the switch-off point to the same value! (Only the operating mode selected in A1 is displayed.) If two factory settings are displayed the one on the left side stands for CUS41 the one on the right side stands for CUS31. Entering a switch-off point selects either a Max contact (switch-off point < switch-on point) or a Min contact (switch-off point > switch-on point), thereby implementing a hysteresis that is constantly required (see "Illustration of the alarm and limit functions "figure). If two factory settings are displayed the one on the left side stands for CUS41 the one on the right side stands for CUS31. Page 29 Coding Field Selection or Range Default in Bold R214 Enter pick-up delay 0s 0 to 2000 s R215 Enter drop-out delay 0s 0 to 2000 s Display Information If the alarm threshold is undershot / overshot, this triggers an alarm with the error message and error current at the transmitter (note alarm delay in field F3). If defined as a Min contact, the alarm threshold must be < switch-off point. If two factory settings are displayed the one on the left side stands for CUS41 the one on the right side stands for CUS31. R216 Enter alarm threshold 9999 FNU 9999 NTU 9999 ppm / 3000 ppm 9999 mg/l / 3000 mg/l 300.0 g/l / 3 g/l 99.99 kg/l 99.99 t/m 200.0 % R217 Display status for limit contactor MAX MIN Display only. R2 (2) Configure limit contactor for temperature measurement LC PV = limit contactor TU (1) LC °C = limit contactor T (2) PID controller (3) Timer (4) Clean = Chemoclean (5) By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings. R221 Switch function of R2 (2) off or on Off On Settings made for the limit contactor are not deleted by switching the function off. R222 Enter switch-on temperature 100.0 °C (212 °F) -5.0 to 100.0 °C (23 to 212 °F) Never set the switch-on point and the switch-off point to the same value! Page 30 Triton TR8 Coding Field Selection or Range Default in Bold R223 Enter switch-off temperature 100.0 °C (212 °F) -5.0 to 100.0 °C (23 to 212 °F) R224 Enter pick-up delay 0s 0 to 2000 s R225 Enter drop-out delay 0s 0 to 2000 s Display Information Entering a switch-off point selects either a Max contact (switch-off point < switch-on point) or a Min contact (switch-off point > switch-on point), thereby implementing a hysteresis that is constantly required. R226 Enter alarm threshold (as absolute value) 100.0 °C (212 °F) -5.0 to 100.0 °C (23 to 212 °F) If the alarm threshold is undershot/overshot, this triggers an alarm with the error message and error current at the transmitter (note alarm delay in field F3). If defined as a Min contact, the alarm threshold must be < switch-off point. R227 Display status for limit contactor MAX MIN Display only. R2 (3) Configure P(ID) controller LC PV = limit contactor TU (1) LC °C = limit contactor T (2) PID controller (3) Timer (4) Clean = Chemoclean (5) By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings. R231 Switch function of R2 (3) off or on Off On Basic PID+B On = PID controller Basic = basic load dosing PID+B = PID controller + basic load dosing Triton TR8 Page 31 Coding Field R232 Enter set point R233 Enter control gain Kp Selection or Range Default in Bold 0.000 FNU 0.000 NTU 0.000 ppm 0.000 mg/l 0.000 g/l 0.000 kg/l 0.000 t/m 0.000 % Display Information The set point is the value to be maintained by the control system. Using this control process, this value is restored upwards or downwards when a deviation occurs. 1.00 0.01 to 20.00 See "P(ID) controller" section. R234 Enter integral action time Tn (0.0 = no I-component) 0.0 min 0.0 to 999.9 min With every Hold, the Icomponent is set to zero. Although Hold can be deactivated in field S2, this does not apply for Chemoclean and timer! R235 Enter derivative action time Tv (0.0 = no D-component) 0.0 min 0.0 to 999.9 min See "P(ID) controller" section. R236 Select controller characteristic inv = inverse dir = direct R237 Select pulse length or pulse frequency len = pulse length freq = pulse frequency curr = current output 2 R238 Enter pulse interval 10.0 s 0.5 to 999.9 s R239 Enter maximum pulse frequency of the adjuster 120 min-1 60 to 180 min-1 Page 32 dir = Max. function inv = Min. function The setting is required depending on the control deviation (upward or downward deviation, see "P(ID) controller" section). Pulse length e.g. for solenoid valve, pulse frequency e.g. for solenoid dosing pump, see "Actuating signal outputs" section. Curr = current output 2 can only be selected if O2 = Contr. This field only appears if pulse length is selected in R237. If pulse frequency is selected, R238 is skipped and entries continue with R239. This field only appears if pulse frequency is selected in R237. If pulse length is selected, R239 is skipped and entries continue with R2310. Triton TR8 Coding Field Selection or Range Default in Bold R2310 Enter minimum switch-on time tON 0.3 s 0.1 to 5.0 s Display Information This field only appears if pulse length is selected in R237. When you select the basic load, you enter the desired dosing quantity. 100% basic load would correspond to: – Constantly on for R237 = len – Fmax at R237 = freq (field R239) – 20 mA at R237 = curr Cleaning only takes place with one cleaning agent (usually water). By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings. Enter basic load 0% 0 to 40 % R2 (4) Configure cleaning function (timer) LC PV = limit contactor TU (1) LC °C = limit contactor T (2) PID controller (3) Timer (4) Clean = Chemoclean (5) R241 Switch function of R2 (4) off or on Off On Settings made for the timer are not deleted by switching the function off. R242 Enter rinsing/cleaning time 30 s 0 to 999 s Settings for Hold and relay are active for this time. R243 Enter pause time 360 min 1 to 7200 min The pause time is the time between two cleaning cycles (see "Timer for cleaning function" section). R244 Enter minimum pause time 120 min 1 to R243 min The minimum pause time prevents constant cleaning if a cleaning trigger is present. R2 (5) Configure cleaning with Chemoclean (for version with four contacts , Chemoclean option and contacts 3 and 4 assigned) LC PV = limit contactor TU(1) LC °C = limit contactor T (2) PID controller (3) Timer (4) Clean = Chemoclean (5) See "Chemoclean function" section. By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings. R2311 Triton TR8 Page 33 Coding Field Selection or Range Default in Bold R251 Switch function of R2 (5) off or on Off On R252 Select type of start pulse Int = internal (time-controlled) Ext = external (digital input 2) I+ext = internal + external I+stp = internal, suppressed by external The cycle for the "int" function is started by the end of the pause time (R257). No real time clock is available. External suppression is required for irregular time intervals (e.g. weekends). R253 Enter pre-rinse time 20 s 0 to 999 s Rinsing with water takes place. R254 Enter cleaning time 10 s 0 to 999 s Cleaning with cleaning agent and water takes place. R255 Enter post rinse time 20 s 0 to 999 s Rinsing with water takes place. R256 Enter number of repeat cycles 0 0 to 5 R253 to R255 is repeated. R257 Enter pause time 360 min 1 to 7200 min The pause time is the time between two cleaning cycles (see "Timer function section). R258 Enter minimum pause time 120 min 1 to R257 min The minimum pause time prevents constant cleaning if an external cleaning start is present. R259 Enter number of cleaning cycles without cleaning agent (economy function) 0 0 to 9 After cleaning with cleaner, up to 9 cleaning sessions can be carried out with water only until the next cleaning session with cleaner takes place. Page 34 Display Information Triton TR8 3.3.8 CONCENTRATION MEASUREMENT The basic unit does not include items in italics. Coding Field K Function group CONCENTRATION K1 Selection of concentration curve, to be used to calculate the display value Selection or Range Default in Bold Display Information Four different concentration curves can be entered in this function group. 1 1 to 4 The curves are independent of each other. Therefore, four different curves can be defined. K2 Selection of table to be edited 2 1 to 4 The modification of a table (curve) is independent from the selected curve in K1. The selected table (curve) will not be overwritten until the input of all table values is complete. K3 Select table option read edit This selection applies to the concentration curve selected in K2. K4 Enter number of reference elements 1 1 to 10 Each element consists of two numeric values. Select element 1 1 to number of element in K4 assign Any element can be edited. K5 K6 Enter turbidity value 0.00 % entire measuring range The function chain K5 ... K7 will be repeated automatically as many times as corresponds to the value in K4. Then the system jumps to K8. K7 Enter concentration value entire measuring range Measuring unit as selected in A2. K8 Message whether or not the table status is ok yes no Only display If not, then set table correctly (all previous settings are kept) or back to measurement mode (this makes the table invalid). Triton TR8 Page 35 3.3.9 SERVICE Coding Field S SERVICE function group Selection or Range Default in Bold Information Service function settings. ENG = English GER = German FRA = French ITA = Italian NL = Dutch ESP = Spanish S+C = Hold during configuration and calibration Cal = Hold during calibration Setup = Hold during configuration None = no Hold S1 Select language S2 Configure Hold S3 Manual Hold off on S4 Enter Hold dwell period 10 s 0 to 999 s S5 Enter SW upgrade release code (Plus Package) 0 0000 to 9999 S6 Enter SW upgrade release code Chemoclean 0 0000 to 9999 Page 36 Display S = setup C = calibration The setting is retained even in the event of a power failure. The code is on the name plate. If an incorrect code is entered, you are taken back to the measurement menu. The number is edited with the PLUS or MINUS key and confirmed with the ENTER key. "1" is displayed if the code is active. You can find the code on the nameplate. If an incorrect code is entered, you are taken back to the measurement menu. "1" is displayed if the code is active. Triton TR8 Coding Field Selection or Range Default in Bold S7 Order number is displayed CUM253-T S8 Serial number is displayed 880CB405G S9 Reset the device to the basic settings no Sens = sensor data Facty = factory settings S10 Perform device test no Displ = display test Coding Field Selection or Range Default in Bold E SERVICE 2 function group Display Information If the device is upgraded, the order code is not automatically adjusted. Sens = last calibration is deleted and is reset to factory setting. Facty = all data (apart from A1 a. S1) are deleted and reset to the factory setting! Display Information Information on the device version Contr = controller (1) Trans = transmitter (2) Main = power unit (3) Rel = relay module (4) E1 Select module E111 E121 E131 E141 Software version is displayed If E1 = contr: instrument software If E1 = trans, main, rel: module firmware E112 E122 E132 E142 Hardware version is displayed Only display function Triton TR8 Page 37 Selection or Range Default in Bold Coding Field Display Information E113 E123 E133 E143 Serial number is displayed Only display function E114 E124 E134 E144 Module ID is displayed Only display function 3.3.10 INTERFACE Selection or Range Default in Bold Coding Field Display Information I INTERFACE function group Communication settings (only for device version HART or PROFIBUS). I1 Enter bus address Each address may only be used once in a network. If a device address ≠ 0 is selected, the current output is automatically set to 4 mA and the device is set to multidrop operation. I2 Display of measuring point Address HART: 0 to 15 or PROFIBUS: 0 to 126 3.4 CALIBRATION FUNCTION GROUP MENUS Measuring chain calibration is performed in this function group. The calibration data are saved in an EEPROM directly in the sensor. For this reason: • Recalibration is not required in the event of a power failure • Recalibration is not required when the transmitter is replaced • Customer-specific recalibration is required, however, when the sensor is replaced Three calibration data records are saved in the sensor for each of the four main operating modes. Factory data record no. 1 not changeable User data record no. 2 changeable User data record no. 3 changeable Page 38 FNU/NTU ppm or mg/l g/l Formazine SiO2 Activated sludge Formazine Kaolin Activated sludge Formazine SiO2 Activated sludge % Residual concrete water Residual concrete water Residual concrete water Triton TR8 Select the desired data record in the SETUP 2 function group in the B4 field. • In the FNU operating mode, the sensor is factory calibrated with formazine traceable to ISO 7027. • In the ppm operating mode, the calibration data records for Kaolin and SiO2 are derived from the FNU data records. • In the % operating mode, the calibration data records are set to the average of various residual concrete waters. They are preset in such a way that correct values are displayed for average clarity. However the settings do not follow a standard currently applicable. • In the g/l operating mode also, the sensor is not calibrated to a fixed value as no standard is directly applicable. You must carry out a calibration because the media of the various applications differ too greatly here. Three-point sensor calibration is the standard calibration. It is absolutely essential: • When commissioning the sensor in sludge applications • When changing to another sludge type Three-point sensor calibration is not necessary: • When commissioning the sensor in the drinking water area (sensor has been calibrated for drinking water applications in the factory). • For residual concrete water. Density measurement for determining the concentration of residual concrete water is based on %–data records. They are preset in such a way that correct values are displayed for average clarity. One-point calibration is often sufficient to adjust the system in the event of deviating values. • When recalibrating with the same sludge type. One-point calibration suffices here if the degrees of lightness and clarity, for example, do not differ too greatly. Note! • Sludge samples tend sediment. Mix the sample well, even during the calibration process, but not to the extent that gas bubbles are formed. • The sensor has to be far enough away from the floor and the wall of the calibration vessel during calibration. The immersion depth must be at least 40 mm. • The characteristic determined during the calibration is stored in the selected data record (Setup 2, B4 field). • Calibration is not possible if data record 1 is selected with the factory setting. • If the calibration data deviate from the standard values by a factor of two or more, a warning (E084) is output. The calibration results are accepted. • If the calibration results are outside the permitted range, a calibration error (E045) is indicated. The calibration results are not accepted. • For every type of calibration, the installation adjustment and the offset are reset to zero and the slope to 1.0. Calibration Menu Options 3–Pt Three-point calibration Corr Three-point correction Edit Edit calibration Refl Installation adjustment 1–Pt One-point calibration Triton TR8 Page 39 Data Calibration data Three-point calibration (3–Pt) You should perform the calibration in the turbidity/solids concentration range in which you plan to measure. The overall calibration characteristic of the measuring chain is determined using three samples of known turbidity or known solids content. Calibration with a very dark, high-absorption medium returns a small slope while light, clear media return big slopes. You can create the requisite probes by diluting a medium sample. In general, very good calibration results are achieved with a concentration gradation of 10 %, 33 % and 100 %. The following condition must be met for the calibration: Sample A > 1.1 x sample B > 1.1 x sample C Figure 20: Preparing 3 point calibration samples A Original Sample B 1 part A + 2 parts water C 1 part A + 9 parts water A different dilution ratio should be selected for very high-absorption media to achieve accurate calibration results. Instead of 100 % – 33 % – 10 %, you can use dilution ratios of 100 % – 20 % – 5 % or 100% – 10 % – 3.3 %. If the scattered light of a light sample hits a sensor that is calibrated for a dark medium, it can happen that the signal value is so high that it is above the calibration curve. Three-point correction (Corr) If the calibration was performed with an unknown sample concentration but with a defined dilution, the correct value determined afterwards in the laboratory is entered here. Edit calibration (Edit) If each of the samples has been determined afterwards in the laboratory, the correction for the calibration is entered here. Installation adjustment (Refl) In installation adjustment, backscatter from the immediate sensor environment is compensated. Installation adjustment must be performed with a medium lower than 2 FNU or 5 ppm. Page 40 Triton TR8 One-point calibration (1–Pt) In the area of TS/concentration measurement, one-point calibration is used to change the conversion factor (field C166). The slopes are not changed. In the FNU range, the two slope values are adjusted by a one-point calibration. This is possible because the conversion factor in the FNU range is always 1 and the editing range is limited to 4000 FNU. In this way, the two curves are always increasing and the calibration remains clear. In the ppm range, the slope values are adjusted up to 500 ppm. For higher values, the conversion factor is changed. Calibration data (Data) Here, you can display the calibration points 1 to 3, slope 1 and 2 and the conversion factor. With three-point calibration, the curves used in the algorithm are adjusted as precisely as possible using the data points acquired. The difference between the ideal curve of the algorithm and the three actual calibration points can be found as a correction factor in the C161, C162 and C163 fields of the data function. The correction values are indicated in %. The values should be as close to 100 % as possible. Values from 70 to 80 % are acceptable. 50 % in one or two calibration points clearly indicates a problematic calibration. A warning (E084) is output here for this reason. This can mean that significant deviations can occur between the calibration points. The calibration points themselves are always retrieved correctly. 3.4.1 CALIBRATION Coding Field C Function group CALIBRATION Selection or Range Default in Bold Display Information Calibration settings. 3-Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection compensation (4) 1-Pt = Single-point calibration (5) Data = Calibration data (6) For data set 1 (B4), only the “Data” function is accessible. The offset is reset with 3 Pt and Edit. Immerse sensor in the calibration solution (sample 1). Immerse the sensor so that there is sufficient distance to the tank wall (no reflection). C1 (1) Triton TR8 Select calibration Page 41 Coding Field Selection or Range Default in Bold C111 Enter concentration of the first calibration solution Value from last calibration Immerse sensor in the calibration solution (sample 2). C112 Enter concentration of the second calibration solution Value from last calibration Immerse sensor in the calibration solution (sample 3). C113 Enter concentration of the third calibration solution Value from last calibration C114 Calibration status is displayed o. k. E. xxx Store calibration results yes no new Select calibration 3-Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection compensation (4) 1-Pt = Single-point calibration (5) Data = Calibration data (6) C115 C1 (2) Page 42 Display Information Immerse the sensor so that there is sufficient distance to the tank wall (no reflection). C112 ≥ 1.1 x C111 Immerse the sensor so that there is sufficient distance to the tank wall (no reflection). C113 ≥ 1.1 x C112 If C114 = E xxx, then only no or new (Exception: calibration warning E84). If new, return to C. If yes / no, return to “Measurement”. Triton TR8 Coding Field Selection or Range Default in Bold C121 Enter correct concentration of the third calibration solution Current value from C113 entire measuring range C122 Calibration status is displayed o. k. Exxx Store calibration results yes no new C1 (3) Select calibration 3-Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection compensation (4) 1-Pt = Single-point calibration (5) Data = Calibration data (6) C131 Enter concentration of the first calibration solution Current value from C111 entire measuring range C132 Enter concentration of the second calibration solution Current value from C112 C132 ≥ 1.1 x C131 C133 Enter concentration of the third calibration solution Current value from C113 C133 ≥ 1.1 x C132 C134 Calibration status is displayed o.k. Exxx C123 Triton TR8 Display Information If the calibration is performed with an unknown sample concentration, but with a definite dilution (1/10; 1/3;1), the laboratory value is to be entered. If C122 = E xxx, then only no or new (Exception: calibration warning E84). If new, return to C. If yes / no, return to “Measurement”. Page 43 Coding C135 Field Selection or Range Default in Bold Store calibration results? yes no new 3-Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection compensation (4) 1-Pt = Single-point calibration (5) Data = Calibration data (6) 0.0 NTU 0.0 to 2.0 NTU 0.0 FNU 0.0 to 2.0 FNU 0.0 ppm 0.0 to 5.0 ppm 0.0 mg/l 0.0 to 5.0 mg/l Display Information If C134 = E xxx, then only no or new (Exception: calibration warning E84). If new, return to C. If yes / no return to “Measurement”. Only for solutions = 2 FNU / 5 ppm! Backscatter from the immediate sensor environment is compensated for clear media. C1 (4) Select calibration C141 Enter correct measured value C142 Calibration status is displayed o.k. Exxx Store calibration results? yes no new If C142 = E xxx, then only no or new (Exception: calibration warning E84). If new, return to C. If yes / no, return to “Measurement”. Select calibration 3-Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection compensation (4) 1-Pt = Single-point calibration (5) Data = Calibration data (6) For FNU: Adaption C164, C165 For ppm, mg/l: up to 500 - adaption C164, C165 above 500 - adaption C166 For g/l, %: Adaption C166. An existing basic calibration (three-point) is corrected by the single-point calibration. C143 C1 (5) Page 44 Only for the ranges NTU, FNU, ppm, mg/l Triton TR8 Coding Field Selection or Range Default in Bold C151 Enter current calibration value Current measured value Entire measuring range C152 Calibration status is displayed o.k. Exxx Store calibration results? yes no new C153 C1 (6) Select calibration Display Information If C152 = E xxx, then only no or new (Exception: calibration warning E84). If new, return to C. If yes / no, return to “Measurement”. 3-Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection compensation (4) 1-Pt = Single-point calibration (5) Data = Calibration data (6) C161 Calibration point 1 is displayed Comparison value Deviation relative to standard sensor (= 100 %) C162 Calibration point 2 is displayed Comparison value Deviation relative to standard sensor (= 100 %) C163 Calibration point 3 is displayed Comparison value Deviation relative to standard sensor (= 100 %) C164 Slope 1 is displayed Current value Slope of characteristic 1 of the sensor C165 Slope 2 is displayed Current value Slope of characteristic 2 of the sensor Triton TR8 Page 45 Coding Field Selection or Range Default in Bold C166 Conversion factor is displayed Current value Display Information Conversion factor of internal turbidity units into displayed unit 3.4.2 OFFSET The settings in the OFFSET function group can be used to calibrate the measurement to a reference measurement. This requires a linear shift of all the measured values, i.e. the adjustment is determined for one measured value, and all others are calculated using the same adjustment. Note! Following a calibration, the offset is automatically set to zero. Selection or Range Default in Bold Coding Field V Function group OFFSET V1 Enter absolute value Current measured value V2 Enter offset Current offset V3 Calibration status is displayed o.k. E xxx V4 Store calibration result? yes no new Display Information If V3 = E xxx, then only no or new. If new, return to V. If yes/no, return to "Measurement". 3.4.3 SLOPE With the settings in the function group SLOPE, a measured value can be adapted to a reference value. All measured values are proportionally adapted over the entire measuring range according to this change. Page 46 Triton TR8 Example: The displayed measured value is 2.5 g/l. It is adapted to the reference value of 2.0 g/l by means of the slope function. The change is 20 %, i.e. all measured values are reduced by 20 % over the entire measuring range. Note! An OFFSET having been edited before is reset to the factory setting. In contrast to the one-point calibration, the edited slope can be reset by setting the slope factor to 1.0. Selection or Range Default in Bold Coding Field N Function group SLOPE N1 Enter absolute value Current measured value N2 Enter slope Current slope N3 Status is displayed o.k. E xxx Store slope? yes no new N4 Triton TR8 Display Information Slope is displayed, can be edited. Page 47 4.0 MAINTENANCE You must carry out maintenance tasks at regular intervals. For this, specify the maintenance times in an operations logbook or operations calendar. The following tasks must be carried out: • Cleaning the transmitter • Checking the measuring function • Cleaning the sensor • Recalibration 4.1 CLEANING THE TRANSMITTER Clean the front of the housing with usual commercial cleaning agents. In accordance with DIN 42 115, the front is resistant to: • Isopropanol • Diluted acids (max. 3%) • Diluted alkalis (max. 5%) • Esters • Hydrocarbons • Ketones • Household cleaners Caution! For cleaning purposes, never use: • Concentrated mineral acids or alkalis • Benzyl alcohol • Methylene chloride • High-pressure steam 4.2 CHECKING THE MEASURING POINT 4.2.1 Transmitter The TR8 Turbidity sensors cannot be simulated as they contain the complete data processing and all the measured values are transmitted to TR8 Analyzer using the digital interface RS 485. Therefore a functional sensor is required for the measuring point test. Method for testing a measuring point: • Check that device is operable and that the display reacts appropriately, e.g. by pressing the PLUS key. • Check the current outputs by carrying out a current simulation (Field O3 (2)). • Measure the sensor operating voltage: approx. 10 to 16 V at terminals 87 (+) and 88 (–). • The cause for an incorrect voltage may be present either at the device or at the sensor. – Replace the sensor. – If the sensor operating voltage is still too low replace the power supply module (The analyzer must be returned to the factory for power supply replacement). Page 48 Triton TR8 • Sensor operating voltage is o.k. but no measured turbidity value even with a new sensor. Replace the transmitter module (The analyzer must be returned to the factory for power supply replacement). 4.2.2 Triton TR8 Sensor 1. Remove the sensor from the medium. 2. Clean and dry the sensor. 3. Check the slope using CUY22 check unit. A stable measured value (between 2 and 6 FNU) must register. The exact value is not of importance but must be repeatable for the same sensor. 4.3 CLEANING THE SENSOR Contamination of the sensor can impact the measurement up to the point that measurement no longer functions, e.g. by: • Thick film build-up on the sensor optics. Film build-ups can cause bad measurement results. • Wiper clogging. This causes bad measurement results. The sensor must be cleaned regularly to ensure reliable measurement. The frequency and intensity of the cleaning depend on the medium. The sensor must be cleaned: • Prior to every calibration • Regularly during the operation • Prior to returning the sensor for repair. 4.4 RECALIBRATION After cleaning check the Triton TR8 in the Calibration Cap the reading should be between 2FNU and 6 FNU. If not then you have the following options in the ''Calibration'' menu of the transmitter: • One-point calibration • Three-point calibration • Installation adjustment • Correction function • Adjust calibration data If you are unable to recalibrate the sensor then the sensor must be replaced. When replacing a sensor pay attention to the following: • Replacing the Triton TR8 sensor All factory calibration data are saved in the sensor. The sensor and the assembly are calibrated together. No additional calibration is required for applications with pure or ultrapure water, if the sensor and the Triton TR8 Page 49 assembly are replaced together. The calibration data of the sensor are automatically transferred to the measuring instrument. 5.0 ORDERING INFORMATION Part# Description Triton TR8 Turbidity Sensor with 7 meter cable 1398000-1 Triton TR8 Turbidity Sensor High Range with 7 meter cable 1398000-2 Triton TR8 Turbidity Sensor with 15 meter cable 1398001-1 Triton TR8 Turbidity Sensor High Range with 15 meter cable 1398001-2 Triton TR8 Turbidity Sensor with 7 meter cable & Automatic Wiper 1398010-1 Triton TR8 Turbidity Sensor High Range with 7 meter cable & Automatic Wiper 1398010-2 Triton TR8 Turbidity Sensor with 15 meter cable & Automatic Wiper 1398011-1 Triton TR8 Turbidity Sensor High Range with 15 meter cable & Automatic Wiper 1398011-2 Triton TR8 Turbidity Sensor with 7 meter cable & Flow Through Assembly 1398100-1 Triton TR8 Turbidity Sensor with 15 meter cable & Flow Through Assembly 1398101-1 Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & Flow Through Assembly 1398110-1 Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & Flow Through Assembly 1398111-1 Triton TR8 Turbidity Sensor with 7 meter cable & De-Bubbler Assembly 1398200-1 Triton TR8 Turbidity Sensor with 15 meter cable & De-Bubbler Assembly 1398201-1 Triton TR8 Turbidity Sensor with 7 meter cable, Automatic Wiper & De-Bubbler Assembly 1398210-1 Triton TR8 Turbidity Sensor with 15 meter cable, Automatic Wiper & De-Bubbler Assembly 1398211-1 (-1) Low Range Sensor < 500 FNU, (-2) High Range Sensor > 500 FNU 5.1 ACCESSORIES Part # 1000223-1 2000268 3200266 1000219 9640004.cond 1000222-2 1000228-1 1000227 Page 50 Model and Description Immersion Assembly ( Cap/Cable feed-through, 1 meter down pipe) Rail Moun3ng Brackets, (2) Quick Release “U” clamps for 2” Guard Rail moun3ng Wall Moun3ng Bracket Flow Through Housing TR8 5 Conductor Cable, per ft. NEMA 4X Junc3on Box, (2) cable glands, terminal strip, PVC box, 6”x3”x2”, LWD Service Kit, for Wiper Arm, rubber wipers and mounting material (3 sets) TR8 Check Unit, used to check Stability of the sensor, (calibration cup) Triton TR8 6.0 TROUBLESHOOTING 6.1 TROUBLESHOOTING INSTRUCTIONS The transmitter constantly monitors its functions itself. If an error occurs which the device recognizes, this is indicated on the display. The error number is shown below the display of the main measured value. If more than one error occurs, you can call these up with the MINUS key. Refer to the "System error messages" table for the possible error numbers and remedial measures. Should a malfunction occur without any transmitter error message, please refer to the "Process-specific errors" or the "Device-specific errors" tables to localize and rectify the error. These tables provide you with additional information on any spare parts required. 6.2 SYSTEM ERROR MESSAGES The system error messages can be called up and selected with the MINUS key. Error No. Display Test and Remedial Action E001 EEPROM memory error E002 Instrument not calibrated, calibration data invalid, no user data, user data invalid (EEPROM error), instrument software not suitable to hardware (controller) E003 Download Error E004 Instrument software version not compatible with module hardware version 1. Switch device off and then on again. 2. Load device software compatible with the Hardware. 3. Load measurementparameter specific device software. 4. If the error persists, send in the device for repair to your local service organization or replace the device. Invalid configuration. Repeat download. Load software compatible with hardware. Load measurement-parameter specific device software. E007 Transmitter malfunction, instrument software not compatible with transmitter version E008 Sensor or sensor connection faulty Triton TR8 Alarm Contact Facty Yes Error Current User Facty No User Auto Cleaning Start Facty User _ 1 Yes No _ Yes No No Yes No No Contact ECD Service. Yes No _ Check sensor and sensor connection (Service). Check initialization of wiper. Yes No No 1 1 Page 51 Error No. Display Test and Remedial Action Alarm Contact Facty E026 Wiper error E045 Calibration aborted Below main parameter measuring range Main parameter measuring range exceeded Below temperature measuring range Temperature measuring range exceeded Below current output range 1 Current output range 1 exceeded Below current output range 2 Current output range 2 exceeded Set point exceeded limit contactor 1 Set point exceeded limit contactor 2 Set point exceeded limit contactor 3 Set point exceeded limit contactor 4 Measuring value outside concentration table Current output 1 range too small Current output 2 range too small E055 E057 E059 E061 E063 E064 E065 E066 E067 E068 E069 E070 E079 E080 E081 E084 Calibration warning E085 Incorrect setting for error current E100 Current simulation active Page 52 Check wiper and test function using manual control if necessary. Repeat calibration Check measurement and connections; check device and measuring cable. Check measured value and current assignment. Error Current User Facty User Auto Cleaning Start Facty User Yes No No No No _ Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No Yes No No No No _ 1 No No _ 1 No No No No No _ 1 No No _ 1 1 Check configuration. Clean sensor, check table. Decrease current output spreading Calibration data are within limits but deviate from standard values by a factor of more than two. If the current range "0 to 20 mA" was selected in field O311, the error current "2.4 mA" may not be set. Triton TR8 Error No. Display Test and Remedial Action E101 Service function active Switch off service function or switch device off and then on again. E102 E106 E116 Manual mode active Download yes Download error E152 PCS alarm E153 Offset Below lower alarm threshold for period exceeding alarm delay Above upper alarm threshold for period exceeding alarm delay Current value undershoots alarm threshold (CC set point) for longer than the set permissible maximum period Current value exceeds alarm threshold (CC set point) for longer than the set permissible maximum period Alarm Contact Facty E154 E155 E156 E157 Wait for download to finish. Repeat download. Check sensor and connection. Adjustment range exceeded Perform manual comparison measurement if necessary. Service sensor and recalibrate. Error Current User Facty User Auto Cleaning Start Facty User No No _ 1 No No No No No No _ _ _ 1 1 1 No No No No No No Yes No No Yes No No Yes No No Yes No No E162 Dosage stop Check settings in the CURRENT INPUT and CHECK function groups. Yes No No E171 Flow in main stream too low or zero Restore flow. Yes No No Check process variables at sending measuring E172 Yes No No instrument. Change range assignment if necessary. Check process variables at E173 Current input < 4 mA sending measuring Yes No No instrument. Check process variables at sending measuring E174 Current input > 20 mA Yes No No instrument. Change range assignment if necessary. _ 1 If this error occurs, there is no possibility of starting a cleaning session (field F8 not applicable with this error). Switch-off limit for current input exceeded Triton TR8 Page 53 6.3 PROCESS SPECIFIC ERRORS Error Possible Cause Sensor / sensor cable defective Value indicated 0.0 Sensor extension line interrupted Check junction boxes and line. Incorrect sensor connection Check connection. Replace module MKT1 for testing. Replace module LSGA (AC power supply unit) or LSGD (DC power supply unit) for testing. Switch instrument off and back on, wiper has to wipe once. Clean optics Instrument input defective Data transfer error Wiper blocked Display value 0.0 Sensor completely blocked Fixed incorrect measured value Impermissible instrument operating state (no response to key actuation) Measuring cable interference Measured value fluctuates Signal output line interference Irregular flow rate / turbulence / air bubbles / big solids particles No or incorrect sensor calibration Display value implausible / no or creeping change of display Test and/or Remedial Measure Test with new or with different functional sensor. Switch instrument off and back on Connect cable screen acc. to connection diagram (do not ground) Check line routing Try separate line routing, grounding screen to PLC/PCS. Choose better place of installation or eliminate turbulences. Possibly use large measured value damping factor. Set gas bubble barrier to 100 %. Calibration with original sample required for concentration or solids concentration Clean sensor. Sensor soiled Use spray cleaning. Use wiper version. Page 54 Equipment, spare parts, personnel Triton TR8 (either type is suitable for rough function test) Sensor simulation see chapter "Maintenance of the entire measuring point". See chapter "Wiring". Repair at manufacturer only. Repair at manufacturer only. Repair at manufacturer only. Use spray cleaning or wiper. EMC problem: check line routing if problem persists, check for possible sources of interference. See chapter "Wiring". Separate signal output, meas. input and supply lines. Measured value damping see field A5. See chapter "Calibration". Remove coarse coats with brush. Remove carbonates and similar coats with 3 % hydrochloric acid. Remove organic coats and grease with oxidation agents and / or grease solvents. See instructions of assembly used for spray cleaning. Wiper upgrade at factory. Triton TR8 Error Display value implausible / no or creeping change of display Possible Cause Test and/or Remedial Measure Wiper rubber defective Replace wiper arm. Sensor installed in “dead” zone or air cushion in assembly or flange Incorrect sensor orientation Incorrect temperature value Controller or timer cannot be activated If temperature display is required: replace sensor. Turbidity measurement itself does not require temperature measurement. No relay module installed Install LSR1-2 or LSR1-4 module. Repair at manufacturer only. Pickup delay setting too long "Hold" function active Controller/limit contact works continuously Controller in "Manual/On" mode Dropout delay setting too long Control loop interruption Line open or short-circuited No turbidity current output signal Fixed current output signal Incorrect current output signal Triton TR8 Frontal “bombardment” of measuring surface with highly viscous solids may result in an adhering coating. Temperature sensor defective Controller switched off Controller in "Manual/Off" mode Controller/limit contact does not work Check installation conditions, move sensor to area with optimum flow conditions. Caution if installed in horizontal lines! Orient sensor: • Measuring surface should face flow in normal media. • Orient meas. surface at 90° to flow in media with high solids concentration Equipment, spare parts, personnel Wiper arm service kit 1000228-1 Activate controller. Choose "Auto" or "Manual/On" mode. Disable or shorten pickup delay. “Automatic Hold” during calibration, “Hold” input activated; “Hold” via keyboard active. Set controller to "Manual/Off" or "Auto". See fields R2xx. Shorten dropout delay. See field R2xx. Check measured value, current output, actuators, chemical supply. Disconnect line and measure directly on instrument. Total load in current loop excessive (>500Ω) Instrument with PROFIBUS PA/DP Current simulation active PA/DP instruments have no current output. Switch off simulation. Impermissible operating state of processor system Switch instrument off and back on. Incorrect current assignment Check current assignment: 0–20 mA or 4–20 mA? Disconnect line and measure line. Keyboard, REL-key See fields R2xx. See fields S2 to S4. Keyboard, REL and AUTO keys mA meter 0–20 mA Ohmmeter See field O3 (2). EMC problem: check installation, screen, grounding if problem persists. Field O311 Page 55 Error Possible Cause Incorrect current output signal Current output table not accepted Total load in current loop excessive (> 500 Ω.) Test and/or Remedial Measure Disconnect line and measure line. Value interval too small Select practical intervals. No central HART module Verify by looking at nameplate: HART = -xxx5xx and -xxx6xx. No or wrong DD (device description) HART interface missing Instrument not registered with HART server Load too low (load > 230 Ω required) HART receiver (e.g. FXA 191) not connected via load but via power supply Incorrect device address (address = 0 for single operation, address > 0 for multi-drop operation) Line capacitance too high Line interferences For further information see “HART® - Field communication Manual” Several devices set to same address Set addresses correctly. Communication not possible with several devices set to same address. No central PA/DP module Verify by looking at nameplate: PA = -xxx3xx /DP = xxx4xx. Upgrade to LSCP module, see chapter "Spare parts". Incorrect instrument software version Commuwin (CW) II: Incompatible CW II and instrument software versions No or incorrect DD/DLL Incorrect baud rate setting for segment coupler in DPV1 server Incorrect station (master) addressed or duplicate address Incorrect station (slaves) address Bus line not terminated Line problems (too long, cross section too small; not shielded, screen not grounded, wires not twisted) For further information, see "PROFIBUS PA/DP – Field communication Manual” No HART communication No PROFIBUS® communication Page 56 Equipment, spare parts, personnel Ohmmeter Upgrade to LSCH-H1 / -H2. Triton TR8 Error Possible Cause No PROFIBUS® communication Bus voltage too low (bus supply voltage typ. 24 V DC for non-Ex) Test and/or Remedial Measure Voltage at instrument’s PA/DP connector must be at least 9 V. Equipment, spare parts, personnel 6.4 INSTRUMENT SPECIFIC ERRORS The following table helps you during the diagnosis and points to any spare parts required. Depending on the degree of difficulty and the measuring equipment present, diagnosis is carried out by: • Trained operator personnel • The user’s trained electrical technicians • Company responsible for system installation/operation Information on the exact spare part designations and on how to install these parts can be found in the "Accessories" section. Error Possible Cause Device cannot be operated, display value 9999 Operation locked No line voltage Supply voltage wrong/too low Connection faulty Display dark, no light-emitting diode active Device fuse defective Power unit defective Central module defective Ribbon cable loose or defective Device gets hot Incorrect measurement turbidity and/or temperature Triton TR8 Voltage wrong/too high Power unit defective Transmitter module defective. Verify the error is not in the cabling or in the sensor, See Process Errors Test and/or Remedial Measure Equipment, spare parts, personnel Press CAL and MINUS keys simultaneously. See "Function of keys" section. Check whether line voltage is present. Compare actual line voltage and nameplate data. Terminal not tightened; insulation jammed; wrong terminals used. Compare line voltage and the nameplate data and replace fuse. Replace power unit, note variant. Replace central module, note variant. Check ribbon cable, replace if necessary. Compare line voltage and nameplate data. Replace power unit. Electrical technician/e.g. multimeter Electrical technician/e.g. multimeter Simulation of the sensor is not possible. Test the input with new or different sensor. Electrical technician Electrical technician/suitable fuse Consult Factory Consult Factory Consult Factory User, electrical technician Consult Factory If test negative: replace module, Consult Factory Page 57 Error Incorrect meas. Turbidity and/or temperature Current output, current value incorrect Possible Cause Data transfer disturbed (EMC) Wrong sensor cable / cable too long Adjustment not correct Load too big Shunt/short to ground in current loop Incorrect mode of operation No current output signal Current output stage defective No function of additional relay defective or loose ribbon cable Test and/or Remedial Measure Check cable routing. Separate sensor cable from supply cables. Maximum cable length with extension is 200 m (656 ft.) Check with installed current simulation, connect mA meter directly to current output. Check whether 0–20 mA or 4–20 mA is selected. Check with installed current simulation, connect mA meter directly to current output. Check ribbon cable seating, renew cable if required. Equipment, spare parts, personnel Connect sensor cable screen to "S" terminal , do not ground. If simulation value incorrect: adjustment in factory or new module required. If simulation values correct: check current loop for load and shunts. If simulation value incorrect: new module required. 6.5 SENSOR TROUBLESHOOTING 6.5.1 Troubleshooting instructions Troubleshooting must take account of the whole measuring system: • Transmitter • Electrical leads and connectors • Assembly • Sensor The possible causes of failure listed in the following table primarily refer to the sensor. Problem No display, no sensor reaction Reading too high or too low Reading greatly fluctuating Measured value jump to 9999 FNU Error 8 Wiper does not find end position Page 58 Check Mains voltage at transmitter? Sensor connected correctly? Medium flow present? Film formation on optical windows? Film formation on optical windows? Gas bubbles present? Sensor calibrated? Check data record. Check with check unit. Gas bubbles present? Check mounting location. Gas bubble formation on the optical windows? Water in sensor? Cable breakage? Incorrect cable connection? Visual inspection: Initial position not in end position? Remedy Connect mains voltage. Set up correct connection. Create flow. Clean the sensor. Clean the sensor. Remove gas bubbles. Calibrate. Change, if necessary. Check in factory. Remove gas bubbles. Select other mounting location. Change orientation. Adjust wiper interval. Increase gas bubble factor in the B9 menu. Send sensor to service. Check cable. Check wiring. Send sensor to service. Triton TR8 6.5.2 Checking the Sensor The sensor may only be checked by trained personnel. You will require a voltmeter. Problem Voltage check Slope check Zero point check Check Check supply voltage at transmitter (sensor connected) Check slope using 1000227-1, check unit Produce zero water by filtration (0.2 μm) Remedy 10 to 16 V between terminals 87 and 88 Stable measured value (1 to 6 FNU) < 0.1 FNU Note! If the values deviate from the set points, carry out trouble-shooting as per the trouble-shooting instructions or contact your sales center. 7.0 ENGINEERING DOCUMENTATION 7.1 SPECIFICATIONS 7.1.1Triton TR8 Sensor Measuring Principle: Measuring Range: Maximum Measured Error: Repeatability: Wavelength: Optical Reference Compensation: Factory Calibration: Temperature/Pressure: Connecting Cable: Maximum Cable Length: Temperature Sensor: Operating Range: Storage Range: Leakage Protection: Wetted Materials: Process Connections: Triton TR8 Nephlelometric 90° NIR Scattered light according to DIN/ISO 7027 (-1) 0.000 to 9999 FNU (-2) 0.00 to 9999 FNU 0.00 to 3000 ppm 0.00 to 9999 ppm 0.0 to 3.0 g/l 0.0 to 300 g/l 0.0 to 20.0 % 0.0 to 200% Less Than 5% of the measured value Less Than 1 % of measured value 880 nm By reference Diodes Formazine standard and SiO2 25 °C/ 6 bar to 50 °C/ 1bar 4 cores shielded 200 m 30 k-ohm NTC -5 – 50 °C -20 – 60 °C IP68 Body: PVC, PPS GF40 Window: Sapphire Wiper: Rubber Cable: Polyolefin Flow Assemblies: PVC, PE G1 and 3/4” NPT Page 59 7.1.2 Triton Analyzer Specifications Inputs Measurements Measurement principle Measured Variables Measuring range (Turbidity) Measuring range (Temp) Binary inputs 1 and 2 Current input Load Outputs Output signal Signal on alarm Load Transmission range Signal resolution Insulation voltage Overvoltage protection Auxiliary voltage output Contact outputs (floating changeover contacts) Limit contactor Alarm Turbidity and temperature The TR8 sensor communicates with the Analyzer through a digital interface. The analyzer receives a standardized turbidity and temperature signal from the TR8 sensor. Turbidity, Solids Content, Temperature 0.000 to 9999 FNU/NTU 0.00 to 9999 ppm 0.0 to 300 g/l 0.0 to 200.0 % -5.0 to +70.0 °C (+23 to +158 °F) Voltage: 10 ... 50 V, Current consumption: max. 10 mA 4 to 20 mA, galvanically isolated Load: 260 Ω for 20 mA (voltage drop 5.2 V) 0/4 to 20 mA, galvanically isolated, active, Terminals 31 & 32 2.4 or 22 mA in case of an error max. 500 Ω adjustable, min. Δ 0.1 FNU, Δ 0.1 (Temp) adjustable, Δ 10 to Δ 100 ppm, Δ 0.1 g/l, Δ 0.1 % % of measuring range max. 700 Digits/mA max. 350 Veff / 500 V DC according to EN 61000-4-5 Output voltage: 15 V ± 0,6 V Output current: max. 10 mA Switching current with ohmic Max. 2 A load (cos ϕ = 1): Switching current with inductive Max. 2 A load (cos ϕ = 0.4): Switching voltage: Max. 250 V AC, 30 V DC Switching capacity with ohmic Max. 500 VA , 60 W DC load (cos ϕ = 1): Switching capacity with inductive Max. 500 VA load (cos ϕ = 0.4): Pick-up/drop-out delay 0 to 2000 s Function (switchable): Latching/momentary contact Alarm delay: 0 to 2000 s (min) Power Supply Supply voltage Power consumption Mains fuse Page 60 Depending on ordered version: 100/115/230 V AC +10/-15 %, 48 ... 62 Hz 24 V AC/DC +20/-15 % Max. 7.5 VA Fine-wire fuse, semi-delay 250 V/3.15 A Triton TR8 Performance Characteristics Measured value resolution Maximum measured error Repeatability Offset range TR8 Turbidity 0.001 FNU, 0.01 ppm, 0.01 g/l, 0.01 % Temperature 0.1 °C Display TR8 ± 2 % of measured value (min. 0.02 FNU) Display (Temp) max. 1.0 % of measuring range Signal Output TR8 1 % of current output range (min. 0.02 FNU) Signal Output (Temp) max. 1.25 % of current output range max. 1 % of measured value TR8 ± 99.99 FNU, ± 99.99 ppm, ± 99.9 g/l, ± 99.9 % Temperature ± 5 °C Environment Ambient temperature range Ambient temperature limits Storage temperature Electromagnetic compatibility Overvoltage protection Degree of protection Relative humidity Triton TR8 -10 to +55 C (+14 to +131 °F) –20 to +60 C (-4 to +140 °F) –25 to +65 C (-13 to +149 °F) Interference emission and interference immunity as per EN 61326: 1997 / A1:1998 As per EN For outputs, binary inputs and current input 61000-4-5:1995 Panel-mounted IP 54 (front), IP 30 (housing) instrument: Field device: IP 65 10 to 95%, not condensing Page 61 7.2 OUTLINE & DIMENSIONAL DRAWING 7.3 WIRING DIAGRAM Optional Relay Card 4-20 mA Page 62 Triton TR8 Triton TR8 Page 63