Sizing Chart

Transcription

Sizing Chart
Sizing Chart
Introduction
The Thoraflex Hybridâ„¢ Sizing Chart incorporates a 15-25% oversize of ring stent
diameter to aortic diameter. Aortic diameter is based on inner vessel (I.D) diameter
measurements therefore no further oversize is required. If outside vessel diameters
(O.D) are measured then an allowance for the vessel wall diameter must be made
before using the sizing chart for device selection.
Vascutek Limited has performed testing and recommends that a 15-25% oversize
and a 40mm distal landing zone will provide optimum sealing within healthy
vessel.1
Catalogue
No.
Graft ID
(mm)
D1
Stent Graft
OD (mm)
D2
Descending Landing
Zone Vessel Internal
Diameter (mm)
D3 *
Stent Graft
Nominal Length
(mm)
L1**
THP2628X100
THP2830X100
THP3032X100
THP3034X100
THP3036X100
THP3238X100
THP3240X100
THP2628X150
THP2830X150
THP3032X150
THP3034X150
THP3036X150
THP3238X150
THP3240X150
26
28
30
30
30
32
32
26
28
30
30
30
32
32
28
30
32
34
36
38
40
28
30
32
34
36
38
40
22 - 24
24 - 26
25 - 27
27 - 29
29 - 31
30 - 33
32 - 34
22 - 24
24 - 26
25 - 27
27 - 29
29 - 31
30 - 33
32 - 34
100
100
100
100
100
100
100
150
150
150
150
150
150
150
* See Diagram 2. Oversize stated is approximately 15-25%.
** Nominal length quoted. Stent graft length may have some slight tolerance variance and can vary based
on individual patient anatomy eg. tortuosity
When oversizing in a dissection clinical judgement must be used on an individual patient basis.
It should be considered when measuring length that the device will take the outer curve of the aorta.
Refer to Diagram 2.
Some movement of the distal ring of the hybrid prosthesis may occur following re-perfusion of the
thoracic aorta please refer to IFU.
8mm 8mm
Stent
Graft
Graft
10mm
Collar
Diagram 1
100mm***
10mm
12mm
D2
D1
240mm
L1
*** Length shown to indicate marker position. Ring configuration will vary.
Diagram 2
L1
40mm
D3
NOTES
References
1. Vascutek design verification testing.
Patent Nos. US 6,740,111 B1 & EP. 0 910 309 B1
is a trademark of Vascutek Ltd.
VASCUTEK, a TERUMO Company
Newmains Avenue, Inchinnan
Renfrewshire PA4 9RR Scotland, UK
Tel: (+44) 141 812 5555
Fax: (+44) 141 812 7170
www.vascutek.com