Brochure 780_final - Lutech Industries
Transcription
Brochure 780_final - Lutech Industries
Tech nic al S p e c i f i c a t i o n s Safety ISO 13485: 2003 approved CE marking according to MDD93/42/EEC Degree of protection Against harmful ingress of water: IP X 1 Operating Environment Power: AC 100-250V, / 50/60Hz. Temperature: 32°F-104°F (0-40°C) Humidity: 15%-85% Patient range: Adult, Pediatric, Neonate Physical Characteristics Dimension: 16.9in x 13.4in x 15.4in (430mm X 340mm X 390mm) G.W: 21lb (9.5kg) N.W: 13.2lb (6kg) Performance Specifications Display: 12.1 inch color TFT touch screen Resolution: 800 X 600 Multi displays selectable, including: Standard display Freeze display Large-font display Trend coexist display Alarm limit display OxyCRG dynamic view display Bed-to-bed display Multi-lead and ECG simultaneous display Trace: 12 waveforms Sweep speed: Sweep speed: 12.5mm/s, Indicator lights: 25mm/s, 50mm/s Interface: Power and Battery Parameter cable interface AC Power input socket Network interface External VGA interface Alternative display Battery: Lithium-ion rechargeable battery Plug & play battery, capacity: 4000mA Maximum 12 hours for charging; Over 4 hours of continuous working time Trend time: 1-120 hours Alarm: User-adjustable high and low limits; Audible and visual alarm Networking: Connects to central monitoring system Recording: Built-in, thermal array Waveform: 2 channels Record mode: Manual, on alarm, time-defined, etc Paper width: 50mm Print speed: 12.5mm/s, 25mm/s, 50mm/s Internal Modules: ECG module, SpO2 module, NIBP module, Main control module and power supply module ECG-Cardio Tee™ 12-leads ECG Analysis Heart Rate Range: Adult: 15-300bpm Pediatric/Neonate: 15-350bpm Accuracy: ± 1% or ± 2bpm Resolution: 1 bpm ECG enlarge width: MON: 0.5-40Hz DIA: 0.05-130Hz OPE: 1-20Hz ECG waveform: 2 channels ST SEGMENT detection: -2.0mV - +2.0Mv (Automatic) Arrhythmia analysis: yes Electrode desquamated indicator: Gain selection: Sweep speed: HR alarming time: PR from ECG: Voice & light indication X1/4, X1/2, X1, X2 Auto 12.5mm/s, 25mm/s, 50mm/s ≤ 12s Adult: 15-300bpm Pediatric/ Neonate: 15-350bpm ECG noise level: -30uV (P-P value) ECG input loop current: 0.1 Ua Input impedance: >5MQ CMRR: >105 Db Time Constant: Monitoring mode >0.3s Diagnostics Mode >3.2s Anti-jamming: Line frequency, Electrosurgical interference and defibrillation Arrhythmia Analysis and pacemaker detection ECG cable: Standard configuration ECG cable for Adult Standard two ECG cables to support 12 leads ECG analysis Alarm setting: Adjustable from lower to upper and automatic memory Waveform: Standard 12-ECG waveforms Nellcor SpO2 Display: Measurement range: Resolution: Accuracy: Alarm range: SpO2 digital, Pulse columnar graphics, Volume waveform, Pulse rate 0-100% 1% ± 2% (90-100%, MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I and MAX-FAST sensors); ± 3% (70-100%, D-YS, DS-100A, OXI- A/N and OXI-P/I sensors) 0-69% unspecified Alarm range: 0-100% Pulse Rate Range: 20-300bpm Resolution: 1bpm Accuracy: ± 3bpm Alarm range: 20-300bpm Exclusive SatSeconds™ alarm management system Much lower false alarms, much less word for users AcuTee™ NIBP Measurement method: Automatic vibration Measurement types: Systolic, Diastolic, Mean Work Mode: Manual/ Automatic/ Continuous Auto measurement time: Adjustable 1 min-480min Measurement unit: mmHg/ Kpa selectable Measurement range: 10-270mmHg Range of Systolic Pressure: Adult: 40-270mmHg Pediatric: 40-200mmHg Neonatal Mode: 40-150mmHg Range of Diastolic Pressure: Adult: 10-215mmHg Pediatric: 10-150mmHg Neonatal Mode: 10-100mmHg Range of Mean Pressure: Adult/Ped. Mode: 40-210mmHg Neonatal Mode: 18-120mmHg Alarm Type: Systolic, diastolic, and mean pressure Accuracy: 5mmHg NIBP pressure range: 0-300mmHG PR from NIBP: 0-300mmHg Over-pressure protection: Hardware and software Both over pressure protection 300 mmHg is the max pressure Higher than the above pressure The safety valve deflate automatically NIBP cuff: Alarm Setting: IBP Measurement mode: Measurement types: Channel: Pressure names: Measurement range: IBP Accuracy: Pulse accuracy: Calibration mode: Dual TEMP Measurement range: Channel: Resolution: Accuracy: Measurement unit: TEMP sensor: Alarm setting: Adult, pediatric or neonate Adjustable from lower To upper and automatic memory Pressure transducer Systolic, Diastolic, Mean 2 Channel AVP, CVP, PAR, LAP, Rap, ICP 30mmHg – 300mmHg 0-100mmHg ± 4mmHg 100mmHg – 300mmHg ± 2% ± 1% or ± 1bpm Zero calibration Lutech TM 32°F-122°F (0-50°C) T1, T2, TD (Temperature Difference) 0.1°C 0.1°C (excluding sensor error) C/F SLECTABLE Skin or Rectal/Esophageal Adjustable form lower to upper and automatic memory RESP R-F (RA-LL) impedance method Method: PR measurement range: Adult: 7-120bpm Neonate: 7-150bpm 1rpm Resolution: 1rpm Accuracy: Adjustable from lower to upper Alarm setting: and automatic memory ARR analysis, RESP 10s-40s Apnea Alarm: EtCO2 Sidestream CO2 CO2 measurements range: Accuracy: Sample Rate: Sample rate accuracy: Start-up time: Respiration rate: Respiration rare accuracy: Response time: Sampling line length: Mainstream CO2 Method: Measurement mode: Measurement range: Resolution: Accuracy: 0-150 mmHg, 0 to 79% 0 to 20kPa (at 7660mmHg) ± 2 mm Hg (0 40mm Hg) ± %5 of reading (41 70mm Hg) ± 8% of reading (71 100 mm Hg) ± 10% of reading (101 150 mm Hg) 50 ml/min. ± 10ml/min 15% <1 min, once the module starts up, it reaches ISO accuracy Mode; minutes after start-up, the module reaches full accuracy mode 0-120bpm ± 2rpm (0-70rpm) ± 5rpm (>70rpm) <2s 7 inches; internal diameter: 0.055 inches; sampling gas flow rate: 150ml/min Infrared Absorption Mainstream 0-150 mm Hg, 0 to 79% 0 to 20kPa (at 760mmHg) 0.1 mmHg 0 to 69 mmHg 0.25mmHg 70 to 150mmHg ± 2 mmHg (0 40 mm Hg) ± 5% of reading (41-70 mm Hg) ± 8% of reading (71-100 mm Hg) ± 10% of reading (101-150 mm Hg) LUTECH INDUSTRIES INC. P h: 6 31 -58 4-6 68 8 TM 105 Rem ingt on B l v d . , S u i t e C , R o n k o n k o m a , N Y 11 7 7 9 U . S . A . Fx : 631- 584- 6699 We b s i t e : w w w. l u t e c h m e d i c a l . c o m E - m a i l : i n f o @ l u te ch m e d i ca l .co m Datalys 780 TM Intensive C ar e M oni tor Datalys 780 TM Anes theti c Gas I nt ensiv e Ca re Mo ni to r... In collaboration with Phase-in of Sweden, the Datalys™ 780 adopts the advanced anesthetic gas module for monitoring 8 types of gas (O2, CO2, N2O, ENF, ISO, DES, SEV, HAL). The module includes automatic recognition of the anesthetic gas, short warm-up time, long service life and support for the MAC value (minimum alveolar concentration). Wh e n e v e r y s e c o n d m at t ers ! The Datalys™ 780 uses CardioTec™ ECG technology, world leading OxiMax technology, high precision AcuTec™ hypertension monitoring technology as a standard configuration. Also incorporated are the world’s best medical technologies for SpO2, ICG, EtCO2, Anesthesia Gas Monitoring from some of the world’s best OEM technology providers, such as NELLCOR, RESPIRONICS, ARTEMA & BIOZ. In essence, the Datalys™ 780 provides medical professionals with powerful monitoring options and functionalities that match the demands of a wide range of care in any hospital environment. EtC O 2 The DatalysTM 780 offers the plug and play EtCO2 from Respironics. Select the Capnostat 5 mainstream sensor for optimal performance in monitoring intubated patients. Small, durable and lightweight, the mainstream sensor provides accurate and reliable monitoring for all intubated patients, from neonates to adults. No calibration is required Select the LoFlo sidestream sensor for monitoring non-intubated patients. Flexible and compact, the LoFlo CO2 sensor provides consistent and reliable monitoring of adult, pediatric Sample rate ≤ 50ml/min (micro-stream). E CG CardioTec™ 12-lead ECG waveform synchronously display High Precision ECG measurement technology, assists in precisely diagnosis of patients. CMRR≥105dB, outstanding ECG anti-interference capability. Support arrhythmia analysis & alarm review. N on- i nv as i v e H em ody nam i c Collaborating with US BIOZ. The Datalys™ 780 offers impedance cardiography for non-invasive continuous hemodynamic monitoring Micro-signal transmit through disposable electrodes. Measurement and display of ICG. DISQ® technology processes impedance signal variation caused by variations in blood volume and Blood flow velocity from heartbeat to heartbeat. Variation of impedance is applied to non-invasive Z MARC™ algorithm for acquiring SV, CO, SVR, contractility Sp O 2 Advanced SpO2 technology, OxiMax® Unique sensor with Digital Memory Chip. Every sensor is calibrated to it’s own specific needs. Desaturation Event Reports in the sensor. Unique LoSaTTM provides highest accuracy range, expanding it from 60% to 100% Exclusive SatSecondsTM alarm management system lowers false alarms and workload for users. Intel l i gent Al ar m I-KloK™ intelligent alarm management allows auto identification of alarm level. Self adjust proper alarm time to reduce false alarms. IPX1 Level Waterproof Auto dust shield Handwriting pen, SD card and USB interface. Wall mount, rolling stand/trolley 4000mAh Li-ion battery, over 4 hours of continuous working time