Brochure 780_final - Lutech Industries

Transcription

Brochure 780_final - Lutech Industries
Tech nic al S p e c i f i c a t i o n s
Safety
ISO 13485: 2003 approved
CE marking according to MDD93/42/EEC
Degree of protection
Against harmful ingress of water: IP X 1
Operating Environment
Power:
AC 100-250V, / 50/60Hz.
Temperature:
32°F-104°F (0-40°C)
Humidity:
15%-85%
Patient range:
Adult, Pediatric, Neonate
Physical Characteristics
Dimension:
16.9in x 13.4in x 15.4in
(430mm X 340mm X 390mm)
G.W:
21lb (9.5kg)
N.W:
13.2lb (6kg)
Performance Specifications
Display:
12.1 inch color TFT touch screen
Resolution:
800 X 600
Multi displays selectable, including:
Standard display
Freeze display
Large-font display
Trend coexist display
Alarm limit display
OxyCRG dynamic view display
Bed-to-bed display
Multi-lead and ECG simultaneous
display
Trace:
12 waveforms
Sweep speed:
Sweep speed: 12.5mm/s,
Indicator lights:
25mm/s, 50mm/s
Interface:
Power and Battery
Parameter cable interface
AC Power input socket
Network interface
External VGA interface
Alternative display
Battery:
Lithium-ion rechargeable battery
Plug & play battery, capacity:
4000mA Maximum 12 hours for
charging; Over 4 hours of
continuous working time
Trend time:
1-120 hours
Alarm:
User-adjustable high and low
limits; Audible and visual alarm
Networking:
Connects to central monitoring
system
Recording:
Built-in, thermal array
Waveform: 2 channels
Record mode: Manual, on alarm,
time-defined, etc
Paper width: 50mm
Print speed: 12.5mm/s, 25mm/s,
50mm/s
Internal Modules:
ECG module, SpO2 module,
NIBP module, Main control
module and power supply module
ECG-Cardio Tee™ 12-leads ECG Analysis
Heart Rate Range:
Adult: 15-300bpm
Pediatric/Neonate: 15-350bpm
Accuracy:
± 1% or ± 2bpm
Resolution:
1 bpm
ECG enlarge width: MON: 0.5-40Hz
DIA: 0.05-130Hz
OPE: 1-20Hz
ECG waveform:
2 channels
ST SEGMENT
detection:
-2.0mV - +2.0Mv (Automatic)
Arrhythmia analysis: yes
Electrode
desquamated
indicator:
Gain selection:
Sweep speed:
HR alarming time:
PR from ECG:
Voice & light indication
X1/4, X1/2, X1, X2 Auto
12.5mm/s, 25mm/s, 50mm/s
≤ 12s
Adult: 15-300bpm
Pediatric/ Neonate: 15-350bpm
ECG noise level:
-30uV (P-P value)
ECG input loop current: 0.1 Ua
Input impedance:
>5MQ
CMRR:
>105 Db
Time Constant:
Monitoring mode >0.3s
Diagnostics Mode >3.2s
Anti-jamming:
Line frequency, Electrosurgical
interference and defibrillation
Arrhythmia Analysis and pacemaker detection
ECG cable:
Standard configuration ECG
cable for Adult
Standard two ECG cables to
support 12 leads ECG analysis
Alarm setting:
Adjustable from lower to upper
and automatic memory
Waveform:
Standard 12-ECG waveforms
Nellcor SpO2
Display:
Measurement range:
Resolution:
Accuracy:
Alarm range:
SpO2 digital, Pulse columnar
graphics, Volume waveform,
Pulse rate
0-100%
1%
± 2% (90-100%, MAX-A,
MAX-AL, MAX-N, MAX-P, MAX-I
and MAX-FAST sensors);
± 3% (70-100%, D-YS, DS-100A,
OXI- A/N and OXI-P/I sensors)
0-69% unspecified
Alarm range: 0-100%
Pulse Rate
Range:
20-300bpm
Resolution:
1bpm
Accuracy:
± 3bpm
Alarm range:
20-300bpm
Exclusive SatSeconds™ alarm management system
Much lower false alarms, much less word for users
AcuTee™ NIBP
Measurement method:
Automatic vibration
Measurement types:
Systolic, Diastolic, Mean
Work Mode:
Manual/ Automatic/ Continuous
Auto measurement time: Adjustable 1 min-480min
Measurement unit:
mmHg/ Kpa selectable
Measurement range:
10-270mmHg
Range of Systolic Pressure:
Adult: 40-270mmHg
Pediatric: 40-200mmHg
Neonatal Mode: 40-150mmHg
Range of Diastolic Pressure:
Adult: 10-215mmHg
Pediatric: 10-150mmHg
Neonatal Mode: 10-100mmHg
Range of Mean Pressure:
Adult/Ped. Mode: 40-210mmHg
Neonatal Mode: 18-120mmHg
Alarm Type:
Systolic, diastolic, and mean
pressure
Accuracy:
5mmHg
NIBP pressure range:
0-300mmHG
PR from NIBP:
0-300mmHg
Over-pressure protection: Hardware and software
Both over pressure protection
300 mmHg is the max pressure
Higher than the above pressure
The safety valve deflate
automatically
NIBP cuff:
Alarm Setting:
IBP
Measurement mode:
Measurement types:
Channel:
Pressure names:
Measurement range:
IBP Accuracy:
Pulse accuracy:
Calibration mode:
Dual TEMP
Measurement range:
Channel:
Resolution:
Accuracy:
Measurement unit:
TEMP sensor:
Alarm setting:
Adult, pediatric or neonate
Adjustable from lower
To upper and automatic memory
Pressure transducer
Systolic, Diastolic, Mean
2 Channel
AVP, CVP, PAR, LAP, Rap, ICP
30mmHg – 300mmHg
0-100mmHg ± 4mmHg
100mmHg – 300mmHg ± 2%
± 1% or ± 1bpm
Zero calibration
Lutech
TM
32°F-122°F (0-50°C)
T1, T2, TD (Temperature
Difference)
0.1°C
0.1°C (excluding sensor error)
C/F SLECTABLE
Skin or Rectal/Esophageal
Adjustable form lower to upper
and automatic memory
RESP
R-F (RA-LL) impedance method
Method:
PR measurement range: Adult: 7-120bpm
Neonate: 7-150bpm
1rpm
Resolution:
1rpm
Accuracy:
Adjustable from lower to upper
Alarm setting:
and automatic memory
ARR analysis, RESP
10s-40s
Apnea Alarm:
EtCO2
Sidestream CO2
CO2 measurements
range:
Accuracy:
Sample Rate:
Sample rate accuracy:
Start-up time:
Respiration rate:
Respiration rare
accuracy:
Response time:
Sampling line length:
Mainstream CO2
Method:
Measurement mode:
Measurement range:
Resolution:
Accuracy:
0-150 mmHg, 0 to 79%
0 to 20kPa (at 7660mmHg)
± 2 mm Hg (0 40mm Hg)
± %5 of reading (41 70mm Hg)
± 8% of reading (71 100 mm Hg)
± 10% of reading (101 150 mm
Hg)
50 ml/min. ± 10ml/min
15%
<1 min, once the module starts
up, it reaches ISO accuracy
Mode; minutes after start-up, the
module reaches full accuracy
mode
0-120bpm
± 2rpm (0-70rpm)
± 5rpm (>70rpm)
<2s
7 inches; internal diameter:
0.055 inches; sampling gas flow
rate: 150ml/min
Infrared Absorption
Mainstream
0-150 mm Hg, 0 to 79%
0 to 20kPa (at 760mmHg)
0.1 mmHg 0 to 69 mmHg
0.25mmHg 70 to 150mmHg
± 2 mmHg (0 40 mm Hg)
± 5% of reading (41-70 mm Hg)
± 8% of reading (71-100 mm Hg)
± 10% of reading (101-150 mm
Hg)
LUTECH INDUSTRIES INC.
P h: 6 31 -58 4-6 68 8
TM
105 Rem ingt on B l v d . , S u i t e C , R o n k o n k o m a , N Y 11 7 7 9 U . S . A .
Fx : 631- 584- 6699
We b s i t e : w w w. l u t e c h m e d i c a l . c o m
E - m a i l : i n f o @ l u te ch m e d i ca l .co m
Datalys 780
TM
Intensive C ar e M oni tor
Datalys 780
TM
Anes theti c Gas
I nt ensiv e Ca re Mo ni to r...
In collaboration with Phase-in of Sweden, the Datalys™ 780
adopts the advanced anesthetic gas module for monitoring 8
types of gas (O2, CO2, N2O, ENF, ISO, DES, SEV, HAL). The
module includes automatic recognition of the anesthetic gas,
short warm-up time, long service life and support for the MAC
value (minimum alveolar concentration).
Wh e n e v e r y s e c o n d m at t ers !
The Datalys™ 780 uses CardioTec™ ECG technology, world leading OxiMax technology, high precision AcuTec™
hypertension monitoring technology as a standard configuration. Also incorporated are the world’s best medical
technologies for SpO2, ICG, EtCO2, Anesthesia Gas Monitoring from some of the world’s best OEM technology
providers, such as NELLCOR, RESPIRONICS, ARTEMA & BIOZ. In essence, the Datalys™ 780 provides medical
professionals with powerful monitoring options and functionalities that match the demands of a wide range of care in
any hospital environment.
EtC O 2
The DatalysTM 780 offers the plug and play EtCO2 from Respironics.
Select the Capnostat 5 mainstream sensor for optimal performance in monitoring intubated
patients.
Small, durable and lightweight, the mainstream sensor provides accurate and reliable monitoring
for all intubated patients, from neonates to adults.
No calibration is required
Select the LoFlo sidestream sensor for monitoring non-intubated patients.
Flexible and compact, the LoFlo CO2 sensor provides consistent and reliable monitoring of adult,
pediatric
Sample rate ≤ 50ml/min (micro-stream).
E CG
CardioTec™ 12-lead ECG waveform synchronously display High
Precision ECG measurement technology, assists in precisely diagnosis
of patients.
CMRR≥105dB, outstanding ECG anti-interference capability.
Support arrhythmia analysis & alarm review.
N on- i nv as i v e H em ody nam i c
Collaborating with US BIOZ. The Datalys™ 780 offers impedance
cardiography for non-invasive continuous hemodynamic monitoring
Micro-signal transmit through disposable electrodes.
Measurement and display of ICG.
DISQ® technology processes impedance signal variation caused by
variations in blood volume and Blood flow velocity from heartbeat to
heartbeat.
Variation of impedance is applied to non-invasive Z MARC™ algorithm
for acquiring SV, CO, SVR, contractility
Sp O 2
Advanced SpO2 technology, OxiMax®
Unique sensor with Digital Memory Chip. Every sensor is calibrated to
it’s own specific needs. Desaturation Event Reports in the sensor.
Unique LoSaTTM provides highest accuracy range, expanding it from
60% to 100%
Exclusive SatSecondsTM alarm management system lowers false alarms
and workload for users.
Intel l i gent Al ar m
I-KloK™ intelligent alarm management allows auto identification
of alarm level.
Self adjust proper alarm time to reduce false alarms.
IPX1 Level Waterproof
Auto dust shield
Handwriting pen,
SD card and USB
interface.
Wall mount, rolling
stand/trolley
4000mAh Li-ion
battery, over 4
hours of continuous working time