2.7 mm LCP Ulna Osteotomy System Powerpoint
Transcription
2.7 mm LCP Ulna Osteotomy System Powerpoint
2.7 mm LCP Ulna Osteotomy System Low profile angular stable fixation for ulna shortening osteotomies. Index – Introduction – Features & Benefits – Indications – X-ray Case Example – Implants & Instrumentation – Technique – – Oblique (45°) Transverse (90°) UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Introduction – The 2.7 mm LCP Ulna Osteotomy System offers the surgeon options to perform either a freehand or more guided shortening of the ulna with specific instrumentation. – The system aims to minimize soft tissue irritation through – improved implant surface – tapered shape and reduced bulk of the plate UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Indications UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Indications The 2.7 mm LCP Ulna Osteotomy System is indicated for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, including osteopenic bone, in the ulna. The 2.7 mm LCP Ulna Osteotomy System is indicated for use in both adults and adolescents (12-21 years) in which growth plates have fused, or in which growth plates will not be crossed by screw fixation. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A X-ray Case Example UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A X-ray Case Example Pre-op Results from case studies are not predictive of results in other cases. Results in other cases may vary. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A X-ray Case Example Post-op Results from case studies are not predictive of results in other cases. Results in other cases may vary. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Features and Benefits UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Features and Benefits Features Benefits Low-profile angular stable plate with rounded edges and tapered ends; available in 6- and 8-holes, SS or Ti Minimize likelihood of soft tissue irritation; ability to bury lag screw in plate (oblique cuts) Combi-holes for 2.7 mm locking and 2.7 mm cortex screws (Outer screw holes for locking screws only) Flexible screw placement: – lag for oblique cuts (most central holes) – neutral – locking Parallel saw blades for transverse (90°) and oblique (45°) cuts of 2.0, 2.5, 3.0, 4.0, and 5.0 mm widths Ability to create osteotomy in one cut Drill templates for pre-drilling of plate fixation holes before osteotomy; correspond to parallel saw blade widths Ensure correct rotational alignment Saw guide Assisted guidance during oblique cuts Compression/distraction instrument for compression or distraction of osteotomy; accepts 2.0 mm K-wires (self-drilling) Aids in obtaining adequate compression to help promote proper bone union – Only compatible with Synthes Power Allows for shortenings >5 mm with freehand technique (no drill template) UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Implants and Instrumentation UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Implants: Plates – 2.7 mm LCP Ulna Osteotomy Plate – 0X.111.900: 6 holes (length 62 mm) – 0X.111.901: 8 holes (length 76 mm) – Three-part center hole: – Plate is slightly bent (2° on each side, radius 45 cm) X = 2: Stainless Steel X = 4: Titanium Alloy (TAN) UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Implants: Screws and Wires – 2.7 mm Cortex Screw Stardrive, T8 StarDrive, self-tapping – – 2.7 mm Locking Screw, T8 StarDrive, self-tapping – – X02.870 – 90 (10 – 30 mm) X02.868 - 24 (8 – 24 mm) 2.0 mm Kirschner Wire with drill tip – Length 100 mm, Stainless Steel – Length 150 mm, Stainless Steel X = 2: Stainless Steel X = 4: Titanium Alloy (TAN) UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Instruments – Drill templates for 2.7 mm LCP Ulna Osteotomy Plate Item Number Shortening 03.111.900 2.0 mm 03.111.901 2.5 mm 03.111.902 3.0 mm 03.111.903 4.0 mm 03.111.904 5.0 mm UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Instruments – Parallel Saw Blades for LCP Ulna Osteotomy Plate: compatible to Synthes Small Battery Drive with Oscillating Saw Attachment Art. No. Osteotomy angle Shortening Art. No. Osteotomy angle Shortening 532.081S Transverse (90°) 2.0 mm 532.091S Oblique (45°) 2.0 mm 532.082S Transverse (90°) 2.5 mm 532.092S Oblique (45°) 2.5 mm 532.083S Transverse (90°) 3.0 mm 532.093S Oblique (45°) 3.0 mm 532.084S Transverse (90°) 4.0 mm 532.094S Oblique (45°) 4.0 mm 532.085S Transverse (90°) 5.0 mm 532.095S Oblique (45°) 5.0 mm UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Instruments – Handle with Quick Coupling – Handle for Torque Limiting Attachment – Torque Limiting Attachment, 0.8 Nm, quick coupling – StarDrive Screwdriver Shaft, T8, 105 mm – StarDrive Screwdriver Shaft, T8, 55 mm – 2.0 mm Drill Bit, with depth mark, quick coupling, 85 mm – 2.7 mm Drill Bit, quick coupling, 100 mm UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Instruments – Depth Gauge for 2.0 mm, 2.4 mm, and 2.7 mm Screws – 2.0 mm Threaded Drill Guide, with depth gauge – 2.7 mm Universal Drill Guide – Holding Forceps with ball tip, soft lock, 156 mm – Reduction Forceps, serrated jaw, ratchet, 144 mm UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Instruments – Saw Guide – Holding Sleeve – Tap for 2.7 mm Cortex Screws – Tap for 2.7 mm Locking Screws – Compression/Distraction Instrument UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Surgical Technique: For an oblique (45°) osteotomy UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Approach 1. Preparation and Patient Postitioning Place the patient in a supine position with the forearm positioned on a hand table in full supination and the shoulder in 90° abduction. The use of a tourniquet and magnifying loops is recommended. Make a longitudinal incision to approach the volar side of the ulna. Plate placement: between distal and middle third of ulna, just proximal to pronator quadratus muscle. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 2. Position and fix drill template Choose the drill template according to the intended amount of shortening. The shape of the drill template is identical to the shape of the short LCP Ulna Osteotomy Plate 2.7. Fix the Drill Template on the bone with the two different lengths KWires available (bicortical fixation). The position of the drill template corresponds to the later plate position (short plate). The drill template should fit properly to the bone, if not plate contouring is necessary. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 3. Apply compression/distraction instrument Position the compression/distraction instrument with the 100 mm 2.0 mm Kirschner wires (02.111.902.01) in the inner part of the clamps. Tighten clamps clockwise to retain the Kirschner wires. For additional support, insert the 150 mm 2.0 mm Kirschner wires (02.111.903.01) into the outer part of the clamps. Note: – Both tips of the compression/distraction instrument should be in contact with the bone – Be sure Kirschner wires do not collide UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps Select parallel saw blade Choose an oblique (45°) cut saw blade indicated for the desired amount of shortening. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 4. Perform oblique osteotomy Attach the saw guide to the drill template for guidance while making the cut. With the Drill template in contact with the bone and the parallel saw blade in motion, create the oteotomy. Remember to irrigate during sawing to avoid excessive heating. Once cut is complete, turn the spindle on the compression/distraction instrument clockwise to reduce the osteotomy. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 5. Remove drill template and place plate After osteotomy completion, remove drill template and place the plate over the K-Wires. Always include the oblong holes with the most proximal and the most distal K-Wire. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 6. Insert cortex screws Remove the K-Wire distal to the osteotomy, measure with depth gauge, and insert the appropriate 2.7 mm cortex screw. Repeat with second distal k-wire and proximal k-wire. Note: The holding forceps with ball tip can be placed in the proximal locking hole to keep plate in position when inserting the most proximal cortex screw. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 7. Insert lag screw Overdrill the near cortex using the 2.7 mm universal drill guide and the 2.7 mm drill bit. Drill the threaded hole using the 2.0 mm end of the 2.7 mm universal drill guide and the 2.0 mm drill bit. Measure the screw length using the depth gauge. Insert appropriate 2.7 mm cortex screw. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 8. Insert locking screws Insert the threaded drill guide into the most distal locking hole until fully seated. Using the 2.0 mm drill bit, drill to the desired depth. Measure the screw length by using the scale on the drill guide or with the depth gauge. Note: In especially hard bone, the use of the tap for 2.7 mm locking screws may be necessary. Insert the appropriate 2.7 mm locking screw. Repeat with additional locking screws as needed. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Surgical Technique: For a transverse (90°) osteotomy UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Approach 1. Preparation and Patient Postitioning Place the patient in a supine position with the forearm positioned on a hand table in full supination and the shoulder in 90° abduction. The use of a tourniquet and magnifying loops is recommended. Make a longitudinal incision to approach the volar side of the ulna. Plate placement: between distal and middle third of ulna, just proximal to pronator quadratus muscle. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 2. Position and fix drill template Choose the drill template according to the intended amount of shortening. The shape of the drill template is identical to the shape of the short LCP Ulna Osteotomy Plate 2.7. Fix the Drill Template on the bone with the two different lengths KWires available (bicortical fixation). The position of the drill template corresponds to the later plate position (short plate). The drill template should fit properly to the bone, if not plate contouring is necessary. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 3. Apply compression/distraction instrument Position the compression/distraction instrument with the 100 mm 2.0 mm Kirschner wires (02.111.902.01) in the inner part of the clamps. Tighten clamps clockwise to retain the Kirschner wires. For additional support, insert the 150 mm 2.0 mm Kirschner wires (02.111.903.01) into the outer part of the clamps. Note: – Both tips of the compression/distraction instrument should be in contact with the bone – Be sure Kirschner wires do not collide UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps Select parallel saw blade Choose a transverse (90°) cut saw blade indicated for the desired amount of shortening. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 4. Perform transverse osteotomy With the drill template in contact with the bone and the parallel saw blade in motion, create the osteotomy. Remember to irrigate during sawing to avoid excessive heating. Once cut is complete, turn the spindle on the compression/distraction instrument clockwise to reduce the osteotomy. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 5. Remove drill template and place plate After complete osteotomy, remove drill template and place the plate over the K-Wires. Always include the oblong holes with the most proximal and the most distal K-Wire. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 6. Insert cortex screws Remove the K-Wire distal to the osteotomy, measure with depth gauge, and insert the appropriate 2.7 mm cortex screw. Repeat with second distal k-wire and proximal k-wire. Note: The holding forceps with ball tip can be placed in the proximal locking hole to keep plate in position when inserting the most proximal cortex screw. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Technique: Surgical steps 8. Insert locking screws Insert the threaded drill guide into the most distal locking hole until fully seated. Using the 2.0 mm drill bit, drill to the desired depth. Measure the screw length by using the scale on the drill guide or with the depth gauge. Note: In especially hard bone, the use of the tap for 2.7 mm locking screws may be necessary. Insert the appropriate 2.7 mm locking screw. Repeat with additional locking screws as needed. UOS In-Service / April 2012 / For Synthes Sales Consultant use only. Do not distribute. DJ8675-A Thank you