Simultaneous measurement of cardiac output by
Transcription
Simultaneous measurement of cardiac output by
British Journal of Anaesthesia 1994; 72: 133-138 EQUIPMENT Simultaneous measurement of cardiac output by thermodilution, thoracic electrical bioimpedance and Doppler ultrasound G. CASTOR, R. K. KLOCKE, M. STOLL, J. HELMS AND I. NIEDERMARK SUMMARY To evaluate the accuracy of two non-invasive techniques for cardiac output (CO) measurement, we have measured CO simultaneously by thoracic electrical bioimpedance (TEB), pulsed Doppler ultrasound (DU) and standard thermodilution methods (TD) under different clinical conditions. Measurements were made in 10patients: (I) during steady state anaesthesia with controlled IPPV ventilation (n = 131), spread over the entire ventilator/ cycle; (II) during apnoea (n = 56); (III) during spontaneous breathing (n = 152) in the intensive care unit. Mean (SD) cardiac output values were: (I) COTD 3.5 (1.0) litre min~1, COTEB 3.4 (0.7) litre min-1, CODU 2.8 (0.7) litre min'1; (II) COTD 3.6 (0.6) litre min-1, COTEB 3.5 (0.4) litre min-1, CODU 2.9 (0.7) litre min'1; (III) COTD 7.7 (1.5) litre min-1, COTEB 7.6 (1.9) litre min-1, CODU 5.2(1.4) litre min'1. The mean percentage deviation of TEB from TD ranged from -2.2% to 1.4% and that of DU from TD was from -16% to -32%. There were no statistically significant differences between TD and TEB, but TD and DU differed significantly during IPPV, apnoea and spontaneous ventilation (P < 0.0001). (Br. J. Anaesth. 1994; 72: 133-138) KEY WORDS Measurement techniques: cardiac output. Thermodilution is the most common clinical method for measurement of cardiac output. However, there are some theoretical and practical limitations to dilution techniques and small but significant associated morbidity and mortality [1, 2]. Consequently, many efforts have been made recently to develop a non-invasive method with similar accuracy and reproducibility as for thermodilution. Two of these alternative methods are thoracic electrical bioimpedance (TEB) and Doppler ultrasound (DU). Thoracic electrical bioimpedance ( TEB) The concept of TEB waveform analysis for estimating stroke volume was introduced in 1966 by Kubicek and colleagues [3]. They assumed that the human thorax is a cylinder with a basic circumference equal to that of the thorax at the level of the xiphoid [4]. This cylinder has an electrical length (L), which is the distance between the voltage- sensing electrodes placed on the neck and the xiphoid; it is perfused homogeneously with blood of specific resistivity, p, (units (Q • cm) which varies with PCV and has a basic impedance, Z o , (Q). The principle of impedance cardiography is that a constant small current (2.5 mA at 70 kHz), applied between the outer current electrodes (1 and 4) on the human thorax, produces changes in the voltage sensed with the inner electrodes (2 and 3) when the thoracic conductivity changes during systole and diastole. The first derivative of the impedance waveform, dZ/dt, is related linearly to aortic blood flow. There exist some modifications of the original assumption and similar basic analysis, such as the formula of Sramek [5], who took a truncated cone instead of a cylinder for the electrically participating tissue of the thorax and who regarded PCV as a constant. The detailed derivation of the equation of Sramek was explained in an article by Bernstein [4]. Doppler ultrasound (DU) The Doppler effect describes the changes in sound wave frequencies when the sound and receiver are moving relative to one another. The waves are directed towards the moving blood cells and then reflected back. This results in a frequency differing from that transmitted and directly proportional to blood flow velocity. Stroke volume can be calculated when the area under the velocity curve and the crosssectional area (CSA) of the aorta are known. A newer pulsed Doppler obviates the need for measuring CSA separately by insonating the ascending aorta with a two-element, pulsed, 2-MHz ultrasound beam from the suprasternal notch. The ultrasound beam comprises a wide beam, which insonates the total lumen of the ascending aorta and GERALD CASTOR*, M.D.; REINHARD K. KxoocEt; M.D.; MARTIN STOLI4, M.D.; JORGEN HELMS$, M.D.; INGRID NIEDERMARK§, M.D.; Department of Anaesthesiology, University of Saarland, 66424 Homburg, Germany. Accepted for Publication: July 15, 1993. Present addresses: * Department of Anaesthesiology, Caritas Krankenhaus, Werkstr. 1, 66763 Dillingen/Saar, Germany. f Department of Cardiology, Marienhospital, 49377 Vechta. $ Department of Neurology, University of Saarland, 66424 Homburg. J Department of Anaesthesiology, Stadtisches Krankenhaus, 67655 Kaiserslautern. 5 Department of Anaesthesiology, Kreiskrankenhaus, 66333 Volklingen. Correspondence to G.C.* BRITISH JOURNAL OF ANAESTHESIA 134 a narrow beam which covers a denned area which lies within this vessel. This enables the meaurement of both CSA and blood flow velocity from the suprasternal notch. The flow lumen area of the wide beam, which totally encompasses the ascending aorta, depends on the angle of insonation and the cross sectional area of the aorta. The average blood flow velocity is measured using the Doppler shift frequency of the wide beam; as this beam covers the whole lumen of the aorta, it is an average value of the whole flow lumen area. So, calculation of CO is independent of the angle of insonation. When the angle increases, the velocity increases and the lumen area decreases proportionally. This independence, which is new compared with other Doppler measurements of CO, is important for the practical use of this method. When compared with standard invasive methods, the results of TEB and Doppler methods range from poor correlation [6-10] to good agreement [11-16]. The purpose of this study was to evaluate the accuracy of CO measurements obtained simultaneously by TEB and DU in comparison with the standard TD method in different clinical conditions. MATERIAL AND METHODS The study was approved by the local Ethics Committee and the patients gave written, informed consent. We studied 10 patients (four male), ASA III and IV, aged 35-65 yr, who had a Swan-Ganz catheter (93A-131-7F, Baxter Edwards Laboratories, Irvine, CA) inserted during neurosurgical removal of intracranial tumour or aneurysm. The catheter was inserted via the right internal jugular vein and placed in the pulmonary artery, as confirmed radiologically. Two pairs of electrodes were applied (5 cm apart) to the neck and lower thoracic aperture at the level of the xiphoid and connected to an impedance cardiograph (NCCOM3-R7, BoMed Med., Irvine, CA). Measurements started after incision of the dura mater at steady-state conditions of arterial pressure and heart rate. TEB and TD measurements (Cardiac Output Computer 9520A, Baxter Edwards Laboratories, Irvine, CA with a closed injectate system) were made simultaneously. Ice-cold 5 % glucose 10 ml was injected via a closed system for TD measurements. Measurement of CO was performed by a time-temperature integration curve of the indicator-blood mixture. As it is the standard method used by Baxter, integration was terminated automatically when the curve returned to 30% of its peak, because this point is related to minimal variability. Deadspace correction was made according to instructions given by Edwards Laboratories. Measurement procedure was as follows: during injection of the ice-glucose until the presentation of the result by the TD computer, COTEB values were derived by beat-to-beat mode, normally four to six values, depending on the heart rate, and the mean taken for comparison. Thereafter, the Doppler transducer (Quantascope-Vital Science, Denver, CO) was placed at the suprasternal notch. Normally, a measuring time of 15—30 s was necessary to obtain one averaged CODU value, depending on the signal quality, which was then taken for comparison with the two other methods. One single TD measurement was compared with the averaged TEB and DU values. Simultaneous measurement of TEB and DU leads to prolonged disturbance of the impedance signal. The metallic Doppler transducer absorbs a large portion of the current between the two inner sensing electrodes of the impedance cardiograph; this reduces the dZ/dtmax signal and thus calculation of the TEB cardiac output results in decreased CO values. The investigator using the Doppler signal was unaware of any of the TD and TEB measurements. CO measurements took place during the entire ventilatory cycle (I) and during endexpiratory apnoea (II). Mechanical ventilation was set at a ventilatory frequency of 12 b.p.m. and a minute volume of 100 ml kg"1. Measurement of TD, TEB and DU was repeated on the next day in the intensive care unit during spontaneous breathing (III) after tracheal extubation. Data are presented as mean (SD) and as mean differences with SD of the differences (SDD). They were analysed for statistical significance using the Wilcoxon matched pairs signed rank test for nonparametric data. Statistically significant differences between TD and TEB and between TD and DU were assigned at P < 0.05. A bias analysis was performed for cardiac outputs obtained using TD and TEB and using TD and DU. According to the statistical method of Bland and Altman [17], we plotted the mean of TD and TEB and the mean of TD and DU against the percentage deviation of TEB from TD and of DU from TD. As confidence and bias are the same when expressed as percentages, we modified the Bland-Altman method using percentage deviations instead of absolute values. Two SD was chosen as confidence interval. In this manner, the limits of agreement of each noninvasive method compared with TD can be taken as bias±2SD. RESULTS During IPPV, 131 simultaneous measurements of TD, TEB and DU were analysed in 10 patients. The mean of all TD CO measurements was 3.5 (SD 1.0) litre min"1, that of TEB 3.4 (0.7) litre mnr 1 and that of DU 2.8 (0.7) litre min-1. During IPPV, the respective mean percentage deviation of TEB from TD was 1.4 (SDD 16.2)% and that of DU from TD - 1 6 (21)% (fig. 1). During apnoea, the mean of all TD measurements was 3.6 (0.6) litre min-1, that for TEB 3.5 (0.4) litre min"1 and that for DU 2.9 (0.7) litre min"1. The mean percentage deviation of TEB from TD was -2.2 (11.2)% and that of DU from TD - 18.7 (16.3)% (fig. 2). In the intensive care unit, during spontaneous ventilation, the mean of TD measurements was 7.7 (1.5) litre min"1, that for TEB 7.6 (1.9) litre mur 1 and that for DU 5.2 (1.4) litre min"1. The mean percentage deviation of TEB from TD was —2.1 (11.0)% and of DU from TD -32.4 (12.4)% (fig- 3). There was no statistically significant difference NON-INVASIVE VS INVASIVE CARDIAC OUTPUT MEASUREMENT 135 8080 • £ 2SD Q Mean Q c o co 0- c o 2SD to 402SD oMean ' I • v -40 > -40 -4 D Q •2SD -80 -80 2 3 4 5 6 7 Mean COTD + COTEB (litre min" 1 ) Mean COTD + CO DU (litre min" 1 ) FIG. 1. Bias plots of CO values obtained by TD and TEB, and by TD and DU during IPPV. Plots represent percentage deviation of TEB from TD, or DU from TD. Solid lines = mean difference between the two methods and two standard deviations, as the random error of variability between the two techniques. 80 -i 80-| 40 - £40-1 Q (- Q •2 SD • 2 SD I CD 0- • Mean o -2SD CO Q Q 0 c o CsO v -40 -J a -80- Mean • 2 SD -80 2 3 4 5 Mean COTD - COTEB (litre mirT1) 2 3 Mean COTD CO DU (litre min" 1 ) FIG. 2. Bias plots of CO values obtained by TD and TEB, and by TD and DU during apnoea. Plots represent percentage deviation of TEB from TD, or DU from TD. Solid lines = mean difference between the two methods and two standard deviations, as the random error of variability between the two techniques. 80-i 40Q £ 40" 3 Q c o 0- 2 SD CO LU h— Mean o- c eviat o 2SD 2SD Mean -40" -402SD -80 5 7 9 11 Mean COTD + COTEB (litre min"1) 13 -80 11 Mean C0 T D + CODU (litre min^1) 13 FIG. 3. Bias plots of CO values obtained by T D and TEB, and by T D and D U during spontaneous breathing. Plots represent precentage deviation of TEB from T D , or D U from T D . Solid lines = mean difference between the two methods and two standard deviations, as the random error of variability between the two techniques. between TD and TEB during the different conditions of measuring, but there was between TD and DU and TEB and DU during IPPV, apnoea and spontaneous ventilation (P < 0.0001). We chose the method of Bland and Altman to average the respective values of TD and TEB and of TD and DU in order to minimize the inaccuracies in the assumption that TD may not be a "gold standard" of measuring CO [18]. As shown in figure 1, agreement between COTEB and CO-rj, was best during IPPV; it was reduced slightly during apnoea and spontaneous breathing. For CODU also, the best agreement was during IPPV, with the largest bias during spontaneous breathing. During IPPV, apnoea and spontaneous breathing, TEB overestimated CO compared with TD at BRITISH JOURNAL OF ANAESTHESIA 136 smaller values of CO, whereas there was an underestimation at greater CO compared with TD during spontaneous breathing. DU seemed to underestimate CO continuously compared with TD during all conditions. DISCUSSION According to Bland and Altman, bias analysis (the mean difference between two methods) is more appropriate than correlation and regression analysis for comparing two methods that measure the same variable. The bias represents the systematic error between two measurement techniques, whereas SD of the bias represents die random error of variability between the two different techniques. Variability in COTD, CO TEB , or CODU measurements can affect the differences between the methods. (For witiiintechnique variabilities, see the work of Wong and colleagues [19].) This variability may be caused by systematic and random errors based on the different principles for determination of CO. TEB and DU measure CO of the left heart, whereas a thermodilution catheter is placed in the right heart. Left and right ventricular stroke volume are not necessarily equal [20, 21]. There is a greater constancy on the left side [22,23], as die influence of ventilation, above all during artificial ventilation, is less on the left ventricle. Right ventricular stroke volume changes significantly during inspiration and expiration as a result of decreased venous return and changing intrathoracic pressure [24, 25]. These noncontinuous flow conditions render the TD method and all dilution methods more inaccurate compared with continuous flow models [26, 27] because the Stewart-Hamilton equation is not correct for varying flow conditions. Mackenzie and co-workers [28] assessed the accuracy and reproducibility of thermodilution in an artificial circulation during continuous or pulsatile flow at several flow rates, using three different cardiac output computers: Instrumentation Laboratories IL 701, Edwards 9520A and Gold SP 1435. Pulsatile instead of continuous flow produced a pronounced and similar increase in the variability of measurement of all three devices, with the SE of the prediction being approximately double that with continuous flow. The accuracy of the measurements was reduced, with flow rates being overestimated. With pulsatile flow, injectate at ice temperature and a 10-ml injectate, the Edwards 9520A computer gave least variability. According to Stetz and colleagues [18], who reviewed the published literature and evaluated the reproducibility and accuracy of the TD method, it is difficult to determine the accuracy of this technique, especially for clinical CO determination. Both the Fick and die dye dilution methods are dilution techniques with the same theoretical errors. With a single TD measurement, a difference of about 22 % is needed, particularly in the intensive care setting, to demonstrate significant trends in CO. As a result, TD measurements in determining CO of 3.5 and 4.5 litre min"1 during therapeutic intervention may represent only reproducibility error, rather dian dierapeutic success. Even with triplicate measurements, changes of 15 % in CO may not be significant. Stetz's group used the standard error of the mean (SEM%) (SEM/average CO) for examination of the reproducibility of the TD technique. It varied from 2.0% to 5.0% for three measurements per determination and from 3.5% to 8.7% for single measurements. The SEM% values predict that the TD method may need to differ by 6.0-15.0% (3XSEM%) for triplicate measurements and by 9.5-26.1% for single measurements, in order to establish a significant change. Other problems with TD are injectate volume and temperature, injection velocity (that may itself produce accelerations of flow in the pulmonary artery system [29]), heat loss in the catheter system and in the vessel wall, and placement of the tiiermistor in the pulmonary artery. As a consequence, thermodilution cannot be regarded as the gold standard. Compared widi TD, TEB slightly overestimates cardiac output in the normal range during spontaneous ventilation and IPPV. In increased cardiac output states there is a tendency to overestimation during spontaneous ventilation, whereas during IPPV, TEB seems to underestimate cardiac output. There are two major problems of die TEB method: the correct signal processing of die critical parameters Zo, LVET and dZ/drmax [30] and die empirically derived equation. The assumption by Sramek [5] that the volume of the diorax, as electrically participating tissue, is calculated as a truncated cone instead of as Kubicek's cylinder, may improve die accuracy [4] or not [31], but does not correct overestimation of stroke volume produced by die TEB mediod in spontaneous breadiing patients. One of die reasons for this overestimation is die assumption diat the ejection velocity during systole is constant during die entire left ventricular ejection time in die equations of Kubicek and colleagues [3] and Sramek [5], whereas in reality die ejection velocity decreases to zero after reaching its peak, represented as the ohmic counterpart dZ/drmax in die impedance-stroke volume calculations. This is supported by the findings of Colocousis, Huntsman and Curreri [32], in animal experiments measuring CO by continuous-wave Doppler technique, tiiat at small stroke volumes die ejection period is abbreviated and die velocity record is symmetrical, while at large values of SV diere is a prolongation of ejection and loss of symmetry. Odier investigators [33,34] observed changes in ejection shape and duration in response to altered haemodynamic conditions or heart rate. During mechanical ventilation, tiiis velocity profile is altered, as TEB overestimates CO compared widi TD in low flow conditions and underestimates CO during sepsis [35]. Alteration of Zo as basic impedance which represents die basic fluid volume of die diorax may be excluded, as our own investigations [13] and diose of Edmunds, Godfrey and Tooley [36] demonstrated no loss of accuracy during different lung volumes. The limitations for TEB are aortic regurgitation, tachyarrhythmiaSj open heart surgery and extreme obesity. Compared with TD, DU underestimates CO con- NON-INVASIVE VS INVASIVE CARDIAC OUTPUT MEASUREMENT sistently, above all in increased cardiac output states during spontaneous ventilation, as described also for transtracheal Doppler [9]. Technical problems arose from the fact that our Doppler device averaged CO over a 12-s interval, which implies more than one ventilatory cycle. Moreover, DU measurements took place after termination of the TD and TEB measurements. Methodological problems in the Doppler method include the assumption that the aorta is circular and that there is no diameter change during systole [37]. As the cross-sectional area is a quadratic function of the radius, small errors in measurement of diameter may produce large errors in CO values [38]. Pulsations of the non-rigid aorta during systole produce 5-17 % changes in CSA from its diastolic to systolic pressure extremes [39]. In the presence of turbulence which may occur with anaemia, tachycardia, etc., the mean velocity measured by the Doppler beam is unrepresentative of the forward flow that represents CO [14]. Another phenomenon, known as aliasing, occurs with the pulsed Doppler in high flow states and makes accurate measurement of blood velocity impossible [6]. Doppler frequency shift can be measured accurately only if it is less than 50 % of the sampling frequency. Practical considerations The DU method is user-dependent and requires a long period of training [40]. It often takes a long time to obtain a satisfactory signal. The recorded Doppler signals selected for calculation of cardiac output should be of optimal quality. With bad positioning of the probe, the large beam does not cover the whole lumen of the aorta and it may be impossible to obtain an adequate signal in some patients [12]. Other sources of error are a bad ultrasound contact through the skin or the electric activity of surgical cautery. Continuous monitoring is not possible, so it is impossible to detect sudden changes in CO. Our results on the accuracy of DU compared with TD relate only to the dual beam Doppler. In contrast, the TEB method is easy to perform. The advantage of using normal ECG electrodes instead of aluminium strips makes it comfortable for the patient and it can be performed by nursing staff after a short period of instruction. Sources of errors in the TEB method include incorrect electrode placement. Small alterations in the position of the sensing inner electrodes 2 and 3 produce changes in CO of 5—10%; decreased distance leads to overestimation, whereas increasing distance produces underestimation of CO. Incorrect input of height and weight of the patient in the computerized system has similar effects. Normally, 2-3 cm errors in the adult range of the length L of the human thorax (25—32 cm), which is implicated in the Sramek equation [5] in the third power, would produce errors of 20-30% in stroke volume calculation. As L is standardized as a percentage of the patient's height H (0.17 H)3 in the Sramek equation, the actual distance between the inner electrodes is less important; the actual error in calculation is reduced to about 10%. Both lead to a constant error in COTEB. 137 REFERENCES 1. Bdhrer H, Wehlage DR, Stuth EEA. Komplikationen der Swan-Ganz-Katheterisierung. Intensivmedizin 1990; 27: 155-162. 2. Elliot CG, Zimmerman GA, Clommer TP. Complications of pulmonary artery catheterization in the care of critically ill patients. Chest 1979; 76: 647-652. 3. Kubicek WG, Karnegis JN, Patterson RP, Witsoe DA, Mattson RH. Development and evaluation of impedance cardiac output system. Aerospace Medicine 1966; 37: 1208-1212. 4. Bernstein DP. A new stroke volume equation for thoracic electrical bioimpedancc: Theory and rationale. Critical Care Medicine 1986; 14: 904-909. 5. Sramek BB. Non-invasive technique of measuring cardiac output by means of electrical impedance. In: Proceedings of the 5th International Conference on Electrical Bioimpedance, Tokyo, 1981; 39. 6. Donovan KD, Dobb GJ, Newman MA, Hockings BES, Ireland M. Comparison of pulsed Doppler and thermodilution methods for measuring cardiac output in critically ill patients. Critical Care Medicine 1987; IS: 853-857. 7. Keim HJ, Wallace JM, Thurston H, Case DB, Dreyer J. Impedance cardiography for determination of stroke index. Journal of Applied Physiology 1976; 41: 797-799. 8. Preiser JC, Daper A, Parquier JN, Contempre, Vincent JL. Transthoracic electrical bioimpedance versus thermodilution technique for cardiac output measurement during mechanical ventilation. Intensive Care Medicine 1989; 15: 221-223. 9. Siegel LC, Fitzgerald DC, Engstrom RH. Simultaneous intraoperative measurement of cardiac output by thermodilution and transtracheal doppler. Anesthesiology 1991; 74: 664-669. 10. Weber J, Heidelmeyer CF, Kubatz E, Bruckner JB. Die Bestimmung des Herzzeitvolumens unter PEEP-Beatmung mit dem nitchtinvasiven Bioimpedanzgerat "NCCOM3" im Vergleich zur Thermodilutionsm-methode. Eine Untersuchung an narkotisicrtcn Hunden. Anaesthesist 1986; 35: 744-747. 11. Castor G, Helms J, Niedermark I, Molter G, Motsch J, Altmayer P. Influence of the ventilatory cycle on cardiac output determination during mechanical ventilation. (Comparison of noninvasive electrical bioimpedance to the standard thermodilution method). American Journal of Noninvasive Cardiology 1991; 5: 57-61. 12. Looyenga DS, Liebson PR, Bone RC, Balk RA, v. Messer J. Determination of cardiac output in critically ill patients by dual beam doppler echocardiography. Journal of the American College of Cardiology 1989; 13: 340-347. 13. Cerny JC, Ketslakh M, Poulos CL, Dechert RE, Bartiett RH. Evaluation of the Velcom-100 pulse Doppler cardiac output computer. Chest 1991; 100: 143-146. 14. Mehta N, Iyvawe VI, Cummin ARC, Baylay S, Saunders ICB, Bennett ED. Validation of a Doppler technique for beat-to-beat measurement of cardiac output. Clinical Science 1985; 69: 377-382. 15. Rose JS, Nanna M, Shahbudin H, Rahimtoola SH, Elkayam U, McKay C, Chandraratna PAN. Accuracy of determination of changes in cardiac output by transcutaneous continuouswave Doppler computer. American Journal of Cardiology 1984; 54: 1099-1101. 16. Spinale FG, Smith AC, Crawford FA. Relationship of bioimpedance to thermodilution and echocardiographic measurements of cardiac function. Critical Care Medicine 1990; 18: 414-^118. 17. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1987; 1:307-310. 18. Stetz CW, Miller RG, Kelly GE, Raffin TA. Reliability of the thermodilution method in the determination of cardiac output in clinical practise. American Review of Respiratory Diseases 1982; 126: 1001-1004. 19. Wong DH, Tremper KK, Stemmer EA, O'Connor D, Wilkur S, Zaccari J, Reeves C, Weidoff P, Trujillo RJ. Noninvasive cardiac output: Simultaneous comparison of two different methods with thermodilution. Anesthesiology 1990; 72: 784-792. 20. Franklin DL, Van Citters RL, Rushmer RF. Balance between BRITISH JOURNAL OF ANAESTHESIA 138 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. right and left ventricular output. Circulation Research 1962; 10: 17-26. Guz A, Hoffman LIE, Charlier A, Wierich WR, Zanger L, Benke J. Simultaneous observations of right and left ventricular stroke volumes in conscious dogs. Federation Proceedings 1961; 20: 133. Hoffman JIE, Guz A, Charlier AA, Wilcken DEL. Stroke volume in conscious dogs: Effect of respiration, posture and vascular occlusion. Journal of Applied Physiology 1965; 20: 865-877. Santamorc WP, Heckman JL, Bove AA. Cardiovascular changes from expiration to inspiration during IPPV. American Journal of Physiology 1983; 245: H307-312. Armcngol J, Man GCW, Balsys AJ, Wells AL. Effects of the respiratory cycle on cardiac output measurements. Reproducibility of data enhanced by timing the thermodilution injections in dogs. Critical Care Medicine 1981; 9: 852-854. Morgan BC, Dillard DH, Guntheroth WG. Effect of cardiac and respiratory cycle on pulmonary vein flow, pressure and diameter. Journal of Applied Physiology 1966; 21: 1276-1280. Cropp GJA, Burton AC. Theoretical considerations and model experiments on the validity of indicator dilution methods for measurements of variable flow. Circulation Research 1965; 18: 26-^8. Scheuer-Leeser M, Morguet A, Rene H, Irmich W. Some aspects to the pulsation error in blood-flow calculations by indicator-dilution techniques. Medical and Biological Engineering and Computing 1977; IS: 118-123. Mackenzie JD, Haites NE, Rawles JM. Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output. British Heart Journal 1986; 55: 14-24. Fronefc A, Ganz V. Measurement of flow in single blood vessels in changing cardiac output by local thermodilution. Circulation Research 1960; 8: 175-179. Nagel JH, Shyn SP, Reddy SP, Hurwitz BE, McCabe PM, Schneiderman N. New signal processing techniques for 31. 32. 33. 34. 35. 36. 37. 38. 39. 40. improved precision of noninvasive impedance cardiography. Annals of Biomedical Engineering 1989; 17: 517-534. De Mey C, Enterling D. Noninvasive assessment of cardiac performance by impedance cardiography: Disagreement between two equations to estimate stroke volume. Aviation, Space and Environmental Medicine 1988; 59: 57-62. Colocousis JS, Huntsman LL, Curreri PW. Estimation of stroke volume changes by Ultrasonic Doppler. Circulation 1977; 56: 914-917. Noble MIM, Trenchard D, Guz A. Changing heart rate on cardio-vascular function in the conscious dog. Circulation Research 1966; 19: 206-213. Rushmer RF. Recent advances in cardiovascular physiology. Anesthesia and Analgesia 1966; 45: 383-389. Castor G, Motsch J, Altmayer P, Helms J, Molter G, Niedcrmark I. Clinical evaluation of cardiac output measurement by thoracic electrical bioimpedance. European Journal of Pharmacology 1989; 36: 38. Edmunds AT, Godfrey S, Tooley M. Cardiac output measured by transthoracic impedance cardiography at rest, during exercise and at various lung volumes. Clinical Science 1983; 63: 107-113. Bernstein DP. Noninvasive cardiac output, Doppler flowmctry, and gold-plated assumptions. Critical Care Medicine 1987; 15: 886-887. Chandrarama A, Nanna M, McKay C, Nimalasuriya A, Swinney R, Elkayam U, Rahimtoola SH. Determination of cardiac output by transcutaneous continuous-wave Ultrasonic Doppler computer. American Journal of Cardiology 1984; 53: 234-237. Greenfield JC, Patel DJ. Relation between pressure and diameter in the ascending aorta of man Circulation Research 1962; 10: 778-781. Loeppky JA, Hoekenga DE, Greene ER, Luft UC. Comparison of noninvasive pulsed Doppler and Fick measurements of stroke volume in cardiac patients. American Heart Journal 1984; 107: 339-346.