A-dec 371-200 Dental Light on an A
Transcription
A-dec 371-200 Dental Light on an A
A-dec 371-200 Dental Light on an A-dec 200 Dental Chair INSTALLATION GUIDE Install the Dental Light 4 1. Insert and fully seat the bearing in the support center post. 2. Place the trim cover on top of the bearing. 3 2 4. Insert the dental light post. 5. Route the wiring out of the post into the support center. 1 5 86.0314.00 Rev D 3. Route wiring down into the support center post. A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Connect the Dental Light Task 1. Prepare the dental light cable in the support center. If the dental light cable has a green wire with a WAGO connector already attached, go to Task 2 on the next page. NOTE back the sheath 80 mm (3") then strip 1 Strip back the green wire 7 mm (1/4"). a WAGO connector to the end of the 2 Attach green wire. To Floor Box 2 86.0314.00 Rev D A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Task 2. Connect the wires from the light to the WAGO connectors with the same color wires. To Dental Light To Floor Box 86.0314.00 Rev D 3 A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Connect the Floor Box Utilities Install the Dental Light Relay Circuit Board Recommended Tool Phillips head screwdriver Secure the dental light relay circuit board with the screw from the top of the power supply. 4 86.0314.00 Rev D A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Connect the Dental Light Jumper Cable and Data Line Get the dental light jumper cable and data line from the kit and make the following connections. 2 Connect the white plug. 3 Connect the data line. 1 Connect the red plug. 86.0314.00 Rev D 5 A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Connect the Dental Light Cable Follow the procedure according to the type of dental light cable used. The circuit board is static sensitive. ESD precautions are required. The circuit board should be installed by an electrician or qualified service personnel. CAUTION Light Cables with a Green Wire Attached If the end of the light cable includes a green wire connected to the red connector, complete the following procedure: Connector with Green Wire Attached 1 Connect the wires to their positions on the relay board. BLK VIO 2 Connect the light cable (from kit). WHT BLK/WHT 6 86.0314.00 Rev D A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Light Cables without a Green Wire Attached If the end of the light cable does not have a connector or does not have a green wire connected to the red connector, complete the following procedures: Cable with No Connector Cable without a Green Wire Attached to the Connector Task 1. 1 Snip off the wires at the end of the sheath. 2 Strip back the sheath 80 mm (3"). 3 Strip back the wires 7 mm (1/4"). 86.0314.00 Rev D 7 A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Task 2. the dental light wires to the adaptor cable with WAGO connectors. 1 Attach Match the wire to the WAGO connector with the same color wire. BLK BLK VIO VIO BLUE GRN 2 Connect the adaptor cables. WHT BLK/WHT the wires from the second adaptor 3 Connect cable to their positions on the relay board. 8 86.0314.00 Rev D A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Install the Dental Light Flexarm Rotation Stop Pins the 1 Lift trim ring. the 2 Insert stop pin. Actual Size of Pin 86.0314.00 Rev D 9 A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Adjust the Dental Light Flexarm Counterbalance (if needed) Recommended Tools • Phillips head screwdriver • 1/2" combination wrench NOTE An optional travel stop limit kit (p/n 90.1044.00) can be installed to limit the upward and downward motion of the flexarm. 1 Remove the two screws and end cap. 2 Remove the screw and end cap. 3 Remove the cover. the light drifts up, turn the nut 4 Ifcounterclockwise. If the light drifts down, turn the nut clockwise. the cover onto the 5 Set flexarm to test the drift. 10 86.0314.00 Rev D A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Adjust the Dental Light Rotation Tension (if needed) Adjust the Dental Light Horizontal Rotation Recommended Tool 5/32" hex key To increase tension, turn the screw clockwise. To decrease tension, turn the screw counterclockwise. Adjust the Dental Light Vertical Rotation Recommended Tool 3/16" hex key To increase tension, turn the screw clockwise. To decrease tension, turn the screw counterclockwise. 86.0314.00 Rev D 11 A-dec 371-200 Dental Light on an A-dec 200 Dental Chair Installation Guide Adjust the Dental Light Focus (if needed) The A-dec 371-200 does not have a focus adjustment. Touchpad Settings For dental light touchpad settings, see the “Touchpad Settings” section of the A-dec 200 Installation Guide (p/n 86.0004.00). Regulatory Information Regulatory information is provided with A-dec equipment as mandated by agency requirements. This information is delivered in the equipment’s Instructions for Use or the separate Regulatory Information and Specifications document. If you need this information, please go to the Document Library at www.a-dec.com. A-dec Headquarters 2601 Crestview Drive Newberg, OR 97132 USA Tel: 1.800.547.1883 Within USA/Canada Tel: 1.503.538.7478 Outside USA/Canada Fax: 1.503.538.0276 www.a-dec.com / www.a-dec.biz A-dec Inc. makes no warranty of any kind with regard to the content in this document including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. ÍvÈ.Ç#.È.00wÎ 86.0314.00 Rev D Copyright 2012 A-dec Inc. All rights reserved. IGgrph4