Manufacturer Specific Considerations for CT Accreditation
Transcription
Manufacturer Specific Considerations for CT Accreditation
Manufacturer Specific Considerations for CT Accreditation Melissa C. Martin, M.S., FAAPM, FACR AAPM Spring Clinical Meeting March 20, 2012 Dallas, TX Phantom Submission Slide provided by Doug Pfeiffer Phantom Submission Slide provided by Doug Pfeiffer Phantom Submission Slide provided by Doug Pfeiffer Phantom Site Scanning Form Page 1 Slide provided by Doug Pfeiffer Phantom Site Scanning Data Form - Page 2 Slide provided by Doug Pfeiffer Phantom Images Slide provided by Doug Pfeiffer Phantom Images Slide provided by Doug Pfeiffer Phantom Images Slide provided by Doug Pfeiffer Phantom Scoring Slide provided by Doug Pfeiffer CT # Accuracy Slide provided by Doug Pfeiffer Contrast-Noise Ratio Slide provided by Doug Pfeiffer Contrast-Noise Ratio Slide provided by Doug Pfeiffer CT # Uniformity Slide provided by Doug Pfeiffer Dosimetry Table 1 - Common Mistakes For mA row – entering mAs, effective mAs or mAs/slice Help site understand difference between these And that they are not all equivalent mA ≠ mAs mAs ≠ eff. mAs mAs ≠ mAs/slice Dosimetry Siemens – eff. mAs (effective mAs) mA * rot _ time Eff . _ mAs = Pitch mA = Eff . _ mAs * Pitch rot _ time Philips – mAs/Slice (similar definition to eff. (mAs / Slice) * Pitch mAs) mA * rot _ time mA = mAs / slice = Pitch rot _ time Toshiba and GE use mA, time , Pitch as separate values Dosimetry Common Mistakes include: Reporting mAs or eff. mAs or mAs/slice in Table 1 Then using mAs or eff. mAs when performing CTDI measurements Example: 200 eff. mAs, pitch .9, rot. time = 0.5 sec In this case, mA = 360 Should perform CTDI measurement with 180 mAs Spreadsheet will use pitch 0.9 and correct for values of effective mAs Dosimetry Common Mistakes include: If site does this incorrectly, spreadsheet will have incorrect values If they perform acquisition with 200 mAs And then use N,T and I such that a pitch of 0.9 results, then CTDIvol reported will be too high If Pitch < 1, CTDIvol reported will be too high If Pitch > 1, CTDIvol reported will be too low Dosimetry Common Mistakes - Reporting collimation incorrectly Admittedly this can be confusing for some scanners Example: Siemens Sensation 64 Scanner user interface says 64 x 0.6 mm Scanner uses z-flying focal spot, which double samples on zaxis of anode to obtain 2X images. Actual beam width is N=32, T = 0.6 mm For Pitch 1.0, table travel will be 19.2 mm/rotation Site sometimes list N=64, T=0.6mm, I= 38.4mm (-> Pitch 1) In spreadsheet, this yields CTDIvol that is half what it should be Dosimetry Common Mistakes - Reporting collimation incorrectly Consult ACR CT Accreditation website for FAQs and clarifications http://www.acr.org/accreditation/computed/ct_faq.asp x Dosimetry Slide provided by Doug Pfeiffer Dosimetry Slide provided by Doug Pfeiffer Dosimetry Slide provided by Doug Pfeiffer Dosimetry Slide provided by Doug Pfeiffer Dose Levels Slide provided by Doug Pfeiffer Dosimetry Slide provided by Doug Pfeiffer Dosimetry Slide provided by Doug Pfeiffer Dosimetry Slide provided by Doug Pfeiffer Pass/Fail Criteria Slide provided by Doug Pfeiffer Phantom Deficiencies Slide provided by Doug Pfeiffer Phantom Deficiencies Slide provided by Doug Pfeiffer Phantom Deficiencies Slide provided by Doug Pfeiffer Phantom Deficiencies Slide provided by Doug Pfeiffer Slide provided by Doug Pfeiffer Toshiba CT Scanners Aquilion Scanners (32, 64 or 320) Slide provided by Erin Angel, Toshiba Medical Systems Toshiba Aquilion 32 or 64 slice CT Scanners The maximum number of axial images able to be acquired simultaneously in a single rotation (Nmax) is 4. The minimum rotation time for a full 360 rotation can be found in the product data sheet or via loading any helical protocol and choosing the lowest available 360 rotation time (note: grayed-out times are not available). Slide provided by Erin Angel, Toshiba Medical Systems Toshiba Aquilion 32 or 64 slice CT Scanners Slice thicknesses available in helical mode include: 0.5mm, 1.0mm, 2.0mm, 3.0mm, 5.0mm, 7.0mm, and 10.0mm. Slice thicknesses available in axial mode include: 0.5mm, 1.0mm, 2.0mm, 3.0mm, 4.0mm, 6.0mm, 8.0mm, 12mm (16mm, 24mm, and 32mm are also available for non-clinical applications). Available beam energies* are 80 kVp, 100 kVp, 120 kVp, and 135 kVp Slide provided by Erin Angel, Toshiba Medical Systems Toshiba Aquilion 16 slice CT Scanners The maximum number of axial images able to be acquired simultaneously in a single rotation (Nmax) is 4. The minimum rotation time for a full 360 rotation can be found in the product data sheet or via loading any helical protocol and choosing the lowest available 360 rotation time (note: grayed-out times are not available). Slide provided by Erin Angel, Toshiba Medical Systems Toshiba Aquilion 16 slice CT Scanners Slice thicknesses available in helical mode include: 0.5mm, 1.0mm, 2.0mm, 3.0mm, 5.0mm, 7.0mm, and 10.0mm. Slice thicknesses available in axial mode include: 0.5mm, 1.0mm, 2.0mm, 3.0mm, 4.0mm, 6.0mm, 8.0mm, 12mm (16mm, 24mm, and 32mm are also available for non-clinical applications). Available beam energies* are 80 kVp, 100 kVp, 120 kVp, and 135 kVp Slide provided by Erin Angel, Toshiba Medical Systems Toshiba Scanners - Location of Technical Parameters Toshiba Scanners - Scan Detail Tab Slide provided by Erin Angel, Toshiba Medical Systems Slide provided by Erin Angel, Toshiba Medical Systems Scan Details for Axial Scanning Slide provided by Erin Angel, Toshiba Medical Systems Recon Details Tab IEC Pitch Value and Table Speed Always use the IEC definition of pitch when reporting to ACR IEC pitch can be calculated either by: Dividing the helical pitch (HP) by the number of detector channels being used (N) Clicking on the HP button on the Scan Details tab to find the IEC pitch value (ex: PF 0.828). The IEC pitch value is called “Pitch Factor” on your Aquilion scanner, also denoted PF. Slide provided by Erin Angel, Toshiba Medical Systems IEC Pitch Value and Table Speed Slide provided by Erin Angel, Toshiba Medical Systems IEC Pitch Values and Table Speed Detailed Pitch Row Helical Pitch (HP) IEC Pitch Factor 16 11 0.688 32 21 0.656 64 41 0.641 Slide provided by Erin Angel, Toshiba Medical Systems IEC Pitch Values and Table Speed Standard Pitch Row Helical Pitch (HP) IEC Pitch Factor 16 15 0.9375 32 27 0.84375 64 53 0.828 Fast PitchRowHelical Pitch (HP)IEC Pitch Factor (PF)16231.437532451.4062564951.484 Slide provided by Erin Angel, Toshiba Medical Systems IEC Pitch Values and Table Speed Fast Pitch Row Helical Pitch (HP) IEC Pitch Factor 16 23 1.4375 32 45 1.40625 64 95 1.484 ACR Tips for Toshiba CT’s Have the scanner calibrated for air and water, at all relevant kVps, prior to testing Make sure you scan the ACR phantom with the small (S) Scan FOV for the image quality tests. Scanning a small size phantom with a large size FOV can lead to CT number and uniformity errors. For best results be sure the phantom is properly centered, utilizing the lasers and a Slide provided by Erin Angel, Toshiba Medical Systems level. ACR Tips for Toshiba CT’s Have your “Average Clinical Parameters” chart filled out in advance and be certain to scan the phantom with the corresponding parameters. Before proceeding with phantom image quality testing, ensure your protocols are optimized for dose while obtaining adequate image quality and that radiation doses do not exceed ACR guidelines. For proper landmarking, note where the center of the slice will be when you prescribe a scan in axial or helical mode Slide provided by Erin Angel, Toshiba Medical Systems Finding the Center Slice Location Axial Mode: If a scan is prescribed, for example, from 24mm to 26mm in 4 x 0.5mm axial mode, the center slice location will be as follows: Slide provided by Erin Angel, Toshiba Medical Systems Helical Mode: If a scan is prescribed, for example, from 24mm to 26mm in 4 x 0.5mm helical mode, the center slice location will be as follows (note: this is different from axial mode): Slide provided by Erin Angel, Toshiba Medical Systems Objectives CONCLUSIONS For all scanners, make sure that the facility has had service perform preventive maintenance recently. Review the protocols used with the technologist prior to performing phantom scans and dosimetry. Contact the vendor help desk for further clarification on a particular scanner. If at all possible, burn the CD’s for ACR Submission prior to leaving the facility. Do not depend on staff at the facility to correctly burn the CD’s. Contact Information: Melissa C. Martin, M.S., FACR, FAAPM Therapy Physics Inc. 879 West 190th St., Ste 400 Gardena, CA 90248 e-mail: Melissa@TherapyPhysics.com website: www.TherapyPhysics.com Phone: 310-612-8127