How to make a list sweep measurement
Transcription
How to make a list sweep measurement
Agilent B2901/02/11/12A Precision Source/Measure Unit How to make a list sweep measurement This material shows how to perform a list sweep measurement through an example of the Photovoltaic Cell IV measurement. Figure 1 illustrates the connection and condition supposed in the example of measuring Photovoltaic Cell using the B2901/02/11/12A. B2901/02/11/12A A V High Force DUT - Low Force Figure 1. Connection and condition supposed in the example Figure 2 shows the timing chart for the list sweep measurement with the front panel operation. In this case, the specified source value is sourced immediately after turning on . Then, when you press , the instrument will make a list sweep measurement. If it is necessary, you can specify any measurement trigger delay time which is the wait time after sourcing each source value and before making a measurement. The measurement time consists of Measurement Speed and some overhead time. Measurement Speed is the parameter specified by the user. Overhead time includes the time to change the measurement range, etc. TP : Trigger Period TMD : Measurement Trigger TM : Measurement Time Delay Time V or I TMD TM :Measurement V2 Value TMD TM V4 Value TMD V3 Value Source Value TMD TM TM V1 Value TP Output Off State Press Output ON Trigger TP Internal Trigger TP TP Internal Trigger Internal Trigger Figure 2. Timing chart for the list sweep measurement Time Agilent B2901/02/11/12A Precision Source/Measure Unit Preparing the list data in advance When you source some complicated waveform with the List Sweep Mode of the B2901/02/11/12A, it is convenient to create the data list on a PC using the software of your choice and then transfer the data list file to the B2901/02/11/12A via USB flash memory. TXT, PRN and CSV file format are available for data list file. Step 1. Develop CSV file which includes the following source list data. (From 0 to 0.7 with 0.1 step, and from 0.71 to 1.30 with 0.01 step, and from 1.4 to 1.7 with 0.1 step) Step 2. Save CSV file as “listdata.csv” to the root folder of your USB flash memory device. Step 3. Attach the USB flash memory device which “listdata.csv” file is saved to the USB port on the front panel of the B2901/02/11/12A. Agilent B2900A Quick I/V Measurement Software is also available to edit the data list for List Sweep Mode. Please refer to the Agilent Web Site below for the detailed information about Agilent B2900A Quick I/V Measurement Software. http://www.agilent.com/find/b2900a Agilent B2901/02/11/12A Precision Source/Measure Unit Performing a list sweep measurement Step 1. Press Step 2. Press source. repeatedly until Single View for Channel 1 is shown in the display. to edit the source function, and then select (1) Press Mode Step 3. Press Press (2) Press VOLTS (V) to edit the source value, and then enter 0 V to set the source value to 0 V. (1) Press Source Step 4. to set the source function to the voltage (2) Enter Source Value to edit the limit value, and then enter 100 mA to set the limit value to 100 mA. (1) Press Limit (2) Enter Limit Value Step 5. Press to configure the measurement parameter, and then select parameter to the current. (1) Press Measure (2) Press AMPS (I) to set the measurement Agilent B2901/02/11/12A Precision Source/Measure Unit Step 6. Press Sub-Panel. to change the keys shown in Assist keys, and then press (1) Press Show Sweep Step 7. Press , then press to show Sweep (2) Sweep Sub-Panel appears to change the keys shown in Assist keys. (1) Press (2) Press More... Step 8. Press to turn on List Sweep Mode. After turning on List Sweep Mode, you can see Source Shape which shows the list sweep mode. (1) Press LIST Step 9. Rotate and press (1) Rotate and press Start value Source Shape to edit Start value and press to edit to open Load List Sweep Data dialogue. (2) Press Load Agilent B2901/02/11/12A Precision Source/Measure Unit Step 10. Rotate and press to select “listdata.csv” file, and then press (1) Rotate and press listdata.csv file to select to load “listdata.csv” file. (2) Press OK to load listdata.csv file Now you have configured List Sweep Source Setup and can see the parameters bellow. Step 11. Press repeatedly until Graph View is shown in the display. Step 12. Press to source the voltage, and then press (The status information will show during the measurement.) to perform a measurement. Step 13. Press to adjust the scale of the graph after finishing the measurement. Now you can see the measurement result on the GUI of the B2901/02/11/12A as bellow. Agilent B2901/02/11/12A Precision Source/Measure Unit Step 14. Rotate and press to edit the Y-axis maximum value, and then enter 0 A to adjust the graph scale. (1) Rotate and press Y-axis maximum value Step 15. Rotate and press to select (2) Enter the scale such as 0 A to edit the Y-axis minimum value, and then enter -80 mA to adjust the graph scale. (1) Rotate and press Y-axis minimum value to select (2) Enter the scale such as -80 mA Now you can see the measurement result scaled on the GUI of the B2901/02/11/12A as bellow. Agilent B2901/02/11/12A Precision Source/Measure Unit Configuring the measurement speed In the default setting, the instrument selects the appropriate measurement speed and range automatically to get the fine accuracy. However, you can also specify these parameters on the GUI of the B2901/02/11/12A to meet a variety of the requirement to the measurement conditions. For example, let’s try to change the measurement speed to NORMAL to make a measurement more carefully. If you select NORMAL, the aperture time is set to 1 PLC. Here, PLC stands for power line cycle and the specified number of power line cycles is used per a measurement. Step 1. Press Step 2. Press repeatedly until Single View for Channel 1 is shown in the display. to edit the measurement speed, and then select NORMAL. (If you can’t see in Assist keys, press (1) Press Speed to set the measurement speed to to change the keys shown in Assist keys.) (2) Press NORMAL Configuring the measurement range operation The parameters which configure the measurement range operation can be displayed in Range Sub-panel in Single View, although Sweep Sub-Panel is shown at this moment. In the default setting, the B2901/02/11/12A performs the current measurement using 1 uA current minimum measurement range with AUTO range operation. With AUTO range operation, the B2901/02/11/12A selects the proper range for the measurement with specified minimum measurement range so that you don’t need to take care about it. To know how to change the measurement range setting, try to configure to use 10 nA current minimum measurement range with AUTO range operation. Step 1. Press to show Range Sub-Panel. (If you can’t see change the keys shown in Assist keys.) (1) Press More... (if necessary) in Assist keys, press (2) Press Hide Sweep to Agilent B2901/02/11/12A Precision Source/Measure Unit Step 2. Rotate and press to edit the current minimum measurement range, and then select to set it to 10 nA. (1) Rotate and press to edit Current Minimum Measurement Range (2) Press 10nA If you’d like to fix the measurement range, you can select FIXED range operation as below. Step 3. Rotate and press to edit the current measurement range operation. Then Select to set the current measurement range operation to FIXED. (1) Rotate and press to edit Current Measurement Range Operation (2) Press FIXED Configuring the measurement trigger delay time The trigger parameters including the measurement trigger delay time and the trigger period can be displayed in Trigger Sub-panel in Single View, although Range Sub-Panel is shown at this moment. In the default setting, the trigger type is set to the automatic trigger type (AUTO) so that you don’t need to specify these trigger parameters. If you’d like to specify the measurement trigger delay time, take the following steps. Step 1. Press to show Trigger Sub-Panel. (If you can’t see change the keys shown in Assist keys.) (1) Press More... (if necessary) in Assist keys, press (2) Press Show Trigger to Agilent B2901/02/11/12A Precision Source/Measure Unit Step 2. Press to edit the trigger type, and then select (1) Press Step 3. Rotate to set the trigger type to MANUAL. (2) Press MANUAL to select Channel 1 Trigger Parameters and set them up as below. (Source Trigger Count: 72, Measurement Trigger Count: 72, Measurement Trigger Delay Time: 100ms) Note) Source and Measurement Trigger Count should be the same number as Sweep Points. Now you’ve configured 100 ms measurement trigger delay time. Configuring the trigger period If you need to control the source and measurement period strictly, the trigger period is useful to make it. In order to specify the trigger period, take the following steps. Step 1. Rotate and press (1) Rotate and press Trigger Type to edit the trigger type, and then select to edit (2) Press TIMER to set the trigger type to TIMER. Agilent B2901/02/11/12A Precision Source/Measure Unit Step 2. Rotate to select Channel 1 Trigger Parameters and set them up as below. (Source Trigger Period 400 ms, Measurement Trigger Period: 400 ms) Note) Source and Measurement Period should be the same as each other. Now you’ve configured 400 ms trigger period. However, please note that FIXED current measurement range operation will be used to control the trigger period strictly. The measurement range is selected by Limit value. In this example, 100 mA measurement range will be used. If using AUTO measurement range operation is prior to controlling the trigger period strictly, you may specify MANUAL trigger type with AUTO source and measurement trigger source by the following steps. Step 3. Press to edit the trigger type, and then select (1) Rotate and press Trigger Type Step 4. Rotate to edit to set the trigger type to MANUAL. (2) Press MANUAL to select Channel 1 Trigger Parameters and set them up as below. (Source Trigger Source: AUTO, Measurement Trigger Source: AUTO) Agilent B2901/02/11/12A Precision Source/Measure Unit Viewing the list of measurement results The measurement results including the measurement time stamp can be referred by the following steps. Step 1. Step 2. If you aren’t on the top of the Function menu, press repeatedly to return to the top level. On the top level of the Function menu Any tree structure can’t be seen In the middle level of the Function menu Some tree structure can be seen If you’d like to see the list of measurement results, press , then press to open Measure Result dialogue. (1) Press Result (2) Press Measure Step 3. Rotate and press (1) Rotate field Step 4. Press type. to select Data Type field. to select Data Type (2) Press to select Data Type to change the keys shown in Assist keys, and then press (1) Press More … (2) Press TIME (t) to select Time as the data Agilent B2901/02/11/12A Precision Source/Measure Unit Step 5. Rotate and press (1) Rotate and press Data field to select Data field. Then rotate to select to scroll the data list. (2) Rotate or press Assist key to scroll the list