Date Project Review Cover Sheet CL 600C /D, 6EX
Transcription
Date Project Review Cover Sheet CL 600C /D, 6EX
Date Planning Department TO: Project Review Cover Sheet CL 600C /D, 6EX FOR: FAX: Review of Drawings Submitted by: 0 0 0 0 FAX: Activity Data: Record Number: Drawings received on: Review completed on: Document Number: Revision: 0 0 PROJECT NUMBER: Contacts: District Manager: Installation Coordinator: Reviewed by: Dear , Attachments: Attachment One - Notes 0 Thank you for the opportunity to review these plans. The purpose of this review is to assist in the development and completion of this project. This review covers only the architectural requirements of our equipment. The checklist outlines the correct and incorrect elements of the plans and those that can not be determined. The notes referenced within the checklist may be found on Attachment One. All boxes marked "N" should be addressed by the architect and the revised documents may be resubmitted for review. The Varian Documents referenced are found in the Installation Data Package. Our review does not check for compliance with building codes and other regulatory agency requirements. Verification of shielding adequacy must be provided by the Physicist of Record for the project. We look forward to answering questions you might have in connection with this project. Sincerely, 0 Project Manager CL 600C /D, 6EX Review review01.xls 12-Jul-05 Varian-Supplied Equipment: Corresponding Checklist: Submitted Documents description: Sheet # Date Sheet # Date Previous Planning Department Activity Description Document # Sheet # Date Sheet # Addressed to: Date review01.xls CL 600C /D, 6EX Review Date 12-Jul-05 Varian Review Checklist - Appendix One item comments Clinac 600 Review Notes 1 INFORMATION COULD NOT BE FOUND ON CONSTRUCTION DOCUMENTS INFORMATION WAS "SHOWN" BUT NOT DIMENSIONED AND/OR SPECIFIED FOR LOCATION OR CONSTRUC 2 3 VERIFY ALL SHIELDING AND ROOM PENETRATIONS WITH THE PHYSICIST OF RECORD. 4 VERIFY RIGGING PATH FROM UNLOADING POINT TO TREATMENT ROOMS. 5 NEED A NOTE 6 LOCATE ALL INSLAB PULLBOXES AND CONDUIT STUB-UPS 7 SHOW CONTROL EQUIPMENT 8 SHOW LOCATION DETAIL 9 SHOW INTERCONNECTION WIRING DIAGRAM DETAILS 10 11 12 review01.xls CL 600C /D, 6EX Review 12-Jul-05 Varian Planning Department Project Drawing Review Clinac 600C Vault Configuration Couch rotation clearance Recommended Acceptable Unacceptable Cannot be determined Gantry ceiling clearance Recommended Acceptable Unacceptable Cannot be determined Refer to Installation Data Package Section Seven FOC = Face of Concrete 9'-6" clear or 10'-0" FOC 7/12/05 Customer Varian Notes: Drawing Doc No. Attach. 1 Location [Page No.] less than 5'-6" clear 861674 [2.26] 861662 [2.23] at least 9'-6" clear 9'-1" clear less than 9'-1" clear Isocenter to back wall distance Recommended 9'-0" to face of concrete Acceptable 8'-0" clear Unacceptable less than 8'-0" clear Cannot be determined 861671 [2.25] 861662 [2.23] Service clearances at stand Recommended Acceptable Unacceptable Cannot be determined 861671 [2.25] 861662 [2.23] Entrance Door & Maze Recommended Acceptable Unacceptable Cannot be determined Console Area Recommended Acceptable Unacceptable Cannot be determined Page 4 861671 [2.25] 861662 [2.23] 8'-9" clear or 10'-0" FOC couch has obstruction OR load/unload not optimum less than 5'-10" clear cklstlow.xls all service doors can open fully some/all doors can be opened and removed some/all doors cannot be accessed 861662 [2.23] 861672 [2.24] 48" by 84" net opening - 72" net width passage 46" by 84" passage - accepted by Varian inadequate for rigging 1100794 [2.27] adequate counter area less than recommended inadequate working area Isocenter Isocenter located from inside two walls Isocenter height identified 4 mv, 6 mv - 4'-2 1/2" Y N 861662 [2.23] 861674 [2.26] CL 600C /D, 6EX Review Page 4 of 9 Varian Planning Department Project Drawing Review Customer Varian Drawing Doc No. Location [Page No.] Page 5 Treatment Room Accessories and Supplies casework Space for Varian-supplied Wedges Space for Blocks Adequate linen/misc. storage space Sink and counter Notes Y N 1100780 [2.30] 1100781 [2.31] Control Console Control area casework - C series - (600C) Counter Height standing ht - 34/36" sitting ht - 29/30" 38" deep counter with 20" deep shelf at least 12" above counter Counter length for: 3' Clinac 5' Varis 3' MLC 5' PV Ded. four-plex: 1 Clinac 1 Varis 1 MLC 1 PV Keyboard drawer: 1 Clinac 2 MLC 1 PV Computer platform: 1 MLC Clearance for Clinac Electronics Cabinet Clinac Electronics Cabinet located to the left of the workstation Grommets for cables File or chart storage/Paper & supplies/Personal property Notes Y N 1100794 [2.27] 1100779 [2.28] Shielding - Review with Physicist Physicist report included with drawings Notes Experimental/Physics Access Routing restricts secondary radiation Location verified with Physicist Primary beam shielding Proper alignment with isocenter 12" wider than beam at each side Y N 1102316 [2.36.0] counterweight no requirements no requirements Secondary shielding location and placement of penetrations restrict secondary radiation Door shielding corresponds to Physicist report Wall and ceiling shielding correspond to Physicist report Notes Y N 1102303 1100728 Notes Y N 1102303 1100728 [2.23.0] Pit Configuration General isocenter referenced to sides of pit Note: "Maintain level floor in 6'-0" radius of isocenter" Rebar shown to avoid anchors (Seismic) Notes Pit construction Pit size: 10'-2"L x 5'-6"W x 12"D Slab thickness: 8" for anchors Seismic details (If required) Notes Y N 861667 [2.34] Y N 861667 [2.34] Base frame installation Installed by Customer Installed by Varian Fill grout specified Grout cure time indicated as seven days Note "Base frame installed into pit by Contractor". Note "Grouting of Base Frame by Contractor". Proper clearances from unloading area to treatment room Rigging clearances require 2-piece breakdown of Baseframe CL 600C /D, 6EX Review Notes Y N 861667 [2.34] 1101809 [2.02] Page 5 of 9 Varian Planning Department Project Drawing Review Mechanical H.V.A.C. Ventilation at room for gantry and stand Ventilation at console Notes Customer Varian Drawing Doc No. Location [Page No.] Page 6 Y N 862530 [2.39] Cooling water Closed Loop system One-Pass Notes Shut-off Valve in secondary loop not required Domestic water back-up option primary source Waste return back-up option primary source Optional flow gauges on supply/return lines located near Clinac 3/4" pipe to Clinac at pit Cap at Clinac connection Utility piping location shown and dimensioned Y N 862530 [2.39] 861667 [2.34] Final Connections Plumbing By Customer "Cooling water connections:" Responsibility note By Varian Notes Y N 862530 [2.39] Electrical Power supply 208VAC three phase- 70 amp, 60Hz(US) 15 KVA supply capacity ground shown at supply transformer or power conditioner Notes Power panel Circuit breaker - GE # 600C70A240V(US) UVR - GE 24VDC # TEDUXVERS Included if above is used Start button Included if above is used Bridge rectifier (24VDC, 0.5 amp) Included if above is used 115/24v (US) control transformer Included if above is used 1 amp fuse(US) Included if above is used Main Circuit Breaker located within 10 feet of console note indicating 8 lin. ft of conductor to be coiled in pull box Notes Y N 861670 [2.38] Y N C-series 861670 [2.38] Cable Access General Runs between equipment pull boxes do not exceed 75 feet conduit runs between pull boxes have 270 degrees of bend max conduit bend radii are at least 6 diameters CL 600C /D, 6EX Review Notes Y N 1100732 [2.36] 1100731 [2.35] Page 6 of 9 Varian Planning Department Pull boxes Access Method Base Frame Console Conduit Base to Console Project Drawing Review Notes W 12" 18" bottom L D W 24" 10" 12" 12" 6" 18" side L 24" 12" Customer Varian Drawing Doc No. Location [Page No.] Page 7 Y N 1100731 [2.35] 1100732 [2.36] 1100788 D 20" 6" Notes Y N three each 4" Dai 1100732 [2.36] Facility Connections Relay Junction Box (12"x12"x6") Two 2" conduit from Relay Junction Box to Base Frame Pull Box 2" conduit & conductor from power panel to base frame pull box Notes In Room Monitor Notes Y N 861670 [2.38] Y N In-Room Monitor conduit (2" dia) and outlet boxes at each end Bracket/Mounting detail for In Room Monitor 1100732 1100790 [2.33] Operating and Safety Devices Conduit routing for required devices Warning light to base frame junction box Door interlocks to base frame junction box Emergency-off switches to base frame junction box Room lights to relay junction box Laser positioning lights to relay junction box Notes Y N 861670 [2.38] Control relay Potter Brumfield - PRD11DG0 - in relay junction box Room lights controlled by relay Lasers controlled by relay Device locations Warning light over entrance door Warning lights visible from all points in room 12 volt and 120 volt Door interlocks Door activators, if required Emergency-off type: Manual reset - normally closed Emergency-off located outside of primary beam path Convenience outlet for each laser positioning light Radiation Monitor and Detector (Customer option) Adequate quantity and location of Room lights Setup lights located on both sides of couch Setup lights dimmable and independent of room lights CL 600C /D, 6EX Review 1100732 [2.36] Notes Y N 1100736 [2.22] 1100733 [2.37] 861670 [2.38] Page 7 of 9 Varian Planning Department Project Drawing Review Lasers supplied by Varian supplied by Customer All lasers mounted rigidly to concrete or steel beams Overhead laser mounted perpendicular or parallel to axis Overhead laser beam port at Isocenter Side lasers beam port shown at Isocenter height Adequate clearance in laser recesses Side lasers protected from bumping, if not recessed 7'-6" minimum height to sagittal laser beam port Notes Service provisions Duplex outlet at stand Eye-bolts above couch (new construction only) Notes Console convenience items Remote Diagnostics phone outlet Varis Network data outlet Telephone outlet Dimmable console area lighting Notes Patient intercom provisions Monitor located near couch Signal cable conduit to console Notes Customer Varian Drawing Doc No. Location [Page No.] Page 8 Y N 1100764 [2.29] Y N 1100733 [2.37] Y N 1100794 [2.27] Y N 1100733 [2.37] (verify req'ts with vendor) CCTV provisions camera(s) by Varian Proper location of camera(s) Power outlets for camera(s) (120v system only) Video cable conduit to console (verify req'ts with vendor) Notes Y N 1100782 [2.32] 1100733 Final Connections Electrical By Customer "Power connection made:" Responsibility note "Signal connections made:" Responsibility note "Varian-supplied cables pulled:" Responsibility note By Varian Notes Y N 861670 [2.38] Finishes Floor "Install finished floor after equipment is rigged" note Anti-static floor specified for Control Console Anti-static floor specified for Treatment Room Notes Rigging By Customer By Varian Proper clearances from unloading area to treatment room Rigging clearances require 4-piece breakdown of Clinac "Install finished floor after equipment is rigged" note 150 sq. feet secure storage area available during installation note Notes CL 600C /D, 6EX Review Y N 861667 [2.37] Y N 861672 [2.39] 1101809 [2.02] Page 8 of 9 Varian Planning Department Project Drawing Review cklstgating.xls Respitory Gating Refer to Installation Data Package Treatment Room Gating Section FOC = Face of Concrete Camera Position 15 degrees right or left of Isocenter (Plan) 40 degrees up from Isocenter Camera positioned between 6 and 14 feet from Isocenter Camera bracket Wall Ceiling Mounted Camera bracket construction detailed Notes Electrical 1" conduit from camera power supply to Gating Equipment Console Standard computer signal cable outlet at each end of conduit Note: Conduit length not to exceed 75'-0" Notes Power Grounded 120V. 60HZ. Power receptical for Camera power supply Power receptical located with in 12" of Camera power supply Standard wall switch for camera power supply receptical Notes 12/5/02 Customer Varian Notes: Drawing Doc No. Attach. 1 Location [Page No.] ] Y N G1.0 G7.0 Y N G1.0 G7.1 Y N G1.0 G7.1 Console Control area casework Gating Workstation located on Control counter 20 Amp dedicated quad receptical Grommets for cables Notes Notes Mounting of Camera power supply by Customer/Contractor Mounting of Camera support Bracket by Customer/Contractor "Varian-supplied cables pulled:" Responsibility note Notes CT Simulator Customer supplied wall mount for Gating camera Support Assembly Customer supplied Curtain rail for Gating Camera cable support Notes CL 600C /D, 6EX Review Y N G1.0 G2.0 Y N G1.0 G2.0 Y N 1.5.0 Page 9 of 9