Performance
Measurement range
-100 to +20°C (-148 to +68°F) dew point
Accuracy
±2°C (±3.6°F) dew point
Response time
5 mins to T95 (dry to wet)
Repeatability
0.5°C (0.9°F) dew point
Calibration
Traceable 13 point calibration
Electrical Specifications
An FM, CSA, ATEX intrinsically safe certified, 2-wire, rugged
impedance dew-point transmitter for continuous measurement in
hazardous area applications. Part number ESBx-NEC Sample Block is
available as an option to the Easidew TX I.S. It comprises NEC Class
1, Division 1 & 2, Groups B, C, and D components for installation in
hazardous areas.
•
FM, CSA, ATEX intrinsically safe certified
2-wire loop powered connection
Dew point or ppm moisture content
IP66 (NEMA 4)
Wide measurement range, calibrated -100 to +20ºC
(-148 to +68ºF) dew point
Operating temperature: -40 to +70ºC (-40 to +158ºF)
Simple Operation
The Easidew TX I.S. is easy to install and operate. It requires a 12
to 28 V DC supply and provides a linear 4-20 mA signal representing
moisture content.
Moisture in Gases
The output can be configured by the customer, using simple
software, to any range within -100 to +20°C (-148 to +68°F) dew
point, or 0-3,000 ppmV in gases. As the Easidew TX I.S. responds
to water vapor pressure in the gas sample, it will give the user an
accurate determination of the actual dew-point at pressure up to 45
MPa (450 barg / 6500 psig). Flow rate is not critical to measurement
accuracy (1 to 5 Nl/min (2.1 to 10.5 scfh)), though speed of
response will be improved by operating at a higher sample flow.
Moisture in Liquids
In addition to its moisture in gases capability the Easidew I.S.
can be used to monitor moisture in non-polar liquids. It is factory
programmed to provide an output in ppmW. This requires the
saturation constants of the liquid to be programmed into the
transmitter either at the factory or by the user, using the application
software, available as a free download from www.michell.com
Issue No: Easidew TX IS NEC_V2_US_0114
4-20 mA (2-wire connection, current source)
User configurable over range
Output
Dew point, moisture content for ppmV and ppmW
Analog output scaled range
Dew point: -100 to +20°C (-148 to +68°F)*
Moisture content in gas: 0-3000 ppmV
Moisture content in liquid: 0-3000 ppmW
Supply voltage
12 to 28 V DC
Load resistance
Max 250 Ω @ 12 V (500 Ω @ 24 V)
Current consumption
20 mA max
Saturation constants
(for moisture in liquids
measurements only)
6-point look-up table for saturation constants
up to 3000 ppmW over the temperature range
0 to +50°C.
Saturation constants for 8 common liquids can
be programmed into the Easidew I.S. via the
application software. Alternatively the user can
program saturation constants manually.
Operating Specifications
Highlights
•
•
•
•
•
Output signal
Operating humidity
0–100% RH
Operating temperature
-40 to +70°C (-40 to +158°F)
Operating pressure
45 MPa (450 barg / 6500 psig) max
Overpressure rating
x2 operating pressure 90 MPa (900 barg /
13053 psig)
Flow rate
1 to 5 Nl/min (2.1 to 10.6 scfh) mounted in
standard sampling block; 0 to 10 m/sec
(0 to 32.8 ft/sec) direct insertion
Temperature coefficient
Temperature compensated across
operating temperature range
Mechanical Specifications
Ingress protection
IP66 in accordance with standard
BS EN 60529:1992, and NEMA 4 in
protection accordance with standard
NEMA 250-2003
Hazardous area certificates
ATEX - II 1 G Ex ia IIC T4 (-20°C ≤ Ta ≤
+70°C)
FM - IS / I / I / ABCD / T4 Ta = +70°C
CSA - IS Class I Div I Groups ABCD T4
IECEx - Ex ia IIC T4 (-20°C ≤ Ta ≤ +70°C)
Housing material
316 stainless steel
Transmitter dimensions
L=132mm x ø27mm (L=5.19” x ø1.06”)
(sensor only)
Filter (sensor protection)
Standard: HDPE Guard <10μm
Optional: 316 stainless steel sintered guard
<80μm
Sensor connection and
material
5/8” - 18 UNF
316 stainless steel
Weight
8 lbs
Interchangeability
Fully interchangeable transmitter
Electrical connections
Hirschmann GDS series (DIN 4350-C)
www.michell.com/us
Please note: Michell Instruments adopts a continuous development program which sometimes
necessitates specification changes without notice. Please contact us for latest version.
Humidity and Dew-Point Instruments
Easidew TX I.S. with NEC Sample Block
Dimensions
Approved galvanic isolators
Condition
Sensor fault
Under-range dew point
Over-range dew point
Output
23 mA
4 mA
20 mA
KFD2-CR-EX1.20200
KFD2-CR-EX1.30200
KFD0-CS-EX1.50P
KFD0-CS-EX2.50P
KFD2-STC4-EX1.H
MTL5041, MTL5040
11"
Diagnostic conditions
(factory programmed)
Technical Specifications (cont.)
BREATHER
JUNCTION BOX
CLASS 1, DIV. 1,
GROUPS B, C & D
REDUCING BUSHING
Accessories and Spare Parts
Pack of 10 replacement HDPE guards
EA2-HDPE
Stainless steel sintered guard replacement
SSG
Easidew communications kit for EA2
EA2-CK
Cable, 0.8m (2.62’) length is standard. For additional cable
length replace the –xx with the length in feet
EA2-CAB-XXF
KFD0-CS-EX1.50P galvanic isolator (for Easidew PRO I.S. Pepperl & Fuchs)
GI-PF-01
10"
DOWTY SEAL,
6500 psig MAX.
INLET
1/8” NPT(F)
CUT-AWAY
VIEW
SAMPLE BLOCK
Electrical Connections
4-20 mA connections 2-wire
Pin 1
4-20 mA
Pin 3
POWER
CONDUIT
SEAL-OFF
3/4” NPT(F)
EASIDEW IS & CABLE
(SUPPLIED SEPARATELY)
4¾"
OUTLET
1/8” NPT
Ordering Codes
To construct the order code, select the relevant feature from the tables below, starting with the base model, which is {Feature
A} and then add on options to create a string: {Feature A} + {Feature B} + {Feature C} + {Feature D}
Order example: EA2-IS-TX + B3 + C1 + D2 + ESBI-NEC
Easidew Transmitter I.S., 5/8” UNF process connection, -100 to +20°C (-148 to +68°F) dp range, stainless steel sintered
guard, ESBI-NEC Stainless steel sample block comprising Class 1, Division 1 & 2, Groups B, C & D malleable iron junction box
with 3/4” NPT female conduit seal-off and breather/drain.
EA2-IS-TX
Base Model {Feature A}
Easidew Transmitter I.S.
EA2-IS-TX
Connection {Feature B}
5/8” UNF process connection
B3
Range {Feature C}
-100 to +20°C (-148 to +68°F) dp range
Non-standard measurement range: v= zero value,
w = full scale value, x = unit (C=°Cdp, F=°Fdp,
P=ppmV) y = Pressure for ppmV conversion in
barg/psig, Z = pressure unit (PG=psia, BG=barg,
BA=bara) blank = 0 barg/0 psig)
B3
C1
D2
ESBI-NEC
Sample Block and Hazardous Area Enclosure
{Feature E}
Stainless steel sample block comprising Class
1, Division 1 & 2, Groups B, C & D aluminum
junction box with 3/4” NPT female conduit seal-off
and breather/drain.
ESBA-NEC
Stainless steel sample block comprising Class 1,
Division 1 & 2, Groups B, C & D malleable iron
junction box with 3/4” NPT female conduit seal-off
and breather/drain.
ESBI-NEC
C1
CX(...)
Protection {Feature D}
Standard HDPE guard (with protection against
fine particulate (<10μm))
D1
Stainless steel sintered guard
D2
Michell Instruments, Inc 319 Newburyport Turnpike, Suite 207, Rowley, MA 01969
Tel: 978 484 0005, Fax: 978 843 7669, Email: us.info@michell.com, Web: www.michell.com/us
© Michell Instruments 2014
CUT-AWAY
VIEW