Installation, Operation and Maintenance Manual Sample Sequencer

Transcription

Installation, Operation and Maintenance Manual Sample Sequencer
Installation, Operation and
Maintenance Manual
Sample Sequencer® IV
SPD 5.5.14 Rev. 7 01/09
SENTRY EQUIPMENT CORP
PO Box 127
Oconomowoc, WI 53066 USA
Phone: 262.567.7256
Fax: 262.567.4523
Email: sales@sentry-equip.com
www.sentry-equip.com
Sample Sequencer®IV
Certification
Certification
The Sample Sequencer IV has been tested and is certified to the following standards:
Product Safety:
Certified to CAN/CSA C22.2 No. 1010.1-92 (Certificate # 1512429)
Certified to ANSI/ISA S82.02.01-99
Low Voltage Directive 73/23/EEC
Required Standards include:
EN 61010-1:2001/IEC 61010-1:2001
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1:
General requirements
Electromagnetic Compatibility:
EMC Directive 89/336/EEC:
Required Standards include:
EN 61326:1997 / IEC 61326:1997 w\A1:1998 & A2:2001
Electrical equipment for measurement, control, and laboratory use EMC requirements - Part 1: General
requirements
Immunity:
Required tests include:
IEC 61000-4-2 ESD
IEC 61000-4-3 Radiated Susceptibility
IEC 61000-4-4 EFT
IEC 61000-4-5 Surges
IEC 61000-4-6 RF Cond.
IEC 61000-4-8 Magnetic Susceptibility
IEC 61000-4-11 Volt dips, interrupts
Emissions:
Required tests include:
EN 55011 Class A limits
EN 61000-3-2 Harmonic
EN 61000-3-3 Voltage Fluctuation (Flicker)
RoHS Compliance:
Directive 2002/95/EC
This product complies with the RoHS Directive.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
Sample Sequencer®IV
Table of Contents
TABLE OF CONTENTS
SPECIFICATIONS ........................................................................................................................1
SAFETY PRECAUTIONS .............................................................................................................3
Use of Hazard Information.................................................................................................3
Definitions ..........................................................................................................................3
QUICK START GUIDE..................................................................................................................5
GENERAL DESCRIPTION ...........................................................................................................7
Modes of Operation ...........................................................................................................7
Inputs and Outputs ..........................................................................................................11
Optional Manifold Sampling Valve...................................................................................12
OPERATION ...............................................................................................................................13
Description of Controls ....................................................................................................13
Screen Descriptions ........................................................................................................15
Operation Screen.................................................................................................16
Main Menu ...........................................................................................................17
Sample/Bypass ....................................................................................................17
Historical Data......................................................................................................18
Setup....................................................................................................................18
Sequence.............................................................................................................19
Timers ..................................................................................................................19
I/O Calibration ......................................................................................................20
Scale ....................................................................................................................20
Input Calibration...................................................................................................21
Output Calibration ................................................................................................21
Test Outputs ........................................................................................................22
Clock ....................................................................................................................22
Point Descriptions ................................................................................................23
Setup Continued ..................................................................................................23
Analyzer Descriptions ..........................................................................................24
Mode ....................................................................................................................24
Valve Assignment ................................................................................................25
Operating Modes .............................................................................................................26
Continuous...........................................................................................................26
Batch....................................................................................................................28
Valve Outputs ..................................................................................................................29
INSTALLATION ..........................................................................................................................30
Unpacking........................................................................................................................30
Mounting ..........................................................................................................................30
Pole Mounting ......................................................................................................31
Wall Mounting ......................................................................................................32
Panel Mount.........................................................................................................33
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
i
Sample Sequencer®IV
Table of Contents
Electrical Connections .....................................................................................................36
Relay Board Connections ................................................................................................36
Power Connection................................................................................................36
Point Identification and Alarm Connections .........................................................37
Track and Hold Board Connections.................................................................................37
CPU Board Connections..................................................................................................38
Valve Output Connections ...................................................................................38
Analog Input Connections....................................................................................38
Analyzer Alarm Connections................................................................................38
Calibration Hold Connections ..............................................................................39
End of Batch Connections ...................................................................................39
MODBUS® Connections......................................................................................39
Starting Up The Sample Sequencer ................................................................................42
MAINTENANCE ..........................................................................................................................43
Cleaning ..........................................................................................................................43
Rechargeable Battery ......................................................................................................43
Fuses ...............................................................................................................................43
Disassembly ....................................................................................................................44
Calibration .......................................................................................................................44
Analog Input Calibration.......................................................................................44
Analog Output Calibration....................................................................................45
TROUBLESHOOTING GUIDE....................................................................................................46
FACTORY ASSISTANCE ...........................................................................................................47
SPARE PARTS LIST ..................................................................................................................48
ACCESSORIES ..........................................................................................................................49
WARRANTY................................................................................................................................50
APPENDIX A - RECOMMENDED PIPING .................................................................................52
APPENDIX B – USING A SEQUENCER WITH HACH SERIES 5000 ANALYZER...................53
APPENDIX C – LCD AVAILABLE CHARACTERS....................................................................56
APPENDIX D – MODBUS® Registers.......................................................................................58
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
ii
Sample Sequencer®IV
Table of Contents
List of Figures
Figure 1 - Sample Sequencer with Continuous Analyzer ................................................................ 8
Figure 2 - Sample Sequencer with Batch Analyzer ......................................................................... 9
Figure 3 – Sample Sequencer with two Analyzers ........................................................................ 10
Figure 4 – Sample Sequencer Front Panel ................................................................................... 14
Figure 5 – Screen Navigation ........................................................................................................ 15
Figure 6 – Operation Screen ......................................................................................................... 16
Figure 7 – Main Menu ................................................................................................................... 17
Figure 8 – Sample/Bypass ............................................................................................................ 17
Figure 9 – Historical Data.............................................................................................................. 18
Figure 10 – Setup Screen ............................................................................................................. 18
Figure 11 – Set Sequence Screen ................................................................................................ 19
Figure 12 – Timers Screen ............................................................................................................ 19
Figure 13 – I/O Calibration ............................................................................................................ 20
Figure 14 – Set Scale.................................................................................................................... 20
Figure 15 – Input Calibration ......................................................................................................... 21
Figure 16 – Selecting Channel to Calibrate................................................................................... 21
Figure 17 – Output Calibration ...................................................................................................... 21
Figure 18 – Selecting Channel to Test .......................................................................................... 22
Figure 19 – Test Outputs............................................................................................................... 22
Figure 20 – Set Clock Screen ....................................................................................................... 22
Figure 21 – Set Sample Descriptions Screen ............................................................................... 23
Figure 22 – Continued Setup Screen ............................................................................................ 23
Figure 23 – Set Analyzer Descriptions Screen.............................................................................. 24
Figure 24 – Mode Selection Screen .............................................................................................. 24
Figure 25 – Valve Assignment Screen .......................................................................................... 25
Figure 26 – Continuous Mode Timing Diagram............................................................................. 27
Figure 27 – Batch Mode Timing Diagram...................................................................................... 29
Figure 28 – Enclosure Dimensions ............................................................................................... 31
Figure 29 – Pole Mounting the Sample Sequencer....................................................................... 32
Figure 30 – Wall Mounting the Sample Sequencer ....................................................................... 32
Figure 31 – Mounting Bracket Dimensions ................................................................................... 33
Figure 32 – Panel Mount Dimensions ........................................................................................... 34
Figure 33 – Panel Mounting .......................................................................................................... 35
Figure 34 – Relay Board ............................................................................................................... 37
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
iii
Sample Sequencer®IV
Table of Contents
Figure 35 – Track and Hold Board ................................................................................................ 38
Figure 36 – CPU Board ................................................................................................................. 39
Figure 37 – Electrical Wiring Diagram ........................................................................................... 40
Figure 38 – Electrical Wiring Diagram ........................................................................................... 41
Figure 39 – Electrical Wiring Diagram ........................................................................................... 41
Figure 40 – Recommended Sample Conditioning......................................................................... 52
Figure 41 – Wiring to a Hach Series 5000 Analyzer ..................................................................... 55
Figure 42 – LCD Character Chart.................................................................................................. 56
Figure 43 – Scrolling Through the Character List.......................................................................... 57
Figure 44 – MODBUS Setup ......................................................................................................... 58
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
iv
Sample Sequencer®IV
Table of Contents
List of Tables
Table 1 – Control Descriptions ...................................................................................................... 13
Table 2 – Power Cord Requirements ............................................................................................ 36
Table 3 – Troubleshooting Guide .................................................................................................. 46
Table 4 – MODBUS® Registers .................................................................................................... 59
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
v
Sample Sequencer® IV
Specifications
SPECIFICATIONS
Specifications are subject to change without notice.
Applications: The Sample Sequencer is designed to support sample stream switching in the batch or
continuous modes.
Maximum Number of Samples: 8
Maximum Number of Analyzers: 2
Power:
•
•
•
Input Voltage: 100 - 240 Vac
Frequency Range: 50/60 Hz
Input Current:
1 A / 115 V
0.7 A / 230 V
Inputs from Analyzer:
• Two analog inputs: 0-20 mA or 4-20 mA (59 Ω load)
• Two digital inputs for end-of-analysis indication.
• Two digital inputs for calibration indication.
• Two digital inputs for analyzer system alarm indication.
Outputs:
•
•
•
•
MODBUS® communication port on an RS-485 network.
Valve outputs (12 VDC) to control a maximum of eight solenoid valves for sample stream
switching.
Ten DPDT relays for point number indication and replicated Analyzer Alarms. Contact ratings:
Maximum operating voltage: 250 VAC, 220 VDC.
Maximum operating voltage: 130 V (CSA certification)
Maximum switching capacity: 30 W, 62.5 VA inductive; 60 W, 125 VA resistive.
Current Outputs: Eight isolated 0 – 20 or 4 – 20 mA outputs.
Isolation: 550 VAC
Maximum Load: 600 Ω
Nominal Accuracy:
• Analog Inputs: ± 0.0015% full scale
• Analog Outputs: ± 0.2% full scale
Resolution:
• Analog Inputs: 0.038 µA/LSB
• Analog Outputs: 0.331 µA/LSB
Data Logging: 255 Records
Keypad: Seven tactile feedback membrane switches.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
1
Sample Sequencer® IV
Display:
Specifications
Four line by 20 character Vacuum Fluorescent Display (VFD).
Enclosure:
NEMA 4X (indoor), IP 66
Dimensions: 8.93” (226.8 mm) H x 9.63” (224.6 mm) W x 6.76” (171.7 mm) D
(with mounting bracket)
Mounting: Panel, surface or pipe mount.
Ambient Temperature: 0-50 C (32-122 F); RH 95% max. non-condensing at 40 C (104 F) max.
Operating Altitude: 0-2000 m (0-6562 ft.) (Consult factory for other altitudes.)
Agency Approvals:
Shipping Weight: 6 lbs (2.72 kg)
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
2
Sample Sequencer® IV
Safety Precautions
SAFETY PRECAUTIONS
Please read the entire manual before attempting to unpack, set up or operate this product. Pay careful
attention to all warnings, cautions and notes. Failure to do so could result in serious personal injury or
equipment damage.
Use of Hazard Information
If multiple hazards exist, the signal word corresponding to the greatest hazard shall be used.
Definitions
Indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury.
CAUTION used without the safety alert symbol indicates a potentially hazardous
situation which, if not avoided, may result in property damage.
NOTE
Information that requires special emphasis
SHALL
This word understood to be mandatory
SHOULD
This word understood to be advisory
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
3
Sample Sequencer® IV
Safety Precautions
Indique une situation imminente et dangeureuse qui doit être évité immédiatement
afin d’éviter tout risque de blessures sérieuses et qui pourrait même entraîner la
mort.
AVERTISSEMENT
Indique une situation potentiellement dangeureuse et qui doit être évité
immédiatement afin d’éviter tout risque de blessures sérieuses et qui pourrait
même entraîner la mort.
PRÉCAUTION
Indique une situation potentiellement dangeureuse et qui doit être évité
immédiatement afin d’éviter tout risque de blessures mineures.
PRÉCAUTION
ATTENTION une utilisation sans le symbole de sécurité indique une situation
potentiellement dangeureuse et que si elle n’est pas évité, pourrait entrainer des
dommages à la propriété.
NOTE
Informations requérant une attention spéciale
DOIT
Ce mot indique une obligation
POURRAIT
Ce mot indique un avertissement
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
4
Sample Sequencer® IV
Quick Start Guide
QUICK START GUIDE
The following guide is intended to help you get started using the Sentry Sample Sequencer™. This
guide does not include detailed explanations and should be used in conjunction with the remainder
of this manual to build a full understanding of the Sample Sequencer™.
• Electrical Connections
Connect all electrical connections necessary for the application. If the analyzer being shared
is a batch type analyzer, an end of analysis signal must be connected to the Sequencer™
“End of Batch” input. Without this signal, the Sequencer™ will never advance to the next
sample.
• Starting and Stopping the Sequencer
With the display showing the Operation Screen, press the ENTER key to toggle Sequencer 1
on and off. Press the ESC key to toggle Sequencer 2 on and off.
• Setting the Sequence
With the display showing the Operation Screen, press the MENU key to display the MAIN
MENU. Using the ENTER and arrow keys, go to SETUP and then SEQUENCE. Edit the
sequence by using the up arrow (▲) and down arrow (▼). Move locations within the
sequence using the right arrow (►) and left arrow (◄). Toggle between sequences using the
ENTER key.
Use the ESC key to exit the SET SEQUENCE screen. Changes to the
sequence are automatically saved as you make them.
• Placing Samples in “Sample/Bypass”
Using the same technique as above, use the keypad to navigate to MAIN MENU,
SAMPLE/BYPASS. In this screen, the character “S” indicates the sample point will be
Sampled. The character “B” indicates the sample point will be Bypassed the next time its
position in sequence is up. Use the ▲ and ▼ keys to change between Sample and Bypass.
Use the ► and ◄ keys to move between sample points. Use the ESC key to exit the
SAMPLE/BYPASS screen.
• Setting Sample Timers
Using the same technique as above, use the keypad to navigate to MAIN MENU, SETUP,
TIMERS. The flashing digits indicate the current field being edited. Use the ▲ and ▼ keys to
change the value. Use the ► and ◄ keys to move between fields. Use the ENTER key to
cycle through the sample points. Use the ESC key to exit the SET TIMERS screen.
• Setting the Mode
Using the same technique as above, use the keypad to navigate to MAIN MENU, SETUP,
MORE, MODE. Use the ENTER key to toggle between Continuous and Batch. Use the ▲
and ▼ keys to move between SEQ1 and SEQ2. Use the ESC key to exit the MODE screen.
• Setting the Valve Assignment
Using the same technique as above, use the keypad to navigate to MAIN MENU, SETUP,
MORE, VALVE ASSIGNMENT. Use the ▲ and ▼ keys to change the valves assigned to
SEQ1. If all eight channels are not assigned to SEQ1, the remaining channels will be
assigned to SEQ2. Use the ESC key to exit the VALVE ASSIGNMENT screen. The SET
SEQUENCE screen will automatically appear after the VALVE ASSIGMENT screen.
• Editing the Sample Names
Using the same technique as above, use the keypad to navigate to MAIN MENU, SETUP, PT
DESC. The flashing character indicates the current character being edited. Use the ▲ and
▼ keys to change the character. Use the ► and ◄ keys to move between characters. Use
the ENTER key to cycle through the sample points. Use the ESC key to exit the SET
DESCRIPTIONS screen.
• Editing Analyzer Descriptions
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
5
Sample Sequencer® IV
Quick Start Guide
Using the same technique as above, use the keypad to navigate to MAIN MENU, SETUP,
MORE, AIT DESCRIPTIONS. The flashing character indicates the current character being
edited. Use the ▲ and ▼ keys to change the character. Use the ► and ◄ keys to move
between characters. Use the ESC key to exit the SET ANALYZER DESCRIPTIONS screen.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
6
Sample Sequencer®IV
General Description
GENERAL DESCRIPTION
The SENTRY Sample Sequencer is a microprocessor-based instrument capable of electronically
switching up to eight sample streams one at a time to an analyzer and of providing signals to
recorders. If a total of eight sample streams are not needed for one process, the remaining points
may be assigned to a second analyzer. A Sentry manifold sampling valve specially designed to obtain
optimum representative samples is available for sample stream switching with the Sample Sequencer.
The Sample Sequencer reduces the number of required analyzers while maintaining equivalent data
accuracy. Reliable sample analysis comparison is obtained by using a Sample Sequencer with one
dedicated analyzer. In addition, significant operating cost savings are obtained by not having to
install, maintain and service multiple analyzers.
The Sample Sequencer is capable of operating in the continuous mode for continuous-type analyzers
or in the batch mode for batch-type analyzers. In addition, the Sequencer is capable of measuring the
analyzer analog output signal and converting the analog signal to digital data to be sent out via the
Sequencer’s MODBUS® communication port. Using the MODBUS® capability, one can connect a
PC via twisted pair cable to one or more Sequencers for remote monitoring, control and data
acquisition.
Track and Hold outputs are available to provide independent, isolated outputs for each sample point.
Contact closures are also available to provide channel number indication as well as replicating
analyzer alarms.
Figures 1 - 3 illustrate a typical system arrangement for sample stream switching with the Sample
Sequencer. Customer-supplied equipment is shown for clarification.
Modes of Operation
The Sample Sequencer has two programmable modes of operation:
BATCH
Used for analyzers that take a sample "bite" (or batch), produce an analytical result and then
provide a contact signal output, in typically one to 20 minutes.
CONTINUOUS
Used for analyzers that continuously produce an analytical result.
NOTE
If using the Sample Sequencer with a Hach analyzer, see Appendix B.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
7
Sample Sequencer®IV
General Description
Figure 1 - Sample Sequencer with Continuous Analyzer
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
8
Sample Sequencer®IV
General Description
Figure 2 - Sample Sequencer with Batch Analyzer
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
9
Sample Sequencer®IV
General Description
Figure 3 – Sample Sequencer with two Analyzers
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
10
Sample Sequencer®IV
General Description
Inputs and Outputs
A variety of inputs and outputs are available on the Sample Sequencer.
Analog Inputs: Two analog inputs are located on the CPU board for reading a 0–20 mA or 4–20 mA
signal from two different analyzers. The Sequencer converts this signal into Engineering Units for
display and outputs the same signal on the Track and Hold board.
Analyzer Alarm Inputs: Two analyzer alarm inputs are located on the CPU board. These inputs are
used to indicate a loss of sample flow to the analyzer. An analyzer alarm will cause the analyzer
description to flash on the operation screen.
When the Sequencer is used in the Continuous mode, an analyzer alarm will immediately close the
active sample valve and open the next sample valve. That sample valve will remain open until the
analyzer alarm clears or one minute elapses. After one minute, the Sequencer will open the next
sample valve. The Sequencer will continue trying to establish flow every minute until the analyzer
alarm clears.
When the Sequencer is used in the Batch mode, an analyzer alarm will not immediately close the active
sample valve. Instead, when the analyzer alarm is received, the Sequencer will start a timeout timer. If
the Sequencer does not receive an End of Batch indication from the analyzer or the alarm does not
clear before that timeout timer, the Sequencer will close the current sample valve and open the next
valve in the sequence. If the End of Batch indication is received before the timeout timer, the
Sequencer assumes the alarm is not a loss of flow and continues with normal operation. The operator
can adjust the timeout timer by setting each sample’s Sample Timer.
Calibration Hold Inputs: Two calibration hold inputs are located on the CPU board. These inputs are
used to indicate the analyzer is performing a calibration. The calibration hold inputs do not perform a
function when the Sequencer is in the Batch mode.
In the Continuous mode, the Sequencer will wait at the active sample with the valve open when a
calibration indication is received. Once the calibration indication clears, the Sequencer will resume
operation entering the Flush state for the sample that was active before the calibration hold indication
was received.
End of Batch Inputs: Two End of Batch inputs are located on the CPU board. These inputs are used to
indicate when a batch type analyzer is finished taking a measurement. Without this input connected,
the Sequencer will never advance to the next sample when the Sequencer is in the Batch mode. The
End of Batch inputs are not used in the Continuous mode.
Valve Outputs: Eight valve outputs are located on the CPU board. These outputs supply 12 VDC to
operate the sample valves. The eight valves may be divided between two analyzers in any
combination. The only requirement is the valves assigned to an analyzer must be consecutive points in
the Sequencer.
Track and Hold Outputs: Eight individually isolated analog outputs are provided on the Track and Hold
board. These outputs are capable of 0-20 mA or 4-20 mA outputs. When a valid reading is taken from
the analyzer for a given sample, the Sequencer updates (“Tracks”) the corresponding signal on the
Track and Hold board. When the sample is not currently being analyzed, the Sequencer “Holds” the
corresponding signal on the Track and Hold board at its last known value.
Relay Outputs: A total of ten individually fused dry contacts are available on the Relay board. Two of
these outputs are used to replicate the two-analyzer alarm inputs while the remaining eight are used for
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
11
Sample Sequencer®IV
General Description
sample point indication. A closed contact indicates the point is currently being sampled and the
analyzer reading corresponds to that sample point.
MODBUS® Communications: The Sequencer is capable of connecting to an RS-485 network for
communicating to a PC or control system via MODBUS®. For more information on this feature, refer to
Appendix D.
Optional Manifold Sampling Valve
Although the Sample Sequencer may be used to control most manifold valve arrangements, the Sentry
Manifold Sample Valve offers a convenient optional accessory to obtain a representative sample. The
patented design incorporates block and bleed design and provides continuously flowing samples with
no areas for cross contamination to occur. The valve is available with a variety of DC and AC solenoid
coils and comes in either four or eight sample line models. Contact Sentry for more information about
this special sampling valve.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
12
Sample Sequencer® IV
Operation
OPERATION
To ensure the protection provided by this equipment is not impaired, this equipment must not
be installed or used in any manner other than that which is specified in this manual.
AVERTISSEMENT
Afin de s’assurer que la protection prévue par cette pièce d’équipement n’est pas défaillante,
cet équipement ne doit pas être installé ou utilisé de manière autre que spécifié dans le manuel.
Description of Controls
Figure 4 and Table 1 describe controls on the Sample Sequencer membrane switch display.
DESCRIPTION
FUNCTION
MENU key
From the operation screen, pressing the MENU key will display the
main menu.
ENTER key
From the operation screen, pressing the ENTER key will toggle SEQ1
on and off.
While navigating through the menu screens, pressing the ENTER key
will select the option indicated by the cursor (►).
When editing sample names, timers and sequences, pressing the
ENTER key will select the next sample point to edit.
ESC key
From the operation screen, pressing the ESC key will toggle SEQ2 on
and off.
While in the menu screens, pressing the ESC key will exit the current
screen. Any changes are automatically saved.
Arrow keys
While navigating through the menu screens, use the arrows keys to
move the cursor to the option you want to choose.
While editing character fields and timers, use the arrow keys to move
between fields and to change the value in the current field.
▲
◄
►
▼
Cursor
Indicates the cursor position on the menu screens.
►
Table 1 – Control Descriptions
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
13
Sample Sequencer® IV
Operation
Figure 4 – Sample Sequencer Front Panel
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
14
Sample Sequencer® IV
Operation
Screen Descriptions
Operation Screen
Main Menu
Sample/Bypass
Sequence
Scale
Historical Data
Timers
Input Cal
Analyzer
Descriptions
I/O Cal
Cloc
k
Output Cal
MODBUS
Setup
Point Description
More
Test
Mode
Valve
Assignment
Figure 5 – Screen Navigation
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
15
Sample Sequencer® IV
Operation
Operation Screen
The operation screen is the main screen on the Sample Sequencer. This screen contains all the
information indicating the state of the Sequencer. Sequencers are started and stopped from this
screen.
Analyzer
Description
Open Sample Valve
Number and Name
S
S
S
S
Status
E
A
E
A
Q
M
Q
M
1
1
S A M P L E
1
P L E
2 0 : 0 0
5 0 0
2
S A M P L E
8
N A
P L E
7
N A ME
5 0
N
0
M
0
A M E
P P B
E
0
Sequencer 1
Display
Sequencer 2
Display
Open Sample
Valve Name
Previous Sample
and Value
Figure 6 – Operation Screen
Figure 6 shows a typical operation screen for the Sample Sequencer. The top two lines display the
status of Sequencer 1 running in the Continuous mode. The bottom two lines display the status of
Sequencer 2 running in the Batch mode.
Starting/Stopping Sequencers: Sequencers are started and stopped from the operation screen. To
toggle SEQ1 on and off, press the ENTER key. To toggle SEQ2 on and off, press the ESC key.
Analyzer Description (Continuous and Batch modes): This is a user definable field that describes the
analyzer being sequenced. The maximum number of characters is four. When an analyzer alarm is
received, this description will flash.
Open Sample Valve Number and Name (Continuous): This displays the valve number and the sample
description for the currently open valve.
Status (Continuous): This displays the state of the Sequencer, time remaining, and value. The three
available states are FLUSH and SAMPLE and OFF. The time displayed is the time remaining in the
current state. The value is only displayed when the Sequencer is in the SAMPLE state.
Open Sample Valve Name (Batch): This displays the sample description of the currently open valve.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
16
Sample Sequencer® IV
Operation
Previous Sample and Value (Batch): This displays the description of the previously sampled point and
its value. The value on the analyzer corresponds to this sample point.
Main Menu
M E N U
►S AM P L E / B Y P A S S
H I S T O R I C A L
D A T A
S E T U P
Figure 7 – Main Menu
Figure 7 shows the main menu screen. The main menu screen is reached by pressing the MENU key
from the operation screen. The cursor indicates the option you wish to choose. Move the cursor by
using the ▲ and ▼ keys. Select the option by using the ENTER key. To exit this screen and return to
the operation screen, press the ESC key.
Sample/Bypass
Sample streams may be temporarily skipped in the sequence by placing the sample point number in
Bypass. To place the point back in the sequence, place the sample point number in Sample. If the
Sample Sequencer is currently on a sample point that you place in Bypass, the Sample Sequencer will
immediately skip to the next point in the sequence. Occasionally, it is necessary to hold an analysis on
one sample point. To accomplish this, simply place all the undesired sample points in Bypass. Sample
points are placed in either Sample or Bypass by using the Sample/Bypass screen.
1
S
Point
Number
2
B
S A M P L E / B Y P A S S
3
4
5
6
7
8
S
S
B
B
B
S
S = Sample
B = Bypass
Figure 8 – Sample/Bypass
Figure 8 shows the Sample/Bypass screen. Use the ▲ key to place a point in “Sample” and the ▼ key
to place a point in “Bypass.” When a point that is placed in “Bypass” is next to be sequenced, the point
will be skipped and the next point that is in “Sample” will be sequenced. The flashing character
indicates the sample point being edited. To move between sample points, use the ◄ and ► keys. To
exit this screen, use the ESC key.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
17
Sample Sequencer® IV
Operation
Historical Data
Time Stamp
Hours : Minutes
Date Stamp
Month / Day
New Data
▲
Old Data
▼
H I S T
1 1 / 2
2
S A
5
O R I C A L
D A T A
9
1 7 : 5 3
M P L E
2
N A M E
0 0 0
Historical Value
Sample Number
and Description
Figure 9 – Historical Data
Figure 9 shows the Historical Data screen. The historical logger is capable of storing the last 255
values. The Date Stamp and Time Stamp are taken the moment the data is logged. Note: The Date
Stamp does not show the year and the Time Stamp is in 24-hour mode. The New Data indication
(▲) indicates more recent data is available for viewing. The Old Data indication (▼) indicates older
data is available for viewing. Use the ▲ key to view the more recent data and the ▼ key to view the
older data. To exit, use the ESC key.
Setup
S E T U P
►S E Q U E N C E
C L O C K
T I M E R S
P T
D E S C
I / O
C A L
MO R E
Figure 10 – Setup Screen
Figure 10 shows the Setup Screen. The cursor indicates the option you wish to choose. Move the
cursor within a column by using the ▲ and ▼ keys. Move the cursor between columns by using the ◄
and ► keys. Select the option by using the ENTER key. To exit this screen and return to the Main
Menu screen, press the ESC key.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
18
Sample Sequencer® IV
Operation
Sequence
The order that the Sample Sequencer opens sample points can be customized. The sequence for
each sequencer is edited from the Set Sequence screen.
S
1
S
5
E
2
E
6
T
S E
3 4 1 2
T
S E
7 5 6 8
Q
3
Q
5
U
4
U
6
E
1
E
7
N
2
N
5
C
3
C
6
E
4
E
8
R
1
1 2 3 4
R
2
5 6 7 8
Figure 11 – Set Sequence Screen
Figure 11 shows the Set Sequence screen. Each sequence contains 16 positions and are freely
editable. The only limitation is the highest point number that can be assigned to Sequencer 1 and the
lowest point number that can be assigned to Sequencer 2. For example, if valves 1 – 4 are assigned to
Sequencer 1, then 4 is the largest sample point number that can be assigned to Sequencer 1 and 5 is
the lowest point number that can be assigned to Sequencer 2. If Sequencer 2 is not displayed, then all
eight sample points are assigned to Sequencer 1. In order to set Sequencer 2, the valve assignment
must be changed in the Valve Assignment screen.
The flashing number indicates the current step in the sequence being edited. To change the number,
use the ▲ and ▼ keys. To move between steps in the sequence, use the ◄ and ► keys. Use the
ENTER key to toggle between Sequencer 1 and Sequencer 2. To exit this screen and return to the
Setup screen, use the ESC key.
Timers
1
T I M E RS
S A M P L E
1
N A M E
S A M P L E
F L U S H
2 0 : 0 0
1 0 : 0 0
Description of
sample point
being edited
Figure 12 – Timers Screen
Figure 12 shows the Timers screen. Each sample point has Sample and Flush timers that can be set
individually. In the Continuous mode, the Flush time is the amount of time the analyzer is flushed out
before the analyzer reading is considered valid. The Sample time is the amount of time the analyzer is
allowed to analyze the sample while the reading is considered valid. The Track and Hold value and
Point Identification are active during the Sample time. The valve for this sample point is open during
both the Sample time and the Flush time.
In the Batch mode, the Flush time is not used. The Sample time serves as a timeout timer that operates
only when the analyzer is in alarm. While in alarm, if an End-of-Batch signal is not received before the
Sample time times out, the Sample Sequencer will close the current sample point and proceed with the
next sample point. Since the Analyzer Alarm input may serve to indicate more than a loss of sample flow,
the Sample time should be set to give the analyzer sufficient time to attempt to complete an analysis and
output an End-of-Batch indication.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
19
Sample Sequencer® IV
Operation
The flashing number indicates the current number being edited. To change the number, use the ▲ and
▼ keys. To move between numbers, use the ◄ and ► keys. To cycle through the eight sample
points, use the ENTER key. To exit this screen and return to the Setup screen, use the ESC key.
I/O Calibration
I / O
C A L I B R A T I O N
T E S T
I N P U T
C A L
O U T P U T
C A L
►S C A L E
Figure 13 – I/O Calibration
Figure 13 shows the I/O Calibration Screen. The cursor indicates the option you wish to choose. Move
the cursor within a column by using the ▲ and ▼ keys. Move the cursor between columns by using
the ◄ and ► keys. Select the option by using the ENTER key. To exit this screen and return to the
Setup screen, press the ESC key.
Scale
S E Q 1
S E Q 2
S C A L E
M I N
M A X
0 0 0 0
5 0 0 0
0 0 0 4
0 0 2 0
P P B
mA
Figure 14 – Set Scale
Figure 14 shows the Set Scale screen. This screen allows the analog input value to be displayed in
engineering units. To display engineering units, the scale should be set to the same range as the
analyzer.
The flashing field indicates the current field being edited. To change the value, use the ▲ and ▼ keys.
To move between fields, use the ◄ and ► keys. The scales can be incremented and decremented by
a factor of 100 by holding the ENTER key while using the ▲ and ▼ keys. The units can be selected
from the built in values PPB, PPM, ug, mA, pH, and uS.
To exit this screen and return to the I/O Calibration screen, press the ESC key.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
20
Sample Sequencer® IV
Operation
Input Calibration
A N A L O G
I N P U T 1
C A L
I N P U T
0 / 4
mA
O F F S E T
1 6 7 7 7 2 1 5
R AW
C O U N T
0 0 0 0 0 0 0 0
Figure 15 – Input Calibration
Figure 15 shows the Analog Input Calibration screen. For more information on calibrating the inputs,
please refer to the Maintenance section of this manual.
Output Calibration
A N A L O G
O U T P U T
C H A N N E L
1
C A L
Figure 16 – Selecting Channel to Calibrate
Figure 16 shows the first screen displayed once OUTPUT CAL is selected from the I/O Calibration
screen. This screen selects the output channel to calibrate. For more information on calibrating the
outputs, refer to the Maintenance section of this manual.
A N A L
C
O U T
O F F
O
H
P
S
G
O U T P U T
C A L
A N N E L
1
U T
0 / 4
mA
E T
1 3 0 6 6
Figure 17 – Output Calibration
Figure 17 shows a typical output calibration screen. For more information on calibrating the outputs,
refer to the Maintenance section of this manual.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
21
Sample Sequencer® IV
Operation
Test Outputs
A N A L O G
O U T P U T
C H A N N E L
1
T E S T
Figure 18 – Selecting Channel to Test
Figure 18 shows the first screen displayed once TEST is selected from the I/O Calibration screen. This
screen selects the Track and Hold output channel to test. To change channels, use the ▲ and ▼ keys.
To select the channel, press the ENTER key. To exit this screen and return to the I/O Calibration
screen, press the ESC key.
A N A L O G
O U T P U T
C H A N N E L
1
7 5 %
T E S T
Figure 19 – Test Outputs
Figure 19 shows the Test Outputs screen, which is displayed after a channel has been selected. To
change the output levels, use the ▲ and ▼ keys. The 5 available output levels are 0%, 25%, 50%,
75%, and 100% of full scale. To exit this screen and return to the Select Channel to Test screen, press
the ESC key.
Clock
The Sample Sequencer contains a real time clock that provides the time and date stamp for the
historical logger. The real time clock is set from the Set Clock screen.
S ET
C L O CK
1 5 : 2 3 : 12
1 2 / 3 1 / 0 2
Figure 20 – Set Clock Screen
Figure 20 shows the Set Clock Screen. The flashing number indicates the number being edited. To
change the number, use the ▲ and ▼ keys. To move between numbers, use the ◄ and ► keys.
Note: The time should be set using the 24-hour mode. To exit this screen and return to the Setup
screen, use the ESC key.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
22
Sample Sequencer® IV
Operation
Point Descriptions
The factory default for each sample point description is in the form of “SAMPLE 1 NAME.” These
descriptions may be changed from the Set Sample Descriptions screen. Refer to Appendix C for a
guide of available characters.
S E T
D E S C R I P T I O N S
1
S A M P L E
1
N A M E
Figure 21 – Set Sample Descriptions Screen
Figure 21 shows the Set Sample Descriptions screen. The flashing character indicates the field being
edited. To change the character, use the ▲ and ▼ keys. To move between characters, use the ◄
and ► keys. To cycle through the eight sample points, use the ENTER key. To exit this screen and
return to the Setup screen, use the ESC key.
Setup Continued
S E T U P
C O N T I N U E D
D E S C R I P T I O N S
MO D E
V A L V E
A S S I G N M E N T
►A I T
Figure 22 – Continued Setup Screen
Figure 22 shows the Continued Setup screen. The cursor indicates the option you wish to choose.
Move the cursor by using the ▲ and ▼ keys. Select the option by using the ENTER key. To exit this
screen and return to the Setup screen, press the ESC key.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
23
Sample Sequencer® IV
Operation
Analyzer Descriptions
The factory defaults for the analyzer descriptions are SEQ1 and SEQ2. These descriptions may be
changed using the Set Analyzer Descriptions screen. Refer to Appendix C for a guide of available
characters.
S ET
A NA L Y Z E R
D E S C R I P T I O N S
S E Q 1
S E Q 2
Figure 23 – Set Analyzer Descriptions Screen
Figure 23 shows the Set Analyzer Descriptions screen. The flashing character indicates the field being
edited. To change the character, use the ▲ and ▼ keys. To move between characters, use the ◄
and ► keys. To exit this screen and return to the Continued Setup screen, use the ESC key.
Mode
The Sample Sequencer is capable of operating in two different modes, depending on the type of
analyzer being used. The continuous mode is for continuous output analyzers such as sodium,
conductivity, pH, O2, some colorimetric, and many others. The batch mode is for analyzers that take a
"bite" or batch of sample and then spend some time analyzing it before outputting a new fixed analog
signal. The mode is changed from the Mode Selection screen.
MO D E
►S E Q 1
S E Q 2
C O N T I N U O U S
B A T C H
Figure 24 – Mode Selection Screen
Figure 24 shows the Mode Selection screen. The cursor indicates the current Sequencer being edited.
To change the mode, use the ENTER key to toggle between BATCH and CONTINUOUS. Use the ▲
and ▼ keys to move between Sequencers. To exit this screen and return to the Continued Setup
screen, use the ESC key.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
24
Sample Sequencer® IV
Operation
Valve Assignment
The Sample Sequencer can sequence eight sample points to an analyzer. If all eight points are not
needed for one analyzer, the remaining sample points may be assigned to a second analyzer.
Assigning all eight points to SEQ1 will suppress information in other screens for SEQ2. The sample
points are assigned to a sequencer using the Valve Assignment screen.
V A L V E
A S S I G N M E N T
S E Q 1
1 - 6
Figure 25 – Valve Assignment Screen
Figure 25 shows the Valve Assignment Screen. This screen assigns the valves to Sequencer 1. The
remaining valves not assigned to Sequencer 1 are automatically assigned to Sequencer 2. Note:
Assigning one valve or seven valves to Sequencer 1 is not allowed. Use the ▲ and ▼ keys to
change the valve assignment. To exit this screen and enter the Set Sequence screen, use the ESC
key.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
25
Sample Sequencer® IV
Operation
Operating Modes
Continuous
This mode is for continuous output analyzers such as sodium, conductivity, pH, O2, some colorimetric, and
many others.
Timing Considerations for the Continuous Mode
Figure 26 illustrates the timing diagram for the continuous mode. The following considerations should be
made when determining the timing for your system.
Analyzer Alarm: An analyzer alarm will immediately close the active sample valve and open the
next sample valve. That sample valve will remain open until the analyzer alarm clears or one
minute elapses. After one minute, the Sequencer will open the next sample valve. The Sequencer
will continue trying to establish flow every minute until the analyzer alarm clears.
Calibration Hold: A calibration hold signal will cause the Sample Sequencer to wait at the active
sample with the valve open. Once the calibration indication clears, the Sequencer will resume
operation entering the Flush state for the sample that was active before the calibration hold
indication was received.
Flush Time: After switching to a successive stream, there will be a settling time for the analyzer to
stabilize to the new sample. The analyzer analog output signal is ignored during this time. The
necessary Flush time depends on the valve manifold design, sample line length from the manifold
arrangement, sample velocity and analyzer response time.
Sample Time: Sample Time is the time period during which a valid analyzer signal is read. Analog
outputs are updated and point identification relay contacts are closed during this time.
The
necessary Sample time should be set long enough to recognize stable readings and to provide
recording devices time to capture the data.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
26
Sample Sequencer® IV
Operation
Figure 26 – Continuous Mode Timing Diagram
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
27
Sample Sequencer® IV
Operation
Batch
The batch mode is for analyzers that take a "bite" or batch of sample and then spend some time analyzing
it before outputting a new fixed analog signal. Typical types are reagent additive, chromatographs, etc.
When these analyzers have completed an analysis there is just one output value until the analyzer's next
cycle is completed. An "End-of-Batch" contact closure signal must be provided at the conclusion of a
batch by the batch type analyzers. The Hach Series 5000 analyzer operates in this manner.
See Appendix B for instructions about connecting a
Hach 5000 analyzer to the Sample Sequencer.
Timing Considerations for the Batch Mode
Figure 27 illustrates the timing diagram for the batch mode. Figure 9 demonstrates how each sample
stream is switched when the analyzer outputs an end-of-batch signal to the Sample Sequencer. If the
Sample Sequencer does not receive an end-of-batch signal, the Sample Sequencer will never
advance to the next sample. Upon receiving an end-of-batch signal from the analyzer, the Sample
Sequencer reads the analyzer value, updates the Track and Hold output, and closes the point
identification relay. Note: The updated Track and Hold output and point identification relay are
for the previous sample point. The open valve number will never match the closed relay point
identification in the batch mode.
The following considerations should be made when determining the timing for your batch mode
system:
End-of-Batch: The Sample Sequencer receives this signal when the analysis is completed
and the analyzer is ready to begin a new analysis.
Analyzer Alarm: The Sample Sequencer receives this signal when the analyzer is in alarm,
such as a loss of sample flow condition. Once the analyzer alarm signal is received the
Sample Sequencer will start the Sample timer and replicate the analyzer alarm for use by
external recorders and control systems.
Sample Time: The Sample time in the batch mode is a timeout timer that operates only when the
analyzer is in alarm. While in alarm, if an End-of-Batch signal is not received before the Sample
time times out, the Sample Sequencer will close the current sample point and proceed with the
next sample point. Since the Analyzer Alarm input may serve to indicate more than a loss of
sample flow, the Sample time should be set to give the analyzer sufficient time to attempt to
complete an analysis and output an End-of-Batch indication.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
28
Sample Sequencer® IV
Operation
Figure 27 – Batch Mode Timing Diagram
Valve Outputs
The Sample Sequencer switches sample streams to an analyzer using the valve outputs on the CPU
board. These outputs provide 12 VDC to control solenoid valves. Sentry offers a manifold sample
valve for use with the Sample Sequencer. With additional hardware, other solenoid valve operating
voltages may be used.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
29
Sample Sequencer® IV
Installation
INSTALLATION
This instrument should be installed by qualified technical personnel to ensure adherence to all
applicable electrical codes.
AVERTISSEMENT
Cet instrument doit être installé par un technicien certifié afin d’assurer que toutes les
normes au code électrique sont respectées.
To ensure the protection provided by this equipment is not impaired, this equipment must
not be installed or used in any manner other than that which is specified in this manual.
AVERTISSEMENT
Afin de s’assurer que la protection prévue par cette pièce d’équipement n’est pas défaillante,
cet équipement ne doit pas être installé ou utilisé de manière autre que spécifié dans le
manuel.
Unpacking
After carefully removing the Sample Sequencer from its shipping carton, inspect it for any damage.
Report damages to the shipper at once.
Mounting
It is recommended the Sample Sequencer be mounted close to the analyzer to permit viewing of both
devices at the same time. This allows the operator to control the Sample Sequencer while viewing the
analyzer.
The Sample Sequencer can mount easily to a panel, wall or pole. When mounting the Sequencer to a
pole, the pole must have an outside diameter of ¾ to 1 ½ inches (19 to 38 mm). All mounting methods
must be capable of supporting at least 24 pounds (10.9 kg).
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
30
Sample Sequencer® IV
Installation
Figure 28 – Enclosure Dimensions
Pole Mounting
1. Obtain two U-bolts and nuts sized for the pole used for installation.
2. Press the back of the mounting brackets against a pole in the proper mounting position, either
vertically or horizontally.
3. Place the curve of one U-bolt behind the pole, and insert the threaded ends of the bolt through
two of the center holes on the mounting bracket. Refer to Figure 29.
4. Secure the two nuts with a wrench, and repeat steps 1-3 on the opposite end of the bracket.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
31
Sample Sequencer® IV
Installation
Figure 29 – Pole Mounting the Sample Sequencer
Wall Mounting
1. Remove the mounting bracket from the back of the Sample Sequencer.
2. Use the bracket as a template or use the dimension information shown in Figure 31 to mark
positions on the wall for four of the eight screw holes – either the vertical or horizontal pair.
Position the device such that a minimum of two (2) mounting screws mount to a stud.
3. Reconnect the mounting bracket to the device.
4. Drill pilot holes in the marked positions for four ¼ -inch (M6) screws.
5. Use four ¼-inch (M6), hex-head or pan-head screws to secure the bracket and device to the
wall.
Figure 30 – Wall Mounting the Sample Sequencer
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
32
Sample Sequencer® IV
Installation
Figure 31 – Mounting Bracket Dimensions
Panel Mount
1. Turn off all power to the Sample Sequencer and disconnect all cables if necessary.
2. Use a Phillips screwdriver to remove the four screws holding the mounting bracket onto the
back of the device. The mounting bracket is not needed for panel mounting.
3. Use a flat-blade screwdriver to loosen the screws of the enclosure lid.
4. Lift the lid off the device and set it aside. Or, store it by screwing the right side of the lid to the
left side of the back enclosure.
5. Use a punch and hammer to remove the three knockouts on the inside rim of the back
enclosure. Position the punch in the center of the knockout for best results.
6. Measure the panel hole using the dimension information given in Figure 32. Mark the panel for
placement of three #10 (M4) pan-head screws.
7. Cut out the panel hole and drill three pilot holes for the screws.
8. Fit the Sample Sequencer into the panel hole and secure it with three #10 (M4) pan-head
screws.
9. Make all wiring connections as described in the following sections.
10. Ensure that the gasket is properly seated in the lid and tighten the screws (10 in-lb, 1.13 N m
maximum). Do not over tighten the screws.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
33
Sample Sequencer® IV
Installation
Figure 32 – Panel Mount Dimensions
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
34
Sample Sequencer® IV
Installation
Figure 33 – Panel Mounting
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
35
Sample Sequencer® IV
Installation
Electrical Connections
Electrical Shock Hazard – All input and output wiring should be installed in such a manner
that access to bare copper is prohibited during normal operation.
AVERTISSEMENT
Avertissement de choc électrique- Tout le filage devra être installé de manière à éviter que
toute partie de fil soit dénudé (sur le cuivre) durant l’utilisation normale
Access holes sized for ½” NPT (15 mm) and ¾” NPT (20 mm) electrical conduit fittings are provided
on the bottom of the enclosure. Sealing type conduit fittings are required to maintain the enclosure’s
rating.
Relay Board Connections
The Sample Sequencer contains three circuit boards stacked in the enclosure. The relay board is
located at the bottom of this circuit board stack. Located on the Relay board are the connections for
point identification contacts, replicated alarm contacts, and the power supply.
Power Connection
The Sample Sequencer contains a universal power supply. The power supply is capable of accepting
an input voltage range of 100-240 VAC and frequency of 50 or 60 Hz. The Sequencer will
automatically operate with power supplied in this range. A power supply lead of 16 AWG (1.3 mm2)
copper wire is recommended.
It is recommended the Sequencer be connected to an electrical outlet or outlet strip with surge
suppression and filtering. A suitable external over-current protection device such as a fuse or circuit
breaker (15 A) is recommended for both the hot (L) and neutral (N) leads. Using this method, the plug
will serve as the disconnect device. The cord and plug shall comply with the following standards:
Location
United States
Canada
European Community
Cord
ANSI/UL 817
CSA C22.2 No. 21 or C22.2 No. 49
IEC 60227 or 60245
Plug
ANSI/UL 498
CSA C22.2 No. 21 and C22.2 No. 42
Appropriate IEC standard
Table 2 – Power Cord Requirements
If desired, the Sample Sequencer may also be hard wired to instrument-quality power using
appropriate certified conduit, fittings, and wiring. A suitable external over-current protection device,
such as a fuse or circuit breaker (15 A), and disconnect device is recommended. The over-current
protection and disconnect devices shall be installed on both the hot (L) and neutral (N) leads. The
disconnect device shall be located near the equipment and marked with appropriate ON(1) OFF(O)
markings as specified by local codes. For fixed wiring methods, the ground conductor shall have green
with a yellow stripe as the insulation. Installation should be performed by qualified personal in
accordance with local codes and procedures.
Figures 34 and 39 show the power connection on the Relay board.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
36
Sample Sequencer® IV
Installation
Point Identification and Alarm Connections
The Sample Sequencer provides individually fused dry contacts to indicate the current active channel.
These connections may be used with a recorder or in situations where analog I/O at the receiving
device is limited. In addition to these point identification contacts, replicated analyzer alarms are also
located on the Relay board. Figures 34 and 39 show typical connections to the Relay board.
Figure 34 – Relay Board
Track and Hold Board Connections
The Track and Hold board provides individually isolated 0/4 – 20 mA signals to recorders or other data
acquisition systems. The Track and Hold board is the middle board in the Sample Sequencer
enclosure. Figures 35 and 38 show typical connections to the Track and Hold board.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
37
Sample Sequencer® IV
Installation
C2
L2
C15
C14
C30
U29
C29
C7
L3
U30
C31
C45
U47
U61
R26
U62
R28
C62
C69
TB1
Q2
VS11
U32
C35
U63
R30
C63
RN16
Q3
VS12
C70
+2-
TB2
+3-
U33
C37
C71
U64
R32
C64
RN17
TB3
Q4
VS13
+4-
U34
C39
U65
R34
C65
RN18
C72
TB4
Q5
VS14
+5-
C73
TB5
U35
C41
R36
C66
RN19
+6-
U28 U20 U12
C52
U59
U58
R13
R23
U66
C28
C44
U43 C60
U57
R12
L9
C43
C51
R35
Q6
VS15
12V
C27
C26
U56
R11
R22
R10
C12
U42 C59
U55
R33
PS8 RN13
L8
C42
C50
U54
R9
R21
R8
C25
C24
VS9
U27
C11
U41 C58
U53
R31
U19
L7
C40
C49
U52
R7
R20
R6
C23
C22
U40 C57
U51
R29
C10
L6
C38
C48
D1
G D S
D1
G D S
D1
G D S
RN15
C21
C20
U50
R5
R19
R4
L5
C36
PS6 RN11
U25
C9
U39 C56
U49
R27
C61
RN14
+1-
C33
U48
R3
R18
R2
Q1
VS10
U31
C47
U46
R25
C19
C18
U38 C55
U37 C54
U45
L4
C34
C46
C53
R1
R17
C17
C16
C32
C8
U17
VS8
U11
U36
C13
C6
PS4 RN9
U23
VS7
U10
RN12
U18
PS7
U26
U44
C5
U15
PS3 RN8
U22
U9
RN10
U16
PS5
U24
U60
U14
U21
VS5
U3
U8
C4
R15
R24
R14
U67
R38
R37
C67
RN20
R16
U68
R40
RN7
PS2
U7
RN5
VS6
U4
D1
G D S
U13
VS3
U6
D1
G D S
RN6
PS1
VS2
D1
G D S
U5
RN4
R39
C68
RN21
C74
Q7
VS16
C75
Q8
VS17
TB6
+7-
TB7
+8-
C76
VS4
VS1
J2
INTERFACE
U2
D1
G D S
2003
C
D1
G D S
RN1 U1
C1
F1
L1
J1
RN3
C3
RN2
TB8
Figure 35 – Track and Hold Board
CPU Board Connections
The CPU board is the top board in the Sample Sequencer enclosure. Multiple functions are contained
on the CPU board. Figure 36 shows the CPU board.
Valve Output Connections
The valve outputs of the Sample Sequencer output 12 VDC, intended for use with the Sentry Manifold
Sample Valve. Other manifold valves may be used, however additional hardware such as interposing
relays may be required. Typical valve connections are shown in Figure 38.
Analog Input Connections
Connect the analog signal from each analyzer to the analog inputs at connector TB1. Terminals
marked 1 are used for analyzer 1. Terminals marked 2 are used for analyzer 2. When using the
Sequencer with only 1 analyzer, connect the analog signal from the analyzer to terminals 1. Typical
analog input connections are shown in Figure 37.
Analyzer Alarm Connections
Analyzer alarm signals are connected at TB2. Terminals marked A1 are used for analyzer 1.
Terminals marked A2 are used for analyzer 2. When using the Sequencer with only 1 analyzer,
connect the alarm signal from the analyzer to terminals A1. Typical analyzer alarm input connections
are shown in Figure 37.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
38
Sample Sequencer® IV
Installation
Calibration Hold Connections
Analyzer calibration signals are connected at TB2. Terminals marked C1 are used for analyzer 1.
Terminals marked C2 are used for analyzer 2. When using the Sequencer with only 1 analyzer,
connect the calibration signal from the analyzer to terminals C1. Typical analyzer calibration input
connections are shown in Figure 37.
End of Batch Connections
Analyzer end of batch signals are connected at TB2. Terminals marked B1 are used for analyzer 1.
Terminals marked B2 are used for analyzer 2. When using the Sequencer with only 1 analyzer,
connect the end of batch signal from the analyzer to terminals B1. Typical analyzer end of batch input
connections are shown in Figure 37.
MODBUS® Connections
MODBUS® connections are made at TB5. The RS-485 network connection is made using a single
twisted pair cable. For more information on using the MODBUS® feature, refer to Appendix D.
ISP
JTAG
C2
INTERFACE
C1
C4
J2
J3
C3
DISPLAY
VS2
L26
L27
L29
R4
RN12
C39 C40
L18
L23
VS4 L19 L20 VS5 L21
L22 VS6
L9
RN16
C43
VS9
VS10
L8
RN15
RN14
FRONT
PANEL
L11
RN17
C45
L10
C44
L12
L15
RN19
C49
L14
L13
RN18
C47
C46
C18
C31
C22
C48
VS12
LED2
RN13 +12V
C50
L16
U9
Q2
Q1
F2
VS11
U5
C42
J6
TB2
+5V
RN8
C32
RN11
C20
C29
C28
RN10
L28
VS3
C30
C27
RN9
L17
C59
C26
C25
C24
C37 C38
C19
Y2
U4
LED1
DISPLAY MOUNTING OUTLINE
C23
C21
Q4
Q3
C53
F3
VS13
F4
VS14
F5
F6
VS15
F7
VS16
F8
VS17
F9
L24
C54
C56 R2
R7 R8
RN29
D2
RN7
C15
C58
U6
C55
R6
R3
C51
R5
C16 C17
RN21
RN20
RN22
RN23 RN24 RN25 RN26 RN27 RN28
R1
L5
C14
C13
C35 C36
L4
RN4
RN6
U8
C52
RN1
RN3
C60
L7
C34
U7
C41
VS1
L3
C11
RN2
VS8
C33
L6
C10
U3
VS7
U2
C57
L25
ASM # 6-03912C REV 0
SERIAL #
SEQUENCER CPU BOARD
RN5
Y1
C8
J5
C7
U1
L2
C9
1
+
C12
D1
F1
L1
J4
BAT1
C6
12V
SENTRY EQUIPMENT CORP. C5J1
C 2003
R9
VS18
L30
L31
L32
TB1
TB3
TB4
TB5
ANALOG ANALOG
INPUT 1 INPUT 2
+ 1 - SHD + 2 -
DISCRETE INPUTS
FOR USE WITH DRY CONTACTS ONLY
B2 COM A2 COM C2 COM
B1 COM A1 COM C1 COM
VALVES
FOR USE WITH 12V DC VALVES
+1-
+2-
+3-
+4-
VALVES
FOR USE WITH 12V DC VALVES
+5-
+6-
+7-
+8-
RS-485
NETWORK
A B COM
Figure 36 – CPU Board
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
39
Sample Sequencer® IV
Installation
Figure 37 – Electrical Wiring Diagram
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
40
Sample Sequencer® IV
Installation
Figure 38 – Electrical Wiring Diagram
Figure 39 – Electrical Wiring Diagram
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
41
Sample Sequencer® IV
Installation
Starting Up The Sample Sequencer
1. Ensure the Sample Sequencer is securely mounted.
2. Wire the Sample Sequencer to the analyzer, sample valves, and other devices as described
in the Installation section of this manual.
3. Apply power.
4. Setup the Sample Sequencer required for your application. This includes setting the sample
timers, operating mode, analog input ranges, and custom sequence. Refer to the Quick Start
Guide and Operation sections of this manual.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
42
Sample Sequencer® IV
Maintenance
MAINTENANCE
Cleaning
The Sample Sequencer is constructed of rugged materials that do not require cleaning. If cleaning
the external surfaces is desired, a soft cloth and water is recommended.
Rechargeable Battery
The CPU board contains a rechargeable lithium ion battery. This battery is not intended for field
replacement. Contact the factory for additional information regarding replacing this battery or
proper disposal of a CPU board containing the lithium ion battery.
Fuses
Potential Fire Hazard – Replace fuses only with the same fuse type and ratings.
AVERTISSEMENT
Risque potentiel d’incendie – Ne remplacer les fusibles qu’avec des fusibles totalement
identiques
Every circuit board in the Sample Sequencer contains re-settable fuses to protect personnel, wiring
and the electronics from excessive current draw. The procedure for resetting one of these fuses after
an overload condition occurs is to remove power and allow the device to cool down.
The dry contacts use a replaceable 2 A fast acting fuse rated for 250 V. These fuses are located on
the relay board.
The power supply uses a 4 A fast acting fuse rated for 250 V.
Replacement fuses for the power supply and dry contacts are available from Sentry. Refer to the
Spare Parts List of this manual for further information. Refer to the Disassembly procedure for
instructions on replacing user replaceable parts such as fuses.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
43
Sample Sequencer® IV
Maintenance
Disassembly
Electrical Shock Hazard – Disconnect all power, including voltages connected to relays,
before servicing or replacing fuses.
AVERTISSEMENT
Risque potentiel choc électrique – Enlever la source d’alimentation incluant les sources de
voltage au relais avant tout entretien ou remplacement des fusibles.
Use the following steps to replace any user serviceable parts:
1. Remove the front cover of the enclosure by removing the four corner screws. Disconnect the
keypad extension cable from the CPU board.
2. Disconnect the interface cable and power supply cable from the CPU board. Remove the
CPU board by removing the four mounting screws.
3. Disconnect the interface cable and power supply cable from the Track and Hold board.
Remove the Track and Hold board by removing the four standoffs located on the board.
4. Replace desire part, such as fuse, circuit board or cable assembly.
5. Replace the Track and Hold board. Secure the board in place using the four standoffs
removed in step 3. Replace the interface cable and power supply cable.
6. Replace the CPU board. Secure the board in place using the four screws removed in step 2.
Replace the interface cable and power supply cable.
7. Replace the front cover. Attached the keypad extension cable to the CPU board. Secure the
front cover in place using the four screws located on the corners of the front cover.
Calibration
Note: The analog circuitry of the Sample Sequencer is calibrated at Sentry prior to shipment.
Re-calibration should be performed by properly trained personnel. Refer to the following procedures
to calibrate the analog circuits of the Sample Sequencer:
Analog Input Calibration
Tools Required: Calibrated milliamp supply.
1. Using the methods discussed in the Operation section of this manual, navigate to the Input Cal
Screen.
2. Remove the existing analog input wires and connect a calibrated milliamp supply to TB1
terminals marked Analog Input 1 of the CPU board.
3. Input 0/4 mA (whichever is desired) on TB1 terminals marked Analog Input 1 of the CPU
board.
4. Use the ▲ and ▼ arrows to change the OFFSET value until RAW COUNT reads between 0
and 500. Note: Raw count should not read 0. When calibrating to 4mA, typical Offset values
range between 1000000 and 1400000.
5. Press the ENTER key.
6. Input 20 mA on TB1 terminals marked Analog Input 1 of the CPU board.
7. Use the ▲ and ▼ arrows to change the GAIN value until RAW COUNT reads between
16776000 and 16777215. Note: Raw count should not read 16777215. When calibrating to
20mA, typical Gain values range between 6500000 and 7500000.
8. Press the ENTER key.
9. Repeat steps 2-7 using channel Analog Input 2 to calibrate channel 2.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
44
Sample Sequencer® IV
Maintenance
Analog Output Calibration
Tools Required: Calibrated multimeter with milliamp measurement capability.
1. Using the methods discussed in the Operation section of this manual, navigate to the Output
Cal Screen.
2. Connect a calibrated multimeter to the + and – terminals of the desired channel on the Track
and Hold board.
3. Use the ▲ and ▼ arrows to select a channel to calibrate. Press the ENTER key to calibrate
the selected channel.
4. Use the ▲ and ▼ arrows to change the OFFSET value until the multimeter reads 4.00 mA (or
0.00 mA) ± 0.01 mA.
5. Press the ENTER key.
6. Use the ▲ and ▼ arrows to change the GAIN value until the multimeter reads 20.00 mA ±
0.01 mA.
7. Press the ENTER key to return to the Channel Selection Screen.
8. Repeat steps 2-7 to calibrate the remaining channels.
9. Press the ESC key to return to the previous screen.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
45
Sample Sequencer®IV
Troubleshooting Guide
TROUBLESHOOTING GUIDE
Symptom
Display doesn’t light
Possible Problem(s)
Power
Incorrect analyzer value
Scale
Calibration
Sequencer doesn’t
recognize end-of-batch
signal from analyzer
Mode
Wiring
Remedy
1. Check the power LEDs on the CPU
board. Are they lit?
NO → proceed to step 2.
YES → contact Service Department.
2. Check the power connections from the
Power Supply to the CPU board. Are
the connections tight?
NO → set cables in place.
YES → proceed to step 3.
3. Check the Power Supply fuse for
continuity. Bad fuse?
YES → replace fuse.
NO → contact Service Department.
1. Verify the scaled output of the analyzer
matches the scaled input of the
Sequencer. Does the problem persist?
YES → proceed to step 2.
2. Perform an analog input calibration.
Does the problem persist?
YES → contact Service Department.
1. Verify the Sequencer is in the Batch
mode. (Refer to Mode in the Operation
section of this manual.) Correct Mode?
NO → change to Batch mode.
YES → proceed to step 2.
2. Verify an End-of-Batch indication is
correctly wired between the analyzer
and Sequencer. Does the problem
persist?
YES → contact Service Department.
Table 3 – Troubleshooting Guide
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
46
Sample Sequencer® IV
Factory Assistance
FACTORY ASSISTANCE
Sentry Equipment Corp has been in the forefront of steam and water sampling equipment technology for
over 70 years. Sentry has always been recognized as the market leader in new sampling technology and
in exceptional customer support.
Please, do not return any equipment before discussing your application problem with a Sentry
representative and obtaining a Return Authorization.
Sentry Equipment Corp has a complete staff of trained Service Department associates to assist
customers with any problems. Call Sentry at 262/567-7256 (Fax: 262/567-4523) to request a service
representative. Please have the unit’s serial number and software date code available when contacting
the Service Department. The software date code can be viewed immediately after power is applied to the
unit.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
47
Sample Sequencer® IV
Spare Parts
SPARE PARTS LIST
Description
Sentry P/N
Sample Sequencer circuit board assemblies:
Relay board
Track and Hold board
CPU board
6-03912A
6-03912B
6-03912G
Sample Sequencer cable assemblies:
Line power to power supply
Power cable
Signal cable
Membrane extension cable
2-05294A
2-05294B
2-05294C
2-05294D
Sample Sequencer Enclosure:
Bracket
Rear Housing
Front Housing Assembly
Front Housing Gasket
Front Housing Screws
4-05588A
2-05321C
6-04165A
4-05588D
4-05588E
Membrane switch assembly
2-05283A
Power Supply
4-05590A
LCD
2-05321B
Fuse, 2 A 250 V Fast Acting (for relay contacts)
Fuse, 4 A 250 V Fast Acting (for power supply)
4-05630A
4-02055F
Terminal Blocks
Relay outputs
Track and Hold outputs
Analog inputs
Discrete inputs
Valve outputs
RS-485 network
4-05631A
4-05631B
4-05631C
4-05631D
4-05631E
4-05631F
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
48
Sample Sequencer® IV
Accessories
ACCESSORIES
Description
Sentry P/N
RS-485 to RS-232 Converter
Sentry Manifold Sample Valve (12 VDC)
Sentry Back Pressure Regulator (20 psi)/Relief Valve (BPR/RV)
Sentry Flow Indicator with control valve (FICV), 580 cc/min
Sentry Flow Indicator with control valve (FICV), 1600 cc/min
Sentry VREL (Variable Pressure Reducing Element valve)
Filter, 40 Micron, 1/4" Tube Fitting
Replacement Filter Element, 40 mesh
6-02627B
7-00927C/D
7-00868A
6-02671D
6-02671B
7-00744A
4-00484K
4-00361L
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
49
Sample Sequencer® IV
Warranty
WARRANTY
Seller warrants products manufactured by it and supplied hereunder to be free from defects in
materials and workmanship for a period of twelve months from date of shipment. If within such period
any such products shall be proved to Seller’s satisfaction to be defective, such products shall be
repaired or replaced at Seller’s option. Seller’s obligation and Buyer’s exclusive remedy hereunder
shall be limited to such repair and replacement and shall be conditioned upon Seller’s receiving
written notice of any alleged defect within 10 days after its discovery and, at Seller’s option, return of
such product to Seller, ex-works Sentry’s factory.
The foregoing warranties are exclusive and in lieu of all other express and implied warranties except in
title, including but not limited to implied warranties of merchantability and fitness for purpose. Seller shall
not be subject to any other obligations or liabilities whatsoever with respect to products manufactured or
furnished by it, or any undertakings, acts or omissions relating thereto.
Warranty Conditions & Limitations
This Warranty shall not apply to any Sentry product which, in the opinion of Sentry Equipment Corp,
has been (a) altered or repaired in a manner affecting the efficiency of performance of the unit or (b)
incorrectly installed or operated or (c) damaged in shipment or (d) damaged by flood or fire or (e) if
the serial number is missing, altered or defaced.
The owner shall be responsible for maintenance of his equipment. Wear or damage caused by lack
of normal maintenance or by misuse of the equipment shall not be considered as defective
workmanship and material.
Sentry reserves the right to make product design changes or improvements without notice and
without imposing any obligation upon itself to install these changes or improvements on its products
previously manufactured.
This warranty is for the sole benefit of the original purchaser and is not transferable unless agreed to
in writing by Sentry Equipment Corp.
Receiving Shipments (including loss or damage by transportation)
It is the customer’s responsibility to check for missing cartons and sign of damage to cartons. If
found, customer should note missing and/or damaged cartons on the delivery receipt and have
delivery receipt signed by the representative of the transportation company. If unpacking discloses
concealed damage from rough handling, the customer should request a concealed damage
inspection from the transportation company.
The Sentry Customer Service Department will aid your organization in any claim proceeding for
shortages or damages in shipment, but it is the receiver’s responsibility to file claim with the carrier for
damage or loss.
Customer Actions For Claims on Products During the Warranty Period
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
50
Sample Sequencer® IV
Warranty
1. Contact the Customer Service Department, Sentry Equipment Corp., Oconomowoc, WI,
Telephone: 262-567-7256, to obtain a Return Material Authorization (RMA) number.
2. You will be sent an “RMA” and a “Decontamination Statement” that is required to be filled
out and returned with the equipment.
3. The following information must appear on the outside of the package:
a.
RMA number marked on outside of box.
b.
Decontamination Statement filled out and attached to outside of box.
4. Return defective equipment FREIGHT PREPAID. Collect shipments will be refused.
5. The factory will not process warranty claims until the customer has properly accomplished
the above items.
6. The Sentry factory may accept the entire claim, a part of the claim or none of the claim if
our inspection of returned parts proves the failure was for reasons other than defective
material or factory workmanship.
Important Notes:
1. Sentry will not be responsible for damage incurred during the return shipment.
2. All returns subject to inspection and a $50.00 evaluation fee.
3. This RMA is not authorization for credit. Credits and/or replacements will be issued upon
evaluation of returned goods.
4. RMA is valid for thirty (30) days from issue date.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
51
Sample Sequencer®IV
Appendix A
APPENDIX A - RECOMMENDED PIPING
It is recommended proper sample conditioning and flow control be maintained to obtain representative
samples. The Electric Power Research Institute (EPRI) guidelines for representative sampling
recommend maintaining constant and continuous sample flow in sample lines. In addition, it is
recommended the sample be cooled to a temperature of 77 F (25 C) prior to analysis. (Note: Some
analyzers may recommend a different sample temperature be used to accelerate the analysis
process.) Appropriate sample coolers or heaters may be required. Consult the pressure and
temperature specifications for your specific analyzer and sampling system. Figure 40 shows the
recommended sample conditioning piping and instrumentation diagram (P & ID) when switching
streams with the Sample Sequencer.
Figure 40 – Recommended Sample Conditioning
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
52
Sample Sequencer® IV
Appendix B
APPENDIX B – USING A SEQUENCER WITH HACH SERIES
5000 ANALYZER
The Sample Sequencer may operate with a Hach Series 5000 analyzer. The Series 5000 instruments
operate in batch-mode style and require the appropriate configuration for use with the Sample
Sequencer. Hach 5000 analyzers have special features that provide a contact closure at the end of
the batch. This is called the “Mark End of Measurement” contact closure. The contact closure causes
the Sample Sequencer to advance to the next sample point. See Batch Mode operation description in
this manual for more information.
Figure 41 illustrates how to connect the Sample Sequencer to a Hach Series 5000 analyzer. The
Hach 5000 analyzer must be configured properly to operate with the Sample Sequencer. The
analyzer must be programmed to use relay #3 and relay #4 for System Alarm and End-ofMeasurement contact closure indications, respectively. Consult the Hach analyzer manual for
information about configuring the System Alarm to relay 3. The following procedure describes how to
configure the End-of-Measurement contact closure on the Hach 5000.
Configuring the Series 5000 (Catalog No. 60000-XX through 60004-XX) for MARK END OF
MEASURE on Relay 4
1. Remove the front cover of the control module. Move the No. 1 DIP switch on SW1 to the ON
position. This will enable the Extended Diagnostic menu. (Switch SW1 is an 8-switch DIP
located near the center of the circuit board.)
2. Return to the front of the control panel and press the TEST key. Using the NEXT key, scroll
through the diagnostic menu to MARK END OF MEAS. Press ENTER and select ENABLE
with the NEXT key. Then, press ENTER. This will activate the MARK END OF MEASURE
contact closure.
3. Turn OFF switch No. 1 on switch bank SW1 of the circuit board. This will return the analyzer
to normal operation but leave the MARK END OF MEASURE function enabled. Note: This
procedure will have to be repeated if a cold start is performed.
4. Reassign the System Alarm to relay 3 by the following: Press the SET UP key. ALARMS will
appear. Press ENTER. Scroll with the NEXT key to RELAY CONFIG and press ENTER.
Scroll to SYS ALRM with the NEXT key and then press ENTER. Scroll with the NEXT key to
RLY 3 and press ENTER.
5. Reassign Alarm 2 to No Relay by the following: Press the SETUP key on the control module.
ALARMS will appear. Press ENTER. Scroll with the NEXT key to RELAY CONFIG and press
ENTER. Scroll to ALARM2 and press ENTER. Scroll with the NEXT key to NO RLY and
press ENTER.
6. Assign relay 4 to Mark End of Measure by the following: Press the SETUP key on the control
module. ALARMS will appear. Press ENTER. Scroll with the NEXT key to RELAY CONFIG
and press ENTER. Scroll to MARK END OF MEAS and press ENTER. Note: MARK END
OF MEAS will not appear unless it has been properly activated per the above steps.
Scroll with the NEXT key to RLY 4 and press ENTER. This will assign the alarm 4 relay to
mark the end of measurement with a contact closure. The contact closure makes the Sample
Sequencer advance to the next sample point.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
53
Sample Sequencer® IV
Appendix B
7. Verify the Sequencer is operating in the batch mode.
Older Hach Series 5000 Analyzers
If you have an analyzer with a catalog model number 5000X-XX, contact Hach Service Department at
1-800-227-4224 regarding using the Sample Sequencer with your analyzer. Updated software
EEPROM may be required to operate correctly with the Sample Sequencer.
If you have an analyzer with a catalog model number 5500X-XX, configure your unit as follows:
Model 5500X-XX Configuration
1. The wiring is different for the model 5500X-XX than shown in Figure C1. The changes are: (a)
Connect the Sample Sequencer End-of-Batch terminals to the Hach terminals labeled NO and
COM on ALRM 2. (b) Connect the Sample Sequencer Analyzer Alarm terminals to the Hach
terminals labeled NO and COM on SYS ALRM.
2. Open the control module door and loosen the screw to allow access to the rear of the display
board. Remove the back cover.
3. Turn DIP switch #1 (S1) ON by pressing it. This puts the analyzer in the Extended Diagnostic
menu.
4. Press the TEST key.
5. Using the NEXT key, scroll through the diagnostic menu to "MARK END of MEAS", then,
press ENTER. This will cause the number 2 alarm contact to close at the end of each analysis
cycle. The Series 5000 now provides a contact closure to advance the Sample Sequencer to
the next sample point each time alarm 2 contact closes. The alarm 2 contact can no longer be
used for an alarm signal.
6. After pressing ENTER, return the #1 DIP switch (S1) to the OFF position. Replace the cover
and close the meter panel. This returns the analyzer to normal operation. Note: If a "cold
start" is initiated or the backup battery fails during a power failure, you must reset the
MARK END of MEAS as outlined above.
7. Verify the Sequencer is operating in the batch mode.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
54
Sample Sequencer® IV
Appendix B
Recommended Sequencer Timer Settings
It is recommended the following times be programmed into the Sample Sequencer when using the
Hach Series 5000 analyzers:
Sample Time.......... 20:00 minutes
Reference the Operation section of this manual for instructions on programming the Sample Timers.
Figure 41 – Wiring to a Hach Series 5000 Analyzer
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
55
Sample Sequencer® IV
Appendix C
APPENDIX C – LCD AVAILABLE CHARACTERS
Figure 42 shows the characters that can be displayed in the Sample Sequencer.
Figure 42 – LCD Character Chart
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
56
Sample Sequencer® IV
Appendix C
When editing fields such as Sample Names and Analyzer descriptions, the ▲ and ▼ keys
are used to scroll through the list of characters. Figure 43 shows how scrolling through the
characters is handled.
Pressing
the ▲ key
Pressing
the ▼ key
Figure 43 – Scrolling Through the Character List
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
57
Sample Sequencer® IV
Appendix D
APPENDIX D – MODBUS® Registers
Multiple Sequencers can communicate with a PC or control system via MODBUS® over a
half duplex (2-wire), RS-485 network.
Figure 44 shows the MODBUS setup screen on the Sequencer.
MO D B U S
S E T U P
A D D R E S S
1
B A U D
1 9 . 2
k b p s
Figure 44 – MODBUS Setup
Each device on the network must be assigned a unique address. The address for each
Sequencer can be in the range of 1 to 247. Use the ▲ and ▼ keys to change the address
setting. To change the address setting in multiples of 25, hold the MENU key while using
the ▲ and ▼ keys. Use the ENTER key to toggle between the address setting and the
BAUD setting.
The Sequencer is capable of operating using a BAUD rate of 19.2 kbps or 9600 bps. To
change the BAUD rate use the ▲ and ▼ keys.
The communications port uses 8 data bits, 1 stop bit, and no parity. The MODBUS® mode
is RTU. These parameters cannot be changed.
The complete listing of MODBUS® registers are listed in Table 4.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
58
Sample Sequencer® IV
Modbus register
400001
400002
400003
400004
400005
400006
400007
400008
400009
400010
400011
400012
400013
400014
400015
400016
400017
400018
400019
400020
400021
400022
400023
400024
400025
400026
400027
400028
Appendix D
Sequencer Register
SEQ1 Status
Minutes Remaining
Seconds Remaining
Open Valve
Analyzer value
Analyzer decimal value * 100
SEQ2 Status
Minutes
Seconds
Open Valve
Analyzer value
Analyzer decimal value * 100
Flush Min Valve 1
Flush Min Valve 2
Flush Min Valve 3
Flush Min Valve 4
Flush Min Valve 5
Flush Min Valve 6
Flush Min Valve 7
Flush Min Valve 8
Flush Sec Valve 1
Flush Sec Valve 2
Flush Sec Valve 3
Flush Sec Valve 4
Flush Sec Valve 5
Flush Sec Valve 6
Flush Sec Valve 7
Flush Sec Valve 8
Modbus register
400029
400030
400031
400032
400033
400034
400035
400036
400037
400038
400039
400040
400041
400042
400043
400044
400045
400046
400047
400048
400049
400050
400051
400052
400053
400054
400055
400056
Sequencer Register
Sample Min Valve 1
Sample Min Valve 2
Sample Min Valve 3
Sample Min Valve 4
Sample Min Valve 5
Sample Min Valve 6
Sample Min Valve 7
Sample Min Valve 8
Sample Sec Valve 1
Sample Sec Valve 2
Sample Sec Valve 3
Sample Sec Valve 4
Sample Sec Valve 5
Sample Sec Valve 6
Sample Sec Valve 7
Sample Sec Valve 8
Sample/Bypass Valve 1
Sample/Bypass Valve 2
Sample/Bypass Valve 3
Sample/Bypass Valve 4
Sample/Bypass Valve 5
Sample/Bypass Valve 6
Sample/Bypass Valve 7
Sample/Bypass Valve 8
Start SEQ1
Stop SEQ1
Start SEQ2
Stop SEQ2
Table 4 – MODBUS® Registers
NOTES:
Registers 400001 through 400006 provide status information corresponding to SEQ1.
Registers 400007 through 400012 provide status information corresponding to SEQ2.
The values located in registers 400001 and 400007 represent the following states:
0 = OFF, 1 = FLUSH, 2 = SAMPLE, 3 = ALARM, 4 = CALIBRATION, 5 = BATCH
The analyzer value is stored in consecutive registers. The first register stores the whole number. The
second register stores the decimal value * 100.
The values located in the Sample/Bypass registers represent the following states:
0 = Bypass, 1 = Sample
Registers 400053 through 400056 should be implemented as momentary pushbuttons. To start or stop
SEQ1 or SEQ2, a 1 should be placed in the corresponding register. After the Sequencer has changed to the
appropriate state, a 0 should be placed in the corresponding register.
SENTRY EQUIPMENT CORP • 262.567.7256 • Fax: 262.567.4523 • sales@sentry-equip.com • www.sentry-equip. com
59