Sample dilution system version DIL-U consisting of dilution probe SP2000H/DIL-U/P Special Features

Transcription

Sample dilution system version DIL-U consisting of dilution probe SP2000H/DIL-U/P Special Features
Embracing Challenge
Sample dilution system version DIL-U
consisting of dilution probe SP2000H/DIL-U/P
and dilution pump unit DIL-U/A (N)
DIL-U/A(N)
Special Features
ƒƒ Lowest dilution ratio 1:1
ƒƒ Ambient air as dilution gas possible
ƒƒ Low dilution gas consumption
Application
The electrically heated M&C dilution system
DIL-U is applicable for such processes, where
the measuring procedure or the handling of
the process gas requires the dilution of the
sample respectively the component to be
measured in a range of 100:1 up to 1:1.
ƒƒ No influence of barometric pressure1
Typical applications:
ƒƒ Compact design
•
toxic gases above TLV value
ƒƒ Gas sample probe SP2000H/...
•
explosive gas mixture above LEL
ƒƒ Incorporated pre heater for dilution gas
•
adapting the concentration of the sample components to the measuring range
of the analyser
ƒƒ Test gas connection at the probe inlet
ƒƒ Completely heated up to 180°C
•
measuring possibility at low sample flow
rates from 1Nl/hr
varying process pressure must be taken into
consideration
•
moisture measurement in flue gas with
IR-analyser
1
The M&C dilution probe (part of this system)
is based on the well tested modular designed
M&C gas sample probe SP2000H. This enables
the user to adapt the probe to nearly every
application, using for example special filter
techniques or special materials.
For test procedure with zero and span gas a
3 way ball valve and a flowmeter is installed.
Version DIL-U/A is suitable for ambient air as
a dilution gas and version DIL-U/N for bottle
gas. Bottle gas requires a pressure regulator,
a pressure gauge and a flowmeter which can
be installed optional in the pump unit.
If the dew point of the diluted sample is
higher then ambient temperature an optional integrated M&C cooler, ECP1000G/SR25,
reduces it to +5 °C.
High dilution ratios cause a very low sample flow. Therefore it is necessary to use an
additional sample bypass 2 to reduce the
response time. This will be done with the
option DIL-U/B with an additional pump and
flowmeter and the cooler as an option.
The heated M&C sample line(s) 3/4-N/M/H
between the probe and the pump unit can
be controlled via an temperature controller
703, built into the pump unit.
The design of the dilution system DIL-U guarantees easy maintenance.
Description
The M&C dilution system DIL-U consists of the
dilution probe SP2000H/DIL-U/P and the dilution pump unit DIL-U (A/N).
The dilution unit, including the critical orifices, is integrated in the sample probe outlet,
heated up with the probe to a stable temperature. The incorporated pre heater heats
up the dilution gas to the probe temperature.
Both steps avoid that the temperature of
sample gas decreases under the dew point.
The test gases for the system calibration enter
the sample probe via an integrated valve.
A vacuum pump integrated in the pump unit
DIL-U sucks the required vacuum, which is
necessary for the function of the critical orifices. A pressure gauge indicates the function of
the dilution system. The flowmeters indicate
the analyser and bypass flow 1. A flow alarm
is available as well.
Technical specifications and illustrations are without
obligation, subject to modifications. 12.99/06.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
5.3
Dimensions
340
345
207
110
To >+30 C
Tsoll
0-180 C
Tu <-30 C
260
12 34 5 6 7 8 9
NL L'
Power
Low temp. alarm
230V, 50Hz
(115V, 60Hz)
Gas connections DN4/6
Dilution gas
Sample
45
Test gas
Dilution gas
Sample
84TE
420
360
Vakuum
Vacuum
Vakuum Ein
Vacuum On
Testgas
Test gas
Bypass2 Pump
On
FM1
S2
-0,2
0,0
-1,0
Druckeinstellung
Pressure set
Bypass1
Filter
FPF-0,1
FM2
M&C
S1
Sensor
FA-1,bi
Y1
Op
Durchfluss
Flow rate
f
Au
en
M1
FM4 Endgas
Span
FM5
6HE
Messgas
Sample
FM1
Endgas
Span
Zu
Closed
-0,8
-0,6
-0,4
Verdünnungsgasdruck
Dilution gas pressure
FM1
Verdünnungsgas
Dilution gas
LA1
Kondensat/Bypass1
Condensate/Bypass1
FM2
Fi1
Kondensat/Messgas
Condensate/Sample
Auf
Open
3,0
1,0
5,0
6,0
0,0
Beh.Leitung
FM3
M2
Zu/Closed
4,0
2,0
PR1
Y2
P2
P1
Dilution unit
DIL-U
Gas connections G1/4"-i
Dimensions in mm
Technical Data
Gas sample dilution system version DIL-U
Dilution gas sample probe SP2000H/DIL-U/P Part No.
Dilution pump unit, standard DIL-U/A Part No.
Dilution pump unit, standard DIL-U/A Part No.
Option: Gas cooler ECP1000G/SR25 Part No.
Option: bypass DIL-U/B Part No.
Insitu sample tube or prefilter
Dilution ratio
Sample flow rate
Dilution gas flow rate
Dilution gas pressure at DIL-U/N
Material of sample contacting parts
Dimensions from dilution pump unit DIL-U
Gas connections
Power supply
Weight
Part-No.
Type
SP2000H/DIL-U/P
20S4950(a) *see to dilution ratio
DIL-U/A
20S4955(a)
DIL-U/N
20S4960(a)
ECP1000G/SR25
20S4965(a)
DIL-U/B
20S4970(a)
optional available, see data sheet 2-1.1b; 2-1.5.1; 2-1.6.1/2; 2-1.9.5
1:1 up to 100:1; adjustable via corresponding orifices
* please specify the required dilution ratio with the order
from approx. 1 Nl/hr up to 50 Nl/hr
from approx. 30 Nl/hr up to 100 Nl/hr
>3 bar g
stainless steel 316Ti, 1.4404, glass, FPM, PTFE
19"-rack, 6U, 84HP, depth = 460 mm
thread G1/4" i
230V 50Hz optional 115V 60Hz (a)
dilution gas sample probe SP2000H/DIL-U/P: 15 kg, dilution pump unit DIL-U: 15 kg
Further technical data see data sheet SP2000, 2-1.1b and 2-1.1.7.