FlexiPump In Service

Transcription

FlexiPump In Service
FlexiPump™ Independent Flushing System
In-service Guide
1
Set up – Nest Accessory
The Nest Accessory can be set up on any counter near a sink. Simply
insert the FlexiPump system into the nest with tubes facing the labeled
front of the accessory. Plug the unit in.
2
Set up – Sink Rail Accessory
The Sink Rail Accessory is sold separately. Adhere it to the inside wall of a clean sink
at the top edge. Lock the back of the FlexiPump system into the Sink Rail track.
3
Set up – on a wall
The Nest Accessory can mount on a wall using the through holes in the back.
Simply hang the nest on a wall near the sink and insert the FlexiPump system.
(Hardware provided by facility)
4
Adding pumps – Pure Station™
Medical Devices Pre-Cleaning System
The FlexiPump system is designed to fit into the corner mount of the
sink, or to mount on the ridges along the side edges of the sink.
5
FlexiPump system cycles
Cycles
•
•
•
6
Forward Flush is used for flushing channels.
There are three options: 30 seconds, 60
seconds, 90 seconds.
Impact Cycle is used ONLY for devices that are
clogged. Tube set must be decontaminated
after each impact cycle.
Speed Indicator
– High speed is for bigger channels, 5mm
or larger.
– Medium speed is for 4mm channels.
– Low speed is for channels smaller than
4mm and for delicate instruments.
Key pad functions
•
•
•
•
7
Impact Cycle - press for clogged devices.
Press button once for two minutes, twice
for four minutes , or three times for six
minutes. NOTE: This cycle is only for
devices with impacted materials within
the channels. The motor goes forward
then reverses, which creates pressure and
suction.
STOP CYCLE - cancels all cycles
Forward Flush - Push button once for 30
seconds, twice for 60 seconds, three
times for 90 seconds.
SPEED Control- Push button once for low
speed (500 ml per minute), twice for
medium speed (750 ml per minute), three
times for high speed (1,000 ml per
minute). WARNING: Do not run the high
speed for delicate/small devices.
Devices with luer locks
•
•
8
Twist the collar onto the device with a
luer to lock it.
Flush the device.
Devices with ports
•
•
•
9
Pull the collar back.
Insert the tip into the device.
Flush for 30 seconds.
Devices without luer locks
Use a cup adapter on these types of devices.
10
Maintenance – tube set
Decontaminate the tube set daily.
•
•
•
•
•
11
Step 1. Use an environmental disinfectant, either ready-to-use or concentrate.
Step 2. Put the drop-down tube into diluted solution.
Step 3. Push the flush button so the entire tube set fills with disinfectant.
Step 4. Allow the disinfectant to sit in the tube for the full required contact
time.
Step 5. When full contact time has expired, push the 30-second flush button
to purge disinfectant out of the tube set.
Maintenance – replacing tube set
It’s important to replace the tube set every month. Here’s why:
• Infection control - hard water minerals can build up on the inside of the tube set.
• Reduced flow rate - the squeeze tube will loose its elasticity with time and use.
• Reduced pump life - the squeeze tube is lubricated. When the lubricant wears
away, the pump will have to work harder, which can reduce its useful life.
Step 1. Press 2 side tabs on cap.
12
Step 2. Remove the cap.
Maintenance – replacing tube set
Step 3. Unthread tubing.
13
Step 4. Re-thread new tubing into
rotor and cap.
Maintenance – replacing tube set
Step 5. Put cap onto base.
14
Step 6. Make sure it snaps down.
Support
877-718-6868
www.pure-processing.com
Sales@pure-processing.com
15