FlexiPump In Service
Transcription
FlexiPump In Service
FlexiPump™ Independent Flushing System In-service Guide 1 Set up – Nest Accessory The Nest Accessory can be set up on any counter near a sink. Simply insert the FlexiPump system into the nest with tubes facing the labeled front of the accessory. Plug the unit in. 2 Set up – Sink Rail Accessory The Sink Rail Accessory is sold separately. Adhere it to the inside wall of a clean sink at the top edge. Lock the back of the FlexiPump system into the Sink Rail track. 3 Set up – on a wall The Nest Accessory can mount on a wall using the through holes in the back. Simply hang the nest on a wall near the sink and insert the FlexiPump system. (Hardware provided by facility) 4 Adding pumps – Pure Station™ Medical Devices Pre-Cleaning System The FlexiPump system is designed to fit into the corner mount of the sink, or to mount on the ridges along the side edges of the sink. 5 FlexiPump system cycles Cycles • • • 6 Forward Flush is used for flushing channels. There are three options: 30 seconds, 60 seconds, 90 seconds. Impact Cycle is used ONLY for devices that are clogged. Tube set must be decontaminated after each impact cycle. Speed Indicator – High speed is for bigger channels, 5mm or larger. – Medium speed is for 4mm channels. – Low speed is for channels smaller than 4mm and for delicate instruments. Key pad functions • • • • 7 Impact Cycle - press for clogged devices. Press button once for two minutes, twice for four minutes , or three times for six minutes. NOTE: This cycle is only for devices with impacted materials within the channels. The motor goes forward then reverses, which creates pressure and suction. STOP CYCLE - cancels all cycles Forward Flush - Push button once for 30 seconds, twice for 60 seconds, three times for 90 seconds. SPEED Control- Push button once for low speed (500 ml per minute), twice for medium speed (750 ml per minute), three times for high speed (1,000 ml per minute). WARNING: Do not run the high speed for delicate/small devices. Devices with luer locks • • 8 Twist the collar onto the device with a luer to lock it. Flush the device. Devices with ports • • • 9 Pull the collar back. Insert the tip into the device. Flush for 30 seconds. Devices without luer locks Use a cup adapter on these types of devices. 10 Maintenance – tube set Decontaminate the tube set daily. • • • • • 11 Step 1. Use an environmental disinfectant, either ready-to-use or concentrate. Step 2. Put the drop-down tube into diluted solution. Step 3. Push the flush button so the entire tube set fills with disinfectant. Step 4. Allow the disinfectant to sit in the tube for the full required contact time. Step 5. When full contact time has expired, push the 30-second flush button to purge disinfectant out of the tube set. Maintenance – replacing tube set It’s important to replace the tube set every month. Here’s why: • Infection control - hard water minerals can build up on the inside of the tube set. • Reduced flow rate - the squeeze tube will loose its elasticity with time and use. • Reduced pump life - the squeeze tube is lubricated. When the lubricant wears away, the pump will have to work harder, which can reduce its useful life. Step 1. Press 2 side tabs on cap. 12 Step 2. Remove the cap. Maintenance – replacing tube set Step 3. Unthread tubing. 13 Step 4. Re-thread new tubing into rotor and cap. Maintenance – replacing tube set Step 5. Put cap onto base. 14 Step 6. Make sure it snaps down. Support 877-718-6868 www.pure-processing.com Sales@pure-processing.com 15