TS-100/TS-110/TS-130 Microprocessor pH/ORP/ION Portable Meter
Transcription
TS-100/TS-110/TS-130 Microprocessor pH/ORP/ION Portable Meter
TS-100/TS-110/TS-130 Microprocessor pH/ORP/ION Portable Meter Operation Manual CONTENTS 1. Specifications .............................................................................................................1 2. Panel Description .....................................................................................................2 2.1 Front Panel ...........................................................................................................2 2.2 Sockets...................................................................................................................2 2.3 Keypad ..................................................................................................................3 3. Operation ....................................................................................................................4 3.1 Measurement ........................................................................................................4 3.2 Block diagram of parameter settings .................................................................5 3.3 Parameter settings ...............................................................................................6 3.4 Auto/Manual data storage (TS-110/130 only)..................................................14 3.5 Data transmitting and display (TS-110/130 only) ...........................................16 3.6 Delete all of the data (TS-110/130 only) ...........................................................20 4. Calibration ................................................................................................................21 4.1 pH calibration.....................................................................................................22 4.2 Operation of pH calibration..............................................................................23 4.3 Auto calibration for preset Buffers (CT / CN) ................................................24 4.4 Any buffer manual calibration (CA)................................................................26 4.5 ION concentration calibration (TS-130 only)..................................................27 4.6 ION concentration calibration (CC)(TS-130 only) .........................................28 5. Error messages ........................................................................................................30 6. Maintenance .............................................................................................................31 6.1 Battery replacement...........................................................................................31 6.2 Electrode clean and maintenance .....................................................................31 7. Modbus protocol and instructions (TS-110/130 only)....................................32 7.1 Function code .....................................................................................................32 7.2 MODBUS name and address table...................................................................33 8. Appendix ...................................................................................................................34 Note: The instrument software and hardware maybe modified to make progress of the better operation and better accuracy. Thus, this may result in difference between this manual and your instrument. We can not guarantee there are no errors in the manual. Therefore, we are sure that you understand that we cannot accept any legal claims resulting from the data, figures or descriptions. 1. Specifications Model Measuring Unit Measuring Range TS-100 TS-110 pH/ORP/TEMP. -2.00~16.00pH -2.00~16.00pH/-2.000~16.000pH ORP TEMP -1999~1999mV -1999~1999mV/-1999.9~1999.9mV pH ORP -10.0~110.0°C −−−− 0.01 pH 1mV 0.1°C TEMP pH ORP ±0.01(±1Digit) ±0.1%(±1Digit) ±0.01/±0.005(±1Digit) ±0.05%(±1Digit) ±0.2°C(±1Digit) −−−− ION Auto-Read(HOLD) Calibration RS232 Interface 1 / 0.1 / 0.01 / 0.001 ppm(mg/l) −−−− TEMP Temperature Compensation 0.001~1999.9ppm(mg/l) 0.01 pH/0.001 pH 1mV/0.1mV ION Accuracy pH/ORP/TEMP./ION pH ION Resolution TS-130 ±0.5%(±1Digit) Yes Automatic with PT1000/NTC30K, auto recognition Manual pH: Tech./ NIST/ ANY/ Define Buffer mode, up to 3 points ION: Ion concentration calibration, up to 3 points (TS-130 only) −−−− One way transmission: To the computer/ printer (ASCII) Two way transmission: Modbus (RTU or ASCII) Data storage Auto power off 500 data sets with time and date stamp Continuously auto data storage & logging /Manual Yes(Programmable time duration for auto-power off) −−−− Ambient Temperature 0~50°C Storage Temperature Display Power Supply Battery lifetime Protection Certification Dimension Weight -20~70°C LCD Display 4 x AA Batteries >1000 hours IP65 CE 184 mm x 98mm x 38mm (L x W x H) 0.35Kg Note: The specifications and appearance of the instrument are subject to change without notice. 1 2. Panel Description 2.1 Front Panel Measurement Setting mode Mode Auto Read Drift Control Measurement Calibration mode Auto Temp. Unit Compensation Data Storage mode Time/Date Low Battery Warning Baud Rate Data display & Transmitting Data Storage Position No. Data Storage ID number 2.2 Sockets RS232 Interface (Optional: 8-30 RS232 Cable) ATC probe Reference sensor/ ATC probe socket Sensor clip holder pH/ORP senor BNC socket 2 2.3 Keypad Power ON/OFF. Press POWER key to turn on the instrument, press it again to turn it off. Press POWER for a while when the power is already on, then the display will show the model type and software version. The user may set up auto turn-off timer. Under a certain of time not touch any keypad, the instrument will auto turn off. Switch to pH, ORP or ION (TS-130 only) mode. Press “Mode” key to back to measuring mode at any time except under calibration mode. Enter calibration mode / Select buffer type Activate/deactivate Auto Read control. Hold the measurement value, press it for activate the function, and press again to cancel it. Increase values / Scroll upwards or leftwards. Decrease values / Scroll downwards or rightwards. Confirm inputs / Execute / Transmitting data manually via RS-232 in measuring mode. Activate manual data recall and mode selection / Deactivate auto data recall. (TS-110/130 only) Activate manual data storage / Deactivate auto data storage Note: The icon of the manual is subject to change without notice. 3 3. Operation 3.1 Measurement Press POWER key to turn on the instrument and it will directly enter the last measurement mode to start measurement. Press POWER key for a while when the power is already on, then the display will show the model type and software version. The user may set up auto turn-off timer. Under a certain of not touch any keypad, the instrument will auto turn off. 3.1.1 Mode Press MODE key to switch to pH, ORP, ION (TS-130 only,measuring mode. (ION mode has 2 units: ppm and mg/l for selection.) Press ENTER key +MODE key to switch on the decimal of resolution. (TS-110/130 only)。 3.1.2 Temperature compensation 1. ATC:Automatically recognition of temperature probe NTC30K or PT-1K and auto temperature measurement and compensation. Under setting mode, the temperature can be adjusted ±5˚C for temperature correction. 2. MTC:Entering into manual temperature compensation mode when there is no ATC probe plugged in. Press upward or downward key to modify the temperature value. Press upward and downward key simultaneously to restore default value 25˚C. 3.1.3 Auto Read Drift Control Under measurement mode, press AUTO READ key to hold the present reading(HOLD), and press ENTER key to confirm the Auto-Read function and to activate new measurement. Wait until the measurement is stable, the instrument hold the value automatically(HOLD) to finish the procedure. To cancel the hold value, press AUTO READ key or MODE key to relieve the Auto-Read hold value. Activate Auto-Read Press AUTO READ key Put the electrode into sample solution, press ENTER for starting reading. The Auto Read symbol twinkles. Wait until the measurement is stable. If the measurement can not be stable in a certain time, the display shows E-03 error message, and the Auto Read symbol disappears. Press ENTER key again to re-start the Auto-Read。 Wait until the measurement becomes stable, the Auto Read symbol stop twinkling. The display shows the pH measurement and holds the value.(HOLD) Relieve Auto-Read Hold Press AUTO READ key or MODE key Note: ORP mode does not apply Auto-Read function. 4 3.2 Block diagram of parameter settings Enter Parameter Setting mode Auto_Power Setting OFF AUTO (On) (TS-110/130 only) SEC (Digital Filter) Set measuement average times ATC Temp. Compansation Data_Clear (Rewritable) Set temperature AUTO (On) OFF Preset: Auto Timer setting Upward and downward key : ENTER key Select RTU or ASCII Select Parity Set baud rate Preset: Auto : Computer/ Modbus Comunication Note: Under any setup mode, press MODE to back to measurement mode at any time. Factory default value: Auto Power-Off: AUTO Temperature compensation: ATC, 0.0°C Data Storage Rewriteable Function: AUTO Communication Settings: ASCII, non, 4800 System Time Settings: 2011, 01:01, 00:06 5 System time setting Set year Set date Set time 3.3 Parameter settings 3.3.1 Entering setting of parameter While the instrument’s power is off, press MODE key and hold the pressing continuously, and press POWER key until the display shows model of the instrument. Release the POWER key and wait until the menu display enter into parameter settings mode, and then release MODE key. Press upward or downward key to select setting menu. To exit the parameter settings mode, press MODE key to back to measurement mode. 3.3.2 Auto power-off settings Under data storage rewritable function menu display, users can set whether to enable the function. The auto power off timer setting range: 10 minutes minimum, 23 hours and 59 minutes maximum. Preset: 10 mins Press ENTER key to enter. Press upward or downward key to select auto power-off, or to turn off this function. Press ENTER key for confirmation. When the value twinkles, press upward or downward key to modify the auto power-off timer. Press ENTER key for confirmation. Enter “Digital Filter” settings 6 3.3.3 Digital filter settings (Sample average of measurements) Enter the setup of Digital filter. You may select the number of sample to be averaged each time to become a reading which is gradually counted in order to increase the stability of measurement. Press ENTER key to enter When the value twinkles, press upward or downward key to modify times of digital filter. Press ENTER key for confirmation. Enter “Auto Temperature Compensation Correction” settings 7 3.3.4 Auto temperature compensation correction settings Enter auto temperature compensation settings. If the ATC probe is connected, it is allowed to correct the temperature value by pressing upward or downward key of ±5°C. Or press upward and downward key simultaneously to zero the temperature modification value. Note: If not ATC mode, it can not enter into auto temperature compensation correction settings. Press ENTER key to enter When the value twinkles, press upward or downward key to modify the temperature value difference. The corrected temperature will show in next menu display. Press ENTER key for confirmation. Back to “Auto Power-Off” settings (TS-100 only) Press MODE key to back to the measurement mode Enter “Data Storage Rewritable Function” settings (TS-110/130 only) 8 3.3.5 Data storage rewritable function (TS-110/130 only) Under data storage rewritable function menu display, users can set whether to enable the function. AUTO means to turn on the function. When data storage is full the data logger will overwrite the new data on the previous one. It is also allowed to select OFF to turn off the function. Press ENTER key to enter. Press upward or downward to select auto data storage rewritable function or to turn off the function. Press ENTER key for confirmation Enter “Communication” settings 9 3.3.6 RS-232 communication settings (TS-110/130 only) Users may according to their communication needs with following table to set transfer mode, (ASCII, RTU), parity (non, odd, even) and transmission baud rate (1200, 2400, 4800, 9600). 1. Communication setting format Mode Single way Data transfer to computer or printer ASCII Transmission 1200, 2400, Baud Rate 4800, 9600 Default data bit 8 Parity non Default stop bit 2 The default address None RS232 Interface Dual way Modbus Protocol RTU ASCII 2400, 4800, 9600 8 7 non, odd, even non odd, even 2 1 1 Note: 1.Because of different transmission formats, two kinds of communication mode can not be operated simultaneously. 2. This Modbus protocol does not include the selection of baud rate 1200. 3. Data bits, stop bits and the instrument’s address are preset value which can not be modified. 10 2. Communication setting procedure Press ENTER key to enter Press upward or downward key to select ASCII or RTU transmission mode. Press ENTER key for confirmation Press upward or downward key to select odd, even or none parity, press ENTER key for confirmation. Press upward or downward key to select baud rate, press ENTER key for confirmation. Enter “System Time” settings 11 3.3.7 System time settings (TS-110/130 only) Users may set the year, month, date, hour, and minute. When replacing the batteries, it is necessary to enter to the setting to update the system time. Press ENTER key to enter. When the value twinkles, press upward or downward key to modify hour. Press ENTER key to confirm it. When the value twinkles, press upward or downward key to modify year. Press ENTER key to confirm it. When the value twinkles, press upward or downward key to modify minute. Press ENTER key to confirm it. When the value twinkles, press upward or downward key to modify month. Press ENTER key to confirm it. When the value twinkles, press upward or downward key to modify date. Press ENTER key to confirm it. 12 Enter “Auto Power-Off” settings Press MODE key to back to measurement mode 3.4 Auto/Manual data storage (TS-110/130 only) Under data storage mode, users may store pH, ORP and ION (TS-130 only) measurement with No. and ID two kinds of classification. The “No.” represents the order of storage sequence. When data is stored to the 500th data, the display will shows “Full” to represent there is no more space for data logger memory. While ID number 1~500 is for grouping the samples which the user requires. The Auto data storage can be set according to the time interval setting, the minimum interval is 5 seconds. If turn on the Data Rewritable Function, when the data memory is full the data logger will overwrite the new data. 3.4.1 Auto data storage Under measurement mode, press and hold ENTER key and press STORE key simultaneously to enter auto storage settings. When auto storage is not activated, press MODE key to exit the menu. Display the starting storage number Press ENTER key to enter Press upward or downward key to modify the grouping ID. Press ENTER key to confirm it. When the value twinkles, press upward or downward key to set the minute interval of auto storage. Press ENTER key to confirm it. When the value twinkles, press upward or downward key to set the second interval of auto storage. Press ENTER key to confirm it. “Auto Store"twinkling represents the auto storage is activating. Press ENTER + STORE to turn off the auto storage function, and back to the measurement mode. 13 3.4.2 Manual data storage 3.4.3 Overflow of data storage Under measurement mode, press STORE key to enter manual data storage mode. Press MODE key to exit at any time. When the all the memory is occupied and the data can not be stored any more, the display twinkles the text for reminder for 5 seconds and back to the measurement mode. Display the starting storage number Press ENTER key to enter Users may delete all of the data according to the ch 3.6 instruction.(When auto rewritable function is turned on, there is no this data full display reminder.) Press upward or downward key to modify the grouping ID. Press ENTER key to confirm it. Back to measurement mode 14 3.5 Data transmitting and display (TS-110/130 only) Note: The output of the instrument is non-insulated signal. It may impact the transmitting measurement value if the instrument connects with other earth-related device. 3.5.1 Time interval auto transmitting 3.5.2 Data display & manual transmitting (Transmit the real-time measurement to a Under measurement mode, press computer or a printer through RS-232) RECALL key to roll up the data Under measurement mode, press and hold display & manual transmitting menu. ENTER key and then press RECALL key There are four kinds of operation mode. to enter auto transmitting menu display. Press ENTER key to enter the menu. After finishing setting, the data is transmitted through RS-232 interface. 1.Display the stored data When the value twinkles, press upward or downward key to set the minute interval of auto transmitting. Press ENTER key to confirm it. Press ENTER key to enter recent stored data menu display. When the value twinkles, press upward or downward key to set the second interval of auto transmitting. Press ENTER key to confirm it. Press upward or downward key to select storage position. Press ENTER key to confirm it. When “Auto RCL” twinkles, it represents the auto transmitting function has been activated. Press ENTER+ RECALL to turn off the auto transmitting mode and back to the measurement mode. Display the data and its ID grouping. Press upward and downward key to directly look up other data and their ID grouping. Press ENTER key to confirm to display this data of measurement value, storage year, date, and time. Back to measurement mode 15 2. Display the last calibration data (1) Under pH measurement mode, press RECALL key to show the last pH calibration data in display. (ex: 3-Point calibration) Press ENTER key to enter. The display shows Buffer 1 calibration data. Press ENTER key to enter. The display shows Buffer 2 calibration data. If no data, ignores the procedure and display the Slope value. Press ENTER key to display the SENSITIVITY value. Press ENTER key to display the ASYMMETRY value. Press ENTER key to display the R2 value. Press ENTER key to enter. The display shows Buffer 3 calibration data. If no data, ignores the procedure and display the Slope value. Press ENTER key to display the cal. data storage year. Press ENTER key to display the Slope value. 16 Press ENTER key to display the cal. data storage date. Press ENTER key to display the cal. data storage time. Press ENTER key Back to measurement mode 17 (2) Under ION measurement mode, press RECALL key to show the last ISE concentration calibration data in display. (TS-130 only) Press ENTER key to enter. The display shows Buffer 3 calibration data. If no data, ignores the procedure and display the Slope value. Press ENTER key to enter. The display shows Buffer 1 calibration data. Press ENTER key to display the mV value of Buffer 1. Press ENTER key to display the mV value of Buffer 3. Press ENTER key to enter. The display shows Buffer 2 calibration data. Press ENTER key to display the Slope value. Press ENTER key to display the SENSITIVITY value. Press ENTER key to display the mV value of Buffer 2. 18 3. Manual stored data transmitting (1) Transmit all the stored data Press ENTER key Press ENTER key to display the R2 value. Transmission completed Back to measurement mode (2) Format of received storage data Press ENTER key to display the cal. data storage year. Press ENTER key to display the cal. data storage date. Note: The output of the instrument is non-insulated signal. It may impact the transmitting measurement value if the instrument connects with other earth-related device. Press ENTER key to display the cal. data storage time. Press ENTER key Back to measurement mode 19 4. Calibration data transmitting (1) Transmit the stored cal. data through RS-232 interface 3.6 Delete all of the data (TS-110/130 only) Under power off status, press & hold STORE key and then press & hold POWER key until the display shows the model of instrument. Release POWER key. While the menu display shows data clean text, release STORE key. Press ENTER key Transmission completed Back to measurement mode Press ENTER key to confirm to delete all the data. To abandon the deletion, please press MODE key. (2) Format of received calibration data TS-110 pH mode TS-130 pH mode TS-130 ION mode 20 Back to measurement mode 4. Calibration Under measurement mode, press CAL key to enter calibration mode, and press MODE key to go back to measurement mode. 4.1 pH Calibration The instrument provides multi-point standard buffer solution calibration. You may decide how many points to calibrate the measurement system.(up to 5-point) The principle is according to “Method of Least Squares”. Apply linear regression to calibration the electrode’s slope and zero point (Asy, Offset or Zero point). When calibrating a electrode, you may calibrate 1 to 3 point by any sequence to provide linear regression for mV and pH multi-calibration of a electrode, and to show the electrode’s slope and zero point(Asy, offset or Zero point) at 25˚C. The electrode’s slope rate which is actual slope divided by theoretical slope and the sensitivity shows in percentage in the display. In addition, the display shows the linear regression determination coefficient, R2, of the electrode and buffer solution to provide you an estimation of an electrode’s regression suitability. According to different combination of standard buffers, the TECH, NIST, Any buffer solution calibration modes are provided. 4.1.1 TECH mode (up to 3 points calibration), CT The electrode is automatically calibrated according to built-in pH4.01, pH7.00, pH10.00 Tech standard buffers. The range of zero point and slope of the electrode is also determined. If one of them is over the range, the display shows error message of zero point and slope failure. (See appendix Table 1, pH/temperature table of TECH standard buffers) 4.1.2 NIST mode (up to 3 points calibration), CN The electrode is automatically calibrated according to pH value and temperature of NIST standard buffers.(Built-in 1.68, 4.01, 6.86, 9.18, 12.45 NIST standard buffer calibration set) The range of zero point and slope of the electrode is also determined. If one of them is over the range, the display shows error message of zero point and slope failure. (See appendix Table 2, pH/temperature table of NIST standard buffers) 4.1.3 Any buffer mode (up to 3 points calibration), CA The electrode measures mV value of different standard solutions. According to theoretic slope and the temperature of standard solutions, the display shows a approximate pH value. Then, you can calibrate the electrode by freely adjust the pH value as those of the standard solutions’. There is not a zero point range failure determination by the instrument but only the slope range determination. If the slope is over the range, the display shows error message of slope failure. 21 4.1.4 Definition of calibration parameter You can calibrate the electrode by one point or up to three points of standard solutions by any sequence. As different calibration point method is applied, the definition of the zero point and slope different. Calibration point Determination One point calibration Two or three point calibration Asy Asy Slope The showed calibration value Zero point (Asy, offset or Zero point)= Asy 1.If not calibrated, Slope = Theoretical slope 2.If calibrated, Slope = Slope of last calibration Zero point (Asy, offset or Zero point)= Asy Slope = Slope* Note: To obtain a new zero point(Asy) and Slope by applying linear regression. 4.2 Operation of pH calibration 4.2.1 Under pH measurement mode, press CAL key to enter pH calibration mode. Press MODE at any time to finish calibration and back to measurement mode. 4.2.2 After entering calibration mode, press CAL key to switch buffer type. Press ENTER for confirmation. Buffer Type:CT1: TECH. Buffer No.1 CN1: NIST. Buffer No.1 CA1: Any Buffer No.1 4.2.3 Within the 3 kinds of buffer type, they all can be done with up to 3 points calibration. You may calibrate the electrode with certain type buffers in any sequence. 22 4.3 Auto calibration for preset Buffers (CT / CN) 4.3.1 After entering to calibration mode, press CAL key to select Ct1 or Cn1. If temperature mode is set as MTC(Manual), the temperature value starts twinkling. Press upward or downward key to adjust the temperature value. Or press upward and downward key simultaneously to back to the preset value--25°C. Press ENTER key to start to read the preset Buffer 1 value. 4.3.2 When reading the value, the “Auto Read” symbol in the display starts to twinkle. Wait until the value is stable or directly press ENTER key to read the present value, then the menu enter into preset buffer 1 value with temperature value. 4.3.3 After entering preset Buffer 1 menu display, press ENTER or wait for 3 seconds to enter the Buffer 2 calibration menu display. 4.3.4 After entering into Buffer 2 menu display, press MODE key to leave the calibration menu to finish it as a single point calibration. The menu display shows slope value; for multi points calibration, clean the pH electrode with distilled water, put the electrode into second buffer solution. If temperature mode is set as MTC (Manual), the temperature value starts twinkling. Press upward or downward key to adjust the temperature value. Or press upward and downward key simultaneously to back to the preset value--25°C. Press ENTER key to start to read the preset Buffer 2 value. 4.3.5 When reading the value, the “Auto Read” symbol in the display starts to twinkle. Wait until the value is stable or directly press ENTER key to read the present value, then the menu enter into preset buffer 2 value with temperature value. 23 4.3.6 After entering preset Buffer 2 menu display, press ENTER or wait for 3 seconds to enter the Buffer 3 calibration menu display. 4.3.7 After entering into Buffer 2 menu display, press MODE key to leave the calibration menu to finish it as a dual point calibration. The menu display shows slope value; for multi points calibration, clean the pH electrode with distilled water, put the electrode into third buffer solution. Press ENTER key to start to read the preset Buffer 3 value for continuing calibration. 4.3.8 After finishing whether single, dual to five points calibration(CT mode up to three points), press ENTER or wait for 3 seconds to show the slope value, the SENSITIVITY value, the ASYMMETRY value, the R2 value. Then after 3 seconds or press ENTER key to go back to measurement mode. 24 4.4 Any Buffer manual calibration (CA) 4.4.1 After entering calibration mode, clean the pH electrode with distilled water, put the electrode into first buffer solution. Press CAL key to select CA1. If temperature mode is set as MTC (Manual), the temperature value starts twinkling. Press upward or downward key to adjust the temperature value. Or press upward and downward key simultaneously to back to the preset value--25°C. Press ENTER key to start to read the preset Buffer 1 value. 4.4.2 When reading the value, the “Auto Read” symbol in the display starts to twinkle. Wait until the value is stable or directly press ENTER key to read the present value, then the menu enter into manual calibration for Buffer 1. 4.4.3 After entering into manual calibration of Buffer 1 menu display (pH value twinkles), press upward or downward key to adjust the value to the selected buffer 1 value. After confirming, press ENTER key to enter into CA2 calibration menu display 4.4.4 Under CA2 calibration menu display, if press MODE key to leave the calibration menu to finish it as a single point calibration. The menu display shows slope value; for multi points calibration, clean the pH electrode with distilled water, put the electrode into second buffer solution. If temperature mode is set as MTC (Manual), the temperature value starts twinkling. Press upward or downward key to adjust the temperature value. Or press upward and downward key simultaneously to back to the preset value--25°C. Press ENTER key to start to read the Buffer 2 value. 25 4.4.5 When reading the value, the “Auto Read” symbol in the display starts to twinkle. Wait until the value is stable or directly press ENTER key to read the present value, then the menu enter into manual calibration for Buffer 2. 4.4.6 After entering into manual calibration of Buffer 2 menu display (pH value twinkles), press upward or downward key to adjust the value to the selected buffer 1 value. After confirming, press ENTER key to enter into CA3 calibration menu display. 4.4.7 Under CA3 calibration menu display, if press MODE key to leave the calibration menu to finish it as a dual point calibration. The menu display shows slope value; for multi points calibration, clean the pH electrode with distilled water, put the electrode into third buffer solution and continue the calibration procedures 4.4.8 After finishing whether single, dual or three points calibration, press ENTER or wait for 3 seconds to show the slope value, the SENSITIVITY value, the ASYMMETRY value, the R2 value. Finally, it will back to measurement mode. 26 4.5 ION concentration calibration (TS-130 only) The instrument applies the easiest and widely used ion-selective electrode direct potentiometry—ISE, to measure the potential reaction of the solution. Establishing the relationship of ion concentration(activity) and electrode potential by calibration, the instrument then is able to start measurement. The measurement and calibration of electrode is necessary to refer to the operation manual of ION selective electrodes(ISE). The method of calibration standard solution usually applies continuous dilution from high concentration standard solution in sequence to get 1000, 100, 10, 1 ppm(mg/l). In order to prevent from the difference of ISE activity and actual concentration, it is usually necessary to add ISAB (Ionic Strength Adjustment Buffer) when calibration for making the sample solution has similar ionic strength. It is recommended to use the calibration standard solution from low to high to prevent contamination of standard solution. If the concentration range of sample solution is known, it should be applied the nearest value of standard solution for calibration. If the sample measurement range and the standard solution calibration range have been confirmed to be in the linear range of the electrode, it is acceptable to make dual points calibration. However, making three of more points calibration can help to confirm whether the measurement is within the linear range or not. The calibration is up to 5 points calibration, and the instrument provides “log(A) = the logarithm of the activity of the measured ion” of linear regression calculation for the electrode potential(mV) and ion concentration(activity). It measures and displays the electrode’s slope under 25°C and the ratio of electrode actual slope with theoretical slope. It also displays the percentage of electrode sensitivity for judgment whether electrode is aged or contaminated. It also displays the linear regression R2 value (determination coefficient) of electrode and standard solution as the judgment of electrode regression suitability. Because the calibration slope is related the temperature, the measurement is compensated with temperature value. However, it is better to measure the standard solution and sample solution within ±2˚C temperature difference. The definition of electrode slope determination: Slope The number of valence 48 ~ 70 +1 monovalent ion -48 ~ -70 -1 monovalent ion 24 ~ 35 +2 ion -24 ~ -35 -2 ion Note: If over range, the error message of slope failure is E-02. 27 4.6 ION concentration calibration (CC)(TS-130 only) For concentration calibration(CC) (up to 5 points calibration), select ppm or mg/l unit under measurement mode. Press CAL key to enter into ION concentration calibration mode. The instrument has built-in 10 sets selection of different standard solution value from: 1.00→2.00→5.00→10.00→20.0→50.0→100.0→200→500→1000. The minimum is 2 points calibration and up to 5 points calibration. Users may press MODE key at any time to finish the calibration and back to measurement mode. 4.6.1 Press CAL key to select CC1. If temperature mode is set as MTC (Manual), the temperature value starts twinkling. Press upward or downward key to adjust the temperature value. Or press upward and downward key simultaneously to back to the preset value--25°C. Press ENTER key to start to read the preset standard solution No.1 value. 4.6.2 After entering into CC1 calibration menu display, the buffer text in the display starts to twinkle. Press upward or downward key to the required value. Press ENTER key to start to measure standard solution No. 1 4.6.3 When reading the value, the “Auto Read” symbol in the display starts to twinkle. Wait until the value is stable or directly press ENTER key to read and to display the standard solution No.1’s mV value. 4.6.4 Under the standard solution 1 menu display, wait for 3 seconds or press ENTER key to enter into CC2 calibration menu display. 4.6.5 Clean the electrode with distilled water, put the electrode into second standard solution Under CC2 calibration menu display. If temperature mode is set as MTC (Manual), the temperature value starts twinkling. Press upward or downward key to adjust the temperature value. Or press upward and downward key simultaneously to back to the preset value--25°C. Press ENTER key to enter into standard solution No.2 menu display. 28 4.6.6 Under standard solution No.2 menu display. The concentration value starts twinkles, press CAL key to select concentration value, press upward or downward key to adjust it to the required value. Press ENTER key to read and to display the standard solution No.2’s mV value. 4.6.7 When reading the value, the “Auto Read” symbol in the display starts to twinkle. Wait until the value is stable or directly press ENTER key to read and to display the standard solution No.2’s mV value. Under the standard solution 2 menu display, wait for 3 seconds or press ENTER key to enter into CC3 calibration menu display. Under CC3 calibration menu display, if press 4.6.8 4.6.9 MODE key to leave the calibration menu to finish it as a dual point calibration. The menu display shows slope value; for multi-point calibration, clean the pH electrode with distilled water, put the electrode into third standard solution and continue the calibration procedures. 4.6.10 After finishing whether dual or three points calibration, press ENTER or wait for 3 seconds to show the slope value, the SENSITIVITY value, the ASYMMETRY value, the R2 value and then back to measurement mode. 29 5. Error messages Messages Reason OFFSET(zero-point electric potential) value over range Dispositions 1. Replace the buffer solutions 2. Maintain the electrode or replace the new electrode, and make another calibration. 1. Replace the buffer solutions SLOPE value exceeds 2. Maintain the electrode or the upper or lower replace the new electrode, and make another limit calibration. Please check whether there is bubble or air in the glass end of The readout is the electrode; maintain the unstable electrode or change a new electrode, and make another calibration. 1. Please adjust the buffer solution temperature to the appropriate temperature the buffer solution range and make another temperature exceeds calibration. the range of 5~50°C 2. Please replace the buffer, or 2. The buffer can not maintain or replace the electrode and make another be identified. calibration. During calibration, Serious error that does not permit any further measuring 30 Please call service engineer. 6. Maintenance 6.1 Battery replacement a. Turn off the instrument power first. b. Use a Philips screwdriver to remove the four screws on the back of the instruments c. Open the battery cover, and remove the old batteries. d. Remove the old batteries, put into the new AA batteries. e. Close the battery cover, and locked the four screws. 6.2 Electrode clean and maintenance Generally speaking, under normal operation, the transmitter produced by our company needs no maintenance expect regular cleaning and calibration of the electrode to ensure accurate and stable measurement and system operation. The cleaning cycle for the electrode depends on the pollution degree of the measurement sample. Normally, it is recommended to make weekly cleaning. The following chart gives introductions of different cleaning methods according to different type of contaminations to provide the operators with reference for cleaning and maintenance. Type of Contaminations Cleaning methods Measuring solutions containing The electrode should be soaked in Pepsin/HCl for several proteins.(Contamination of the hours. METTLER-TOLEDO 9891 Electrode Cleaner is junction) recommended. Measuring solution containing The junction should be soaked in Thiourea/HCl solution sulfides.(The junction becomes until being bleached. METTLER-TOLEDO 9892 black) Electrode Cleaner is recommended. Short rinsing of the electrode with acetone and Contamination by grease or ethanol. organic substance Acid and alkaline soluble contaminations Rinsing the electrode with 0.1mol/l NaOH or 0.1mol/l HCl for a few minutes. Apply clean water to flash the electrode after above cleaning steps and immerse the electrode in 3M KCl solution for 15 minutes at least, and then calibrate the electrode. The electrode should only be rinsed and never rubbed or mechanically cleaned, since this would lead to electrostatic charges. This could cause an increase in the response time. In cleaning the platinum electrode, the platinum ring of the electrode can be rubbed gently with a wet soft piece of cloth. ※ The frequency of electrode cleaning depends on the type and degree of contamination. However it is recommended that the electrode be cleaned once a week. 31 7. Modbus protocol and instructions (TS-110/130 only) 7.1 Function code In the original Modbus protocol definition, there are many command functions on the substance of duplication. Base on the reason, TS110/130 only supports some of the function codes. When starting communication, the Master sends the request through function code to tell the Slave to perform which actions. The Slave responds function which is the same as that sent by the Master to indicate the Slave has already responded to the Master and has performed the action. The range of function code is 1~255. In between, the function code 128~255 are defined as abnormal responding function code. The function codes which the slaves normally respond are the same as those from the Master. The highest bit(MSB) is 0. When the slaves respond abnormal message, the lowest bit(LSB) of function codes is the same as those of the master, however, the highest bit(MSB) of the function codes are set to be 1. It means that there occurs a error when executing the function codes. The following table lists the function codes supported by TS-110/130 Function code 01H Read Coil Status 03H Read Holding Registers 05H Definition Operation Reads the ON/OFF status of discrete coils in the slave Read the status of one or more discrete contacts Read the binary contents of holding registers in the slave Read one or more data value register Writes a single coil to either ON or OFF. Set the status of discrete individual contacts to the specified address Writes a value into a single holding register Set a single 16-bit data register Writes each coil in a sequence of coils to either ON or OFF Set the status of multiple discrete continuous contacts to the specified address Writes values into a sequence of holding registers Write Multiple Registers Set multiple sequence of 16-bit data register Write Single Coil 06H Write Single Register 0FH Write Multiple Coils 10H 08H Diagnostics Diagnostics function 32 The assessment of network communication capability 7.2 MODBUS name and address table Function Code:03H, 06H, 10H Modbus response (setup parameter) Logic Number Informatio Description of data Item address of Byte n type transmission Equipment’s 0001H 2 USHORT 1 ID 0002H 0005H 0006H Meter’s model Communicatio n protocol Serial transmission speed (Baud rate) 6 USHORT 2 USHORT 2 USHORT ASCII Code 0:RTU 1:ASCII 1:2400 2:4800 0008H USHORT 0009H USHORT Minute USHORT Hour USHORT Day USHORT Month 000AH 000BH Parity Real-time clock* Note 1 TS110 TS130 1 2 3:9600 0: None 1: Even 2: Odd Second 0007H Default value 2 12 000CH USHORT 0 2011-01-010 0:00:00 Year 000DH USHORT The actions without * sign only support for function code 03H. The actions with * sign support function code 03H, 06H, 10H. USHORT data range from 0 to 65535, SHORT data range from -32768 to 32767. FLOAT is a 4 data bits IEEE format float. The data range follows is the same. Function code: 03H Modbus response (measurement parameter) Logic Number Informatio Description of data Item address of Byte n type transmission Number of TS110/TS130 only measurement 0031H 2 USHORT has one channel channels pH ORP(mV) ˚C Sign byte 0032H 6 CHAR Default value Note 1 ASCII code ppm (TS-130 only) mg/l(TS-130 only) 0035H pH/ORP/ION measurement 4 FLOAT pH/ORP/ION(TS-13 0 only) measurement Data affected by sign byte 0037H Temperature measurement 4 FLOAT Temperature measurement Data affected by sign byte 33 8. Appendix Table 1 Tech. Buffers CT Mode TEMP℃ Buffer 4.01 Buffer 7 Buffer 10 5 3.999 7.087 10.241 10 3.998 7.053 10.155 15 3.999 7.031 10.116 20 4.002 7.011 10.047 25 4.006 6.996 9.998 30 4.011 6.985 9.952 35 4.018 6.976 9.925 40 4.031 6.971 9.874 45 4.047 6.969 9.843 50 4.055 6.969 9.810 Table 2 NIST. Buffers(DIN 19266) CN Mode TEMP℃ Buffer 1.68 Buffer 4.01 Buffer 6.86 Buffer 9.18 Buffer 12.45 5 1.668 4.004 6.951 9.395 13.207 10 1.670 4.000 6.923 9.332 13.003 15 1.672 3.999 6.900 9.276 12.810 20 1.675 4.001 6.881 9.225 12.627 25 1.679 4.006 6.865 9.180 12.454 30 1.683 4.012 6.853 9.139 12.289 35 1.688 4.021 6.844 9.102 12.133 40 1.694 4.031 6.838 9.068 11.984 45 1.700 4.043 6.834 9.038 11.410 50 1.707 4.057 6.833 9.011 11.705 34 SUNTEX INSTRUMENTS CO., LTD. 13F, No. 31, Lane 169, Kangning St., Xizhi Dist., New Taipei City, Taiwan (R.O.C.) Tel: 886-2-2695-9688 Fax: 886-2-2695-9693 e-mail: suntex@ms1.hinet.net 5040C TS-100_110_130/ Technical data subject to alternations/ Quality Systems ISO 9001/201112 www.suntex.com.tw/en