Microbiological Safety Cabinet

Transcription

Microbiological Safety Cabinet
Microbiological Safety Cabinet
TOP SAFE CLASSE II
 Microbiological Safety Cabinet
 Continuous monitoring of Safety
parameters
 Alarms for low air flow and wrong
front window position
 Sloped front and back wall for the
most comfortable access
 Front access for filter maintenance
and service
 Easy retrofittable option kits
Top Safe Cabinets are supplied in three
different sizes (1.2mt, 1.5mt and 1.8mt).
These last generation Microbiological
Safety Cabinets have been certified according
to the most stringent safety standards (EN12469-2000).
The internal design, the air flow aerodynamics and
monitoring, the built-in safety devices and the very accurate
manufacturing, guarantees the highest performances at the
most stringent safety levels, as specified by EN12469 standard and have been tested by
the most prestigious European certification bodies for Safety Cabinets.
Main specifications
• Microprocessor controlled motor blower, with volumetric sensor for exhausted air flow
monitoring
• Highest air flow stability both in case of transitional disturbances or to progressive filter
clogging
• Continuous monitoring of front barrier air flow for the highest operator safety
• Low barrier alarm
• Free voltage contact for remote monitoring of exhaust fan.
• Bar-graph analogical display for the easiest reading of air speed value (barrier and down flow
air speed display for regular, pre-alarm and alarm conditions)
Mechanical and functional specifications
• Sloped front design for the highest operational comfort. Sloped back side of the working
chamber for the best down flow distribution
• Utilities inlets from the top of the cabinet.
S-SAFEHOODENGL-R0-0704
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Stainless Steel internal surfaces with 2B finishing (including spillage tray). Solid work surface
(3 sections) and special designed front grill.
Hinged multilayer safety glass window (max opening at 120°)
Comfortable 200mm front opening
Front closing panel
Retrofittable options through lateral sides.
Exposed exhaust Hepa filter for easy visual integrity chck.
Filter change and maintenance from the front of the cabinet.
Exhaust transitions easily installable.
Key operated. The key can be removed when the unit is in SAFE mode, in order to avoid
unwanted operation. In case of power failure, the cabinet is re-set to original working
conditions.
Self check cycle performed when cabinet is switched on. High speed rinse and set up cycle
performed, before reaching the SAFE operating mode.
Visual display of SAFE conditions. Pre-warning before actual alarm conditions are reached
(visual and acoustic alarms)
Soft touch control with keys for standard service utilities. Interconnected UV and fluorescent
lights.
H14 class High Efficiency Particulate Air filters with 99.999% efficiency on .3micron particles
Exhaust and recirculating flow rates ensure 25 air changes/min in the working area (30% 70%
split)
Front barrier air speed ≥ 0.5mt/sec
Aperture protection Factor (Apf) ≥ 1.5 x 10 exp 5
Cleanability Index CC grade, according to EN12469 standard
Light intensity on work surface ≥ 800 lux.
Noise level, complies with EN 12469:2000, determined according to EN ISO
3744 and ISO 11201
230V, 50Hz
Max power (for each power point) 3Amps.
Microprocessor equipped with analogical watch dog.
CE certification according to Machinery Directive 89/392/EEC, 91/368/EEC, 93/44/EEC
93/68/EEC.
TECHNICAL CHARACTERISTICS TOPSAFE
Model
TOPSAFE Class II
External Dimension (LxWxH) mm
Weight (Kg)
Internal Dimension camera di lavoro
(LxPxA) mm
Power (excl. int. power point) (KW)
Power point
Fluorescent Light
Fans
Work Surface Sections
OPTIONALS
Standard Options (incuded)
Elapsed time meter
Service power point
Solenoid valve for gas
Gas and vaacum taps
UV light kit (optional) connections
S-SAFEHOODENGL-R0-0704
1.2
1.5
1.8
1380x795x1450
180
1230x580x710
1685x795x1450
220
1530x580x710
1990x795x1450
290
1840x580x710
0,7
1
2x30W
1
3
1
1
2x36W
1
3
1,4
2
2x58W
2
4
Standard Options (not included)

Prefilter grid

Retrofittable kits for gaseous fluids

Pass-through holes for additional utilities

Second gas line kit with solenoid valve

UV light kit with or without timer

Passive exhaust transition for formaldheyde fumes eshaust or, alternatively, activated charcoal filter with
relevant accessories

Formaline vaporiser

Active exhaust transition (with fan) for ducted configuration

Remote active exhaust transition (with fan) for exhaust through negative presure duct (to be used with Passive
exh. Transition)

Thimble
Notes: ehaust transitions can be delivered according to specific customer requirements.
S-SAFEHOODENGL-R0-0704
s@feflow family
Microbiological Safety Cabinet Class II , Type A ,B3 convertible
Designed to provide protection to the operator,the product and the environment,against risks due to airborne contamination
THE ULTIMATE SAFETY CABINET
S@feflow is Microbiological Safety Cabinet (MSC) Class II, type A,B3 convertible, designed to provide protection to the
oprator, the product and the environment, against risks due to airbome contamination.
Ventilation of the cabinet includes recirculating air flow, filtered through high efficiency filter (HEPA) and a make-up air,
passing through the front opening and then discharged through an exhaust HEPA filter into the environment.
Recirculated and make-up air flows are maintained at a pre-set value by a closed loop microprocressor, which automatically
monitors and controls the air volumes, assisted by an advanced volumetric sensor.
S@feflow is a MSC certified according to EN 12469.
S@feflow has a sloped front, ergonomically designed to maximise the cabinet's operating comfort and user friendliness.
An electrically operated sliding sash can be moved up and down at different opening heights : at a finger tip touch, an
electronic device automatically searches for the optimised height of the front opening.
Operating mode:
•Type A configuration:The air leaving the main HEPA filter at a constant speed produces a laminar downflow Class 100
enviroment.An air barrier enters through the front opening at a speed of 0.45 m/s and is mixed with the downflow air
underneath the work surce.70% of the air is recirculated 30% is ducted through the port on top of the unit.
Type B3 configuration: The operating principle is the same but the unit is ducted to the outsite of the laboratory .
Moreover the sash is hinged at the top, allowing easy cleaning of the internal surface of the glass.The microprocessor has a
large, high resolution display , and all functions available both for operator and service staff, and are easily set through
multi-task switches.
Main functions includes:
•display of speed/volume values of air (downflow and front barrier)
•alarms: airflow and windows
•introduction of two different passwords to enter operational or service mode
•automatic fumigation cycle (with automatic set up of front sliding sash at defined heights in the different phases of the
fumigation cycle).
•UV cycles control
•Filters life
•UV lamp life
S-SAFEFLOWFAMILY-R0-0704
Page 1 of 2
BIOAIR Instruments s.r.l. is part of
group of company
TECHNICAL DATA S@FEFLOW FAMILY
S@feflow 0.9
Work chamber dimensions (lxdxh):
879x480x650 mm
Overall dimensions (lxdxh):
1055x792x1455 mm
Weight:
200 kg
Power supply:
230V-50Hz
Power requirement:
1.750 W
Noise Level
≤57dB (A)
Lighting
≥800 lux
S@feflow 1.2
1183x480x650 mm
1360x792x1455 mm
270 Kg
230V-50Hz
1.780 W
≤57 dB (A)
≥800 lux
S@feflow 1.8
1793x480x650 mm
1970x792x1455 mm
350 Kg
230 V 50 Hz
2.370 W
≤57 dB (A)
≥800 lux
HOW TO ORDER
Code
LD
LD
LD
LD
LD
LD
LD
80000
60000
70000
80001
60001
70001
60200
Item
S@feflow 0.9 with perforated working surface EN 12469 version
S@feflow 1.2 with perforated working surface EN 12469 version
S@feflow 1.8 with perforated working surface EN 12469 version
S@feflow 0.9 with solid working surface EN 12469 version
S@feflow 1.2 with solid working surface EN 12469 version
S@feflow 1.8 with solid working surface EN 12469 version
S@feflow 1.2 BS with perforated working surface and with dual exhaust HEPA filter
ACCESSORIES FOR S@feflow 0.9, 1.2 and 1.8
Code
AK
AS
AK
AS
AS
AZ
17000
18000
16000
16000
17000
33000
Item
UV-Lamp for S@feflow 0.9
Support stand for S@feflow 0.9
UV-Lamp for S@feflow 1.2 (EN 12469 and BS version) and S@feflow 1.8
Support stand for S@feflow 1.2 (EN 12469 and BS version)
Support stand for S@feflow 1.8
Formaldehyde vapourizer for all S@feflow models
Other accessories for S@feflow 1.2
AZ 16100
Transition adapter with extraction fan (without filter) according to customer's specifications
AZ 16300
Passive transition adapter
AZ 16400
Extraction open hood (thimble without fan)
AZ 16400
Extraction charcoal filter (with charcoal filter) only for S@feflow 1.2 EN 12469 Version (not for BS version)
S-SAFEFLOWFAMILY-R0-0704
Page 2 of 2
BIOAIR Instruments s.r.l. is part of
group of company
AURA 2000 M.A.C family
Microprocessor Automatic Control
Microbiological Safety Cabinet with microprocessor based control
Personnel protection, product protection and ambient safety
from exposure to biological agents
Class II microbiological safety cabinet for handling of low - moderate risk-hazarded agents; certified LNE (AFNOR)
EN 12469-2000
BIOHAZARD
The AURA 2000 M.A.C. is a Class II Microbiological Safety Cabonet 70% of fhe HEPA filtered air is therefore reciculated with the
remaining 30% being expelled through a dedicated exhaust HEPA filter. The cabinet is used to reduce the exposure.The AURA
2000 M.A.C has a new, electronic airflow control system.A microprocessor based module provides automatic control of the
airflow throughout the whole life of the HEPA filter.
The microprocessor based module contains the alarm system which warns the operator when recirculated and exhaust airflow
rate deviate from the fixed operation range and when the front windows is not firmly closed.The soft-touch keyboard together
with the alphanimeric display, bargraphs and pilot lamps give the operators a full controll control and monitoring of all
functions.Laminare flow and face velocites are permanently
displayed while the UV lamp,cabinet conditions ,filters, run hours and blackout events are displayed by touching the
multifunction keys.
The microprocessor module also controls the lights ,electrical power outlet, gas solenoid valve and 24 hours clok with UV time
function.
All of the contaminated plenum inside the cabinet is surrounded by a negative pressure to prevent contaminated air leaks,as
required by the latest international standard
•The microprocessor control module is located inside the front panel in a non contaminated zone
•The air flow control system provides a constant air flow unit the pressure across the HEPA filter emissing in reaches a level that
the fan cannot compensate
•The digital display and air velocity bargraph indicators give a continuous monitoring of the air flow conditions.
•Visible and audible alarms are actived when the laminar and exhaust air flow rates fall or exceed the fixed thresholds and when the fro
is not closed securely.
•Sof touch keys on the control panel provide control of the lighiting,UV lamp,electrical power outlet and the solenoid valve which regulat
supply line.
•Variable speed centrifugal fan with built-in thermal cut-out
•Electrical equipment accoding to EN 61010-1 International Standards and EMC directives.
•Ligting >800 lux.
•Noise level complies with EN 12469:2000, determined according to EN ISO 3744 and ISO 11201
•Cabinet outer surfaces of rolled steel with paint finish.
•Inner chamber made of AISI 304 stainless stell with radiussed cormers and full access to exposed surfaces for ease of cleaning.
•Removable work surface made of AISI 304 stainless steel with perforated work area and shaped front grille to prevent
obstruction of the safety air barrier intake.
S-AURA200MACFAMILY-R0-0704
Page 1 of 2
BIOAIR Instruments s.r.l. is part of
group of company
•Sump made of AISI 304 stainless steel,with coved comers for easy remaval of spilt liquids.
TECNICAL DATA AURA 2000 MAC FAMILY
Work chamber dimensions
(LxDxH):
Overall dimensions (LxDxH):
Weight:
Power supply:
Power requirement:
Lighting
Noise
AURA 2000 M.A.C.3
AURA 2000 M.A.C.4
AURA 2000 M.A.C.6
880x580x670 mm
1.185x580x670 mm
1.795x580x670 mm
1.145x785x1.460 mm
1.450x785x1.460 mm
2.060x785x1.460 mm
175 kg
225 kg
310 kg
230 V 50 Hz
230 V 50 H
230 V 50 Hz
1.750 W
2.200 W
2.370 W
800 LUX
800 LUX
800 LUX
Level noise level comply with EN 12469:2000, determined according to EN ISO 3744
and ISO 11201
HOW TO ORDER
Code
LD00100
LD50300
LD90100
Item
AURA 2000 M.A.C. 3
AURA 2000 M.A.C 4 LNE (AFNOR) EN 12469-2000 approved
AURA 2000 M.A.C 6
Accessories for Aura 2000 M.A.C. 3/4/6:
Code
Item
AS14000
AK17000
AS15000
AK15000
AS19000
AC10000
Support stand for AURA 2000 M.A.C
UV lamp for AURA 2000 M.A.C 3
Support stand for AURA 2000 M.A.C.4
UV lamp for AURA 2000 M.A.C.6
Drawer for all AURA 2000 M.A.C models
Support stand for AURA 2000 MAC 6
Other accessories for Aura 2000 MAC 4:
Code
AZ15100
AZ15300
AZ15200
Item
Exhaust kit with motorblower for AURA 2000 MAC 4( filter excluded)
Exhaust Adapter for AURA 2000 MAC
Exhaust passive adapter for AURA 2000 MAC
An additional (HEPA or charcoal) filter can be installed with item AZ15100
S-AURA200MACFAMILY-R0-0704
Page 2 of 2
BIOAIR Instruments s.r.l. is part of
group of company
AURA 2000 M.A.C BS
Microbiological Safety Cabinet with microprocessor based control and dual exhaust hepa filter
Personnel protection, product protection and ambient safety from exposure to biological agents
Class II microbiological safety cabinet for handling low to moderate risk-hazarded agents; certified BS 5726 dual exhaust HEPA.
CERTIFIED BS 5726
BIOHAZARD
The AURA 2000 BS is a Class II Microbiological Safety Cabinet. 70% of the HEPA filtered air is therefore recirculated with the
remaining 30% being expelled through two online dedicated exhaust HEPA filters. The cabinet is used to reduce the
exposure of both product and personnel to airborne biological agents. The AURA 2000 B.S. has a new, electronic airflow
control system. A microprocessor based module provides automatic control of the airflow throughout the whole usable life of
the HEPA filters. The microprocessor based module contains the alarm system which warns the operator when recirculated
and exhaust airflow rate deviate from the fixed operation range and when the front window is not firmly closed. The softtouch keyboard together with the alphanumeric display, bargraphs and pilot lamps give the operators full control and
monitoring of all functions. Laminar flow and face velocities are permanently displayed while the UV lamp, cabinet
conditions, filters, run hours and blackout events are displayed by touching the multifunction keys. The microprocessor
module also controls the lights, electrical power outlet, gas solenoid valve and 24 hour clock with UV timer function. All of
the contaminated plenum inside the cabinet is surrounded by a negative pressure to prevent contaminated air leaks as
required by the latest international standard.All of the contaminated plenum inside the cabinet is surrounded by a negative
pressure to prevent contaminated air leaks,as required by the latest international standard
•Variable speed centrifugal fan with built-in thermal cut-out.
•Electrical equipment according to IEC 1010-1 International Standards and EMC directives.
•Lighting > 800 lux.
•Noise level < 62 dB (A)
•Cabinet outer surfaces made of rolled steel with paint finish.
•Inner chamber made of AISI 304 stainless steel with radiussed corners and full access to exposed surfaces for ease of
cleaning.
•Removable work surface made of AISI 304 stainless steel with perforated work area and shaped front grille to prevent
obstruction of the safety air barrier intake.
•Sump made of AISI 304 stainless steel, with covred corners for easy removal of spilt liquids.
•The microprocessor control module is located inside the front panel in a non contaminated zone.
•The air flow control system provides a constant air flow until the pressure across the HEPA filter in reaches a level that the
fan cannot compensate
•The digital display and air velocity bargraph indicators give a continuous monitoring of the air flow conditions
•Visibile and audible alarms are activated when the laminar and exhaust air flow rates fall or exceed the fixed thresholds and
when the front window is not losed securely.
•Soft touch keys on the control panel provide control of the lighting, UV lamp, electrical power outlet and the solenoid valve
which regulates the gas supply line.
•Two on line exhaust HEPA filters.
•Two dedicated D.O.P. ports.
S-AURA200BS-R0-0704
Page 1 of 2
BIOAIR Instruments s.r.l. is part of
group of company
TECHNICAL DATA AURA 2000 MAC BS
Work chamber dimensions (LxDxH):
Overall dimensions (LxDxH):
Weight:
Power supply:
Power requirement:
Noise Level
Lighting
1185 x 580 x 670 mm
1450 x 785 x 1440 mm
225 kg
230 V 50 Hz
2.200 W
≤62 dB (A)
≥800 LUX
HOW TO ORDER
Code
LD50200
AK15000
AS15000
AC10000
Item
Microbiological Safety Cabinet (Class II)
UV lamp for Aura 2000
Support stand for Aura 2000
Drawer for model AURA 2000
S-AURA200BS-R0-0704
BS
BS
BS
BS
Page 2 of 2
BIOAIR Instruments s.r.l. is part of
group of company
AURA VF
Laminar downflow workstation
Probably the highest Apf (Aperture protection factor) in its category for the safest handling
of products
The aura VF has been designed and tested for the following applications:
Manipulation of non pathogenic micro-organisms Animal and plant Cell-cultures Molecular biology
Drug preparation Assembly of optical and electronic components Food Industry
LAMINAR DOWNFLOW
CLASS 100
Maximum operator, product and environmental protection on the Aura VF is obtained with a design identical to the one typical
of biosafety cabinets.
The unit works on a recirculatory principle with air passing through the main HEPA filter before entering the work area where
Class 100 conditions are obtained. Air coming through the front aperture is then mixed with the downflow air without affecting
the product, and recirculated; 30% of it is exhausted into the environment through an exhaust HEPA filter.
70% of the air is continually recirculated through the main HEPA filter. This ensures an environment conforming with Federal
standard 209E class 100 and British standard 5295 class B.
At the same time the operator and the environment are protected from any possible contaminant present in the working area.
•Construction: epoxy coated mild steel
•Base tray: AISI 304 stainless steel; autoclavable
•Glazing: tempered glass
•Filters types: main and exhaust filters:
HEPA (high efficiency particulate airfilter) with efficiency of 99,999% tested, and conform to EN 1822
•CE marked
•Fan: centrifugal with thermal cut-out.
•Control panel: soft-touch membrane, on/off switch, fan switch, elapsed time meter, light switch.
•Lighting: > 800 lux, fluorescent lights
•Noise level < 60dB(A) Aura VF product specifications
•Laminar downflow at a constant airspeed of 0.40 m/s
•Face velocity: 0.45 m/s
•Recirculatory mode with 70% recirculated air through main HEPA filter and 30% exhausted air through high efficiency exhaust
filter
•Electronic control.
•Double plenum with dynamic air-tightness
•Standard Optionals (included):
S-AURAVF-R0-0704
Page 1 of 3
BIOAIR Instruments s.r.l. is part of
group of company
front closing panel, internal 6A socket, vacuum tap
•Available as single HEPA filter configuration (downflow main HEPA only), no exhaust HEPA filter.
S-AURAVF-R0-0704
Page 2 of 3
BIOAIR Instruments s.r.l. is part of
group of company
TECHNICAL DATA AURA VF 48
Work chamber dimensions (LxDxH):
Overall dimensions (LxDxH):
Weight:
Power supply:
Power requirement:
Noise level:
Lighting:
1190 x 580x 675 mm
1350 x 850 x 1415 mm
170 kg
230V - 50Hz
1.700 W
< 60 d B (A)
> 700 lux
HOW TO ORDER
Code
LV10000
Item
AURA VF 48 con doppio filtro HEPA
ACCESSORIES FOR AURA VF MODEL
Code
AK10000
AZ50000
AS31000
AC10000
Item
UV lamp for AURA VF 4
Solenoid gas valve for AURA VF 48
Support stand for AURA VF 48
Drawer for AURA VF 48
S-AURAVF-R0-0704
Page 3 of 3
BIOAIR Instruments s.r.l. is part of
group of company
AURA HZ
Horizontal laminar air flow cabinet class 100 for product protection.
HZ cabinets are designed according to the most advanced electronic and aerodinamic
technologies in order to ensure a Contamination Controlled Environment of the highest
quality
Main applications for AURA HZ are:
• Media preparation
• Plant and animal tissue culture
• Pharmaceutical preparations
• Assembling of optical or microelectronic components
• Micromechanics
LAMINAR AIRFLOW
CABINET
CLASS 100 HORIZONTAL
Automatic air control: AURA HZ is a horizontal laminar air flow cabinet, Class 100, for product protection from external
contamination and cross-contamination inside the working chamber. The air speed is, for the highest product safety,
permanently controlled.
An integral vane anemometer is installed in the unit as a standard option.
Venturi-effect side walls: special tempered-glass side walls are installed inside the air flow, to prevent any leakage from the
outside.
Operating mode: the air flow, leaving the HEPA filter at a constant speed, produces an extremely clean - better than Class 100
- environment in the work area, preventing any contamination to enter from the ambient outside.
High efficiency washable polyurethane prefilters remouve coarse particles before the air reaches the HEP
A filter.
•Electrical equipment according to International Standards and EMC directives
•Lighting › 800 lux
•Noise level ‹ 60dB(A)
•Electronic control air speed
•Centrifugal motorblower
•Bargraph display for air speed value
•Soft touch keys on the control panel provide control of the lighting, motorblower, UV light.
•Visible and acoustic alarms are provided for air speed alarm and filter clogging.
•Two sizes to satisfy all applications
•Cabinet outer surfaces made of cold rolled steel with paint finish
•Working surface made of grade 3 stainless steel
•Side panels in tempered glass
•HEPA filter manufactured in accordance with MIL-F51068 and BS-3928, with an efficiency higher than 99.999% on 0.3µ
particles, DOP tested
•Micromesh membrane on HEPA filter downstream surface for perfect airspeed distribution
•CE marked
S-AURAHZ-R0-0704
Page 1 of 2
BIOAIR Instruments s.r.l. is part of
group of company
Work chamber dimensions (LxDxH)
Overall dimensions (LxDxH)
Weight
Power supply
Power requirement
Noise level:
Lighting
HZ 48
HZ 72
mm 1120x610x732
mm 1350x1063x1353
128 kg
230V - 50Hz
0,42 kW
< 60 dB
> 700 lux
mm 1730x610x732
mm 1960x1063x1353
194 kg
230V - 50Hz
0,8 kw
< 60 dB (A)
>700 Lux
HOW TO ORDER
Codice
Item
LH20000
LH30000
AURA HZ 48 Stainless steel work surface
AURA HZ 72 Stainless steel work surface
ACCESSORIES FOR AURA HZ MODELS
Code
AS22000
AS23000
AZ32001
AZ31001
AZ32000
AZ31000
AZ60000
AZ61000
AC10000
Item
Support stand for Aura HZ 48
Support stand for Aura HZ 72
Double tap for vacuum
Single tap for vacuum
Double tap for gas
Single tap for gass
Double power point
Single power point
Drawer for all models of Aura HZ
S-AURAHZ-R0-0704
Page 2 of 2
BIOAIR Instruments s.r.l. is part of
group of company