Apotome - Robarts Research Institute
Transcription
Apotome - Robarts Research Institute
TRAINING MANUAL Zeiss Apotome Wide Field Microscope July 2014 TABLE OF CONTENTS A. Start-Up Procedure ...................................................................................... 1 B. Image Acquisition ..... ................................................................................... 4 Single fluorophore image acquisition........................................................................ Multidimensional Acquisition ................................................................................... Saving your Image...................................................................................................... ROI ............................................................................................................................. Z-stack........................................................................................................................ 4 6 8 8 8 C. Transmitted Light ........................................................................................ 9 D. Display Image Properties ........................................................................... 10 E. Exporting your Image................................................................................... 10 F. Apotome ........................................................................................................14 Principles.................................................................................................................... 11 Acquiring an Image ................................................................................................... 12 Configurations ........................................................................................................... 14 G. Shut Down Procedure ...................................................................................15 A. Start-Up Procedure 1. Turn on the mercury burner (#1) 2. Turn the stage box on (MCU) - back right (#2) 3 3. If you are using the Apotome function, turn the Apotome bon on (#3). If not, go to the next step. 1 4. Turn the microscope on (at the right side of the microscope) 2 5. Now turn the computer on and log in 6. Start the AxioVision Software by double clicking on the icon in the desktop Main screen: tool bar work area Image Window Robarts Research Institute Apotome Training Manual July 2014 B. Image Acquisition This system has the capability to acquire multichannel, Z-stack and time-lapse images. There are pre-set channels that you can modify it. 1- After turning all the components of the microscope on, select the objective that you want to use Here are the specifications of the objectives in this system: Objective NA Immersion Working Distance Coverglass thickness 10X Plan Neofluar 0.3 Dry 5.2mm 0.17mm 20X Plan Neofluar 0.5 Dry 2.0mm 0.17mm 40X LD Plan Neofluar 0.6 Dry 2.9mm 0 -1.5mm 63X Apochromat 1.4 Oil 0.19mm 0.17mm 2- Add the appropriate immersion (see table above) and load your sample with the coverslip facing down 3- At the toll bar, select the appropriate filter cube according to your fluorophore (green, red or DAPI) 4- Select EYEPIECE to visualize your image in the microscope Robarts Research Institute Apotome Training Manual July 2014 B. Image Acquisition Continued... If you want to change the filter cube to visualize a different dye, you can do it at the main tool bar in the AxioVision software (see step #3 in previous page) or by pressing the reflector button in the microscope (last 2 buttons on the right back side of the microscope) 4- After focusing your sample, close the fluorescence shutter to protect your sample from bleaching. You can do this by pressing the FL button (first button on the right side of the microscope) or in the software by clicking on the SHUTTER icon (toolbar) 5- To switch to the camera, click on CAMERA in the software (toolbar) 6- Click on LIVE in the main tool bar to visualize the image at the computer screen To Adjust the brightness: If you want to adjust the exposure time (brightness) for your dye, select CAMERAS AxioCamHR in the work area window. You can click on MEASURE to let the software find the best exposure time for your image or change it yourself by using the Exposure bars. Measure is always too dark or too bright for your purposes you can adjust the level e.g. set to 90% for a darker image (underexposed) or 110% to overexpose using the lower slider control. If you have Auto Snap box checked the exposure will be adjusted automatically before each single snap. Checking the Auto Live box enables continuous exposure time adjustment in the live image. Robarts Research Institute Apotome Training Manual July 2014 B. Image Acquisition Continued... To Adjust the number of pixels: If you want to change the number of pixels of your image, go to CAMERAS - AXIOCAMHR in the work area, select the FRAME tab and choose the number of pixels that suit your experiments. We usually use 1388X1040 pixels as default. Click on LAST to go back to the same number of pixels of your previous image. For single fluorophore images: If you only have one dye in your sample, at this point all you have to do is click on SNAP at the main tool bar. The image will be storage and you must save it before closing the software. For samples with multiple labels, use the Multidimensional acquisition function. Multidimensional Acquisition The multidimensional acquisition was designed to acquire multichannel, Z-stack and time lapse images. 1-In the work area window, select MULTIDIMENSIONAL ACQUISITION 2- In the Experiment tab, select the type of image you want to acquire (Z-stack, time lapse...). You don’t need to select anything if all you are doing is acquiring a multichannel image 3-Click on the C tab to open the available channels By default, the software has 5 different channels already ready to use. If you don’t need some of them, just right click on the channel. This command will inactivate the channel selected. Robarts Research Institute Apotome Training Manual July 2014 B. Image Acquisition Continued... Multidimensional Acquisition continued... Channel Filter Set Excitation Emission Beam Spliter Designed for Blue 488001-0000000 BP 365/12 LP 397 FT 395 DAPI, Hoechst, Alexa Fluor 405 and 350 FITC 488010-0000000 BP 450-490 BP 515-565 FT 510 GFP, FITC, Alexa-Fluor 488 Rhodamin (Cy3) 488020-0000000 BP 546/12 BP 575-640 FT 560 Rhodamin, Cy3, Alexa Fluor546, 555, 568 Far-Red (Cy5) 488050-0000000 BP 640/30 BP 690/50 Cy5, Alexa Fluor 633, 647 Transmitted light 424931-0000000 - - FT 660 - Bright field image The first thing you should do it to adjust the exposure time for each individual channel before acquire the final image. To do this, select the channel you want to adjust, for example DAPI, click on MEASURE. The camera will find the optimum exposure level for your sample. If you are happy with the result, click on OK. If not, change the exposure time manually. Do the same procedure for all the channels you have and then press START to acquire a multi channel image. AxioVision will display the overlay image with all the channels. If you want to visualize only some of theses channels, on the bottom left on the image window, click on the channel(s) that you want to inactivate. If you want to change the color of the channel, right click on it and choose a different color. Robarts Research Institute Apotome Training Manual July 2014 B. Image Acquisition Continued... To save an image To save this image, go to FILE-SAVE AS, choose the format of the image (zvi, tif, jpeg or bmp) and save it. You can then restart the whole process to acquire new images. To find out all the information about your image, click on INFOVIEW in the bottom left tab of the image window. To capture an ROI If you want to capture a region of interest (ROI) rather than the full frame, click on LIVE and then REFRESH OVERVIEW in the frame tab. A small preview of your image will show up in the frame tab window. Use this image as your reference and change the size of the selected area (green window) to crop your image. To go back to the regular full size image, click on FULL. To capture an Z-stack To acquire a Z-stack image, go to MULTIDIMENSIONAL ACQUISITION and select Z-stack under the experiment tab. Select the appropriate filter cube for the dye you want to use as a reference. Under the Z-tack tab, select start/stop as MODE. Now, press the ... button on the right side of the STOP button. A window displaying your live image will be open. Change the focus all the way down (counter-clock wise) and click on OK. This command will set the bottom of your image. Now click on the ... button on the right side of the START button. Change the focus all the way up (clock wise) and click on OK. This command will set the top of your image. Robarts Research Institute Apotome Training Manual July 2014 B. Image Acquisition Continued... To capture an Z-stack continued... How far up and down you should go is a matter of your judgment. Try to avoid the planes with high background (for example the plane closer to the coverslip). After selecting the bottom and top of your sample click on OPTIMAL DISTANCE to set the optimal interval between the stacks according to the optics you are using. If you prefer you can enter this number manually. Click START to acquire a Z-stack image. Tip: If you are combining multichannel and Z-Stack acquisition, AxioVision will normally acquire all the z images for channel 1 then all the z images for channel 2 and so on. This gives the fastest acquisition but may suffer from slight z registration problems. However if you check All Channels per Slice AxioVision will acquire all channels for the first z position and then move on to the next z position and so on. This will give exact z registration but usually increases acquisition time. C. Transmitted light If you want to take pictures of only bright field images, select TRANSMITTED at the main tool bar and click live. Adjust the exposure time or click on measure in the camera menu (workarea - left side). If after clicking measure you think your image is not brigh or dark enough, you can change the percentage (AxioCAmHR: Exposure) to a higher or lower number, for example 120% instead 100%. Make sure to also select the Auto-snap option. When you are ready to acquire the final image (SNAP), the software will adjust the exposure time to the percentage you have selected. Robarts Research Institute Apotome Training Manual July 2014 D. Image Display Properties You may want to adjust the contrast and brightness of the displayed image. For that, click on the Image Properties function and select DISPLAY tab. You use this window to modify the way in which your image is displayed. The histogram shows the number of pixels at each intensity level for each color (RGB). Contrast, brightness and gamma can be controlled via the sliders or by dragging the curve on the histogram. Click on Linear for the default display curve. Try selecting Min/Max to stretch the contrast to the histogram maximum and minimum. Tip: Best Fit is similar to Min/Max but it will stretch the contrast a little further since it ignores the extreme tails of the histogram. This is set at 5% by default but can be changed. Image Properties E. Exporting your images If you just want to export a single image then you can simply use Save As in the File menu and select the file format. However if you want to export multidimensional images it becomes more complicated as each image in the ZVI file must be exported as a separate image. In this case you need to use the function described below. First ensure that the image to be exported is the active document (i.e. click on the bar at the top) then use Export in the File menu. Refer to the online help for a detailed description of this function, we will just cover the main points. Insert a Base name and a Save in location – all images contained within the ZVI will be exported to a file with a name starting with the base name. If you click Create project folder a folder will be created with the Base name which will contain all exported images. Mode allows you to export some of the image e.g. a specified channel, time, z position or all images. Generate merged images will combine multichannel images as overlaid color images. File Type - select the output file format. Convert to 8 bit (only seen when saving as a *.tif) will convert a higher bit image down to an 8 bit TIFF. Gray scale will convert a color image to a black and white image. Apply display mappings will retain the current display property settings Robarts Research Institute Apotome Training Manual July 2014 F. Apotome Principles In thick specimens, the signal from the focal plane is compromised by the out of focus light coming from planes above and bellows this plane. The result is a blurred, low contrast image. In this condition, is not possible to render a three-dimensional projection of the specimen. Some techniques were developed to remove the out of focus light, either optically (Confocal) or mathematically (Apotome). In the Apotome, a grid is inserted in the illumination pathway of a conventional fluorescence microscope. A plane-parallel glass plate is placed in front of the grid and it tilts back and forth projecting the grid in the image in 3 different positions. For each depth, 3 raw images with the projected grid superimposed in 3 different positions are taken. The software uses the projected images of the grid to calculate and removes the out of focus light. The principle is, if the grid is visible that information is in focus. If not, that signal corresponds to out of focus light and it is removed from the final image. The software combines the 3 images into 1 for each stack. GRID LIGHT SOURCE Image 1 sample plane-parallel plate GRID LIGHT SOURCE Image 2 sample plane-parallel plate GRID Image 3 LIGHT SOURCE sample Robarts Research Institute Apotome Training Manual July 2014 plane-parallel plate F. Apotome Acquiring an image using the ApoTome slider As seen in the principles of the structure illumination method, the idea of removing the out of focus light of wield field images is specially helpful for Z-stack images. Therefore, the first step is setting the configurations for your Z-stack (bottom and top, interval and how many channels). You can do this part without inserting the Apotome Slider in but in order to calculate the best exposure time, you MUST insert the slider. Since the software removes the out of focus light of the processed image, this image will look much dimmer that the non-processed one (regular wide field). You need to compensate for that by measuring the best exposure time with the grid in. 1- Gently insert the apotome slider (move all the way in). You will hear a beep 2- Open the Apotome window (main tool bar) if you click on MEASURE, the software will calculate the best exposure time for your sample if you click on Refresh Overview, your image will be displayed in this window. You can then crop it. select GRID VISIBLE select OPTICAL SECTIONING VH = for inverted and high magnification objectives VL = for inverted and low magnification objectives Our Apotome is calibrated for the 63X objective. Therefore, you should always select the VH grid option. If you wish to use another objective, the grid needs to be changed and a new calibration needs to be done. Robarts Research Institute Apotome Training Manual July 2014 F. Apotome Acquiring an image using the ApoTome slider Make sure that in the tab EXTRAS, the DISPLAY NORMALIZATION option is checked. This option will rescale how the image is displayed, adjusting the intensity. X You are now ready to acquire your image. Press START in the Z-stack tab. The system will acquire 3 images for each focal plane, changing the position of the grid (3 different positions). These images will be used to calculate the out of focus light and remove it from the final image. If you want to continue using the Apotome mode for your next images, just keep the slider in, find a different cell/area and repeat the process. If you want to go back to the widefield mode, pull the Apotome slider out (gently) and close the Apotome window. You will have to set new exposure time for your channels. Regarding the Apotome Filter and Averaging, if your final processed image shows the grids, you can apply a weak filter and noise reduction of minimum 2 to improve it. WIDE FIELD X APOTOME Robarts Research Institute Apotome Training Manual July 2014 F. Apotome Configurations How the configuration affects your final image: Acquisition Mode These options influence the way the image acquisition is done. Ø No Processing: In this mode, only one image is acquired, with the grid in its current position. No processing takes place. This function is particularly important if, with automated multichannel image acquisition, you want to acquire an additional image in transmitted light. In this case you need to deactivate the acquisition of three raw images with subsequent processing, as fringe projection does not take place in transmitted light. Ø Optical Sectioning: In this mode three images are acquired, combined online into an optical section, and displayed. Ø Conventional Fluorescence: In this mode three raw images are acquired and combined into an image resembling the widefield situation. This mode allows comparisons between widefield image and optical section. Ø Raw Data Mode: In this mode all raw images (i.e. the individual images with the imprinted grid positions) are kept after acquisition. The advantage of this mode is that it allows you to switch between the algorithms (Optical Sectioning or Conventional Fluorescence) after image acquisition. Switching takes place in the Properties window. In addition, the different filter operations or image- vs. display normalization (usually gives result images with better dynamic range) or even further improve resolution by deconvolution can be done after acquisition. Filter Depending on the sample, bleaching rates of the fluorescent dye, exposure time, vibrations in the setup etc., residual stripes may be visible in the resulting image. These lines can be removed using the filter. The options Off, Low, Middle and Strong are available. Any repetitive lines or stripes appear as points along a vertical line in Fourier space. The filter can remove a defined group of these points out of the Fourier spectrum. · Off: No filter operation · Weak: Suppression of the 3rd and 6th order of the fundamental grid frequency · Medium: Suppression of the fundamental, the 3rd and 6th order of the fundamental grid frequency · Strong: Suppression of the fundamental, 2nd, 3rd and 6th order of the fundamental grid frequency Start with the setting Low. If residual lines can still be identified in the image, increase the setting step by step. Robarts Research Institute Apotome Training Manual July 2014 F. Apotome Configurations continued... Averaging (Noise Reduction) This function can be used to achieve noise suppression through the sequential acquisition of several images and the subsequent averaging of the gray values. It is possible to acquire and average between two and five images. This ApoTome 2 function not only acquires and averages a number of images, but also displaces the position of the grid lines in each case. This means that the function also enables you to minimize residual lines in the resulting image when working with samples that bleach out easily. G. Shut Down Procedure 1- Save all your images 2- Before shutting everything down, check the online calendar to see if there is anybody coming after you. If the next user is coming within 1 hour after you, close the software, log off and leave the other components on. If the next user is coming with more than 1 hour after you, proceed with the shut down steps listed bellow. 2- Close the software and shut computer down 3- Remove your sample and clean the objective 4-Turn the microscope off 5- Turn the Apotome box off (if used) 6- Turn the Fluorecence lamp off 7- Turn the stage box off (in the back) 8- Cover the microscope 9- Sign the user sheet Robarts Research Institute Apotome Training Manual July 2014