On Deck Thermal Cycler

Transcription

On Deck Thermal Cycler
On Deck Thermal Cycler
for Hamilton Liquid Handling Platforms
Fully automated. Seamlessly integrated.
On deck thermal cycling with the ODTC® from INHECO.
Simple, on-deck integration
XX Designed for on-deck integration with all Hamilton liquid handling workstations
XX On-deck footprint of the ODTC is equivalent to 2.5 SBS positions;
controlling unit & power supply external
XX Up to 2 ODTC® fit on a 7-track carrier; up to 4 ODTC® fit on a STAR Line instrument
Fast PCR runs and highest performance guaranteed
XX Thermal cycling with INHECO´s innovative 3D vapor chamber technology (VCM®)
XX Excellent well-to-well temperature uniformity, high heating and cooling rates,
rapid transition into plateau temperature
Easily upgrade Hamilton liquid handling systems on site
XX Integrate the ODTC® at any time without any deck modification needed
Streamline your genomic workflow
XX Pipette directly into PCR plates located on the ODTC®
XX Lid can be opened horizontally at any time during heating/cooling
XX Molecular biological workflows, even if multi-step processes, can now be fully automated
Hamilton 96 well PCR FramePlate®
A dual component PCR plate designed for automation
XX Improved consistency of PCR results due to dual component
technology (Polycarbonate frame and Polypropylene well)
XX Decreased sample evaporation at corner wells due
to rigid frame design
XX Optimal heat transfer as a result of low profile thin wall design of wells
XX Reliable use with CO-RE Gripper because of rigid frame design
Discover the possibilities
Technical data
�
ODTC® dimensions: (WxDxH)
156.5 mm x 248 mm x 124 mm or
Possible applications
XX End-point PCR: easily combine Hamilton`s PCRsetup Starlet &
ODTC® from INHECO to merge these two steps of the PCR workflow
in a single instrument
137 mm on the 7-track carrier
�
Temperature uniformity (based on
functional sample measurements)
@ 55°C: ± 0.1°C
XX NGS library preparation: the excellent uniformity of the temperature
in the ODTC® ensures high quality libraries as a basis for successful NGS
@72°C: ± 0.1 °C
@95°C: ± 0.15 °C
XX Sanger sequencing: prepare and perform Sanger sequencing
reactions on one compact instrument
�
Temperature range: 4°-99°C
(heated lid temperature adjustable
between 5 K above ambient
Automate any molecular biological workflow requiring
heating or cooling steps with an ODTC® integrated
on a Hamilton liquid handling workstation.
temperature up to 115°C)
�
Performance data:
Temperature accuracy
± 0.3 °C @ 55 °C
Average heating rate max 4.4 °C/s
Average cooling rate max 2.2 °C/s
Runtime < 40 min for 40 cycles
® 384
ODTC
now
available
�
Plate Format 96 and 384
other formats/disposables on request
�
Automated plate ejection mechanism
�
SiLA compliant
P/NDescriptionPackaging
819072
Kit for ODTC® 961
Includes ODTC® & controller, router, hardware integration kit including modified
shaded left cover, software & manuals, 1 case Hamilton 96 well FramePlate® 819416
Kit for ODTC® 96 Tr (clear)
1
®
Includes 2 ODTC & controller, router, hardware integration kit including modified
clear left cover, software & manuals, 1 case Hamilton 96 well FramePlate®
814302
Hamilton 96 well FramePlate® 10 pcs/bag
Full skirted, low profile, clear well design, Free of DNA, DNase, 50 pcs/case
RNase and PCR inhibitors
Visit INHECO at: www.inheco.com
© 2015 Hamilton Bonaduz AG. All rights reserved.
ODTC and VCM are registered trademarks of INHECO
Web: www.hamiltoncompany.com
800-648-5950
USA:
Email: infoservice@hamiltonrobotics.com
Lit. No. F-1305-03/03
United States
Tel: +1 775 858 3000
China
Tel: +86 21 6164 6567
United Kingdom & Ireland
Tel: +44 (0)121 717 0199
France
Tel: +33 (0)169 75 16 16
Brazil
Tel: +55 (11) 9677 4093
Italy
Tel: +39 039 689 33 93
Denmark, Norway,
Sweden, Finland
Tel: +46 8410 273 73
Germany, Switzerland,
Austria, Benelux
Tel: +49 (0)89 552 649 0
To find a subsidiary or distributor in your area, please visit hamiltoncompany.com/contact-us/distributors.