Spraytec roll fold 31/8
Transcription
Spraytec roll fold 31/8
REAL-TIME, HIGH SPEED HIGH CONCENTRATION SPRAY DROPLET SIZER THEORY Malvern Spraytec For real time, high speed measurements of high concentration liquid droplets in sprays, aerosols and inhalers. Particle size measurement using laser diffraction has been developed and applied by Malvern Instruments for more than 30 years. Malvern’s Mastersizer is the industry standard for a range of laboratory measurements. Malvern’s Spraytec is built around the same principle, but adapted for the specific needs for the characterization of sprays and aerosols. Malvern’s Spraytec combines the proven technique of laser diffraction particle characterization with patented high concentration data inversion routines for the analysis of sprays with high or rapidly changing concentrations. Spray measurement requires an accurate, highspeed measurement device immune to material characteristics.The Malvern Spraytec provides scientists with the ability to obtain precise measurement data from all types of sprays. Sealed to prevent moisture, dust and gas ingress, the robust instrument housing allows operators to easily access the droplet stream in a wide variety of experimental or production environments. Aerosol and spray researchers face challenges every day. Malvern’s Spraytec makes meeting these challenges easier. Aerospace Composite Technologies Ltd, Icing Research Wind Tunnel High Speed Measurements up to 2500Hz with Flexible Triggering Options Spraytec offers real time spray measurement at speeds of up to 2500Hz, giving a time resolution of one measurement every 400 microseconds.The measurement speed is software configurable. Measurements can consist of single detector scans or averages of multiple scans, to provide the most accurate data possible.The high temporal resolution offered by the Spraytec reveals fine temporal fluctuations in spray patterns and allows measurement of short duration sprays such as metered dose inhalers, dry powder inhalers and fuel sprays, that are missed by slower systems. Multiple measurement triggers are provided, making it easy to synchronize measurements with spray actuation in any experiment configuration. Since sprays can be of very short duration, random, or intermittent at high repetition rates, correct triggering of the measurement of such sprays is important to ensure that data is captured consistently. Malvern’s Spraytec offers four triggering modes; allowing measurements to be triggered manually, through automatic spray detection based on changes in light transmission, via low concentration triggering based on detection of a predefined light scattering signature, or by external triggering with a signal from another device. A variety of statistical displays enables quick data analysis and comprehension, without being forced to export data into additional analysis programs. Spraytec Software – “Frame-by-Frame” Analysis,Time History Utility, Data Filtering The software also allows operators to include or exclude measurement sequences, allowing analysis to avoid periods between sprays, outside of acceptable obscuration or signal limits, or otherwise undesired. Malvern’s Spraytec software provides a user-friendly interface to monitor and control all data reduction and display functions.The time history utility enables a “frame-by-frame” inspection of a spray sequence, that shows exactly how the droplet size distribution evolves throughout the time frame in question. Fig 1 shows a time history of a fuel spray of 0.444 seconds in duration. During this period, 112 separate measurements were made. Fig 1 Fig 2 Fig 2 shows a time history file for a 15-millisecond puff from a dry powder inhaler.The control variables shown are the Dv50, the Dv10 and the Dv90 as well as the transmission. High Concentration Measurement Capability – Allows the Measurement of Sprays at Obscurations of up to 95% The phenomenon known as “multiple scattering” occurs when the light scattered by droplets in the spray encounters further droplets and is further scattered at larger angles, before reaching the detector. Uncorrected, this leads to a substantial overstatement of the fine droplets in the distribution. A patented multiple scattering algorithm, automatically corrects for this change in signal and allows successful measurements in extremely high concentrations with light obscuration as high as 95% (5% transmission). The multiple scattering algorithm has been validated by measuring progressively higher concentrations of a suspension of Aluminum Oxide.The corrected curve at 95% obscuration (5% transmission) over-plots the measurement obtained at 16% obscuration (84% transmission). Fit for Purpose – Flexible, Robust, and Simple in Design Sprays are wet, messy and can occupy large volumes. How does Spraytec deal with these fundamental problems? Large clearances between transmitter and receiver modules accommodate sprays with large cone angles and minimize contamination of the optics. Sealed transmitter and receiver modules and air-purgeable optics further minimize contamination by spray products.Transmitter and receiver modules are designed so that mounting on custom rigs is achievable with minimal engineering effort. A range of laser diameters is available to improve spatial resolution when characterizing extremely brief highly concentrated sprays, for example, fuel injection nozzles. For applications that require additional protection against dust, water or flammable materials a range of NEMA enclosures is available. A choice of optical units are available (see specifications for options) which allows the user to select the best balance between minimum measurable droplet size and working distance.The software also includes an imaging correction facility that allows the correction of data, where, for any reason, the droplets being measured are outside the working distance of the lens. SPECIFICATIONS Applications Malvern’s Spraytec instruments are used in a wide range of industries: Industry Application Engineering Design and optimisation of spray nozzels – for the pharmaceutical, food, cosmetic, paint, chemical, fire protection, power generation, agriculture and lubrication industries Formulation, development and testing of MDI’s, DPI’s and nasal sprays Diesel and Fuel injection studies, airbag development Simulation of extreme climates Pharmaceutical Automotive Engineering Aerospace Engineeing Specifications Optical Unit Measurement Principle Light Source Laser Beam Diameter Measurement Rate Concentration Range Accuracy Precision Mie Scattering Laser Diode 670nm 2 10mm, 6mm and 3mm 1/e 1 to 2500Hz (giving 1 measurement every 400 microseconds) Sprays with obscurations from 2% to 95% 3% (using RS-3 reference standard reticle) 0.5% (using RS-3 reference standard reticle) Droplet Size Range Maximum Working Distance 0.5µm to 200µm 1.0µm to 400µm 2.25µm to 850µm Air Purge 100mm 150mm 300mm 200mm 675mm 450mm Optics can be purged with clean air to3 protect them from contamination during -1 measurement. Minimum supply of 2ft min must be completely oil and water free Non-NEMA enclosures, water and dust proof to NEMA 4/12. Hazardous gases NEMA 7, NEC Class 1 Groups A, B, C and D Sealing Standards Lens Focal Length Software and Data Processing Minimum Specification of PC Operating System Data Analysis Data Display Time History Display Data Export Pentium II 300MHz processor, 32Mbytes RAM, 3Gbyte hard disc and CD–ROM Screen resolution VGA 800x600 (SVGA 1024x768 is recommended) Windows NT™ 4.0 and above Model-independent analysis gives droplet size distribution Patented Multiple Scattering Algorithm for operation at high concentrations (US Patent No. 5,619,324) Statistical analysis for QA/QC User configurable SPC charting and report formats “Frame-by-frame” data display Direct to Excel or similar spreadsheet in text file format (ASCII) Accessories XYZ aerosol mounting unit Weights and Dimensions Module Weights Transmitter module Receiver module Bench Options Short Bench Long Bench Accurate positioning in X, Y and Z axes is accomplished using this mount 5.75kg 7.50kg 300mm spray gap. Height of laser above bench 250mm 500mm spray gap. Height of laser above bench 250mm Custom Benches are available at extra cost depending on material, finish and size X5 X7 X6 Setting standards of excellence – our commitment to you The people at Malvern Instruments are innovators, not just in product design, but in every area of business. Malvern has invested to achieve ISO9001 with TickIt accreditation – we accept nothing short of excellence. As many of our systems are used in the toughest regulatory markets, product validation and development traceability are key commitments to our customers. Malvern trained specialists are available in more than 50 countries to assist with applications development and to advise on and analyze ‘difficult’ samples. Our laboratory facilities in North and South America, Europe and Asia routinely run thousands of customer samples every year. Malvern Instruments Limited Enigma Business Park Grovewood Road Malvern Worcestershire UK WR14 1XZ Tel: +44 (0)1684 892456 Fax: +44 (0)1684 892789 Our innovative approach to customer service is illustrated by the development of after-sales remote diagnostics. Malvern Instruments’ service specialists can access and control systems via standard telephone lines in order to minimize down time and reduce cost. Malvern Instruments SA Parc Club de L’Université 30, Rue Jean Rostand 91893 Orsay Cedex France Tél: +33 (1) 69 35 18 00 Fax: +33 (1) 60 19 13 26 In many industries particle size analysis has become a key QC parameter. The need to obtain data as close to the line as possible, and to react quickly to that data, is increasingly important. Malvern’s in-process division applies its expertise to meet this challenge through the development and production of on-line particle characterization systems for in-process applications. Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments’ policy of continual product development we are unable to guarantee the accuracy of this, or any other document after the date of publication.We therefore disclaim all liability for any changes, errors or omissions after the date of this publication. Malvern Instruments GmbH Rigipsstraße 19 71083 Herrenberg Germany Tel: +49 (0) 7032 97770 Fax: +49 (0) 7032 77854 Malvern Instruments (Asia Pacific) 59 Jalan Maarof Bangsar Baru 59000 Kuala Lumpur Malaysia Tel/Fax: 603 283 6139 Malvern Instruments Nordic AB Box 15045, Vallongatan 1 750 15 UPPSALA Sweden Tel: +46 (0) 18 55 24 55 Fax: +46 (0) 18 55 11 14 www.malvern.co.uk Ref: MRK349 For a full description of all Malvern Instruments’ products and services please contact your nearest office. Designed & produced by C•Force Communications Limited, Hungerford, UK. With laboratory systems and in-process products, Malvern Instruments meets your particle analysis needs from Lab to Line. Malvern Instruments Inc. 10 Southville Road Southborough MA 01772 USA Tel: +1 (508) 480-0200 Fax: +1 (508) 460-9692