DATA SHEET OXYPHAN
Transcription
DATA SHEET OXYPHAN
MEDICAL MEMBRANES DATA SHEET OXYPHAN® Intended use: Cardiac therapy Capillary membrane Type PP 50/200 Sheet No. supersedes 485/9090/832 485/9090/832 Chemical composition of 11/13 of 08/12 Test Ref.No: Polymer Polypropylene Solvent residues 100 ppm 702/206 50 10 µm µm 701/018 300 20 µm µm 701/018 200 µm Physical properties Wall thickness Outer diameter m m Inner diameter (nominal) Maximum pore size (calculated from BP) 0.2 µm Bubble point (Isopropylalcohol, 23°C) 3.1 bar 701/005 Tensile strength 150 cN 701/010 Elongation at break 400 % 701/010 Explosion pressure 3.0 bar 701/005 Implosion pressure 3.5 bar 701/022 75 30 ml/[cm² x min x bar] 701/036 Membrane performance characteristics Nitrogen flux m Available make-up configurations Single layer mats or cross-wound mats The information contained in this data sheet reflects the company’s knowledge and experience at the time of issue. No guarantee can, however, be given as to its completeness. Neither must it be construed to embody any liability on the part of Membrana GmbH beyond the company’s “General Conditions of Sale”. For questions related to handling and manufacturing please refer to our separate “Instructions for handling and processing (OXYPHAN®)”. This product is delivered non-sterile. m = mean value ® = reg.trade mark Membrana GmbH Oehder Straße 28, D – 42289 Wuppertal, Germany Postal address: Postfach 20 01 51, D – 42201 Wuppertal Phone (+49) (202) 6099-1 Fax (+49) (202) 60 70 296 www.membrana.com