DATA SHEET OXYPHAN

Transcription

DATA SHEET OXYPHAN
MEDICAL MEMBRANES
DATA SHEET
OXYPHAN®
Intended use:
Cardiac therapy
Capillary membrane
Type PP 50/200
Sheet No.
supersedes
485/9090/832
485/9090/832
Chemical composition
of 11/13
of 08/12
Test Ref.No:
Polymer
Polypropylene
Solvent residues

100
ppm
702/206
50
10
µm
µm
701/018

300
20
µm
µm
701/018

200
µm
Physical properties
Wall thickness
Outer diameter
m
m
Inner diameter (nominal)
Maximum pore size
(calculated from BP)

0.2
µm
Bubble point
(Isopropylalcohol, 23°C)

3.1
bar
701/005
Tensile strength

150
cN
701/010
Elongation at break

400
%
701/010
Explosion pressure

3.0
bar
701/005
Implosion pressure

3.5
bar
701/022
75
30
ml/[cm² x min x bar]
701/036
Membrane performance characteristics
Nitrogen flux
m

Available make-up configurations
Single layer mats or cross-wound mats
The information contained in this data sheet reflects the company’s
knowledge and experience at the time of issue. No guarantee can, however, be given as to its completeness. Neither must it be construed to
embody any liability on the part of Membrana GmbH beyond the
company’s “General Conditions of Sale”.
For questions related to handling and manufacturing please refer to
our separate “Instructions for handling and processing
(OXYPHAN®)”.
This product is delivered non-sterile.
m = mean value
® = reg.trade mark
Membrana GmbH
Oehder Straße 28, D – 42289 Wuppertal, Germany
Postal address: Postfach 20 01 51, D – 42201 Wuppertal
Phone (+49) (202) 6099-1
Fax
(+49) (202) 60 70 296