[ Care and Use ManUal ] Sirocco Protein PreciPitation Plate
Transcription
[ Care and Use ManUal ] Sirocco Protein PreciPitation Plate
[ Care and Use Manual ] Sirocco Protein Precipitation Plate Step 2 – Add Crash Solvent The Sirocco™ plate was designed to improve the automation and throughput of protein precipitation in drug metabolism/ADME/toxicology labs. Add precipitation solvent to all wells. Use a ratio of at least 3:1 solvent The Sirocco Protein Precipitation Plate is a unique combination of novel to sample. Acetonitrile is the recommended precipitation solvent. The 96-well plate tip design with proprietary new filter membranes. Sirocco plate is designed to prevent solvent from leaking through the plate at this step. A straightforward 7-step procedure designed to enable precipitate free Step 3 – Load Sample sample filtrates suitable for high-throughput LC/MS analysis. Add 20-200 µL of sample (plasma, urine, etc.) rapidly into the wells. Rapid delivery of sample into the crash solvent creates mixing and initiates precipitation process. Step 1 – Preparation Place the Sirocco Plate on a collection plate for processing. This protects the valve tips during the processing steps. Sirocco Protein Precipitation Plate 96-Well Collection Plate Sirocco Protein Precipitation Plate 1 [ Care and Use Manual ] Step 4 – Seal Step 6 – Filter Apply vented cap mat, using a cap roller to ensure secure, uniform sealing. After mixing, place the Sirocco plate on a Vacuum Manifold (e.g. Waters This will prevent splashing and cross-talk during mixing. vacuum manifold P/N 186001831), and apply 8-5” Hg (271-508 mbar) of vacuum for up to three minutes. A higher vacuum setting can be used if flow does not occur at 10” Hg. Cap Mat Note: The Sirocco proprietary valve tip design prevents liquid flow through the device until a minimum 5” Hg (169 mbar) of vacuum is applied. Step 5 – Mix Note: The Sirocco Plate can be filtered by using centrifiguation (@ 2,000 x g Shake (mix) the loaded Sirocco plate/collection plate assembly on a vortexer for 1 min.) in place of vacuum device. (e.g., Multi-Tube Vortexer from VWR) at medium setting (e.g., setting 5 on a 1-10 scale) for 0.5 to 1.0 minute. Note: There is no need to remove the vented cap mats or valve tips from the Step 7– Analysis Sirocco Plate before filtration. This eliminates the risk of cross-talk between Discard the Sirocco Plate and remove the collection plate containing clear wells. The valves and vented cap mat are designed to open under vacuum filtrate from the vacuum device, seal and transfer to the LC/MS autosam- to allow controlled flow during filtration. pler or store for later analysis. Note: Excessive mixing at too high a setting (e.g., setting 8 on a 1-10 Note: There is no need to remove cap mats or valve tips from Sirocco Plate. scale) could cause cross-talk. Under mixing (e.g., setting 2 on a 1-10 scale) The Sirocco Plate is discarded after filtration. will cause clogging or a cloudy filtrate. Xevo™ TQ MS ACQUITY® SM BSM Sirocco Protein Precipitation Plate 2 [ Care and Use Manual ] Sales Offices Ordering Information Description Sirocco Protein Precipitation Plate (includes 1 plate and 1 cap mat) Sirocco Protein Precipitation Plate (includes 5 plates and 5 cap mats) Volume Qty/Pkg Part Number – 1/pkg 186003873 – 5/pkg 186002448 Austria and European Export (Central South Eastern Europe, CIS and Middle East) 43 1 877 18 07 Italy 39 02 265 0983 Australia 61 2 9933 1777 Mexico 52 55 5524 7636 Belgium 32 2 726 1000 The Netherlands 31 76 508 7200 Brazil 55 11 5094-3788 Canada 1 800 252 4752 x2205 China 86 21 6879 5888 CIS/Russia +497 727 4490/290 9737 Japan 81 3 3471 7191 Korea 82 2 6300 4800 Norway 47 6 384 60 50 Poland 48 22 6393000 Puerto Rico 1 787 747 8445 Singapore 65 6593 7100 Czech Republic 420 2 617 1 1384 Spain 34 936 009 300 Denmark 45 46 59 8080 Switzerland 41 56 676 70 00 Finland 09 5659 6288 Taiwan 886 2 2543 1898 France 33 1 30 48 72 00 United Kingdom 44 208 238 6100 Germany 49 6196 400600 Hong Kong 852 2964 1800 Hungary 36 1 350 5086 India and India Subcontinent 91 80 2837 1900 Sweden 46 8 555 11 500 All other countries: Waters Corporation U.S.A. 1 508 478 2000 1 800 252 4752 www.waters.com Ireland 353 1 448 1500 © 2010 Waters Corporation. Waters, The Science of What’s Possible, Xevo and ACQUITY are trademarks of Waters Corporation. Pall, , and Sirocco are trademarks of Pall Corporation. VW R is a trademark of VW R International. August 2010 720003377EN Rev B KK-PDF 3 Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com